K Number
K142500
Device Name
One Way Soft Valve
Date Cleared
2015-01-29

(146 days)

Product Code
Regulation Number
870.4400
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The One Way Soft Valve is intended to be used for the prevention of retrograde flow in ¼/' or smaller lines subject to arterial pressure during cardiopulmonary bypass procedures for use up to six hours.
Device Description
The One Way Soft Valve is a transparent one-way valve with ¼" inlet and outlet ports. The One Way Soft Valve prevents retrograde flow. A directional arrow is printed on the valve to provide the user with a visual indication of the allowable flow direction within the valve is constructed entirely of PVC and is to be used in lines subject to arterial pressures.
More Information

No
The device description and performance studies focus on the mechanical function of a one-way valve and do not mention any AI or ML components or capabilities.

Yes
The device is described as preventing retrograde flow in lines during cardiopulmonary bypass procedures, which is a therapeutic intervention. It is also compared to a predicate device, the Quest Medical Retroguard Arterial Safety Valve, which is associated with medical procedures.

No
The device is described as a one-way valve for preventing retrograde flow during cardiopulmonary bypass procedures, which is a therapeutic function, not a diagnostic one. Its purpose is to control fluid movement, not to detect, diagnose, or monitor a medical condition.

No

The device description explicitly states it is a physical valve constructed entirely of PVC, intended for use in tubing lines. It is a hardware device, not software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states the device is for preventing retrograde flow in lines during cardiopulmonary bypass procedures. This is a direct intervention on the patient's circulatory system during a surgical procedure.
  • Device Description: The description details a physical valve designed to control fluid flow within a medical circuit connected to a patient.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens (blood, tissue, etc.) in vitro (outside the body) to provide information about a patient's health status, diagnose conditions, or monitor treatment. The performance studies are focused on the mechanical and functional performance of the valve itself within a simulated or actual bypass circuit, not on analyzing biological samples.

IVD devices are used to perform tests on samples taken from the body to provide diagnostic or other health-related information. This device is a component used during a medical procedure on the patient.

N/A

Intended Use / Indications for Use

The One Way Soft Valve is intended to be used for the prevention of retrograde flow in ¼/' or smaller lines subject to arterial pressure during cardiopulmonary bypass procedures for use up to six hours.

Product codes

MJJ

Device Description

The One Way Soft Valve is a transparent one-way valve with ¼" inlet and outlet ports. The One Way Soft Valve prevents retrograde flow. A directional arrow is printed on the valve to provide the user with a visual indication of the allowable flow direction within the valve is constructed entirely of PVC and is to be used in lines subject to arterial pressures.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Terumo Cardiovascular Systems has conducted in-vitro performance evaluations for the purpose of demonstrating that the One Way Soft Valve device is substantially equivalent to the predicate Quest Medical Retroguard Arterial Safety Valve.
Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the subject device to the predicate devices. Substantial equivalence is demonstrated with the following in-vitro performance evaluations:

  • One Way Soft Valve Tubing Connection (Bond) Strength Evaluation
  • One Way Soft Valve Mechanical/Structural Integrity/Burst Strength Evaluation
  • One Way Soft Valve Opening Pressure Evaluation
  • One Way Soft Valve Negative Pressures and Retrograde Flow Evaluation
  • One Way Soft Valve Circulation and Hemolysis Evaluation

The Terumo Cardiovascular Systems Corporation One Way Soft Valve exhibits substantially equivalent performance to the predicate, the Quest Medical Retroguard Arterial Safety Valve-K922356. Performance testing was conducted with substantially equivalent results and is presented in this submission.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K922356

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.4400 Cardiopulmonary bypass blood reservoir.

(a)
Identification. A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.(b)
Classification. Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

0

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing, ribbon-like element beneath them. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 29, 2015

Terumo Cardiovascular Systems Corporation Garry A. Courtney Director, Regulatory Affairs 28 Howe Street Ashland, Massachusetts 01721

Re: K142500

Trade/Device Name: One Way Soft Valve Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II Product Code: MJJ Dated: January 2, 2015 Received: January 5, 2015

Dear Garry A. Courtney,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

1

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.A. Hillebrenner

for

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Image /page/2/Picture/0 description: The image shows the Terumo Medical Corporation logo. The logo consists of a green circle with a stylized "T" inside, followed by the word "TERUMO" in green, sans-serif font. A registered trademark symbol appears to the right of the word "TERUMO".

SECTION 4 – Indications for Use

One Way Soft Valve

510(k) Number (if known): K 142500

Device Name: One Way Soft Valve

Indications For Use:

The One Way Soft Valve is intended to be used for the prevention of retrograde flow in ¼/' or smaller lines subject to arterial pressure during cardiopulmonary bypass procedures for use up to six hours.

| Prescription Use XX
(Part 21 CFR 801 Subpart D) | OR | Over-The-Counter Use ____________________
(Part 21 CFR 807 Subpart C) |
|----------------------------------------------------------------------------------------------------|----|--------------------------------------------------------------------------|
| (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF
NEEDED) | | |

Concurrence of CDRH, Office of Device Evaluation (ODE)

3

Image /page/3/Picture/0 description: The image shows the logo for Terumo. The logo consists of a circle with the letter T inside of it, followed by the word "TERUMO" in all capital letters. The logo is green.

SECTION 5 – 510(k) Summary One Way Soft Valve

Submitter Information5-2
Device Names/Classification5-2
Identification of Predicate Device5-2
Device Description5-2
Comparison of Indication5-2
Principles of Operation, Technology, and Design5-3
Performance Evaluations5-4
Substantial Equivalence Statement5-4
Additional Safety Information5-4
Conclusion for 510(k) Summary5-4

4

Submitter Information:

This submission was prepared in August 2014 by: Suzanne Grenier, RAC. COA Regulatory Affairs Senior Specialist Terumo Cardiovascular Systems Corporation 125 Blue Ball Road Elkton, MD 21921 Telephone: 443-745-8443

This submission was prepared for:

Terumo Cardiovascular Systems Corporation 28 Howe Street Ashland, MA 01721 Facility Registration No. 1212122

Device Names/Classifications:

Proprietary NameClassification NameCommon Name
One Way Soft ValveCardiopulmonary BypassCPB Check Valve,
Blood Reservoir (ProCode: MJJ)Retrograde Flow, In-Line

Identification of Predicate Device:

The device submitted in this 510(k) maintains characteristics that are substantially equivalent in indications for use, design, technology including features, materials, and principles of operation to the Quest Medical Retroguard Arterial Safety Valve - K922356.

Device Description:

The One Way Soft Valve is a transparent one-way valve with ¼" inlet and outlet ports. The One Way Soft Valve prevents retrograde flow. A directional arrow is printed on the valve to provide the user with a visual indication of the allowable flow direction within the valve is constructed entirely of PVC and is to be used in lines subject to arterial pressures.

Comparison of Indication

The One Way Soft Valve product and the predicate, Quest Medical Retroguard Arterial Safety Valve product, are intended to be used to prevent retrograde flow during cardiopulmonary bypass procedures for up to six hours.

Subject Device Indication: The One Way Soft Valve is intended to be used for the prevention of retrograde flow in ¼" or smaller lines subject to arterial pressure during cardiopulmonary bypass procedures for use up to six hours.

Predicate Device Indication: Per the IFU, the Quest Medical Retroguard Arterial Safety Valve is "for prevention of retrograde flow when used with centrifugal pump."

The differences between the intended use for the subject device and the predicate device do not constitute a new intended use. The wording differences are details of device specifications (e.g. dimensions), specified circuit placement (e.g. lines subject to arterial pressures, line size), and

5

Image /page/5/Picture/0 description: The image shows the word "TERUMO" in green font. To the left of the word is a green circle with a "T" inside. The font is sans-serif and the word is in all capital letters.

circuit component choices (e.g. pump choice), which do not significantly impact the overall intended use of the valves in preventing retrograde flow during cardiopulmonary bypass procedures. These differences are not critical to the overall intended use of the device preventing retrograde flow during cardiopulmonary bypass procedures. Additionally, the differences do not affect the safety and effectiveness of the device when used as labeled. Performance testing was conducted with substantially equivalent results and is presented in this submission.

Principles of Operation, Technology, and Design:

The One Way Soft Valve is a single-use, disposable device that prevents retrograde flow in lines subject to arterial pressure by utilizing a closed path, mechanical, one-way valve. The design allows flow through the valve in one-direction and limits the flow in the opposing direction. The operating principles of the Ouest Medical Retroguard Arterial Safety Valve are identical to that of the device subject to this submission, the One Way Soft Valve.

The primary differences in design between the One Way Soft Valve and the predicate device are found in the material selection, the type of check-valve used, port dimensions, and overall prime volume.

  • The first of these differences is the material choice. The One Way Soft Valve is o constructed of a flexible PVC housing and a flexible PVC valve, while the predicate valve is constructed of a rigid ABS housing and flexible silicone valve. Flexible Polyvinylchloride (PVC) is a common material used in Cardiopulmonary Bypass Circuits. Performance testing results are presented within this submission. The subject devices were tested to clinically relevant pressures to ensure that they could withstand both positive and negative pressures and still maintain desired performance.
  • The second difference is the type of check-valve used. The One Way Soft Valve uses a 0 sleeve-like valve, while the predicate uses a duckbill valve design. Both designs allow flow through the valve in one-direction and limit flow in the opposing direction.
  • The third difference is the adapter (or port) dimensions. The One Way Soft Valve has a o smaller port size (1/4" compared to 3/8"), which allows the One Way Soft Valve to be used in smaller diameter lines.
  • Lastly, the One Way Soft Valve has a smaller prime volume than the predicate device. 0 The One Way Soft Valve has a prime volume of approximately 9 mL. The predicate device has a prime volume of approximately 20 mL. Submission of this 510(k) was primarily motivated by the clinical need for a smaller, more versatile valve for use in Cardiovascular Procedure Convenience Kits.

Testing was completed that demonstrates the One Way Soft Valve is as safe and effective as the predicate. Any differences in design between the One Way Soft Valve and the predicate do not raise different questions of safety and effectiveness.

6

Image /page/6/Picture/0 description: The image shows the logo for Terumo. The logo consists of a green circle with a white "T" inside of it. To the right of the circle is the word "TERUMO" in green, block letters.

Performance Evaluations:

Terumo Cardiovascular Systems has conducted in-vitro performance evaluations for the purpose of demonstrating that the One Way Soft Valve device is substantially equivalent to the predicate Quest Medical Retroguard Arterial Safety Valve.

Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the subject device to the predicate devices. Substantial equivalence is demonstrated with the following in-vitro performance evaluations:

  • One Way Soft Valve Tubing Connection (Bond) Strength Evaluation 0
  • 0 One Way Soft Valve Mechanical/Structural Integrity/Burst Strength Evaluation
  • o One Way Soft Valve Opening Pressure Evaluation
  • 0 One Way Soft Valve Negative Pressures and Retrograde Flow Evaluation
  • o One Way Soft Valve Circulation and Hemolysis Evaluation

The Terumo Cardiovascular Systems Corporation One Way Soft Valve exhibits substantially equivalent performance to the predicate, the Quest Medical Retroguard Arterial Safety Valve-K922356. Performance testing was conducted with substantially equivalent results and is presented in this submission.

Substantial Equivalence Statement:

The One Way Soft Valve device and the predicate device. the Quest Medical Retroguard Arterial Safety Valve - K922356, are substantially equivalent in intended use, principles of operation and technology, design and materials, and performance. Any noted differences between the subject device and the predicate devices do not raise new issues of safety and effectiveness.

Additional Safetv Information:

  • Sterilization conditions have been validated to provide a Sterility Assurance Level (SAL) of . 106. Terumo further asserts that the ethylene oxide residues will not exceed the maximum residue limits at the time of product distribution.
  • Terumo maintains biocompatibility studies as recommended in the FDA General Program o Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing." [External Communicating Devices, Circulating Blood, Limited Exposure (≤ 24 hours) Contact Duration]. The blood contacting materials were found to be biocompatible.

Conclusion:

In summary, Terumo deems the Terumo Cardiovascular Systems Corporation One Way Soft Valve device is substantially equivalent to the predicate device, the Quest Medical Retroguard Arterial Safety Valve - K922356. with respect to intended use, duration of use, design, materials. principles of operation, and performance. It is further noted that any recognized differences do not raise any new issues of patient/user safety or product effectiveness.