K Number
K142517
Device Name
SoftVue
Date Cleared
2014-10-31

(53 days)

Product Code
Regulation Number
892.1560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SoftVue™ is indicated for use as a B-mode ultrasonic imaging of a patient's breast when used with an automatic scanning curvilinear array transducer. The device is not intended to be used as a replacement for screening mammography.

Device Description

SoftVue™ is indicated for use as a B-mode ultrasonic imaging system for imaging of a patient's breast using a curvilinear array transducer that completely surrounds the breast. The modification of SoftVue™ will allow the system to generate colorized and grayscale relative stiffness ultrasound images in addition to the grayscale B-mode images generated by the unmodified SoftVue™ system. This new feature will allow the user to be able to determine whether a region of interest is harder or softer than the surrounding tissue. No clinical diagnostic claims are being made.

SoftVue™ is comprised of the following subsystems: the Transducer, Table/Housing Assembly, Water Conditioning System, Computer Control System, Image Reconstruction System, Power System, and the Data Acquisition System.

Soft Vue™ has a built-in curvilinear transducer that is used to acquire ultrasound data. Data are acquired from a patient lying prone on the table with their breast submerged in an imaging chamber filled with warm (body temperature) water. The breast is positioned in the center of the transducer. A camera, located at the bottom of the imaging chamber provides a live video feed to the system operator to aid in positioning the patient's breast. Once the scan is initiated, the transducer collects data that are processed to produce a series of B-mode ultrasound image slices that can be stacked to yield a volumetric ultrasound image of the breast. Soft Vue™ also outputs colorized and grayscale relative stiffness image stacks that provide the radiologist with additional reference information.

The Water Control System is used to de-gas and warm the water that is used as the image acquisition medium. The Computer Control System controls all of the functionality of the other subsystems. The reconstruction engine processes the image data acquired by the transducer ring into B-mode ultrasound images.

The system includes a barcode reader and a touchscreen display (user interface). The touchscreen display allows the user to perform an imaging procedure. Patient information is entered into the system using either the QWERTY keyboard on the touchscreen display, using the barcode reader, or the user can import the patient information from a DICOM modality worklist on an externally networked RIS server. Device errors and warnings are displayed on the touchscreen display.

SoftVue™ outputs the images to an externally networked PACS server which allows the images to be stored until they are reviewed on a workstation.

AI/ML Overview

The provided text does not contain detailed acceptance criteria for a specific device performance study beyond safety testing and general system requirements. It describes modifications to an existing device ("SoftVue") and a small-scale clinical study for "image characterization" of a new feature (color relative stiffness images). The study is not presented as a formal performance study demonstrating device efficacy against predefined acceptance criteria.

However, I can extract information related to the device's technical specifications and the qualitative observations from the clinical study, which could be interpreted as a form of "reported performance" against an implicit goal of providing "qualitative tissue stiffness information."

Here's an attempt to structure the information based on your request, highlighting what is implicitly or explicitly stated and noting where information is absent:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied/Explicit)Reported Device Performance
Safety:
Max. Mechanical Index (MI) ≤ 1.90.431 (Pass)
Max. ISPTA.3 ≤ 94 mW/cm²15.1 mW/cm² (Pass)
Max. Soft Tissue Thermal Index (TIS) (No specific threshold given, but implicit pass for safety)0.0006 (Implicitly Pass, as no issues raised and general safety standards met)
Conformity to IEC 60601-1:2005, IEC 60601-1-2:2007, IEC 62304:2006, IEC 60601-2-37:2007All safety standards were met.
System Functionality (General):
All software unit and system requirements metAll requirements were met and no new issues of safety or effectiveness were raised.
Subsystem and System design requirements metAll requirements were met and no new issues of safety or effectiveness were raised.
Human Factors/Usability requirements metAll requirements were met and no new issues of safety or effectiveness were raised.
Intended Use requirements metAll requirements were met and no new issues of safety or effectiveness were raised.
Image Characterization (Qualitative Tissue Stiffness Information):
Phantom Study: Correlation of color stiffness images with known phantom inclusion stiffness7 phantom inclusions: 1 cancer and 3 fibroadenomas (known stiff) appeared red; 3 cysts (known soft) appeared blue. This showed complete concordance with known properties.
Clinical Study: Qualitative assessment of lesion stiffness in correlation with pathology (where known/biopsied)13 imaged masses (4 cancers, 4 fibroadenomas, 5 cysts).- All 4 cancers were characterized as "stiff" (red/green, redder than background).- 2 fibroadenomas were "mixed," 1 was "stiff" (red), and 1 was "soft" (blue).- 4 cysts were "soft" (blue/green, bluer than background), while 1 was "mixed." This demonstrates SoftVue's ability to qualitatively measure tissue stiffness.
Image Quality: Improve image quality of B-mode images (implicit acceptance of improved image quality)"Updates to the pre-processing of the attenuation image result in better gain correction and equivalent contrast for the B-mode image.""Updates to the sound speed image parameters results in better delay correction and equivalent contrast for the B-mode image." (These are described as improvements, implying they met the goal of improved quality.)

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size:
    • Phantom Study: 1 anthropomorphic breast phantom with 7 inclusions.
    • Clinical Study (Image Characterization): 10 patient exams, yielding a total of 13 imaged masses.
      • Breast cup size ranged from B to DDD.
      • Breast densities: 5 scattered, 2 heterogeneous, 3 dense (one extremely dense).
  • Data Provenance:
    • Phantom Study: Not explicitly stated, but typically labs or manufacturers produce phantoms.
    • Clinical Study: Prospective. "Patient data were collected as part of the clinical study that Delphinus currently has in process under pre-IED # I120143." The modified SoftVue system was installed at the Alexander J Walt Breast Center at the Karmanos Cancer Institute in Detroit, Michigan, indicating U.S. origin.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • Phantom Study: The "ground truth" for the phantom was based on its known, predetermined characteristics (stiff/soft inclusions). No human experts were required to establish this ground truth, but the qualitative assessment against the known properties would have been done by the study team.
  • Clinical Study:
    • Number of Experts: One expert.
    • Qualifications: "board certified radiologist, Dr. Peter Littrup."
    • Role: Reviewed all patient images from the SoftVue system and guided the identification of lesions.

4. Adjudication Method for the Test Set

  • Phantom Study: Not applicable; the ground truth was inherent to the phantom's design.
  • Clinical Study: No formal adjudication method involving multiple experts is described for the clinical image analysis. The "ground truth" (lesion identification and likely pathology) was guided by "Previous biopsies and/or imaging." The evaluation of the SoftVue color stiffness images was performed by a single board-certified radiologist (Dr. Peter Littrup), matching the SoftVue appearance/assessment against the lesion pathology and characteristics derived from prior imaging/biopsy.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.
  • This study was for image characterization of a new feature, not explicitly a comparative effectiveness study of human readers with/without AI assistance. The device is an imaging system, and the new feature provides "additional reference information for radiologists," but its impact on reader performance or an effect size is not reported.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, in a qualitative sense for the new feature. The device now generates and outputs color and grayscale relative stiffness images. The phantom study, in particular, could be seen as a standalone evaluation of the algorithm's ability to represent stiffness, as it directly correlated the device's output (color) to the known physical properties of the phantom inclusions without human interpretation as a primary outcome. The clinical observations by the radiologist then confirmed that these images provided "qualitative tissue stiffness information."

7. The Type of Ground Truth Used

  • Phantom Study: Known, predetermined physical characteristics (stiffness, size) of inclusions within an anthropomorphic breast phantom.
  • Clinical Study: A combination of "previous biopsies and/or imaging" to identify and characterize lesions (appearance, size, location) and their pathology (cancer, fibroadenoma, cyst).

8. The Sample Size for the Training Set

  • The document describes modifications to an existing device ("SoftVue") and the evaluation of a new feature for that device (color and grayscale relative stiffness images). It does not mention the use of a "training set" in the context of machine learning or AI algorithm development for these new features. The focus is on verifying the system's performance and characterizing the new image type. Therefore, there is no explicit mention of a training set sample size. The "image characterization" study (both phantom and clinical) serves more as a validation/demonstration of the new feature's output.

9. How the Ground Truth for the Training Set Was Established

  • As no training set is mentioned for the development of the new feature related to stiffness imaging, this question is not applicable based on the provided text. The document refers to updates in pre-processing and sound speed image parameters to improve image quality, which are likely engineering/physics-based improvements rather than AI model training.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol consisting of three overlapping human profiles facing to the right, with flowing lines extending from the bottom of the profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 31, 2014

Delphinus Medical Technologies, Inc. % Ms. Andrea Wallen-Gerding Director of Quality & Regulatory Affairs 46701 Commerce Center Drive PLYMOUTH MI 48170

Re: K142517

Trade/Device Name: SoftVue Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: September 10, 2014 Received: September 11, 2014

Dear Ms. Wallen-Gerding:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Smh.7)

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K142517

Device Name SoftVue

Indications for Use (Describe)

SoftVue™ is indicated for use as a B-mode ultrasonic imaging of a patient's breast when used with an automatic scanning curvilinear array transducer. The device is not intended to be used as a replacement for screening mammography.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Diagnostic Ultrasound Indications for Use

System: SoftVue

Clinical ApplicationMode of Operation
General(Track 1 Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging &OtherFetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Breast)P
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under this appendix

Additional Comments: SoftVue is intended for ultrasonic breast examinations

{4}------------------------------------------------

1.0 Submitter Information

  • 1.1 This Premarket Notification is submitted by:
    Delphinus Medical Technologies, Inc. 46701 commerce Center Drive Plymouth, Michigan 48170

  • 1.2 Contact Information:
    Contact Name: Andrea N. Wallen-Gerding 734-233-3984 Office Telephone: Fax: 734-207-3165 E-mail: awallen@delphinusmt.com

  • 1.3 Date: September 5, 2014

2.0 Device Name

  • Trade/ Proprietary Name: SoftVue 2.1
  • 2.2 Common Name:
    • System, Imaging, Pulsed Echo Ultrasonic A
    • A Transducer, Ultrasonic, Diagnostic
  • Classification Name: 2.3
    • 21 CFR § 892.1560: Ultrasonic pulsed echo imaging system A
    • 21 CFR § 892-1570: Diagnostic ultrasonic transducer A

3.0 Predicate Device

The predicate device is identified as SoftVue™ manufactured by Delphinus Medical Technologies. SoftVue™ received market clearance under 510(k) number K123209.

The Toshiba Aplio 500 (K133761) was used as a secondary predicate device for elastography image comparison.

4.0 Device Description

SoftVue™ is indicated for use as a B-mode ultrasonic imaging system for imaging of a patient's breast using a curvilinear array transducer that completely surrounds the breast. The modification of SoftVue™ will allow the system to generate colorized and grayscale relative stiffness ultrasound images in addition to the grayscale B-mode images generated by the unmodified SoftVue™ system. This new feature will allow the user to be able to determine

{5}------------------------------------------------

whether a region of interest is harder or softer than the surrounding tissue. No clinical diagnostic claims are being made.

SoftVue™ is comprised of the following subsystems: the Transducer, Table/Housing Assembly, Water Conditioning System, Computer Control System, Image Reconstruction System, Power System, and the Data Acquisition System.

Soft Vue™ has a built-in curvilinear transducer that is used to acquire ultrasound data. Data are acquired from a patient lying prone on the table with their breast submerged in an imaging chamber filled with warm (body temperature) water. The breast is positioned in the center of the transducer. A camera, located at the bottom of the imaging chamber provides a live video feed to the system operator to aid in positioning the patient's breast. Once the scan is initiated, the transducer collects data that are processed to produce a series of B-mode ultrasound image slices that can be stacked to yield a volumetric ultrasound image of the breast. Soft Vue™ also outputs colorized and grayscale relative stiffness image stacks that provide the radiologist with additional reference information.

The Water Control System is used to de-gas and warm the water that is used as the image acquisition medium. The Computer Control System controls all of the functionality of the other subsystems. The reconstruction engine processes the image data acquired by the transducer ring into B-mode ultrasound images.

The system includes a barcode reader and a touchscreen display (user interface). The touchscreen display allows the user to perform an imaging procedure. Patient information is entered into the system using either the QWERTY keyboard on the touchscreen display, using the barcode reader, or the user can import the patient information from a DICOM modality worklist on an externally networked RIS server. Device errors and warnings are displayed on the touchscreen display.

SoftVue™ outputs the images to an externally networked PACS server which allows the images to be stored until they are reviewed on a workstation.

Intended Use 5.0

SoftVue™ is indicated for use as a B-mode ultrasonic imaging system for imaging of a patient's breast when used with an automatic scanning curvilinear array transducer. The device is not intended to be used as a replacement for screening mammography.

Predicate Device Comparison 6.0

Delphinus Medical Technologies claims that the SoftVue device is substantially equivalent to the SoftVue system cleared by the FDA in K123209. Delphinus Medical Technologies claims substantial equivalence because the proposed device has an equivalent intended use, manufacturing materials, operating principles, physical and operational specifications as compared to the predicate device.

{6}------------------------------------------------

The SoftVue device and the predicate device utilize B-mode grayscale ultrasound images to achieve their intended use. Both the modified SoftVue device and the unmodified predicate SoftVue device are table-top systems that have automatic scanning transducers to image breast tissue.

The specific details regarding similarities and differences between the modified SoftVue device and the unmodified SoftVue device have been identified and explained in the Comparison Table section provided in Section 10 of this submission. A brief summary of the similarities and differences between the modified SoftVue device and the unmodified SoftVue device is included below. The differences noted between the modified SoftVue device and the unmodified SoftVue device do not present any new issues related to safety and effectiveness.

Similarities:

  • The modified and unmodified SoftVue devices use an automated transducer to acquire A 2D images of a patient's breast.
  • A The modified and unmodified SoftVue devices use broadband transducers.
  • A The modified and unmodified SoftVue devices have a single operating frequency of 3 MHz.
  • A Both systems acquire and process B-mode grayscale images of a patient's breast.
  • Both systems position the patient in a prone position lying on their examination table A with the patient's in a pendulous position within an imaging chamber.
  • A Both systems position the patient's breast in a fluid environment to eliminate the need for breast compression and facilitate the transmission of ultrasound waves.

Differences:

  • A The differences between the modified and unmodified Soft Vue systems are listed in Table 1 below.
Item #ModificationReason for Change
1Added mesh to the spillage chamber.Prevent debris from entering/ getting trapped anddegrading valves.
2Added a second membrane moduleto degasserImprove degassing efficiency and maintain thesame flow rate of water.
3Replaced transmit power supplieswith medical grade power suppliesConform to 60601-1:2005
4Replaced in-line filter with medicalgrade versionConform to 60601-1:2005
5Added Ethernet opto-isolatorConform to 60601-1:2005
6Replaced vacuum pump circuitbreaker with lower amp circuitbreakerConform to 60601-1:2005
7Added thermistor interface boardBetter match reservoir and imaging chamber

Table 1: Changes made to Modified SoftVue System

{7}------------------------------------------------

Item #ModificationReason for Change
8Added cleaning toolstemperature sensors to more practical values
9Parallelized user login and system initializationAllows the user to view the status of the initialization process.
10Added ability to query external RIS serverAllows user to upload patient information for an exam.
11Updated GUIUpdated to accommodate other changes made such as allowing research users to select between running a research or clinical exam, RIS query, and initialization status.
12Added acquisition PCBA temperature monitoringSafety feature
13Transferred full control of motors to control computerStreamline system operation
14Added ability to generate and output color and grayscale relative stiffness imagesAdditional reference information for radiologists
15Improved image quality by updating pre-processing of attenuation image (not displayed) and updating sound speed image parameters.Improve image quality.• Updates to the pre-processing of the attenuation image result in better gain correction and equivalent contrast for the B-mode image.• Updates to the sound speed image parameters results in better delay correction and equivalent contrast for the B-mode image.
16Fixed a bug in firmware on the digital image processing boardImprove acquisition reliability
17Updated the Master Clock and PCIe Interface (MCP) Board to make clock signals more reliableImprove initialization reliability

7.0 Summary of Non-Clinical Testing

  • 7.1 The function and performance of the modified SoftVue device has been evaluated through non-clinical design verification and validation testing. Testing includes system performance and simulated use tests. The results of the evaluation tests demonstrate that the modified SoftVue device successfully meets the requirements of its intended use.
  • 7.2 Delphinus Medical Technologies conducted performance evaluations to verify that SoftVue's subsystems successfully meet predetermined specifications and product performance requirements. Results of the testing performed demonstrate that SoftVue's subsystems meet the system and performance requirements necessary for its intended use. A brief list of some of the testing performed is included below.
    • 7.2.1 Software Unit and System Verification and Validation Testing - Software was tested at the unit and system level to ensure that it met the software's design and

{8}------------------------------------------------

intended use requirements. All requirements were met and no new issues of safety or effectiveness were raised.

  • 7.2.2 Sub-system Verification Testing - SoftVue subsystems that were modified from the original SoftVue system that received clearance to market underwent subsystem verification to ensure that they met their respective design requirements. All requirements were met and no new issues of safety or effectiveness were raised.
  • 7.2.3 System Verification Testing - Since there were modifications made to the SoftVue system as a whole, system verification testing was conducted to ensure that the modified SoftVue system met the design requirements. All requirements were met and no new issues of safety or effectiveness were raised.
  • 7.2.4 Human Factors/Usability Testing: Usability testing was conducted per the FDA Guidelines and ANSI/AAMI HE75:2009. All requirements were met and no new issues of safety or effectiveness were raised.
  • 7.2.5 Design Validation Testing - Testing was conducted to ensure that the modified SoftVue device met the user needs and intended use requirements. All requirements were met and no new issues of safety or effectiveness were raised.
  • 7.3 Soft Vue has undergone acoustic output testing per IEC 60601-2-37:2007. Testing was conducted by Acertara Acoustic Laboratories, an independent testing laboratory, located in Longmont, Colorado. SoftVue meets all Track 1 acoustic output requirements. The results of acoustic output testing are listed in Table 2 below. A copy of the test report can be found in Appendix 09-E of Section 9 (Declaration of Conformity) of this submission.
FeatureTrack 1 Exposure LevelSoftVue LevelPass/Fail
Max. Mechanical Index(MI)1.90.431Pass
Max. ISPTA.394 mW/cm215.1 mW/cm2Pass
Max. Soft Tissue ThermalIndex (TIS)0.0006

Table 2: Global Maximum Acoustic Output Values

  • 7.4 In addition to the tests listed above, the modified SoftVue device has also undergone testing to the safety standards listed in Table 3 below.
Table 3: Safety Testing Performed on SoftVue
----------------------------------------------------------
Standard # and DateStandard Title
IEC 60601-1:2005; Corr.1:2006; Corr. 2:2007Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988;Amendment 1, 1991-11, Amendment 2, 1995
IEC 60601-1-2:2007Medical Electrical Equipment - Part 1-2: General Requirements for Safety -Collateral standard: Electromagnetic Compatibility - Requirements and Tests(Edition 2:2001 with Amendment 1:2004)
IEC 62304:2006Medical device software - Software life cycle processes

{9}------------------------------------------------

Standard # and DateStandard Title
IEC 60601-2-37:2007Medical electrical equipment - Part 2-37: Particular requirements for the basicsafety and essential performance of ultrasonic medical diagnostic andmonitoring equipment.

Summary of Clinical Testing 8.0

  • 8.1 No clinical testing is required to be performed using SoftVue; however, image characterization testing using clinical data has been completed per study parameters suggested by the FDA during the Submission Issue Q-Sub Meeting (Q130909) held on August 23, 2013. The parameters are listed below.
    • 10 - 20 patients

    • A Range of breast densities from Fatty to Dense Breasts
    • Small to Large Breasts A
    • Lesion sizes 5 mm to 15 mm or 20 mm A
  • 8.2 The study was conducted in order to demonstrate SoftVue's ability to provide qualitative tissue stiffness information with the color relative stiffness images. The invivo comparisons are not meant to guide a statistical study, but are solely intended to provide an illustrative example of in-vivo stiffness imaging using the modified SoftVue System.
  • 8.3 As part of the evaluation, a phantom analysis was also included in the test protocol to establish a baseline for the clinical image evaluations. An anthropomorphic breast phantom was chosen for this initial analysis because its characteristics are predetermined such that the stiffness properties of the entire phantom and its inclusions are accurately known. The phantom contains stiff and soft inclusions of varying sizes. Each of the masses in the phantom have known stiffness characteristics that have been used to demonstrate that the color of the masses within the color stiffness images output by Soft Vue correlate to the stiffness of the masses within the anthropomorphic breast phantom. The appearance of each mass was used to qualitatively characterize the mass stiffness based on relative color differences.
    • 8.3.1 The phantom scan yielded images of 7 inclusions. Among the 7 phantom inclusions, 1 cancer and 3 fibroadenomas were found to be stiff (appearing red) compared to the background material while the 3 cysts were found to be soft (appearing blue) in complete concordance with the known properties of the phantom.
  • 8.4 Delphinus Medical Technologies conducted a small scale clinical study to collect clinical images from the modified Soft Vue system. The images collected include Bmode images as well as color and grayscale relative stiffness images. Patient data were collected as part of the clinical study that Delphinus currently has in process under pre-IED # I120143. A modified SoftVue system has been installed at the Alexander J Walt Breast Center at the Karmanos Cancer Institute located in Detroit, Michigan.
  • 8.5 The image evaluations were conducted per protocol 100-00745, SoftVue Color Stiffness Image Characterization Protocol (Appendix 20-A).

{10}------------------------------------------------

  • 8.6 The patient data were chosen from the clinical study on the basis of having a suspicion or a biopsy proven lesion such that the lesions in question could be identified by:
    • 8.6.1 Appearance (echogenicity)
    • 8.6.2 Size (dimensions)
    • Location (clock position) 8.6.3
  • All patient images from the SoftVue system were reviewed by board certified 8.7 radiologist, Dr. Peter Littrup. Previous biopsies and/or imaging were used to guide the identification of the lesions in the SoftVue images using clock position, appearance, size and location for guidance.
  • 8.8 A set of 10 representative in-vivo comparisons of SoftVue's B-mode and color relative stiffness images was carried out to demonstrate SoftVue's ability to qualitatively measure tissue stiffness with clinical data, using the B-mode images for initial mass identification.
    • A total of 10 SoftVue patient exams were used in the analysis. Some patients 8.8.1 had multiple lesions so that the final data set consisted of 13 imaged masses consisting of 4 cancers, 4 fibroadenomas, and 5 cysts.
    • Breast cup size ranged from B to DDD with densities consisting of 5 scattered, 2 8.8.2 heterogeneous and 3 dense (one of which was extremely dense) indicating that a range of breast size and density was sampled.
  • 8.9 All 4 cancers were characterized as "stiff" by SoftVue's color stiffness images. Two fibroadenomas were found to be mixed (range of colors), 1 was stiff (red) and 1 was found to be soft (blue). Of the 5 cysts, 4 were found to be soft, while 1 was found to be mixed. A summary of the results are included in Table 4 below.
Mass#Case#Study #Breast SizeBreastDensityLesionPathologyReportedLesionPositionAverageLesionSize (cm)Soft VueAppearanceSoftVueStiffnessAssessment
11SV021CDenseCancer8:002.5HypoechoicRed / green (redderthan background onaverage)Stiff
22SV022DDScatteredFibroadenoma1:303.0HypoechoicYellow / green(redder thanbackground onaverage)Stiff
33SV045CScatteredCancer10:001.8HypoechoicRed (redder thanbackground onaverage)Stiff
44SV077_1BExtremelyDenseCyst2:001.4AnechoicBlue/green (bluerthan background onaverage)Soft
54SV077_2BExtremelyDenseCyst12:001.9AnechoicBlue/green (bluerthan background onaverage)Soft

Table 4: Clinical Image Analysis Summary

{11}------------------------------------------------

Mass#Case#Study #Breast SizeBreastDensityLesionPathologyReportedLesionPositionAverageLesionSize (cm)SoftVueAppearanceSoftVueStiffnessAssessment
64SV077_3BExtremelyDenseCyst12:002.0AnechoicBlue/green (bluerthan background onaverage)Soft
75SV079CHeterogeneousCyst9:001.4AnechoicBlue/green(bluer thanbackground onaverage)Soft
86SV089DScatteredCancer9:301.7HypoechoicRed (redder thanbackground onaverage)Stiff
97SV090DDD26yrs(Hetero)*Fibroadenoma10:004.3HypoechoicBlue/Green (rangeof color, nodominant color)Mixed
108SV113DDScatteredCyst2:001.8AnechoicBlue/green (rangeof color, nodominant color)Mixed
119SV114DScatteredCancer10:00 RA1.2HypoechoicRed/green (redderthan background onaverage)Stiff
1210SV117_1D21 yrs(Dense)*Fibroadenoma5:001.9HypoechoicBlue/green/red(range of color, nodominant color)Mixed
1310SV117_2D21 yrs(Dense)*Fibroadenoma8:002.6HypoechoicBlue (range ofcolor, no dominantcolor)Soft

9.0 Conclusion

The modified SoftVue device performs as intended and is substantially equivalent to the unmodified SoftVue device with respect to intended use, design, principles of operation, technology, materials, and performance. Any noted differences do not raise any new issues of safety and effectiveness.

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.