(57 days)
Wound Matrix TF is intended for the management of wounds including:
- · Partial and full thickness wounds;
- · Pressure ulcers;
- Venous ulcers:
- · Diabetic ulcers;
- Chronic vascular ulcers;
- · Tunneled, undermined wounds;
- · Surgical wounds (donor sites/grafts, post-Mohs' surgery, post-laser surgery, podiatric, wound deliscence);
- · Trauma wounds (abrasions, lacerations, second-degree burns, and skin tears);
- · Draining wounds.
The device is supplied sterile and is intended for one-time use.
Wound Matrix TF is an animal-sourced, acellular wound dressing that is derived from porcine liver tissue. The liver tissue undergoes perfusion decellularization and the resulting wound dressing is comprised primarily of collagen type I. The device is intended for use in the management of wounds. Wound Matrix TF is terminally sterilized in its packaging and is hydrated, moist and flexible when its packaging is opened. The dressing is available in sizes ranging from 1 cm x 2 cm to 10 cm x 25 cm, and may be trimmed or cut as required.
The provided text does not describe acceptance criteria or a study proving that the device meets acceptance criteria in the way typically expected for a medical device's performance claims (e.g., accuracy, diagnostic sensitivity/specificity).
Instead, this document is a 510(k) summary for a wound matrix (Wound Matrix TF), which is a type of medical device generally used for wound management, not a diagnostic AI device or a device with quantifiable performance metrics like sensitivity/specificity. The "Performance Data" section specifically states: "Verification and validation testing was conducted to confirm that bioburden, endotoxin and device dimensions were within the specified values in accordance with the risk assessment."
Therefore, it's impossible to create the requested table and answer many of the questions as they pertain to AI/diagnostic performance. However, I can extract the relevant information regarding the device itself and its "performance data" as described.
Here's the information based on the provided document, addressing the prompts where applicable and noting where the information is not present or not relevant:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance/Confirmation |
|---|---|
| Bioburden within specified values | Confirmed to be within specified values in accordance with the risk assessment. |
| Endotoxin within specified values | Confirmed to be within specified values in accordance with the risk assessment. |
| Device dimensions (including fenestrations) | Confirmed to be within specified values in accordance with the risk assessment. |
| Minimum thickness specification (0.1 mm) | The only modification was a reduction in the minimum thickness specification from 0.3 mm to 0.1 mm (implies met). |
2. Sample size used for the test set and data provenance
- Sample size: Not specified. The document mentions "Verification and validation testing" but does not give specific sample sizes for bioburden, endotoxin, or dimensional testing.
- Data provenance: Not applicable in the context of diagnostic/AI data. The tests are laboratory/manufacturing quality control tests on the device itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of experts: Not applicable. The "ground truth" for bioburden, endotoxin, and dimensions would be established by scientific/engineering standards and measurements, not by expert consensus in the diagnostic sense.
- Qualifications of experts: Not applicable.
4. Adjudication method for the test set
- Adjudication method: Not applicable. This is not a diagnostic device involving human interpretation or adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance
- MRMC study: No, not applicable. This is not an AI diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone study: No, not applicable. This is not an AI algorithm.
7. The type of ground truth used
- Type of ground truth: Laboratory measurements and engineering specifications for bioburden, endotoxin, and physical dimensions.
8. The sample size for the training set
- Sample size for training set: Not applicable. This is not an AI device.
9. How the ground truth for the training set was established
- How ground truth was established: Not applicable. This is not an AI device.
Summary of Device and its "Performance Data" Context:
The device, "Wound Matrix TF," is an animal-sourced, acellular wound dressing derived from porcine liver tissue. The 510(k) summary focuses on demonstrating "substantial equivalence" to a predicate device (Miromatrix Wound Matrix, K140510) and a reference device (Cook Biotech Oasis® Wound Matrix, K061711).
The "performance data" provided refers to verification and validation testing conducted to confirm manufacturing and material quality attributes such as bioburden, endotoxin levels, and device dimensions. The primary change from its previous clearance (K140510) was the addition of fenestrations and a reduction in minimum thickness. These changes were assessed via a risk assessment to confirm continued substantial equivalence to the predicate device, rather than through a clinical performance study with human-subject outcomes or diagnostic accuracy metrics.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name encircling a symbol of a stylized caduceus, which is a staff entwined by two snakes. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the seal. The caduceus symbol is composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 27, 2015
Miromatrix Medical Incorporated % Mr. Stephen Rhodes Biologics Consulting Group Incorporated 400 North Washington Street, Suite 100 Alexandria, Virginia 22314
Re: K143426
Trade/Device Name: Wound Matrix TF Regulatory Class: Unclassified Product Code: KGN Dated: December 30, 2014 Received: December 31, 2014
Dear Mr. Rhodes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K143426
Device Name Wound Matrix TF
Indications for Use (Describe)
Wound Matrix TF is intended for the management of wounds including:
- · Partial and full thickness wounds;
- · Pressure ulcers;
- Venous ulcers:
- · Diabetic ulcers;
- Chronic vascular ulcers;
- · Tunneled, undermined wounds;
- · Surgical wounds (donor sites/grafts, post-Mohs' surgery, post-laser surgery, podiatric, wound deliscence);
- · Trauma wounds (abrasions, lacerations, second-degree burns, and skin tears);
- · Draining wounds.
The device is supplied sterile and is intended for one-time use.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
5. 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for Wound Matrix TF is provided below.
| Device Common Name: | Animal-derived, extracellular matrix wound care product |
|---|---|
| Device Proprietary Name: | Wound Matrix TF |
| Submitter: | Miromatrix Medical, Inc.18683 Bearpath TrailEden Prairie, MN 55347www.miromatrix.com |
| Prepared By | Stephen P. RhodesBiologics Consulting Group, Inc.400 N. Washington St., Suite 100Alexandria, VA 22314Phone: 301-579-3700Fax: 703-548-7457Email: srhodes@bcg-usa.com |
| Contact: | Jeff RossVP Product DevelopmentMiromatrix Medical, Inc.Phone: 763-458-8801Email: jross@miromatrix.com |
| Date Prepared: | November 29, 2014 |
| Classification Regulation: | Unclassified |
| Panel: | General & Plastic Surgery |
| Product Code: | KGN |
| Predicate Device:Reference Device: | K140510, Miromatrix Wound MatrixK061711, Cook Biotech Oasis® Wound Matrix |
Indication for Use:
Wound Matrix TF is intended for the management of wounds including:
- Partial and full thickness wounds; ●
- . Pressure ulcers;
{4}------------------------------------------------
- Venous ulcers;
- Diabetic ulcers; ●
- Chronic vascular ulcers; ●
- Tunneled, undermined wounds; ●
- Surgical wounds (donor sites/grafts, post-Mohs' surgery, post-laser surgery, podiatric, wound dehiscence);
- Trauma wounds (abrasions, lacerations, second-degree burns, and skin tears); ●
- Draining wounds. ●
The device is supplied sterile and is intended for one-time use.
Device Description:
Wound Matrix TF is an animal-sourced, acellular wound dressing that is derived from porcine liver tissue. The liver tissue undergoes perfusion decellularization and the resulting wound dressing is comprised primarily of collagen type I. The device is intended for use in the management of wounds. Wound Matrix TF is terminally sterilized in its packaging and is hydrated, moist and flexible when its packaging is opened. The dressing is available in sizes ranging from 1 cm x 2 cm to 10 cm x 25 cm, and may be trimmed or cut as required.
Comparison of the Technical Characteristics with the Predicate Device:
The only modification made the device since its previous clearance in K140510 is the addition of fenestrations and a reduction in the minimum thickness specification from 0.3 mm to 0.1 mm.
Performance Data:
Verification and validation testing was conducted to confirm that bioburden, endotoxin and device dimensions were within the specified values in accordance with the risk assessment.
Substantial Equivalence:
The addition of fenestrations and reduced minimum thickness is achieved by the physical cutting of the matrix under quality controlled processes. Wound Matrix TF is substantially equivalent to the predicate device with respect to the indications for use, materials, manufacturing processes, packaging, sterilization, and shelf life.
Conclusion/Summary of SE argument
The substantial equivalence of Wound Matrix TF was established based on a declaration of conformity to design control requirements and on an assessment of the effects of the changes to the predicate device based on a risk assessment. Wound Matrix TF is substantially equivalent to the Miromatrix Wound Matrix predicate device with respect to material composition, device characteristics and intended use.
N/A