K Number
K142428
Manufacturer
Date Cleared
2014-11-13

(76 days)

Product Code
Regulation Number
876.4480
Panel
GU
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ShockPulse-SE Lithotripsy System is intended to be used for the fragmentation of urinary tract calcul in the kidney, ureter and bladder.

Device Description

The ShockPulse-SE Lithotripsy System (CYBERWAND II) is the next generation of the CyberWand System. The ShockPulse-SE is an electromechanical device, whose intended purpose is to fragment and aspirate calculi. The hand piece consists of an ultrasonic transducer containing piezoelectric elements which are driven by a generator operating at approximately 21,000 Hz. The transducer is lightweight and incorporates two contact switches that activate Standard Power and High Power generator output. Activation will either require continuous pushing of the button or a double click to latch activation on; a single click of either button would then turn activation off. There are markings near the buttons to indicate function. Suction control is integrated into the transducer housing. With a thumb wheel that can rotate approximately 20 degrees, the flow can be varied from "full" on to "off." There are markings on the transducer to indicate which direction increases and decreases suction flow through the transducer lumen. There are families of probes that function similar to the probes of the current CyberWand System (K120303) by transmitting stress-waves from the transducer to the calculi to break up kidney-stones. Hollow probes permit simultaneous suction. The direct patient-contact material is stainless steel, which is the same material used in the predicate system.

AI/ML Overview

The provided document is a 510(k) Summary for the ShockPulse-SE Lithotripsy System (CYBERWAND II). It details the device's technical specifications, indications for use, and a comparison to its predicate device, the CyberWand Dual Action Ultrasonic Lithotripsy System (K120303). The document primarily focuses on demonstrating substantial equivalence through performance data, safety considerations, and compliance with various standards.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria / Performance GoalReported Device Performance
Stone Breakage Performance
CyberWand II 3.76 probe drills through a stone at least 10% faster than CyberWand I 3.76 Dual Probe.The CyberWand II 3.76 probe drills through a stone at least 10% faster than the CyberWand I 3.76 Dual Probe. (Met)
CyberWand II 3.76 probe removes mass at least 20% faster than CyberWand I 3.76 Dual Probe.The CyberWand II 3.76 probe removes mass at least 20% faster than the CyberWand I 3.76 Dual Probe. (Met)
CyberWand II 1.65 probe has less impact force than LithoClast pneumatic ureteral probe.The CyberWand II 1.65 probe has less impact force than the LithoClast pneumatic ureteral probe. (Met)
Smallest probe size (0.97 mm) equivalent stone breakage.Supported by stone breakage measurements and identification of a reference device (K111058, Med-Sonics Corp. UreTron Multi Probe Lithotripter) with an 0.8 mm ultrasonic probe. (Met by comparison to a cleared device)
Transducer and Torque Wrench Life Cycle (50 Cycles)
Performance (Mass Removal, Drill Rate, Probe Tip Displacement) does not degrade.Evaluated, implied to be met as it states performance does not degrade over 50 simulated use cycles.
Functionality (Suction Control, Button Activation Switches) remains.Evaluated, implied to be met as it states every aspect of the transducer is functional after 50 simulated use cycles.
Safety (Electrical Safety - Patient Leakage Current) remains.Evaluated, implied to be met as it states every aspect of the transducer is functional after 50 simulated use cycles.
Appearance Quality (Black Anodize, Laser Etching) does not degrade.Evaluated, implied to be met as it states every aspect of the transducer is functional after 50 simulated use cycles.
Probe Reliability and Reusability
Probes are functional after a single simulated use (5x typical use time).Probes are functional after a single simulated use, capable of 5x typical use time. (Met - Note: Probes are to be labeled single use)
IEC 61847 Evaluation
Output characteristics of CyberWand II system compared to CyberWand I 3.76 mm dual probe.Demonstrated.
Less probe tip displacement than predicate (safe and substantially equivalent).Less probe tip displacement (101 microns peak-to-peak) compared to the predicate CyberWand I (102 microns peak-to-peak). (Met)
Tissue Perforation Testing (Safety)
CyberWand consistently induces less kidney tissue damage than LithoClast Ultra.The CyberWand consistently induced less kidney tissue damage than the LithoClast Ultra using a visual assessment methodology (on predicate device). ShockPulse SE system designed and tested to have equal or less maximum tip amplitude than CyberWand I, and tested in porcine model with no significant tissue damage. (Met by lineage and animal study)
Smallest probe (0.97 mm) equivalent tissue perforation potential.Supported by identification of a reference device (K111058, Med-Sonics Corp. UreTron Multi Probe Lithotripter) with an 0.8 mm ultrasonic probe. (Met by comparison to a cleared device)
Torque Wrench Performance
Consistent torque specification of 35-40 inch pounds.Demonstrated a consistent torque specification of 35-40 inch pounds. (Met)
Generator and Footswitch Life Cycle
Generator useful life of 7 years.Demonstrated a useful life of 7 years for the generator. (Met)
Footswitch durability and IPX6 rating.Demonstrated footswitch durability and IPX6 rating. (Met)
Shipping and Handling Durability
System functionality after rigorous simulated shipping and handling.Demonstrated system functionality after being run through rigorous simulated shipping and handling. (Met)
Noise Evaluation
Less noise output compared to CyberWand I.Demonstrated less noise output of the CyberWand II compared to the CyberWand I. (Met)
Surface Temperature Profile
Surface temperature stabilizes at an acceptable level (IEC 60601-1) with aspiration.Demonstrated that the surface temperature of the transducer stabilizes at an acceptable level while running with aspiration. (Met)
Surface temperature stabilizes at a lower temperature than CyberWand I (with and without aspiration).Demonstrated that the surface temperature of the CyberWand II transducer stabilizes at a lower temperature than the CyberWand I while running with and without aspiration. (Met)
Setup Steps
Fewer setup steps for CyberWand II compared to CyberWand I (37.5% fewer).Demonstrated the number of setup steps is less for the CyberWand II as compared to the CyberWand I (37.5% fewer steps). (Met)
Usability Verification and Validation
Ability to fragment simulated urinary tract calculi (3.76, 1.65, 0.97 probes).Demonstrated the ability for the CyberWand II 3.76, 1.65, and 0.97 to fragment simulated urinary tract calculi. (Met)
Usability of IFU and labeling.Demonstrated the usability of the CyberWand II IFU and labeling. (Met)
Meets customer needs.Demonstrated the CyberWand II meets customer needs. (Met)
Electrical Safety (IEC 60601-1:2005 3rd Ed.)Complies with IEC 60601-1:2005 3rd Edition. All testing conducted by Intertek. (Met)
EMC (IEC 60601-1-2 ed3.0 and CISPR 11:2009+A1:2010)Complies with IEC 60601-1-2 ed3.0 and CISPR 11:2009+A1:2010. All testing conducted by Intertek. (Met)
Risk Management (ISO 14971)Compliance with ISO 14971 established by inspection (desktop audit) of the Risk Management File. (Met)
Software Validation (FDA Moderate Level of Concern)Performed according to FDA's Moderate Level of Concern recommendations. (Met)
Sterilization Validation
SPL-PD376 probe (EO sterilization)Packaging and sterilization validations apply to SPL-PD376 probes, confirmed by MOOG Medical Device Group. (Met by comparison to predicate data K120303)
5-year shelf life sterile packaging (SPL-PD376 probe)Assured with accelerated and real-time aging and subsequent testing; supported by data in K132795. (Met)
Cleaning efficacy for non-sterile componentsValidated manual cleaning method demonstrated. (Met)
Steam efficacy (US steam sterilization cycle)Demonstrated efficacy at the US steam sterilization cycle (4 minutes at 132 degrees C, 20 minute dry time) for various components. (Met)
Dry time validation (20 minute dry time for US steam sterilization)Demonstrated adequate 20 minute dry time for a US steam sterilization cycle for various components. (Met)

The document describes several performance tests and validations to demonstrate the ShockPulse-SE Lithotripsy System (CYBERWAND II)'s safety and effectiveness and its substantial equivalence to the predicate device (CyberWand Dual Action Ultrasonic Lithotripsy System, K120303).

Summary of Studies and Information:

2. Sample Size Used for the Test Set and Data Provenance:

  • CyberWand II Performance Analysis for Stone Breakage:
    • No specific sample size (number of stones, trials) is mentioned for the quantitative stone drilling and mass removal tests. The comparison states "at least 10% faster" and "at least 20% faster," implying a statistically significant difference was observed, but the underlying data points are not provided.
    • Data Provenance: Not specified, but generally in vitro (bench-top) or ex vivo (simulated use) given the nature of the tests (drilling into stones, mass removal).
  • Retropulsion Study:
    • No specific sample size is mentioned. It compares the CyberWand II 1.65 probe to a "competitively marketed lithotripter" (LithoClast pneumatic).
    • Data Provenance: Implied in vitro or bench-top measurement, as it quantifies "impact force."
  • Tissue Perforation Testing (Dr. Evans' Pig Kidney Test):
    • No specific sample size (number of kidneys or trials) is explicitly stated for the "Pig Kidney Test." It mentions "The CyberWand consistently induced less kidney tissue damage..." The ShockPulse SE system was "tested in simulated use conditions by several physicians in the porcine model."
    • Data Provenance: Ex vivo (freshly harvested porcine kidney) for the initial comparison with the predicate device. The subsequent test for ShockPulse SE was ex vivo in a porcine model.
  • Life Cycle Analyses (Transducer, Torque Wrench, Generator, Footswitch):
    • Sample size: "50 simulated use cycles" for transducer and torque wrench. No specific number of units tested is provided for generator (7 years) or footswitch (durability).
    • Data Provenance: Bench-top, simulated use testing.
  • Probe Reliability and Reusability Analysis:
    • Sample size: Refers to "functional after a single simulated use" for "5x typical use time," but the number of actual probes tested is not stated.
    • Data Provenance: Bench-top, simulated use testing.
  • IEC 61847 Evaluation:
    • No specific sample size (number of probes or trials) for output characteristics or probe tip displacement.
    • Data Provenance: Bench-top measurements.
  • Usability Verification and Validation:
    • Sample size: Not specified for stone fragmentation (simulated calculi), IFU usability, or customer needs. "Several physicians" were involved in the porcine model for tissue perforation, which might feed into "customer needs" indirectly.
    • Data Provenance: Bench-top for stone fragmentation; likely qualitative assessment for IFU and customer needs validation, possibly involving real users/physicians.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

  • Tissue Perforation Testing: "Dr. Evans' Pig Kidney Test" and "several physicians" were involved in the porcine model for the ShockPulse SE.
    • Qualifications: "Dr. Evans" implies a qualified expert, and "physicians" indicates medical professionals. Specific specializations (e.g., urologists) and years of experience are not provided.
  • For other performance tests like stone breakage, retropulsion, life cycle, EMC, electrical safety, usability, the "ground truth" is typically objective measurement against engineering specifications and validated test methods rather than expert consensus on a subjective finding.

4. Adjudication Method for the Test Set:

  • Given that most tests are objective engineering or bench-top measurements (e.g., drill rate, mass removal, impact force, displacement, temperature, electrical safety, noise levels, torque, cycle counts), adjudication methods like 2+1 or 3+1 (common for subjective image interpretation) are not applicable.
  • For the "Dr. Evans' Pig Kidney Test," a "visual assessment methodology" was used to compare tissue damage. The document doesn't specify if multiple observers were used or an adjudication process.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, a "multi-reader multi-case (MRMC) comparative effectiveness study" focusing on human readers improving with AI vs. without AI assistance was not conducted. This is an electrohydraulic lithotripter for physical stone fragmentation, not an AI-assisted diagnostic or image interpretation device. The studies focus on device performance, safety, and functionality.

6. Standalone (Algorithm Only Without Human-in-the Loop Performance) Study:

  • This question is not applicable as the device is a medical instrument (lithotripter), not an algorithm or AI software for diagnosis or image analysis. The device's "standalone" performance is evaluated through the bench-top and simulated use tests described (e.g., stone breakage, life cycle, safety, EMC).

7. Type of Ground Truth Used:

  • Objective Measurements: For most tests (stone breakage, retropulsion, displacement, life cycle, noise, temperature, torque, electrical safety, EMC), the ground truth is established by objective, quantitative measurements against predefined engineering specifications, national/international standards (e.g., IEC 60601-1, CISPR 11), and internal performance targets derived from the predicate device.
  • Comparisons to Predicate/Reference Devices: Performance (e.g., stone breakage, displacement, tissue damage, noise) is often compared directly to the predicate device (CyberWand I) or other legally marketed devices (LithoClast, Med-Sonics Corp. UreTron Multi Probe Lithotripter K111058), which serve as a benchmark for substantial equivalence.
  • Expert Visual Assessment: For tissue perforation (Dr. Evans' Pig Kidney Test), ground truth was based on a "visual assessment methodology" of kidney tissue damage.
  • Compliance with Standards: For regulatory aspects (electrical safety, EMC, risk management, software validation, sterilization validation), ground truth is compliance with specified regulatory standards and guidance documents.

8. Sample Size for the Training Set:

  • This question is not applicable. The ShockPulse-SE Lithotripsy System is a mechanical/electronic medical device, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The device's design is based on engineering principles and iterative development, not data-driven machine learning.

9. How the Ground Truth for the Training Set Was Established:

  • This question is not applicable as there is no "training set" for this type of device. The design and performance validation rely on engineering specifications, physical testing, and comparison with predicate devices and established standards.

§ 876.4480 Electrohydraulic lithotriptor.

(a)
Identification. An electrohydraulic lithotriptor is an AC-powered device used to fragment urinary bladder stones. It consists of a high voltage source connected by a cable to a bipolar electrode that is introduced into the urinary bladder through a cystoscope. The electrode is held against the stone in a water-filled bladder and repeated electrical discharges between the two poles of the electrode cause electrohydraulic shock waves which disintegrate the stone.(b)
Classification. Class II. The special control for this device is FDA's “Guidance for the Content of Premarket Notifications for Intracorporeal Lithotripters.”