K Number
K142677
Date Cleared
2015-03-27

(189 days)

Product Code
Regulation Number
866.3361
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bruker Daltonics, Inc. MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization – time-of-flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens.

The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast infections.

Device Description

The Bruker Daltonics, Inc. MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization – time-of-flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens. The system measures the time (in the nanosecond range) between pulsed acceleration and the corresponding detector signal, the speed is converted into an exact molecular mass. The mass-to-charge ratio of an ion is proportional to the square of its drift time. Highly abundant microbial proteins (mainly ribosomal proteins) result in a mass spectrum with characteristic mass and intensity distribution. It is specific for many bacteria and is interpreted as a molecular fingerprint to identify the test organism.

AI/ML Overview

Here's an analysis of the provided text to extract the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" in a formal table format with numerical targets. However, based on the performance data presented, the implicit acceptance criteria for identification confidence (log(score) ≥ 2.00) are assessed. The reported performance for "Overall Isolate Performance" (Table 6) shows the aggregate performance across all claims.

Acceptance Criteria (Implicit)Reported Device Performance (Overall Isolate Performance - Table 6)
Identification with high confidence (log(score) ≥ 2.0)3817 / 3966 (96.27%) (high resolution species from reference algorithm)
Percentage of high resolution species correctly identified at high confidence (log(score) ≥ 2.0)96.27%
Percentage of high & low resolution species correctly identified at high or low confidence (log(score) ≥ 1.7) (Total positive identifications)99.02% (calculated as (3817 + 392 + 107 + 13) / (3966 + 406) = 4329 / 4372) which is 99.02% if only considering identified positives and not the incorrect IDs)
Incorrect MBT-CA ID (≥ 1.7) / No ID (< 1.7) should be minimized43 out of 4399 data points (approx. 0.98%) for all isolates

2. Sample Sizes and Data Provenance

  • Test Set Sample Size: 4,395 fresh and stored organisms, generating 4,399 data points for the method comparison study (which serves as the primary test set for demonstrating performance against a reference). This also includes 3,802 replicate testing results reported separately.
  • Data Provenance: The method comparison study was performed at six (6) US clinical test sites and an in-house laboratory. The organisms included both fresh and stored organisms. The document indicates that due to the rarity of some organisms, replicates were tested by multiple sites.

3. Number of Experts and Qualifications for Ground Truth of Test Set

The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") for establishing the ground truth of the test set organisms. However, it indicates:

  • "The interim reference laboratory stored all organisms included in the study and sent all organisms to the sequencing reference laboratory for sequencing and protein sequencing when requested."
  • This implies that a reference laboratory, likely staffed by expert microbiologists and molecular biologists, was responsible for confirming the identity of the organisms used as ground truth, primarily through sequencing and protein sequencing.

4. Adjudication Method for the Test Set

The document does not mention a specific adjudication method like "2+1" or "3+1" for discrepancies in the test set. The ground truth was established by a "sequencing reference laboratory for sequencing and protein sequencing." This suggests that the final identification was based on these molecular methods, rather than an expert consensus process between multiple readers.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an automated mass spectrometry system for identifying microorganisms. The evaluation focuses on the device's ability to identify organisms compared to a reference method (sequencing/protein sequencing), not on how it assists human readers or improves their performance.

6. Standalone (Algorithm Only) Performance

Yes, a standalone performance study was done. The entire Method Comparison section (pages 20-25) describes the performance of the MALDI Biotyper CA System (the algorithm/device) in identifying microorganisms without human intervention in the identification process other than following the described workflow. The log(score) system is an output of the algorithm.

7. Type of Ground Truth Used

The primary type of ground truth used for the test set was molecular sequencing (sequencing and protein sequencing), which is considered a highly accurate and definitive method for microorganism identification. This was handled by a "sequencing reference laboratory."

8. Sample Size for the Training Set

The document does not explicitly state the sample size for the training set. It refers to a "reference spectra library (database)" against which the test organism's spectrum is compared. This library is implicitly the training data, but its specific size or composition is not detailed in terms of number of unique isolates or spectra.

9. How the Ground Truth for the Training Set Was Established

The document states: "Calculates matches by comparing a new spectrum against each single reference entry of a reference database." It also mentions "MALDI Biotyper for Clinical Applications (MBT-CA)" as the database. While it doesn't describe the exact process for establishing ground truth for each entry within that database, it's generally understood that such reference libraries are built using highly characterized strains whose identities have been confirmed through gold-standard methods (e.g., 16S rRNA gene sequencing, whole-genome sequencing, or polyphasic taxonomy). The context implies that the internal reference library's entries serve as the established ground truth for comparison.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

BRUKER DALTONICS, INC MARKUS KOSTRZEWA VICE PRESIDENT CLINICAL MASS SPECTROMETRY R&D FAHRENHEILSTRASSE-4 BREMEN 28359 DE

March 27, 2015

Re: K142677 Trade/Device Name: MALDI Biotyper CA System Regulation Number: 21 CFR 866.3361 Regulation Name: Mass spectrometer system for clinical use for the identification of microorganisms Regulatory Class: II Product Code: PEX Dated: February 27, 2015 Received: March 2, 2015

Dear Dr. Kostrzewa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Uwe Scherf - S for

Sally Hojvat, M.Sc., PhD. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142677

Device Name

MALDI Biotyper CA System

Indications for Use (Describe)

The Bruker Daltonics, Inc. MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization – time-of-flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens.

The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast infections.

BACTERIA

Achromobacter xylosoxidans Acinetobacter baumannii complex [4] Acinetobacter haemolyticus Acinetobacter johnsonii Acinetobacter junii Acinetobacter lwoffii Acinetobacter radioresistens Acinetobacter ursingii Actinomyces meyeri Actinomyces neuii Actinomyces odontolyticus Actinomyces oris Aerococcus urinae Aerococcus viridans Aeromonas salmonicida Aeromonas sp[7] Alcaligenes faecalis Anaerococcus vaginalis Bacteroides fragilis Bacteroides ovatus group Bacteroides thetaiotaomicron group Bacteroides uniformis Bacteroides vulgatus group Bordetella group[3] Bordetella hinzii Brevibacterium casei Brevundimonas diminuta group Burkholderia cepacia complex [13] Burkholderia gladioli Burkholderia multivorans Campylobacter coli Campylobacter jejuni Campylobacter ureolyticus Capnocytophaga ochracea Capnocytophaga sputigena

{3}------------------------------------------------

Chryseobacterium gleum Chryseobacterium indologenes Citrobacter amalonaticus complex Citrobacter freundii complex Citrobacter koseri Clostridium difficile Clostridium perfringens Corynebacterium amycolatum Corynebacterium aurimucosum group Corynebacterium bovis Corynebacterium diphtheriae Corynebacterium glucuronolyticum Corynebacterium jeikeium Corynebacterium kroppenstedtii Corynebacterium macginleyi Corynebacterium minutissimum Corynebacterium propinquum Corynebacterium pseudodiphtheriticum Corynebacterium riegelii Corynebacterium striatum group Corynebacterium tuberculostearicum Corynebacterium ulcerans Corynebacterium urealyticum Corynebacterium xerosis Cronobacter sakazakii group Cupriavidus pauculus group Delftia acidovorans group Dermacoccus nishinomiyaensis Edwardsiella tarda Eikenella corrodens Elizabethkingia meningoseptica group Enterobacter aerogenes Enterobacter amnigenus Enterobacter cloacae complex Enterococcus avium group Enterococcus casseliflavus Enterococcus faecalis Enterococcus faecium Enterococcus gallinarum Enterococcus hirae Escherichia coli Finegoldia magna Fusobacterium canifelinum Fusobacterium necrophorum Fusobacterium nucleatum Gardnerella vaginalis Gemella haemolysans Gemella sanguinis Granulicatella adiacens Haemophilus haemolyticus Haemophilus influenzae Haemophilus parahaemolyticus group Haemophilus parainfluenzae Hafnia alvei

{4}------------------------------------------------

Kingella kingae Klebsiella oxytoca / Raoultella ornithinolytica Klebsiella pneumoniae Kocuria kristinae Kytococcus sedentarius Lactococcus garvieae Lactococcus lactis Leuconostoc mesenteroides Macrococcus caseolyticus Micrococcus luteus Moraxella sg Branhamella catarrhalis Moraxella sg Moraxella nonliquefaciens Moraxella sg Moraxella osloensis Morganella morganii Myroides odoratimimus Myroides odoratus Oligella ureolytica Oligella urethralis Pantoea agglomerans Parabacteroides distasonis Pasteurella multocida Pediococcus pentosaceus Peptoniphilus harei group Peptostreptococcus anaerobius Plesiomonas shigelloides Porphyromonas gingivalis Prevotella bivia Prevotella buccae Prevotella denticola Prevotella intermedia Prevotella melaninogenica Propionibacterium acnes Proteus mirabilis Proteus vulgaris group Providencia rettgeri Providencia stuartii Pseudomonas aeruginosa Pseudomonas fluorescens group Pseudomonas oryzihabitans Pseudomonas putida group Pseudomonas stutzeri Rhizobium radiobacter Rothia aeria Rothia dentocariosa Rothia mucilaginosa Salmonella sp Serratia liquefaciens Serratia marcescens Serratia plymuthica Serratia rubidaea Staphylococcus aureus Staphylococcus auricularis Staphylococcus capitis Staphylococcus caprae

{5}------------------------------------------------

Staphylococcus carnosus Staphylococcus cohnii Staphylococcus epidermidis Staphylococcus equorum Staphylococcus felis Staphylococcus haemolyticus Staphylococcus hominis Staphylococcus lugdunensis Staphylococcus pasteuri Staphylococcus pettenkoferi Staphylococcus pseudintermedius Staphylococcus saccharolyticus Staphylococcus saprophyticus Staphylococcus schleiferi Staphylococcus simulans Staphylococcus vitulinus Staphylococcus warneri Stenotrophomonas maltophilia Streptococcus agalactiae Streptococcus anginosus Streptococcus constellatus Streptococcus dysgalactiae Streptococcus gallolyticus Streptococcus gordonii Streptococcus intermedius Streptococcus lutetiensis Streptococcus mitis / oralis group Streptococcus mutans Streptococcus pneumoniae Streptococcus pyogenes Streptococcus salivarius Sutterella wadsworthensis Vibrio parahaemolyticus Vibrio vulnificus Yersinia enterocolitica Yersinia pseudotuberculosis

YEAST

Candida albicans Candida boidinii Candida dubliniensis Candida duobushaemulonii Candida famata Candida glabrata Candida guilliermondii Candida haemulonis Candida inconspicua Candida kefyr Candida krusei Candida lambica Candida lipolytica Candida lusitaniae Candida metapsilosis Candida norvegensis

{6}------------------------------------------------

Candida orthopsilosis Candida parapsilosis Candida pararugosa Candida pelliculosa Candida tropicalis Candida valida Cryptococcus gattii Cryptococcus neoformans var grubii Cryptococcus neoformans var neoformans Geotrichum candidum Geotrichum capitatum Kloeckera apiculata Pichia ohmeri Saccharomyces cerevisiae Trichosporon asahii

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Bruker. The word "BRUKER" is written in a bold, sans-serif font. Above and below the word are blue lines that form a circular shape, with a blue dot at each end of the lines. The logo is simple and modern.

510(k) SUMMARY

Date of Summary:March 27, 2015
Product NameMBT-CA System
SponsorBruker Daltonics, Inc.40 Manning Road,Billerica, MA 01821
CorrespondentBruker Daltonik GmbHMarkus Kostrzewa, Vice President Clinical Mass Spectrometry R&DFahrenheitstrasse 4Bremen, 28359Phone: +49 (0)421-2205 1258Fax: +49 (0)421-2205 205Email: Markus.Kostrzewa@bruker.com
Device Identification
Trade or Proprietary Name:MALDI Biotyper CA System
Common or Usual Name:System, mass spectrometry, maldi tof, microorganism identification,cultured isolates
Product Code:DEV
Product Code:PEX
Regulation Section:21 CFR 866.3361
Device Class:Class II
Panel:Microbiology

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Bruker. The logo consists of the word "BRUKER" in bold, black letters. Above the word is a blue atom symbol, which is a common symbol used to represent science and technology. The atom symbol has three orbits with two dots on each orbit.

Substantial Equivalency

The Bruker Daltonics, Inc. MBT-CA System is substantially equivalent to the MS (K124067) and the Bruker Datonics, Inc. MBT-CA System (K13083). Table 1 compares the characteristics of the MBT-CA System (New Device) and the Vitek MS (Predicate Device).

Table 1. Substantial Equivalency Table

Similarities
NEW DEVICEPRIMARY PREDICATE DEVICEPREDICATE DEVICE
CharacteristicBruker Daltonics, Inc. MBT-CA SystemVitek® MSBruker Daltonics, Inc. MBT-CA
(TBD)(K124067)System (K130831)
Product CodesPEXPEXPEX
Intended useThe Bruker Daltonics, Inc. MALDI Biotyper CA System is a massspectrometer system using matrix-assisted laserdesorption/ionization - time-of-flight (MALDI-TOF) for theidentification of microorganisms cultured from humanspecimens.The MALDI Biotyper CA System is a qualitative in vitrodiagnostic device indicated for use in conjunction with otherclinical and laboratory findings to aid in the diagnosis ofbacterial and yeast infections.The Vitek® MS is a mass spectrometer systemusing matrix-assisted laserdesorption/ionization-time-of- flight (MALDI-TOF) for the identification of microorganismscultured from human specimen.The VITEK MS is a qualitative in vitro diagnosticdevice indicated for use in conjunction withother clinical and laboratory findings to aid inthe diagnosis of bacterial and yeast infections.See "Differences"
Sample typeIsolated colony from any patient sample source.Acceptable media:• Columbia blood agar with 5% sheep blood• Trypticase soy agar with 5% sheep Blood• Chocolate agar• MacConkey Agar• Columbia CNA agar with 5% sheep blood• Brucella Agar with 5% horse blood• CDC anaerobe Agar with 5% sheep blood• CDC anaerobe 5% sheep blood Agar with phenylethyl alcohol• CDC anaerobe laked sheep blood Agar with kanamycin andvancomycin• Bacteroides bile esculin Agar with amikacin• Clostridium difficile Agar with 7% sheep blood• Sabouraud-Dextrose Agar• Brain Heart Infusion Agar• Campylobacter Agar with 5 Antimicrobics and 10% SheepBlood• Bordet Gengou Agar with 15% sheep bloodIsolated colony from any patient samplesource.Acceptable media:• Columbia blood agar with 5% sheep blood• Trypitcase soy agar with 5% sheep Blood• Chocolate polyvitex agar• Campylosel agar• MacConkey Agar• Modified Sabouraud dextrose Agar• ChromID CPSIsolated colony from any patient samplesource.Acceptable media:• Columbia blood agar with 5% sheepblood• Trypticase soy agar with 5% sheepBlood• Chocolate agar• MacConkey Agar
Type of TestAutomated Mass Spectrometry SystemAutomated Mass Spectrometry SystemAutomated Mass Spectrometry System
Similarities
CharacteristicNEW DEVICEBruker Daltonics, Inc. MBT-CA System(TBD)PRIMARY PREDICATE DEVICEVitek® MS(K124067)PREDICATE DEVICEBruker Daltonics, Inc. MBT-CASystem (K130831)
Matrixα-Cyano-4-hydroxycinnamic acidα-Cyano-4-hydroxycinnamic acidα-Cyano-4-hydroxycinnamic acid
Method ofTestingBacteria & Yeast: Direct testingIf after initial analysis the log(score) is reported at <2.00,organisms may be processed using the Extraction (Ext)procedure or extended Direct Transfer (eDT, 70% aqueousformic acid) procedure. If eDT procedure still yields log(score)<2.00, organism may be processed via Ext procedure.Bacteria & Yeast: Direct testingBacteria: Direct testingIf after initial analysis the log(score) isreported at <2.00, organisms areprocessed using the Extraction procedure.
Result ReportingOrganism identification is reported with high confidence if thelog(score) is ≥2.00.An organism identification is reported with low confidence ifthe log(score) is between 1.70 and <2.00.A single identification is displayed, with aconfidence value from 60.0 to 99.9, when onesignificant organism or organism group isretained."Low-discrimination" identifications aredisplayed when more than one but not morethan four significant organisms or organismgroups are retained.When more than four organisms or organismgroups are found, or when no match is found,the organism is considered unidentified.Organism identification is reported withhigh confidence if the log(score) is ≥2.00.An organism identification is reported withlow confidence if the log(score) is between1.70 and <2.00.
MatchingAlgorithmCalculates matches by comparing a new spectrum against eachsingle reference entry of a reference database.Uses a proprietary process called "massbinning." In this process, the spectrumbetween 3,000 and 17,000 Daltons are dividedinto 1300 pre-definedintervals called "bins". Next, an algorithmbased on supervised machine learning knownas the "Advanced Spectrum Classifier", is usedto determine howinformative each bin was in differentiating thatspecies from all other species in the database.Calculates matches by comparing a newspectrum against each single referenceentry of a reference database.
Recorded massrange2,000 - 20,000 m/z2,000 - 20,000 m/z2,000 - 20,000 m/z
Similarities
CharacteristicNEW DEVICEBruker Daltonics, Inc. MBT-CA System(TBD)PRIMARY PREDICATE DEVICEVitek® MS(K124067)PREDICATE DEVICEBruker Daltonics, Inc. MBT-CA System (K130831)
CalibrationBruker US IVD Bacterial Test Standard (BTS)See "Differences"Bruker US IVD Bacterial Test Standard (BTS)
MALDI Target PlateUS IVD 48 Spot Target• 48 positions reusable steel targetsSee "Differences"US IVD 48 Spot Target• 48 positions reusable steel targets
MALDI-TOF MS instrumentsBruker microflex (benchtop)See "Differences"Bruker microflex (benchtop)
DatabaseMALDI Biotyper for Clinical Applications (MBT-CA)See "Differences"MALDI Biotyper for Clinical Applications (MBT-CA)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Bruker Corporation. The logo consists of the word "BRUKER" in a bold, sans-serif font, with a stylized atom symbol above it. The atom symbol is composed of two overlapping ellipses, with blue dots representing electrons orbiting a central nucleus. The overall design is clean and modern, reflecting Bruker's focus on scientific and technological innovation.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Bruker, a company that specializes in scientific instruments. The logo features the company name in bold, sans-serif font, with the letters stacked on top of each other. Above the name is a stylized atom symbol, consisting of three overlapping elliptical orbits with blue dots representing electrons. The logo is simple, modern, and easily recognizable.

Differences
CharacteristicNEW DEVICEBruker Daltonics, Inc. MBT-CA System(TBD)PREDICATE DEVICEVitek® MS(K124067)PREDICATE DEVICEBruker Daltonics, Inc. MBT-CA System(K130831)
Culture AgeBacteria and yeasts growth should be between18h to 48h (+12h storage at RT)Specific requirements:• Bordetella: Incubation on BG agar shouldnot be longer than 24h (+12h storage at RT).• Campylobacter: Incubation can beprolonged to 72h (+12h storage at RT).• Streptococcus pneumoniae: Incubationshould not be longer than 24h (+12h storageat RT) due to possible autolysis.Bacteria and yeast growth should bebetween 24 to 72 hours.Bacteria growth should be between 18h to 36h
CalibrationBruker US IVD Bacterial Test Standard (BTS)E. coli ATCC 8739See "Similarities"
MALDI TargetUS IVD 48 Spot TargetVITEK MS-DS Target SlidesSee "Similarities"
Plate• 48 positions reusable steel targets• 48 positions disposable plastic targetsSee "Similarities"
MALDI-TOF MSBruker microflexShimadzu AXIMA® Assurance MSSee "Similarities"
instruments(benchtop)(floor standing)See "Similarities"
DatabaseMALDI Biotyper for Clinical Applications (MBT-CA)VITEK® MS V2.0 Knowledge BaseSee "Similarities"

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Bruker Corporation. The logo features the word "BRUKER" in bold, sans-serif font, with the "B" slightly larger than the other letters. Above the word, there is a stylized graphic of an atom, with two intersecting elliptical orbits around a central point, representing the nucleus. The orbits and the central point are in a light blue color.

Differences
CharacteristicNEW DEVICEBruker Daltonics, Inc. MBT-CA System(TBD)PREDICATE DEVICEVitek® MS(K124067)PREDICATE DEVICEBruker Daltonics, Inc. MBT-CA System(K130831)
Intended UseThe Bruker Daltonics, Inc. MALDI Biotyper CA System is a mass spectrometer system using matrix-assistedlaser desorption/ionization - time-of-flight (MALDI-TOF) for the identification of microorganisms culturedfrom human specimens.The MALDI Biotyper CA System is a qualitative in vitrodiagnostic device indicated for use in conjunction withother clinical and laboratory findings to aid in thediagnosis of bacterial and yeast infections.See "Similarities"The Bruker Daltonics, Inc. MALDI Biotyper CASystem is a qualitative in vitro diagnostic massspectrometer system for the identification of Gram-negative bacterial colonies cultured from humanspecimens using matrix-assisted laserdesorption/ionization - time-of-flight (MALDI-TOF)mass spectrometry technology.The MALDI Biotyper CA System is indicated for usein conjunction with other clinical and laboratoryfindings to aid in the diagnosis of Gram-negativebacterial infections.

These differences do not affect substantial equivalence of the MBT-CA System and MBT-CA System (K130831). All systems are mass spectrometers using matrix-assisted laser desorption – time-of-flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens. The differences noted above do not impact the intended use and do not raise questions as to the safectiveness of the test (new) device.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the Bruker logo. The logo consists of the word "BRUKER" in bold, sans-serif font, with a stylized graphic above it. The graphic features two intersecting, curved lines that form a circular shape, with small circles at the ends of the lines, resembling an atom or molecule. The logo is simple and modern, conveying a sense of scientific precision and innovation.

Intended Use

The Bruker Daltonics, Inc. MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization – time-of-flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens.

The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast infections.

The following organisms are claimed:

Bacteria:

Achromobacter xylosoxidansCupriavidus pauculus groupPropionibacterium acnes
Acinetobacter haemolyticusDelftia acidovorans groupProteus mirabilis
Acinetobacter johnsoniiDermacoccus nishinomiyaensisProteus vulgaris group
Acinetobacter juniiEdwardsiella tardaProvidencia rettgeri
Acinetobacter lwoffiiEikenella corrodensProvidencia stuartii
Acinetobacter radioresistensElizabethkingia meningoseptica groupPseudomonas aeruginosa
Acinetobacter ursingiiEnterobacter aerogenesPseudomonas fluorescens group
Acinetobacter baumannii complex [4]Enterobacter amnigenusPseudomonas oryzihabitans
Actinomyces meyeriEnterobacter cloacae complexPseudomonas putida group
Actinomyces neuiiEnterococcus casseliflavusPseudomonas stutzeri
Actinomyces odontolyticusEnterococcus faecalisRhizobium radiobacter
Actinomyces orisEnterococcus faeciumRothia aeria
Aerococcus urinaeEnterococcus gallinarumRothia dentocariosa
Aerococcus viridansEnterococcus hiraeRothia mucilaginosa
Aeromonas salmonicidaEnterococcus avium groupSalmonella sp
Aeromonas sp[7]Escherichia coliSerratia liquefaciens
Alcaligenes faecalisFinegoldia magnaSerratia marcescens
Anaerococcus vaginalisFusobacterium canifelinumSerratia plymuthica
Bacteroides fragilisFusobacterium necrophorumSerratia rubidaea
Bacteroides uniformisFusobacterium nucleatumStaphylococcus aureus
Bacteroides ovatus groupGardnerella vaginalisStaphylococcus auricularis
Bacteroides thetaiotaomicron groupGemella haemolysansStaphylococcus capitis
Bacteroides vulgatus groupGemella sanguinisStaphylococcus caprae
Bordetella group[3]Granulicatella adiacensStaphylococcus carnosus
Bordetella hinziiHaemophilus haemolyticusStaphylococcus cohnii
Brevibacterium caseiHaemophilus influenzaeStaphylococcus epidermidis
Brevundimonas diminuta groupHaemophilus parainfluenzaeStaphylococcus equorum
Burkholderia gladioliHaemophilus parahaemolyticus groupStaphylococcus felis
Burkholderia multivoransHafnia alveiStaphylococcus haemolyticus
Burkholderia cepacia complex [13]Kingella kingaeStaphylococcus hominis
Campylobacter coliKlebsiella pneumoniaeStaphylococcus lugdunensis
Campylobacter jejuniKlebsiella oxytoca /Staphylococcus pasteuri
Raoultella ornithinolytica
Campylobacter ureolyticusKocuria kristinaeStaphylococcus pettenkoferi
Capnocytophaga ochraceaKytococcus sedentariusStaphylococcus pseudintermedius
Capnocytophaga sputigenaLactococcus garvieaeStaphylococcus saccharolyticus

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Bruker. The logo consists of the word "BRUKER" in a bold, sans-serif font, with a stylized atom symbol to the right. The atom symbol is made up of three overlapping circles, with a small dot in the center of each circle. The logo is simple and modern, and the atom symbol suggests that the company is involved in science or technology.

Chryseobacterium gleumLactococcus lactisStaphylococcus saprophyticus
Chryseobacterium indologenesLeuconostoc mesenteroidesStaphylococcus schleiferi
Citrobacter amalonaticus complexMacrococcus caseolyticusStaphylococcus simulans
Citrobacter koseriMicrococcus luteusStaphylococcus vitulinus
Citrobacter freundii complexMoraxella sg Branhamella catarrhalisStaphylococcus warneri
Clostridium difficileMoraxella sg MoraxellanonliquefaciensStenotrophomonas maltophilia
Clostridium perfringensMoraxella sg Moraxella osloensisStreptococcus agalactiae
Corynebacterium amycolatumMorganella morganiiStreptococcus anginosus
Corynebacterium bovisMyroides odoratimimusStreptococcus constellatus
Corynebacterium diphtheriaeMyroides odoratusStreptococcus dysgalactiae
Corynebacterium glucuronolyticumOligella ureolyticaStreptococcus gallolyticus
Corynebacterium jeikeiumOligella urethralisStreptococcus gordonii
Corynebacterium kroppenstedtiiPantoea agglomeransStreptococcus intermedius
Corynebacterium macginleyiParabacteroides distasonisStreptococcus lutetiensis
Corynebacterium minutissimumPasteurella multocidaStreptococcus mutans
Corynebacterium propinquumPediococcus pentosaceusStreptococcus pneumoniae
CorynebacteriumpseudodiphtheriticumPeptoniphilus harei groupStreptococcus pyogenes
Corynebacterium riegeliiPeptostreptococcus anaerobiusStreptococcus salivarius
Corynebacterium tuberculostearicumPlesiomonas shigelloidesStreptococcus mitis / oralis group
Corynebacterium ulceransPorphyromonas gingivalisSutterella wadsworthensis
Corynebacterium urealyticumPrevotella biviaVibrio parahaemolyticus
Corynebacterium xerosisPrevotella buccaeVibrio vulnificus
Corynebacterium aurimucosum groupPrevotella denticolaYersinia enterocolitica
Corynebacterium striatum groupPrevotella intermediaYersinia pseudotuberculosis
Cronobacter sakazakii groupPrevotella melaninogenica

Yeast:

  • Candida albicans Candida boidinii Candida dubliniensis Candida duobushaemulonii Candida glabrata Candida famata Candida guilliermondii Candida haemulonis Candida inconspicua Candida kefyr Candida krusei Candida lambica Candida lipolytica Candida lusitaniae Candida metapsilosis Candida norvegensis Candida orthopsilosis
  • Candida parapsilosis Candida pararugosa Candida pelliculosa Candida tropicalis Candida valida Cryptococcus gattii Cryptococcus neoformans_var_grubii Cryptococcus neoformans var neoformans Geotrichum candidum Geotrichum capitatum Kloeckera apiculata Pichia ohmeri Saccharomyces cerevisiae Trichosporon asahii

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Bruker Corporation. The logo consists of the word "BRUKER" in bold, black letters. Above the word is a blue atom symbol with two electrons orbiting the nucleus. The atom symbol is stylized and has a modern look.

Methodology

Biochemical methods are currently the most commonly used methods for the identification of microorganisms. Organisms are tested against a range of reagents and organism identification is based on a microorganism's reaction to these reagents.

The MBT-CA System uses a different methodology for organism identification based on unique protein patterns of the microorganisms obtained from mass spectrometry. The test organism's spectrum (a pattern of mass peaks) is compared with a reference spectra library (database). Using biostatistical analysis, a probability ranking of the organism identification is generated. The probability ranking is represented as a log(score) between 0.00 and 3.00. Organism identification is reported with high confidence if the log(score) is ≥2.00. An organism identification is reported with low confidence if the log(score) is between 1.70 and <2.00.

Some MBT-CA identifications displayed are non-clinically validated organisms. In the interest of public health, these organisms are displayed but are grayed out in the MBT-CA report as a means of directing the required additional laboratory testing. These results are not reported; identifications must be confirmed using alternate laboratory methods. Results for non-clinically validated organisms cannot be transmitted from the MBT-CA to the laboratory information system.

Organisms to be identified with the MBT-CA System should be isolated for purity on appropriate isolation media.

Direct Transfer (DT): An individual colony from a subculture plate is transferred to a selected position on an US IVD 48 Spot Target plate (target) and overlaid with US IVD HCCA portioned (matrix). The standard solvent (50% acetonitrile / 47.5% H2O / 2.5% trifluoroacetic acid) in the matrix solution extracts proteins (mainly ribosomal proteins, which are present in high concentration) from the microorganisms. When dried matrix crystallizes, the inoculated target is ready to be analyzed on the MBT-CA System. If after initial analysis the log(score) is reported as <2.00, organisms can be processed using the extended Direct Transfer (eDT) procedure or the Extraction (Ext) procedure and analysis repeated. If eDT is employed and log(score) is reported as <2.00, reanalysis via the Extraction procedure may be used.

extended Direct Transfer (eDT):

If DT analysis yields a (log(score) <2.00) result, an individual colony from a subculture plate may be transferred to a selected position on a target and overlaid with 70% aqueous formic acid solution. The target is air-dried and then matrix is overlaid. When dried matrix crystallizes, the inoculated target is ready to be analyzed on the MBT-CA System. If a high confidence result is not achieved (log(score) is reported at <2.00), organisms can be processed using the Extraction procedure and analysis repeated.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for Bruker. The logo consists of the word "BRUKER" in bold, black letters. Above the word is a blue graphic that resembles an atom with electrons orbiting around it. The atom graphic is stylized with curved lines and blue dots.

Extraction procedure (Ext): If after initial analysis and eDT procedure the log(score) is reported at <2.00, organisms are processed using the Extraction procedure and analysis repeated. For this purpose, isolated colonies from the subculture plate are extracted in accordance with MBT-CA System user manual. Afterwards they are transferred to the target and treated as described above.

MALDI-TOF Analysis:

Samples are analyzed using MALDI (matrix-assisted laser desorption/ionization) TOF (time-offlight) mass spectrometry. The matrix transfers protons onto the extracted proteins and absorbs UV light. A laser in the MALDI- TOF mass spectrometer irradiates the matrix sample composite, causing evaporation and release of positively charged intact proteins and peptides ("soft" ionization technique). These ions are electrostatically accelerated over a short distance and arrive in the flight tube at a mass-dependent speed. As different proteins/peptides have different masses, ions arrive at the detector at different times (time-of-flight). The system measures the time (in the nanosecond range) between pulsed acceleration and the corresponding detector signal, the speed is converted into an exact molecular mass. The massto-charge ratio of an ion is proportional to the square of its drift time.

Highly abundant microbial proteins (mainly ribosomal proteins) result in a mass spectrum with characteristic mass and intensity distribution. It is specific for many bacteria and is interpreted as a molecular fingerprint to identify the test organism.

Data acquisition is controlled with MBT-CA Software. The spectrum of the unknown organism is first transformed into a peak list. Using a biostatistical algorithm, this peak list is compared to the reference peak lists of organisms in the reference library (database) and a log(score) is generated. A higher log(score) indicates a higher degree of similarity to the organism in the reference library. Organism identification is reported with high confidence if the log(score) is ≥2.00. An organism identification is reported with low confidence if the log(score) is between 1.70 and <2.00.

The log(score) ranges, defined in the MBT-CA System, are indicative of the probability of organism identification. Results should be reviewed by a trained microbiologist and final organism identification should be based on all relevant information available. This information includes but is not limited to: Gram staining, colony morphology, growth characteristics, sample matrix, etc.

Performance Data

Precision/Repeatability:

Validation of the whole MALDI Biotyper CA System was performed on six (6) working days with two (2) runs/day following manufacturer's instructions for use. Ten (10) test organisms were tested in triplicate via Direct Transfer (DT) and extended Direct Transfer (eDT) in each run. If a replicate yielded a log(score) <2.00, the test organism was repeated in triplicate via Extraction. The study also tested multiple sources of system variability including two (2) test operators, two

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for Bruker. The logo consists of the word "BRUKER" in bold, sans-serif font. Above the word is a blue graphic that resembles an atom with two orbiting electrons. The graphic is stylized and abstract, with the electrons forming a circular path around the central nucleus.

(2) microflex LT/SH instruments and two (2) target plates. Overall results from the precision/repeatability study are presented below.

# samples# samples# samples# samples ≥2.0 ID
Test organismmeasured≥2.0 ID (DT)≥2.0 ID (eDT)(DT/eDT+Ext)
Brevibacterium casei3636 (100%)36 (100%)36 (100%)
Enterococcus faecalis3634 (94%)36 (100%)36 (100%)
Micrococcus luteus3621 (58%)36 (100%)36 (100%)
Staphylococcus aureus3636 (100%)36 (100%)36 (100%)
Staphylococcus epidermidis3636 (100%)36 (100%)36 (100%)
Streptococcus agalactiae3634 (94%)36 (100%)36 (100%)
Candida albicans3618 (50%)30 (83%)36 (100%)
Candida parapsilosis366 (17%)32 (89%)36 (100%)
Candida tropicalis3634 (94%)35 (97%)36 (100%)
Saccharomyces cerevisiae3620 (56%)27 (75%)36 (100%)

Table 2: Overall Precision per Test Organism

Based upon the data presented, the study confirms repeatability and precision of the MALDI Biotyper CA System independent from:

  • System Operators
  • microflex LT/SH instruments
  • Target plates

Limit of Detection/Dynamic Range:

The Limit of Detection study was conducted to estimate the dynamic range (in terms of sample amount) of Gram-positive bacteria and yeasts to be identified on the MALDI Biotyper CA System. Six (6) frequently occurring clinically relevant test organisms [three (3) Gram-positive and three (3) yeast] were chosen for this study. [NOTE: Due to the nature of yeast organisms, dynamic range studies using known yeast concentration was not feasible for the Direct Transfer procedure].

Turbidity measurements of stock suspensions containing microbial material were performed at a wavelength of 600 nm. To determine the amount of cfu/μL the stock suspensions of each testorganism were diluted in a series of 1:10 dilutions resulting in a final dilution of 10′ (Grampositive bacteria) and 10° (yeasts). 10 µL from the final diluted test-suspensions were transferred to TSA isolation media plates and incubated for 18-24h at (37±2)°C for Gram-positive bacteria and at (29±2)°C for yeasts, respectively. To account for random errors, the determination of each suspension's concentration in cfu/μL containing microbial material was done in triplicate. All suspensions were tested in replicates of eight (8) via each testing methodology (DT, eDT, Ext). Study results concluded that the estimated dynamic range for the Direct, extended Direct and Extraction procedure are as follows:

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for Bruker. The logo consists of the word "BRUKER" in bold, black letters. Below the word, there is a blue atom-like graphic with three elliptical orbits intersecting each other. The orbits are connected by small blue dots.

DTeDTEXT
Test OrganismLower limit[cfu/µL]Upper limit[cfu/µL]Lower limit[cfu/µL]Upper limit[cfu/µL]Lower limit[cfu/µL]Upper limit[cfu/µL]
Enterococcus faecalis1.2 x 1066.0 x 1073.6 x 1061.8 x 1083.6 x 1061.8 x 108
Enterococcus faecium4.5 x 1074.5 x 1072.1 x 1061.1 x 1082.1 x 1061.1 x 108
Staphylococcus aureus3.5 x 1051.8 x 1084.1 x 1042.1 x 1084.1 x 1052.1 x 108
Candida albicansN/AN/A2.0 x 1052.0 x 1062.0 x 1061.0 x 107
Candida parapsilosisN/AN/A2.5 x 1052.5 x 1062.5 x 1061.3 x 107
Saccharomyces cerevisiaeN/AN/A1.5 x 1051.5 x 1061.5 x 1057.5 x 106

Media and Colony Stability

With the inclusion of Gram-negative microaerophilic, Gram-negative anaerobic, Gram-positive aerobic and anaerobic and yeast organisms, a study on the following media was conducted to confirm acceptability of the recommended agar and the stability of the colony for up to 12 hours at room temperature after initial plate incubation prior to analysis.

  • Chocolate Agar (CHOC) .
  • Columbia CNA agar with 5% sheep blood (CNA) .
  • Brucella Agar with 5% horse blood (BRU) ●
  • CDC anaerobe Agar with 5% sheep blood (CDC) .
  • CDC anaerobe 5% sheep blood Agar with phenylethyl alcohol ● (CDC/PEA)
  • CDC anaerobe laked sheep blood Agar with kanamycin and vancomycin (CDC/LKV)
  • Bacteroides bile esculin Agar with amikacin (BBE)
  • . Clostridium difficile Agar with 7% sheep blood (CDA)
  • Trypticase Soy Agar with 5% sheep blood (TSA)
  • Sabouraud-Dextrose Agar (SDA)
  • Brain-Heart Infusion Agar (BHI) ●
  • . Campylobacter Agar with 5 Antimicrobics and 10% sheep blood (CAMPY BAP)
  • . Bordet Gengou Agar with 15% sheep blood (BGA)
  • Columbia Blood Agar with 5% sheep blood (CBA) .

Testing was conducted using three (3) Gram-positive bacteria, three (3) yeasts, five (5) anaerobic bacteria, two (2) Campylobacter and three (3) Bordetella species at varying incubation time points in replicates of eight (8). After initial testing, isolates were further tested at room temperature after twelve (12) hours post-incubation.

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for Bruker Corporation. The logo consists of the word "BRUKER" in bold, sans-serif font, stacked above a stylized graphic. The graphic features a blue atom-like design with two intersecting elliptical orbits around a central point, suggesting scientific or technological themes.

The study results confirmed the acceptability of all culture media tested with the following parameters:

  • Bacteria and yeasts growth should be between 18h to 48h (+12h storage at room temperature (RT)).
    Specific requirements:

  • Bordetella: Incubation on BG agar should not be longer than 24h (+12h storage at RT).

  • Campylobacter: Incubation on CAMPY BAP can be prolonged to 72h (+12h storage at RT).

  • Streptococcus pneumoniae: Incubation should not be longer than 24h (+12h storage at RT) due to possible autolysis.

Organism Stability Prior to MBT-CA Analysis

This study was conducted to assess Gram-positive and yeast isolate stability on the target plate prior to matrix overlay via Direct Transfer (DT), extended Direct transfer (eDT) and Extraction (Ext) procedure. Two (2) Gram-positive organisms were inoculated eight times and overlaid with matrix at five (5) different time points. Five (5) yeasts organisms were inoculated eight times and overlaid with matrix at (4) different time points. Extracts of Gram-positive bacteria and yeasts were stored at room temperature and inoculated eight times and overlaid with matrix at five (5) different time points. All testing was performed in duplicate.

The study results confirmed that Gram-positive bacteria and yeast organisms are stable on the target plate for up to 60 minutes and 30 minutes respectively prior to matrix addition. Extracts of Gram-positive bacteria and yeasts are stable at room temperature for up to 24 hours and 4 hours respectively.

Sample Stability Overlaid with Matrix

This study was conducted to assess test organism stability overlaid with matrix after inoculation on the target plate. For this study, six (6) organisms were tested (three (3) Gram-positive and three (3) yeast organisms). All organisms were subcultured and aging experiments were conducted at two (2) temperatures and two (2) different relative humidity conditions to stress the plate. Plates were inoculated with the test organism via DT, eDT and Ext procedure. Plates were then read immediately (0h) and then incubated at each test condition and analyzed at three (3) additional time points (4 hours, 8 hours and 24 hours).

The study results confirmed that organisms overlaid with matrix on the target plate are stable for up to 24 hours when stored at room temperature.

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the logo for Bruker Corporation. The logo consists of the word "BRUKER" in bold, black letters. Above the word is a stylized graphic of three overlapping blue orbits with a blue dot at the end of each orbit. The logo is simple and modern, conveying a sense of technology and innovation.

Reproducibility:

The reproducibility study for Gram-positive aerobic bacteria, Gram-negative microaerophilic bacteria, Gram-positive anaerobic bacteria, Gram-negative anaerobic bacteria and yeast organisms was carried out to confirm day-to-day reproducibility and precision of the MALDI Biotyper CA System at different clinical study sites. The study was conducted for five (5) days with two (2) runs (two (2) operators) each day per clinical site. The sources of variability tested were:

    • Two (2) operators/each clinical study site
    • Three (3) clinical study sites
    • At least four (4) target plates/each clinical study site
    • Four (4) microflex LT/SH instruments

Ten (10) well-characterized organisms were chosen for this study and tested in duplicate via Direct Transfer and extended Direct Transfer procedure in accordance with product instructions. When the DT and/or eDT log(score) was <2.00, per product instructions, the test organism was tested following Extraction procedure in duplicate.

Blinded Test OrganismReproducibilityPanel≥2.0 ID(DT)≥2.0 ID(eDT)≥2.0 ID(DT+eDT+Ext)
Enterococcus faecalisREPRO-160/60 (100%)60/60 (100%)60/60 (100%)
Staphylococcus epidermidisREPRO-258/60 (97%)59/60 (98%)60/60 (100%)
Streptococcus agalactiaeREPRO-358/60 (97%)55/60 (92%)60/60 (100%)
Bacteroides fragilisREPRO-460/60 (100%)60/60 (100%)60/60 (100%)
Fusobacterium necrophorumREPRO-556/60 (93%)53/60 (88%)58/60 (97%)
Clostridium perfringensREPRO-653/60 (88%)58/60 (97%)59/60 (98%)
Propionibacterium acnesREPRO-753/60 (88%)49/60 (82%)60/60 (100%)
Candida albicansREPRO-833/60 (55%)41/60 (68%)60/60 (100%)
Saccharomyces cerevisiaeREPRO-95/60 (8%)0/60 (0%)31/60 (52%)
Cryptococcus neoformans vargrubiiREPRO-1031/60 (52%)44/60 (73%)53/60 (88%)
TOTAL467/600 (78%)479/600 (80%)561/600 (94%)

Table 3: Overall Reproducibility Panel Testing per Test Organism using ≥2.0 MBT-CA log(scores)

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the logo for Bruker Corporation. The logo consists of the word "BRUKER" in bold, black letters. Above the word is a stylized graphic of three overlapping blue circles, resembling an atom or a stylized orbit. The circles are connected by two small blue dots.

94% of test organisms were correctly identified with a log(score) ≥2.00 result. In addition, no isolates were falsely identified. Thus, data confirm reproducibility and precision of the whole MALDI Biotyper CA System independent from:

  • Clinical Site
  • System operator
  • microflex LT/SH instrument
  • Target plate

Challenge Panel:

A panel of 55 organisms (24 Gram-positive aerobic bacteria, 1 Gram-negative microaerophilic bacterium, 4 Gram-negative anaerobic bacteria, 6 Gram-positive anaerobic bacteria, 20 yeasts) was tested at four (4) study sites. Fifty-three (53) of the organisms included in the panel were selected from stored organisms tested during the clinical study. Two (2) were selected from strain collections. The study reference laboratory, prepared the panel. Organism identifications were blinded to test sites. Each site tested the challenge panel member via Direct Transfer and extended Direct Transfer procedure in accordance with product instructions. If DT and/or eDT result yielded a log(score) <2.00, the organism was retested using the Extraction procedure.

Table 4: Challenge Panel Study Summary

Test procedureSite ASite BSite CSite DTOTAL
≥2.0 ID≥2.0 ID≥2.0 ID≥2.0 ID≥2.0 ID
DT method46/55 (84%)47/55 (85%)35/55 (64%)*36/55 (65%)164/220 (75%)
eDT method47/55 (85%)50/55 (91%)44/55 (80%)*43/55 (78%)184/220 (84%)
Ext method54/55 (98%)53/55 (96%)49/55 (89%)37/55 (67%)193/220 (88%)
MBT-CA workflow54/55 (98%)55/55 (100%)49/55 (89%)46/55 (84%)204/220 (93%)
  • One sample was incorrectly identified due to a mixed culture.

93% of test organisms were correctly identified with a log(score) ≥2.00 result applying MBT-CA workflow. Testing of the challenge panel confirms intra laboratory performance of the MALDI Biotyper CA System.

Method Comparison:

To demonstrate performance of the MALDI Biotyper CA (MBT-CA) System, a method comparison study was performed at six (6) US clinical test sites and in-house laboratory. 4,395 (generating 4,399 data points) fresh and stored organisms were tested on the MALDI Biotyper CA System in accordance to manufacturer's instructions for use. All organisms included in the study were subcultured for purity. Testing on the MBT-CA System was done from a fresh isolated colony. Due to the rarity of some organisms, replicates of these rarer species were tested by multiple testing sites to generate additional data to support performance of the MBT-CA System. Results from the 3,802 replicate testing results were analyzed separately from the method comparison isolates.

{21}------------------------------------------------

All organisms included in the study were also sub-cultured on to an agar slant or appropriate media for isolation and shipped to the study interim reference laboratory. The interim reference laboratory stored all organisms included in the study and sent all organisms to the sequencing reference laboratory for sequencing and protein sequencing when requested.

The following Gram-negative, Gram-positive and yeast isolates are included in the reference library (please refer to K130831 for previously claimed Gram-negative organisms)

Table 5: Claimed Organisms

Organisms
Acinetobacter haemolyticus
Acinetobacter johnsonii
Acinetobacter junii
Actinomyces meyeri
Actinomyces neuii
Actinomyces odontolyticus
Actinomyces oris
Aerococcus urinae
Aerococcus viridans
Aeromonas salmonicida
Anaerococcus vaginalis
Bacteroides fragilis
Bacteroides ovatus group
Bacteroides thetaiotaomicron group
Bacteroides uniformis
Bacteroides vulgatus group
Bordetella group[3]
Bordetella hinzii
Brevibacterium casei
Brevundimonas diminuta group
Campylobacter coli
Campylobacter jejuni
Campylobacter ureolyticus
Candida albicans
Candida boidinii
Candida dubliniensis
Candida duobushaemulonii
Candida famata
Candida glabrata
Organisms
Candida guilliermondii
Candida haemulonis
Candida inconspicua
Candida kefyr
Candida krusei
Candida lambica
Candida lipolytica
Candida lusitaniae
Candida metapsilosis
Candida norvegensis
Candida orthopsilosis
Candida parapsilosis
Candida pararugosa
Candida pelliculosa
Candida tropicalis
Candida valida
Capnocytophaga ochracea
Capnocytophaga sputigena
Chryseobacterium gleum
Chryseobacterium indologenes
Clostridium difficile
Clostridium perfringens
Corynebacterium amycolatum
Corynebacterium aurimucosum group
Corynebacterium bovis
Corynebacterium diphtheriae
Corynebacterium glucuronolyticum
Corynebacterium jeikeium
Corynebacterium kronnenstedtii

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the logo for Bruker Corporation. The logo consists of the word "BRUKER" in bold, sans-serif font, stacked above a stylized graphic. The graphic features a series of interconnected, overlapping circles or ellipses, resembling an atom or a network. The color scheme is primarily black for the text and a light blue for the graphic elements.

Organisms

Organisms
Corynebacterium macginleyi
Corynebacterium minutissimum
Corynebacterium propinquum
Corynebacterium pseudodiphtheriticum
Corynebacterium riegelii
Corynebacterium striatum group
Corynebacterium tuberculostearicum
Corynebacterium ulcerans
Corynebacterium urealyticum
Corynebacterium xerosis
Cronobacter sakazakii group
Cryptococcus gattii
Cryptococcus neoformans var grubii
Cryptococcus neoformans var neoformans
Cupriavidus pauculus group
Delftia acidovorans group
Dermacoccus nishinomiyaensis
Edwardsiella tarda
Elizabethkingia meningoseptica group
Enterobacter amnigenus
Enterococcus avium group
Enterococcus casseliflavus
Enterococcus faecalis
Enterococcus faecium
Enterococcus gallinarum
Enterococcus hirae
Finegoldia magna
Fusobacterium canifelinum
Fusobacterium necrophorum
Fusobacterium nucleatum
Gardnerella vaginalis
Gemella haemolysans
Gemella sanguinis
Geotrichum candidum
Geotrichum capitatum
Granulicatella adiacens
Haemophilus haemolyticus
Haemophilus influenzae
Haemophilus parahaemolyticus group
Kingella kingae
Kloeckera apiculata
Kocuria kristingae
Organisms
Kytococcus sedentarius
Lactococcus garvieae
Lactococcus lactis
Leuconostoc mesenteroides
Macrococcus caseolyticus
Micrococcus luteus
Moraxella sg Moraxella nonliquefaciens
Myroides odoratimimus
Myroides odoratus
Oligella ureolytica
Oligella urethralis
Parabacteroides distasonis
Pediococcus pentosaceus
Peptoniphilus harei group
Peptostreptococcus anaerobius
Pichia ohmeri
Plesiomonas shigelloides
Porphyromonas gingivalis
Prevotella bivia
Prevotella buccae
Prevotella denticola
Prevotella intermedia
Prevotella melaninogenica
Propionibacterium acnes
Pseudomonas oryzihabitans
Pseudomonas stutzeri
Rhizobium radiobacter
Rothia aeria
Rothia dentocariosa
Rothia mucilaginosa
Saccharomyces cerevisiae
Serratia plymuthica
Serratia rubidaea
Staphylococcus aureus
Staphylococcus auricularis
Staphylococcus capitis
Staphylococcus caprae
Staphylococcus carnosus
Staphylococcus cohnii
Staphylococcus epidermidis
Staphylococcus equorum

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows the Bruker logo. The logo consists of the word "BRUKER" in bold, black letters. Above the word, there is a blue graphic that resembles an atom with electrons orbiting around it. The graphic is stylized and abstract, with the blue lines forming interconnected loops.

Organisms

Organisms
Staphylococcus haemolyticus
Staphylococcus hominis
Staphylococcus lugdunensis
Staphylococcus pasteuri
Staphylococcus pettenkoferi
Staphylococcus pseudintermedius
Staphylococcus saccharolyticus
Staphylococcus saprophyticus
Staphylococcus schleiferi
Staphylococcus simulans
Staphylococcus vitulinus
Staphylococcus warneri
Streptococcus agalactiae
Streptococcus anginosus
Streptococcus constellatus
Streptococcus dysgalactiae
Streptococcus gallolyticus
Organisms
Streptococcus gordonii
Streptococcus intermedius
Streptococcus lutetiensis
Streptococcus mitis / oralis group
Streptococcus mutans
Streptococcus pneumoniae
Streptococcus pyogenes
Streptococcus salivarius
Sutterella wadsworthensis
Trichosporon asahii
Vibrio parahaemolyticus
Vibrio vulnificus

Tables 6 - 8 below show the overall isolate performance.

Table 6: Overall Isolate Performance - claim 2

Overall Performance - claim 2
MBT-CA RESULTREFERENCE ALGORITHM
high resolutionspecieslow resolutionspecies / genusNegativeTotal
Organism ID ≥ 2.0(High Confidence)3817392184227
Organism ID (≥1.7; <2.0)(Low Confidence)107139129
- INCORRECT MBT-CA ID (≥1.7)- NO ID (<1.7)421n/a43
Total3966406274399
PositiveNegative
highresolutionhigh & lowresolution
96.27%99.02%n/a

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image shows the Bruker logo. The word "BRUKER" is written in bold, black letters. Above the word is a graphic of a stylized atom with two electrons orbiting the nucleus. The electrons are represented by blue dots.

Table 7: Overall Bacteria Performance

Overall Performance BACTERIA
MBT-CA RESULTREFERENCE ALGORITHM
high resolutionspecieslow resolutionspecies / genusNegativeTotal
Organism ID ≥ 2.0(High Confidence)3079389173485
Organism ID (≥1.7; <2.0)(Low Confidence)5212973
- INCORRECT MBT-CA ID (≥1.7)- NO ID (<1.7)251n/a26
Total3156402263584
PositiveNegative
highresolutionhigh & lowresolution
97.47%99.27%n/a

Table 8: Overall Yeast Performance

Overall Performance YEAST
MBT-CA RESULTREFERENCE ALGORITHM
high resolutionspecieslow resolutionspecies / genusNegativeTotal
Organism ID ≥ 2.0(High Confidence)73831742
Organism ID (≥1.7; <2.0)(Low Confidence)551056
- INCORRECT MBT-CA ID (≥1.7)- NO ID (<1.7)170017
Total81041815
PositiveNegative
highresolutionhigh & lowresolution0.00%
91.03%97.91%0.00%

{25}------------------------------------------------

Statement of Safety and Efficacy

The data presented clearly demonstrate the safety and efficacy of the Bruker Daltonics, Inc. MBT-CA System as compared to the reference algorithm, when the instructions for use are followed.

§ 866.3361 Mass spectrometer system for clinical use for the identification of microorganisms.

(a)
Identification. A mass spectrometer system for clinical use for the identification of microorganisms is a qualitative in vitro diagnostic device intended for the identification of microorganisms cultured from human specimens. The device is comprised of an ionization source, a mass analyzer, and a spectral database. The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software.
(2) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols.
(3) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(4) As part of the risk management activities performed as part of your 21 CFR 820.30 design controls, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(5) Premarket notification submissions must include details on the appropriate end user device training program that will be offered while marketing the device.