K Number
K142881
Date Cleared
2015-02-13

(134 days)

Product Code
Regulation Number
872.3750
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DMRC Dual Cure Orthodontic Band Cement is a dual-cure adhesive intended for use as an orthodontic band cement for bonding of orthodontic bands to enamel

Device Description

DMRC Dual Cure Orthodontic Band Cement is a resin based material containing a dual-cure adhesive for cementation of orthodontic bands. The DMRC Dual Cure Orthodontic Band Cement is a polymer based filling and restorative material; it contains a base and catalyst for dual curing process as orthodontic band cement.

AI/ML Overview

The provided text is related to a 510(k) premarket notification for a dental adhesive device, DMRC Dual Cure Orthodontic Band Cement. This type of document is for regulatory clearance of a medical device and thus does not include information on acceptance criteria for software, detailed clinical study designs, or AI performance metrics.

Therefore, I cannot provide the requested information for the following reasons:

  1. The document is not about AI software or a device that uses AI. It concerns a dental adhesive for orthodontic bands, which is a physical material, not a digital diagnostic or treatment support tool.
  2. The document does not detail acceptance criteria for software performance. The "acceptance criteria" mentioned in the context of the device are for its physical properties and biocompatibility, not for accuracy, sensitivity, or specificity of an algorithm.
  3. No study is mentioned that proves the device meets AI-specific acceptance criteria. The "studies" mentioned are non-clinical bench tests (Vickers Hardness, Ultradent Bracket Shear Bond Strength, Work and Set times, ISO 10993 cytotoxicity) to demonstrate performance comparable to a predicate device. These are standard for material-based medical devices.
  4. There is no mention of sample sizes for test sets, data provenance, ground truth experts, adjudication methods, MRMC studies, standalone algorithm performance, or training set details, as these are all relevant to AI/software performance evaluation, which is not applicable to this device submission.

In summary, the provided text does not contain any information relevant to the acceptance criteria or study design for an AI-powered medical device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem consists of a stylized human figure with three faces, representing the department's focus on health and well-being. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 13, 2015

Danville Materials, LLC Ms. Dong Hua Regulatory Affair Director and QA Manager 3420 Fostoria Way, Suite A-200 San Ramon, CA 94583

Re: K142881

Trade/Device Name: DMRC Dual Cure Orthodontic Band Cement Regulation Number: 21 CFR 872.3750 Regulation Name: Bracket Adhesive Resin and Tooth Conditioner Regulatory Class: II Product Code: DYH Dated: January 13, 2015 Received: January 15, 2015

Dear Ms. Hua:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Dong Hua

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142881

Device Name

DMRC Dual Cure Orthodontic Band Cement

Indications for Use (Describe)

DMRC Dual Cure Orthodontic Band Cement is a dual-cure adhesive intended for use as an orthodontic band cement for bonding of orthodontic bands to enamel

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Danville Engineering & Materials. The logo consists of a square graphic on the left and the company name on the right. The square graphic is a dark blue color and contains a white design that resembles a sun or starburst. The company name is written in a bold, black font, with the word "Danville" appearing on the top line and "ENGINEERING & MATERIALS" appearing on the bottom line in a smaller font.

www.danvillematerials.com

3420 Fostoria Way Suite A-200 San Ramon, CA 94583 USA

510(K) Summary

This summary of the Traditional 510(K) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807.92

A. Applicant's Name and Address

  • A Name: Danville Materials LLC
  • A Address: 3420 Fostoria Way Suite A-200 San Roman, CA 94583 USA
  • A Contact Person: Dong Hua
  • A Regulatory Affair Director and QA Manager Title:
  • A Phone: 800-827-7940/925-973-0710. Ext. 212
  • 925-973-0764 Fax:

  • Date Summary Prepared: January 13, 2015 A

B. The Name of the Device:

  • A Trade/Proprietary Name: DMRC Dual Cure Orthodontic Band Cement
  • A The common name of the device: Dental adhesive, bracket and tooth conditioner, Resin
  • A The Classification Name: Bracket adhesive resin and tooth conditioner per 21 CFR 872.3750, product Code DYH has been classified under section 513 of the Act as a Class II device

C. Legally Marketed Predicate Device to Which Substantial Equivalence (SE) is claimed:

  • A K073697 Transbond Supreme Adhesive By 3M Unitek

D. Description of the Device:

DMRC Dual Cure Orthodontic Band Cement is a resin based material containing a dual-cure adhesive for cementation of orthodontic bands. The DMRC Dual Cure Orthodontic Band Cement is a polymer based filling and restorative material; it contains a base and catalyst for dual curing process as orthodontic band cement. This product can be used for patients of all ages. The materials used in DMRC Dual Cure Orthodontic Band Cement are the same as used by our predicates, Transbond Supreme Adhesive By 3M Unitek (K073697), which is the similar dental adhesive Product. The raw chemical materials have been widely used by numerous manufacturers in the medical/dental industry.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Danville Engineering & Materials. The logo features a square graphic on the left, with a stylized design of radiating lines. To the right of the graphic, the word "Danville" is written in a bold, sans-serif font. Below "Danville", the words "ENGINEERING & MATERIALS" are written in a smaller, sans-serif font.

www.danvillematerials.com

3420 Fostoria Way Suite A-200 San Ramon, CA 94583 USA

  • A Indication For Use: The DMRC Dual Cure Orthodontic Band Cement is a dual-cure adhesive intended for use as an orthodontic band cement for bonding of orthodontic bands to enamel.
  • E. A comparison of the Transbond Adhesive Products by 3M Unitek and the DMRC Dual Cure Orthodontic Band Cement to determine SE:
    • A The equivalence to the predicate device is supported by the physical performance testing
    • A Chemicals, function of each component of the product are identified
    • A Similarities in the Indications for Use:
Product510(K) NumberClassification NameIndications For Use
3M Unitek'sTransbond SupremeAdhesiveK073697Bracket Adhesive Resinand Tooth ConditionerLight cure orthodontic adhesive designedfor bonding brackets and other bondableappliances to etched enamel
DMRC Dual CureOrthodontic BandCementK142881Bracket Adhesive Resinand Tooth ConditionerA dual-cure adhesive intended for use asan orthodontic band cement for bondingof orthodontic bands to enamel
  • A Discussion of Non-Clinical and Clinical Tests performed for Determination of Substantial Equivalence:
    DMRC Dual Cure Orthodontic Band Cement is a resin based material containing a dualcure adhesive for cementation of orthodontic bands. The materials used in DMRC Dual Cure Orthodontic Band Cement are the same as used by our predicate, Transbond Supreme Adhesive By 3M Unitek (K073697), which is a similar dental adhesive product. The raw chemical materials have been widely used by numerous manufacturers in the medical/dental industry.

Non-Clinical Testing performed on DMRC Dual Cure Orthodontic Band Cement included the following:

  • Vickers Hardness
  • Ultradent Bracket Shear Bond Strength ●
  • Work and Set times for Orthodontic Products
  • ISO 10993 testing for Cytotoxicity

All testing performed on the DMRC Dual Cure Orthodontic Band Cement was derived from the risk assessment which evaluated the effects of the feature changes. The efficacy or suitability to the intended purpose of DMRC Dual Cure Orthodontic Band

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Danville Engineering & Materials. The logo consists of a square graphic on the left, which contains a pattern of lines radiating from a central point. To the right of the graphic is the word "Danville" in a bold, sans-serif font. Below "Danville" are the words "ENGINEERING & MATERIALS" in a smaller, sans-serif font.

www.danvillematerials.com

3420 Fostoria Way Suite A-200 San Ramon, CA 94583 USA

Cement has been demonstrated by a combination of in-house testing and side-by-side comparisons to the predicate device currently on the market. Results of our bench testing indicate that DMRC Dual Cure Orthodontic Band Cement performs as well as the predicate device currently on the market.

Discussion of Clinical Tests performed:N/A

F. Conclusion:

Our records indicate that our predicates have been used by dentists and large group practices in the United States and purchased by a large number of international distributors.

In conclusion, the subject device, DMRC Dual Cure Orthodontic Band Cement has been designed and manufactured with the intended use and claims for the product in mind. The bench testing contained in our submission demonstrates that there are no differences in their technological characteristics, thereby not raising any new issues of safety or effectiveness. The DMRC Dual Cure Orthodontic Band Cement is as safe and effective as the predicate device, and may be released to the market.

§ 872.3750 Bracket adhesive resin and tooth conditioner.

(a)
Identification. A bracket adhesive resin and tooth conditioner is a device composed of an adhesive compound, such as polymethylmethacrylate, intended to cement an orthodontic bracket to a tooth surface.(b)
Classification. Class II.