K Number
K143437
Manufacturer
Date Cleared
2015-06-25

(206 days)

Product Code
Regulation Number
880.5570
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Droplet® Pen Needles are intended for use with pen injector device for the subcutaneous injection of insulin.

Device Description

The Droplet® Pen Needles are sterile, single use needles designed to be used with commercially available pen-injectors for the subcutaneous injection of insulin. Pen needles are used by consumers, caregivers and healthcare professionals. The pen needle assembly consists of a double-ended cannula that is assembled into an injection molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient end and the cartridge end of the cannula are lubricated using a silicone based lubricant for ease of injection and rubber septum penetration. There is an inner needle shield assembled over the patient end of the cannula to protect the needle point from damage and accidental needle sticks. There is also an outer protective container. Each pen needle assembly is protected with a peel away seal to provide a sterility barrier. To use a pen needle, the user needs to remove the seal, remove the outer protective cap and attach it to the pen injector. Then the user removes the inner protective cap to expose the needle and make an injection. Following the injection the user inserts the used pen needle into the outer cap to remove the pen needle from the pen injector and dispose of it immediately. The pen needle is individually packaged and sterilized with Gamma radiation. It is intended for single use only. Droplet® Pen Needles are available in the following lengths and gauges: 4 mm x 32G, 5 mm x 32G, 6 mm x 32G, 8 mm x 32G, 5 mm x 31G, 6 mm x 31G, 8 mm x 31G, 10 mm x 29G, 12 mm x 29G.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Droplet® Pen Needles, extracted from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Test ParameterAcceptance Criteria (from ISO 11608-2:2012)Reported Device Performance
MaterialsThe needle shall be made of tubing materials specified in ISO 9626.Meets standard
DimensionsThe needles shall fit the test apparatus specified in item 7.3 of ISO 11608-2.Meets standard
Determination of flow rate through the needleThe needle was tested in accordance with Annex A to ISO 11608-2 to determine flow rate through the needle.Meets standard
Bond between hub and needle tubeThe union of the hub and needle tube shall not break when tested in accordance with Clause 9 of ISO 11608-2.Meets standard
Freedom from defectsThe needle tube shall fulfill the requirements of ISO 7864, 11.3.Meets standard
LubricationThe needle tube should be lubricated at both the patient end and the cartridge end. The lubricant shall not, under normal or corrected-to-normal vision, be visible as droplets of fluid on the outside surface of the needle tube.Meets standard
Dislocation of measuring point at patient endDislocation of the cannula point at the patient end shall be in accordance with Table 2 (not provided in this document) below when tested as per Clause 8 (of ISO 11608-2).Meets standard
Determination of functional compatibility with needle-based injection systemsCompatibility with any NIS shall be claimed only after testing in accordance with Clause 11.Meets standard
Ease of assembly and disassemblyAttachment of the needle shall be possible without removing the needle from its opened unit packaging. Compliance is checked according to the requirements of Clause 11.Meets standard
Pre-conditioning of needlesAll requirements of the standard related to preconditioning of needles were met.Meets standard
Biocompatibility - CytotoxicityBased on ISO 10993-1 requirements.Passed
Biocompatibility - SensitizationBased on ISO 10993-1 requirements.Passed
Biocompatibility - Intracutaneous ReactivityBased on ISO 10993-1 requirements.Passed
Biocompatibility - HaemocompatibilityBased on ISO 10993-1 requirements.Passed
Biocompatibility - System toxicity (acute)Based on ISO 10993-1 requirements.Passed
Biocompatibility - Subchronic toxicity (subacute toxicity)Based on ISO 10993-1 requirements.Passed
Sterility (Gamma Radiation)To ensure sterility of the device over its shelf life, relevant sterility testing is conducted. (Validated in accordance with ISO 11137-1 to achieve SAL 10-6)Passed all relevant sterility tests successfully.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample sizes used for each specific test in the "Non-Clinical Performance Data" section. It broadly states that "Verification/Validation testing was done according to the requirements of ISO 11608-2:2012 as summarized below. All testing met the applicable requirements." Similarly, for biocompatibility, it states "The selection of biocompatibility tests was made based on the requirements of ISO 10993-1. All the tests have been successfully passed." The sterility testing states "Droplet® Pen Needles have passed all relevant sterility tests successfully."

The data provenance is for a manufacturing company located in Poland (HTL-Strefa S.A.). The data would be considered prospective as it's generated for the purpose of a 510(k) submission, meaning the tests were conducted specifically for this device to demonstrate its performance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The tests performed are primarily engineering and laboratory-based, adhering to international standards (ISO), rather than relying on human expert assessment for "ground truth" in the way clinical studies or image interpretations might.

4. Adjudication Method for the Test Set

This information is not applicable/not provided. The testing described is objective, standards-based, and performed in a laboratory/technical setting, rather than through expert human adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study done. This document describes a medical device (pen needles) that does not involve AI or human readers in an interpretive capacity.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

There was no standalone (algorithm only) performance study done. This device is a physical medical instrument, not a software algorithm.

7. The Type of Ground Truth Used

The ground truth used for the performance data in this submission is based on established international standards and objective measurement criteria. For example:

  • Physical and mechanical properties are compared against the specifications in ISO 11608-2:2012 (e.g., materials per ISO 9626, bond strength per Clause 9, freedom from defects per ISO 7864).
  • Biocompatibility is evaluated against the requirements of ISO 10993-1.
  • Sterility is validated according to ISO 11137-1.

8. The Sample Size for the Training Set

This information is not applicable/not provided. This device is a physical medical product, not an AI/machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/not provided, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 25, 2015

HTL-Strefa S.A. Ms. Aleksandra Prazmowska-Wilanowska QA/RA/ Director Adamówek 7 95-035 Ozorków Poland

Re: K143437

Trade/Device Name: Droplet® Pen Needles Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: May 21, 2015 Received: May 26, 2015

Dear Ms. Prazmowska-Wilanowska:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Prazmowska-Wilanowska

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143437

Device Name Droplet® Pen Needles

Indications for Use (Describe)

The Droplet® Pen Needles are intended for use with pen injector device for the subcutaneous injection of insulin.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K143437

As required by the Safe Medical Devices Act of 1990 and in accordance with 21 CFR ട്ല807.92(a).

[807.92 (a)(1,2)]

Date Summary Prepared:June 11, 2015
Submitted By:HTL-STREFA S.A.Adamówek 7,95-035 OzorkówPOLANDPhone: +48 42 270 00 10Fax: +48 42 270 00 20
Contact Person:Aleksandra Prazmowska-WilanowskaQA/RA Directorola.prazmowska-wilanowska@htl-strefa.pl
Trade Name:Droplet® Pen Needles
Models:4mm x 32G, 5mm x 32G, 6mm x 32G, 8mm x 32G, 5mm x 31G, 6mm x 31G, 8mm x 31G, 10mm x 29G, 12mm x 29G
Common Name:Insulin Pen Needle
Classification name:Hypodermic Single Lumen Needle
Device Classification:Class II, 21 CFR § 880.5570Product Code: FMI
Review Panel:80 General Hospital

Predicate Devices: [807.92(a)(3)]

The legally marketed devices to which substantial equivalence is claimed are:

ManufacturerTrade Name510(k) Number
Becton Dickson & Co.BD Ultra-FineK031200
Becton Dickson & Co.BD Ultra-FineK100005

Reference devices:

ManufacturerTrade Name510(k) Number
Novo Nordisk Inc.Novofine® PlusK133738
Disetronic Medical SystemsPenfine® Insulin InjectionPen NeedleK013782

{4}------------------------------------------------

Description of Device: [807.92(a)(4)]

The Droplet® Pen Needles are sterile, single use needles designed to be used with commercially available pen-injectors for the subcutaneous injection of insulin. Pen needles are used by consumers, caregivers and healthcare professionals. The pen needle assembly consists of a double-ended cannula that is assembled into an injection molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient end and the cartridge end of the cannula are lubricated using a silicone based lubricant for ease of injection and rubber septum penetration. There is an inner needle shield assembled over the patient end of the cannula to protect the needle point from damage and accidental needle sticks. There is also an outer protective container. Each pen needle assembly is protected with a peel away seal to provide a sterility barrier. To use a pen needle, the user needs to remove the seal, remove the outer protective cap and attach it to the pen injector. Then the user removes the inner protective cap to expose the needle and make an injection. Following the injection the user inserts the used pen needle into the outer cap to remove the pen needle from the pen injector and dispose of it immediately. The pen needle is individually packaged and sterilized with Gamma radiation. It is intended for single use only. Droplet® Pen Needles are available in the following lengths and gauges: 4 mm x 32G, 5 mm x 32G, 6 mm x 32G, 8 mm x 32G, 5 mm x 31G, 6 mm x 31G, 8 mm x 31G, 10 mm x 29G, 12 mm x 29G.

Indications for Use: [807.92(a)(5)]

The Droplet® Pen Needles are intended for use with pen injector device for the subcutaneous injection of insulin.

Technological Characteristics: [807.92(a)(6)]

A comparison of characteristics of Droplet pen needles and the predicate devices is shown in the table below:

Device NameSubject DevicePredicateDevice #1Predicate Device#2
ManufacturerHTL STREFA S.A.Becton DickinsonBecton Dickinson
510(k) NumberPendingK031200K100005

{5}------------------------------------------------

Intended useintended for usewith pen injectordevice for thesubcutaneousinjection of insulinintended for usewith pen injectordevice forthe subcutaneousinjection of insulinintended for usewith pen injectordevice forthe subcutaneousinjection of drugs,including insulinand exenatide
Operation principleManualManualManual
DesignNeedle assembly –cannula, needle hub,inner protective cap,outer protective cap,sealNeedle assembly –cannula, needlehub, innerprotective cap,outer protectivecap, sealNeedle assembly –cannula, needlehub, innerprotective cap, outerprotective cap, seal
Product CodeFMIFMIFMI
Length4mm, 5mm,6mm,8mm,10mm,12mm5mm, 8mm,12.7mm4mm
MaterialsNeedletubeAISI 304Grade StainlessSteelAISI 304Grade StainlessSteelAISI 304Grade StainlessSteel
HubPolypropylenePolypropylenePolypropylene
PrimaryContainerandNeedleShieldPolypropylenePolypropylenePolypropylene
BiocompatibilityConforms toISO10993-1Conforms toISO10993-1Conforms toISO10993-1
SterilizationGamma radiation(validated inaccordance withISO 11137-1 toachieve SAL 10-6)Gamma radiationGamma radiation

{6}------------------------------------------------

Size characteristics of the reference devices:

ManufacturerTrade NameSize
Disetronic Medical SystemsPenfine® Insulin InjectionPen Needle29G x 6mm
30G x 6mm
31G x 6mm
29G x 10mm
30G x 10mm
31G x 10mm
29G x 12mm
30G x 12mm
31G x 12 mm
Novo Nordisk Inc.Novofine® Plus32G x 6mm

The reference devices are provided for purposes of sizes and lengths that are not available with the predicate devices. The reference devices have same or similar IFUs as the subject device.

Non-Clinical Performance Data: [(807.92(b)(1)]

Verification/Validation testing was done according to the requirements of ISO 11608-2:2012 as summarized below. All testing met the applicable requirements.

Test ParameterRequirement – ISO 11608-2:2012Result
MaterialsThe needle shall be made of tubingmaterials specified in ISO 9626.Meets standard
DimensionsThe needles shall fit the test apparatusspecified in item 7.3 of ISO 11608-2.Meets standard
Determination of flow ratethrough the needleThe needle was tested in accordance withAnnex A to ISO 11608-2 to determineflow rate through the needle.Meets standard
Bond between hub and needletubeThe union of the hub and needle tubeshall not break when tested in accordancewith Clause 9 of ISO 11608-2.Meets standard
Freedom from defectsThe needle tube shall fulfill therequirements of ISO 7864, 11.3.Meets standard
LubricationThe needle tube should be lubricated atboth the patient end and the cartridge end.The lubricant shall not, under normal orMeets standard

{7}------------------------------------------------

Test ParameterRequirement – ISO 11608-2:2012Result
corrected-to-normal vision, be visible as droplets of fluid on the outside surface of the needle tube.
Dislocation of measuring point at patient endDislocation of the cannula point at the patient end shall be in accordance with Table 2 below when tested as per Clause 8 (of ISO 11608-2).Meets standard
Determination of functional compatibility with needle-based injection systemsCompatibility with any NIS shall be claimed only after testing in accordance with Clause 11.Meets standard
Ease of assembly and disassemblyAttachment of the needle shall be possible without removing the needle from its opened unit packaging. Compliance is checked according to the requirements of Clause 11.Meets standard
Pre-conditioning of needlesAll requirements of the standard related to preconditioning of needles were met.Meets standard

Biocompatibility testing

The following biocompatibility testing has been conducted on the Droplet® Pen Needles. The selection of biocompatibility tests was made based on the requirements of ISO 10993-1. All the tests have been successfully passed.

TestResult
CytotoxicityPassed
SensitizationPassed
Intracutaneous ReactivityPassed
HaemocompatibilityPassed
System toxicity (acute)Passed
Subchronic toxicity (subacute toxicity)Passed

{8}------------------------------------------------

Sterility testing

Droplet® Pen Needles are sterilized with Gamma radiation. To ensure sterility of the device over its shelf life, relevant sterility testing is conducted. Droplet® Pen Needles have passed all relevant sterility tests successfully.

Clinical Performance Data: [(807.92(b)(2)] Clinical data were not provided.

Conclusion: [(807.92(b)(3)]

Droplet® Pen Needles are substantially equivalent in the intended use, technology/principle of operation, materials and performance to the predicate devices and do not raise any new questions of safety and effectiveness.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).