K Number
K142351
Date Cleared
2015-08-25

(368 days)

Product Code
Regulation Number
862.1825
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 25-Hydroxy Vitamin DS EIA assay is intended for the quantitative determination of 25-hydroxyvitamin D [25(OH)D] and other hydroxylated metabolites in human serum or plasma. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in an adult population.

Device Description

The 25-Hydroxy Vitamin De EIA is a manual assay test and does not require the use of an automated system. The whole assay is performed in a microtitre plate and requires a user to perform each step. The 25-Hydroxy Vitamin De assay is an enzymeimmunoassay for the quantitation of 25(OH)D and other hydroxylated metabolites in serum or plasma. 25uL of each calibrators, controls and samples are diluted with biotin labelled 25(OH)D. The diluted samples are incubated in microtitre wells which are coated with a highly specific sheep 25(OH)D at room temperature before aspiration and washing. Enzyme (horseradish peroxidase) labelled avidin, is added and binds selectively to complexed biotin and, following a further wash step, colour is developed using a chromogenic substrate (TMB). The absorbance of the stopped reaction mixtures are read in a microtitre plate reader, colour intensity developed being inversely proportional to the concentration of 25(OH)D.

AI/ML Overview

The provided document describes the performance characteristics of the 25-Hydroxy Vitamin D EIA assay, but it focuses on analytical and comparative performance rather than a study proving the device meets general acceptance criteria in a clinical setting with human readers. The document is for an in-vitro diagnostic (IVD) device, so the criteria and studies described are relevant to laboratory performance.

Here's an analysis of the acceptance criteria and supporting studies based on the provided text, adapted for an IVD context where appropriate:

1. A table of acceptance criteria and the reported device performance

For IVD devices, "acceptance criteria" are usually internal performance goals set by the manufacturer to demonstrate analytical and clinical performance required for regulatory clearance. The document lists several performance characteristics and provides the corresponding measurements for the candidate device and, in some cases, the predicate device.

Performance CharacteristicAcceptance Criteria (from text or implied)Reported Device Performance (Candidate Device)
TraceabilityTraceable to ID-LCMS/MS 25(OH) Vitamin D RMP (VDSP standard)Traceable to ID-LCMS/MS RMP (VDSP human serum panel) and certified by CDC VDSP
Stability (Kit)8 months when stored at 2-8°C8 months when stored at 2-8°C
Stability (Open Vial)(Implied: adequate for practical use)Reconstituted Biotin 25(OH)D Solution: 8 weeks (2-8°C, dark) Unused Ab. coated plate strip: 8 weeks (2-8°C, foil pouch) Calibrators, Controls: 8 weeks (2-8°C) Wash Solution: 8 weeks (room temp, after prep)
Limit of Blank (LoB)(Implied: sufficiently low)1.3 ng/mL (4.5 nmol/L)
Limit of Detection (LoD)(Implied: sufficiently low)2.7 ng/mL (6.9 nmol/L)
Limit of Quantitation (LoQ)(Implied: sufficiently low)4.8 ng/mL (12.0 nmol/L)
Cross-Reactivity (25(OH)D3)(Implied: high expected, near 100%)95%
Cross-Reactivity (25(OH)D2)(Implied: high expected, near 100%)109%
Interference (Triglycerides)No interference up to 475 mg/dLNo interference up to 475 mg/dL
Interference (Bilirubin conjugated)No interference up to 20 mg/dLNo interference up to 20 mg/dL
Interference (Haemoglobin)No interference up to 400 mg/dLNo interference up to 400 mg/dL
Interference (HAMA)No interference up to 1000 ng/mLNo interference up to 1000 ng/mL
Interference (Rheumatoid Factor)No interference up to 800 IU/mLNo interference up to 800 IU/mL
Interference (Red Blood Cells)No interference up to 0.4%No interference up to 0.4%
Interference (Vitamin D Binding Proteins)No interference up to 2000 ng/dLNo interference up to 2000 ng/dL
Interference (Total Protein)No interference up to 9.2 g/dLNo interference up to 9.2 g/dL
Interference (Cholesterol, Total)No interference up to 500 mg/dLNo interference up to 500 mg/dL
Interference (Biotin)No interference up to 200 nmol/LNo interference up to 200 nmol/L
Method Comparison (vs. Predicate) Slope0.85 to 1.150.88 (Passing Bablok), 0.82 (Linear Reg)
Method Comparison (vs. Predicate) Intercept+/- 7 ng/mL3.23 ng/mL (Passing Bablok), 4.86 ng/mL (Linear Reg)
Method Comparison (vs. Predicate) Correlation (r)≥ 0.900.97
Method Comparison (vs. Reference) Slope0.88 to 1.02 (95% CI from study)0.96 (Passing Bablok), 0.97 (Deming Reg)
Method Comparison (vs. Reference) Intercept-1.62 to 1.88 ng/mL (95% CI from study)-0.11 ng/mL (Passing Bablok), -0.84 ng/mL (Deming Reg)
Method Comparison (vs. Reference) Correlation (r)0.947 (from study)0.947
Precision (Total %CV)(Implied: acceptable for diagnostic assay)3.7% to 11.5% across 11.7 to 65.1 ng/mL
Linearity (Regression Coefficient R²)(Implied: close to 1.00)1.00
Linearity (Max Deviation)(Implied: acceptable limits)-8.8% (>20ng/mL), 2.37 ng/ml (<20ng/mL)
Reportable Range6.5 to 100 ng/mL6.5 to 100 ng/mL (16.3 to 250 nmol/L)

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Precision/Reproducibility: 10 serum samples, assayed using 3 lots of reagents in duplicate, twice per day for a minimum of 20 days (n ≥ 80 replicates per sample). The origin of these samples is not specified.
  • Linearity/Assay Reportable Range: Samples containing varying concentrations of 25-hydroxyvitamin D were assayed in replicate of four. The number of unique samples for this study is not explicitly stated, but they were patient samples (high patient sample diluted with a low patient sample). Origin not specified.
  • Analytical Specificity (Cross-reactivity): Vitamin D serum samples were spiked with various cross-reactants. Number of samples not specified. Not specified if retrospective or prospective. Origin not specified.
  • Analytical Specificity (Interference): Substances spiked into the assay. Number of samples/experiments not specified. Not specified if retrospective or prospective. Origin not specified.
  • Method Comparison (vs. Predicate Device):
    • Sample Size: n = 195 (patient samples).
    • Provenance: Not specified (country/retrospective/prospective).
  • Method Comparison (vs. Reference Measurement Procedure - RMP):
    • Sample Size: n = 109 independent native serum samples.
    • Provenance: Samples were value assigned by the Ghent reference method procedure. Not specified if retrospective or prospective, or country of origin for the samples themselves.
  • Matrix Comparison:
    • Sample Sizes: SST (n=28), EDTA plasma (n=38), Sodium Heparin plasma (n=28), Lithium Heparin plasma (n=28), Citrate plasma (n=28).
    • Provenance: Not specified (country/retrospective/prospective).
  • Expected Values/Reference Range:
    • Sample Size: 280 apparently healthy adult individuals.
    • Provenance: Retrospective. Individuals were "living in geographical diverse regions of the United States to represent a broad spectrum of UV light exposure".

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

For this IVD device (25-Hydroxy Vitamin D EIA assay), the "ground truth" is established by reference methods or the true concentration of the analyte, not typically by expert interpretation from medical images or clinical judgment directly.

  • Method Comparison (vs. Reference Measurement Procedure): The "ground truth" for this comparison was established by the Ghent reference method procedure for 25-Hydroxy Vitamin D, and the assay is traceable to the isotope dilution-liquid chromatography/tandem mass spectrometry (ID-LCMS/MS) 25(OH) Vitamin D reference method procedure (RMP) which was used in assigning the target value for the Vitamin D Standardization Program (VDSP single donor human serum panel, itself traceable to NIST SRM 2972). This indicates a highly standardized and validated analytical method, which serves as the "expert" or definitive reference for concentration. The "experts" are the developers and maintainers of these reference methods, typically highly qualified analytical chemists or metrologists, not clinical experts for interpreting results in a diagnostic imaging sense.
  • Other studies: For precision, linearity, and specificity, the "ground truth" is inherent to the prepared samples or known concentrations, not established by human experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Adjudication methods like "2+1" or "3+1" are typically used in studies involving human interpretation (e.g., radiology reads) where disagreements need to be resolved. This is not applicable to the analytical performance studies of an IVD assay where quantitative results are compared to a reference method or known values. Therefore, none was used in the sense of clinical expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was done, as this device is an in-vitro diagnostic assay for quantitative laboratory measurement, not an AI-assisted diagnostic imaging tool that involves human readers interpreting images.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This device is a standalone assay. It provides a numerical result for 25-Hydroxy Vitamin D concentration. The performance studies detailed (precision, linearity, LoD, LoQ, method comparisons, interference, cross-reactivity) are all evaluations of this standalone analytical performance, quantifying its accuracy and reliability in measuring the analyte in patient samples. There is no "human-in-the-loop" once the sample is processed by the assay; a user performs the assay steps, but the result generated is direct from the assay's chemical reactions and spectrophotometric reading.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth used primarily consists of:

  • Reference Measurement Procedures (RMP): Specifically, the ID-LCMS/MS method (traceable to NIST SRM 2972) and the Ghent reference method procedure. These are highly accurate and standardized analytical methods for determining the true concentration of 25-Hydroxy Vitamin D.
  • Known Concentrations: For studies like linearity, LoB, LoD, LoQ, interference, and cross-reactivity, ground truth is established by preparing samples with known, precise concentrations of the analyte or interfering substances.

8. The sample size for the training set

This document describes a premarket notification (510(k)) for a medical device (an in-vitro diagnostic assay). The concept of a "training set" is typically associated with machine learning or AI models. For an IVD assay, raw materials and assay design are developed through R&D, and then validated with studies like those described. There isn't a "training set" in the sense of data used to train an algorithm. The development and optimization of the assay's components (e.g., antibodies) would occur during the manufacturing process, but the document doesn't provide details on sample sizes used during that phase.

9. How the ground truth for the training set was established

As there isn't a "training set" in the AI/ML context for this IVD assay, this question is not directly applicable. The "ground truth" for the analytical performance studies (as described in point 7) was established using reference measurement procedures and known concentrations.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The eagle is depicted in a simple, line-art style, and the text is in a sans-serif font.

August 25, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

IMMUNODIAGNOSTIC SYSTEMS LTD. MICK HENDERSON RA OFFICER 10 DIDCOT WAY, BOLDON BUSINESS PARK BOLDON, TYNE & WEAR NE35 9PD, GREAT BRITAIN

Re: K142351

Trade/Device Name: 25-Hydroxy Vitamin De EIA Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: II Product Code: MRG Dated: August 14, 2015 Received: August 19, 2015

Dear Mick Henderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142351

Device Name 25-Hydroxy Vitamin DS EIA

:

Indications for Use (Describe)

The 25-Hydroxy Vitamin DS EIA assay is intended for the quantitative determination of 25-hydroxyvitamin D [25(OH)D] and other hydroxylated metabolites in human serum or plasma. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in an adult population.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number,"

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

510k Number510(k) SUMMARYK142351
Introductiondetermination of substantial equivalence.According to the requirements of 21CFR807.92, the followinginformation provides sufficient detail to understand the basis for a
SubmitterImmunodiagnostic Systems Limited10 Didcot WayBoldon Business ParkBoldonTyne and WearNE35 9PDUnited Kingdom
Contact Person: Mick HendersonPhone: +44 191 5190660Fax: +44 191 5190760Email: mick.henderson@idsplc.com
Secondary Contact: Mr. R. PrebulaHogan Lovells US LLP,Columbia Square,555 Thirteenth Street, NW,Washington DC
20004WashingtonPhone: (202)637 6548Fax: (202)637 5910Email: randy.prebula@hoganlovells.com
Date prepared:18 August 2015
Device Name
Proprietary names:25-Hydroxy Vitamin Dª EIA
Common names:Vitamin D test
Classification:Regulatory Class:21CFR862.1825 Vitamin D Test SystemII
Product Code:MRG
IDS, OCTEIA 25-OH Vitamin D EIA, (K021163)Predicate Device Name:

{4}------------------------------------------------

A. Test Principle:

The 25-Hydroxy Vitamin De EIA is a manual assay test and does not require the use of an automated system. The whole assay is performed in a microtitre plate and requires a user to perform each step.

The 25-Hydroxy Vitamin De EIA assay

The 25-Hydroxy Vitamin De assay is an enzymeimmunoassay for the quantitation of 25(OH)D and other hydroxylated metabolites in serum or plasma. 25uL of each calibrators, controls and samples are diluted with biotin labelled 25(OH)D. The diluted samples are incubated in microtitre wells which are coated with a highly specific sheep 25(OH)D at room temperature before aspiration and washing. Enzyme (horseradish peroxidase) labelled avidin, is added and binds selectively to complexed biotin and, following a further wash step, colour is developed using a chromogenic substrate (TMB).

The absorbance of the stopped reaction mixtures are read in a microtitre plate reader, colour intensity developed being inversely proportional to the concentration of 25(OH)D.

B. Indications For Use:

For In Vitro Diagnostic Use

The 25-Hydroxy Vitamin De EIA assay is intended for the quantitative determination of 25-hydroxyvitamin D [25(OH)D] and other hydroxylated metabolites in human serum or plasma. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in an adult population.

Rx Only

C. Comparison with the predicate:

Similarities compared to the chosen predicate device (OCTEIA 25-hydroxy vitamin D EIA)

PredicateK021163Candidate Device
Intended UseThe OCTEIA 25-hydroxyvitamin D EIA (25-OH D) kit isan assay for the quantitativemeasurement of 25-Hydroxyvitamin D in serum and plasma.Same
Indications for useResults are to be used inconjunction with other clinicaland laboratory data to assist theSame

{5}------------------------------------------------

clinician in the assessment ofvitamin D sufficiency
Analyte25-hydroxy vitamin DSame
Reagent storage2-8°CSame
CaptureAntibody coated micro plateSame
CalibratorLow charcoal stripped humanserum containing 25-hydroxyvitamin D and sodium azide as apreservativeSame
Sample volume25µLSame
Unitsnmol/LSame
Kit componentsEIA kit consisting of Anti-25-OH D coated plate, Biotinylated25-OH D, Standards (7 levels),kit controls (2 levels), AvidinHRP solution, substrate solution,stop solution and wash solutionSame
CalibrationFull standard curve to be runwith all assaysSame
Calibration intervalPer assay runSame
Quality ControlTwo (2) controls providedSame
Conversion of unitsnmol/L x 0.40 = ng/mLng/ml. x 2.5 = nmol/LSame

Differences compared to the chosen (FDA cleared; marketed) predicate device

PredicateK021163Candidate Device
AntibodiesSheep anti 25-hydroxy vitaminD.new source
Analytical sensitivity5 nmol/LNA
Functional Sensitivity(Limit of Quantitation(LoQ)NA4.8ng/mL (12 nmol/L)
Limit of Detection(LoD)NA2.7ng/mL (6.9 nmol/L)
Limit of Blank(LoB)NA1.3ng/mL (4.5nmol/L)
PrecisionIntra-assay Precision n = 105.3% to 6.7% in theconcentration range 39 to165nmol/LInter-assay Precision n = 114.6% to 8.7% in theconcentration range 40 to132nmol/LWithin Run Precision n = 881.9% to 3.7% in theconcentration range 11.7 to65.1 ng/mL (29.3 to 183nmol/L)Total Precision n = 883.7% to 11.6% in theconcentration range 11.7 to65.1ng/mL (29.3 to 183nmol/L)

{6}------------------------------------------------

Specificity,SpecificitySpecificity
Interfering substances25-Hydroxy vitamin D325-Hydroxy vitamin D3
& Cross Reactivity100%તે તે જેની જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે
PerformancePredicateK021163Candidate Device
Specificity,Interfering substances& Cross ReactivityInterference:HaemoglobinTested up to 1470 mg/dLBilirubinTested up to 513 µmol/LLipidTested up to 5.6 mmol/LtriglycerideInterference:HaemoglobinNo interference up to400mg/dLBilirubin, conjugatedNo interference up to 20mg/dLTriglyceride (Intra Lipid)No interference up to475mg/dLTotal ProteinNo interference up to 9.2g/dLHAMANo interference up to1000ng/mLRed Blood CellsNo interference up to 0.4%Rheumatoid FactorNo interference up to800IU/mLVitamin D Binding Protein(Gc globulin)No interference up to 2000ng/mLCholesterol, TotalNo interference up to500mg/dLBiotinNo interference up to200nmol/L
PerformancePredicateK021163Candidate Device
Cross Reactivity25-Hydroxy vitamin D275%Cross Reactivity25-Hydroxy vitamin D2109%
24,25-DiHydroxy vitamin D3$\ge$ 100%24-25 Di- Hydroxy vitamin D395%
24,25-(OH)2D391%
Cholecalciferol (D3)< 0.01%3-epi-25-OH Vitamin D3-1%
Cholecalciferol (D2)< 0.30%3-epi-25-OH Vitamin D2-1%
1,25-(OH)2D35%
1,25-(OH)2D284%
Paricalcitol17%
25(OH)D395%
25(OH)D2109%
Vitamin D3 (Cholecalciferol)0%
Vitamin D2 (Ergocalciferol)6%
PerformancePredicateK021163Candidate Device
Reference range/Expected Values47.7 – 144 nmol/LNon Supplemented11.2 to 45.9ng/mL28.0 to 114.6 nmol/LSupplemented15.4 to 86.8ng/mL38.5 to 217.1 nmol/LOverall12.3 to 49ng/mL30.7 to 122.5 nmol/L
Method comparisonAgainst a recognisedradioimmunoassayn = 180AC57 = 1.01(x) + 0.7Correlation coefficient (r) =0.9Against Predicate devicen = 195Passing Bablok regression:25-Hydroxy Vitamin DS= 0.88x (25-Hydroxy Vitamin D) +3.2 ng/mL (+ 8.1 nmol/L)Pearson correlation coefficient,r: 0.94
Reportable Range2ng/mL to 152ng/mL6.5 to 100ng/mL
LinearityMean Measured / Expected102%Range individual dilutions88% to 125%Linear regression of theobserved concentrations versusthe expected concentrations:Observed = 1.02 x (Expected)+ 0.23 ng/mLObserved = 1.02 x (Expected)+ 0.58 nmol/LRegression coefficient R²:=1.00Maximum deviation; -8.8% forsamples > 20 ng/mL and 2.37ng/ml for samples < 20 ng/mL
Sample matrix(primary tube type)Serum or Plasma (EDTA orHeparin)Serum (standard samplingtubes or tubes containingserum separating gel) orplasma (EDTA, lithiumheparin, sodium heparin orsodium citrate)

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

D. Performance Characteristics :

    1. Analytical performance:
    • a. Precision/Reproducibility:

Precision was determined in accordance with CLSI EP5-A2, "Evaluation of Precision Performance of Quantitative Measurement Methods". 10 serum samples were assayed using 3 lots of reagents in duplicate, twice per day for a minimum of 20 days (n ≥ 80 replicates per sample).

ng/mL
EP Evaluator
Within RunBetween RunBetween DayTotal
n =MeanSD%cvSD%cvSD%cvSD%cv
IQC18819.10.41.9%0.63.2%0.10.4%0.73.7%
IQC28843.70.81.9%0.94.3%0.00.0%2.04.6%
IQC38865.12.33.5%4.06.2%1.72.6%4.97.6%
S18811.70.43.4%1.311.0%0.00.0%1.411.5%
S28824.50.62.5%1.35.3%0.00.0%1.45.8%
S38840.71.02.5%2.15.1%0.00.0%2.35.7%
S48873.22.73.7%4.05.5%0.60.8%4.96.6%
S58821.70.52.4%0.94.1%0.00.0%1.04.7%
S68851.01.93.7%3.46.6%0.00.0%3.97.6%
S78828.30.82.7%1.76.2%0.00.0%1.96.7%

Results from one representative lot is summarized in the table below (n=88)

b. Linearity/assay reportable range:

Linearity was evaluated based on a modified version of CLSI EP-6A, "Evaluation of the Linearity of Quantitative Measurement Procedures". Samples containing varying concentrations of 25-hydroxyvitamin D were assayed in replicate of four. The resulting mean concentrations were compared to predicted concentrations. Samples were prepared by diluting a high patient sample with a low patient sample prior to assay. The linear regression (weighted) of the Observed concentrations versus the expected concentrations is:

Observed = 1.02 x (Expected) + 0.23 ng/mL Observed = 1.02 x (Expected) + 0.58 nmol/L Regression coefficient R == 1.00 Maximum deviation; -8.8% for samples > 20 ng/mL and 2.37 ng/ml for samples < 20 ng/mL

The reportable range of the assay is 6.5 - 100 ng/mL (16.3 - 250 nmol/L). Any value that reads below 6.5 ng/mL (16.3nmol/L) should be reported as "< 6.5 ng/mL (16.3 nmol/L)".

{10}------------------------------------------------

  • Traceability, Stability, Expected values (controls, calibrators, or methods): C.

Traceability:

The 25-Hydroxy Vitamin D assay is traceable to the isotope dilution-liquid chromatography/tandem mass spectrometry (ID-LCMS/MS) 25(OH) Vitamin D reference method procedure (RMP) which was used in assigning the target value for the Vitamin D Standardization Program (VDSP single donor human serum panel. The ID-LCMS/MS RMP is traceable to the National Institute of Standards and Technology Standard Reference Material (SRM) 2972.

The 25-Hydroxy Vitamin DS assay is certified by the CDC Vitamin D Standardization Certification Program (VDSP) http://www.cdc.gov/labstandards/hs.html

Stability

This kit is stable for 8 months when stored at 2-8℃.

Kit componentAfter opening or preparation (open vial stability)
Biotin 25(OH)D Solution8 weeksafter reconstitutionStore at 2-8°C in the dark immediately after use
Unused Ab. coated plate strip8 weeksStore at 2-8°C in foil pouch withdesiccant sachet
Calibrators,Controls8 weeksStore at 2-8°C after opening
Wash Solution8 weeksStore at room temperature (18-25 °C) after preparation

d. Detection limit:

The limit of blank (LoB), limit of detection (LoD) and limit of quantitation (LoQ) were determined based on guidance from CLSI EP17-A "Protocols for the determination of limits of detection and limits of quantitation".

Sensitivity25OHD Levels
LoB1.3 ng/mL (4.5 nmol/L)
LoD2.7 ng/mL (6.9 nmol/L)
LoQ4.8 ng/mL (12.0 nmol/L)

e. Analytical specificity:

Cross-reactivity of endogenous 25(OH) vitamin D metabolite. Endogenous 25(OH) vitamin D metabolites were spiked into vitamin D serum samples and analyzed with the 25-Hydroxy Vitamin Do assay.

{11}------------------------------------------------

Cross ReactantSpike Conc. (ng/mL)Mean % Cross Reactivity
25(OH)D310.095%
20.0
25(OH)D210.0109%
20.0
24,25-(OH)2D34.091%
6.0

Cross-reactivity of exogenous synthetic 25(OH) vitamin D metabolite. Exogenous synthetic 25(OH) vitamin D metabolites were spiked into vitamin D serum samples and analyzed with the 25-Hydroxy Vitamin De assay.

Cross ReactantSpike Conc.(ng/mL)Mean% CrossReactivity
3-epi-25(OH) D3100-1%
3-epi-25(OH) D2100-1%
1,25-(OH)2D325%
1,25-(OH)2D22084%
VitaminD3 (Cholecalciferol)10000%
VitaminD2 (Ergocalciferol)1006%
Paricalcitol*10017%
  • Paricalcitol interferes with the 25-Hydroxy Vitamin De EIA test. Paricalcitol, when tested, caused a positive bias in sample result.

The following substances do not interfere in the 25-Hydroxy Vitamin DS EIA assay when the concentrations presented in the following table are below the stated threshold.

Potentially InterferingThreshold
AgentConcentration
Triglycerides475 mg/dL
Bilirubin (conjugated)20 mg/dL
Haemoglobin400 mg/dL
HAMA1000 ng/mL
Rheumatoid Factor800 IU/mL
Red Blood Cells0.40%
Vitamin D BindingProteins2000 ng/dL

{12}------------------------------------------------

Total Protein9.2g/dL
Cholesterol, Total500 mg/mL
Biotin200nmol/L
    1. Comparison studies:
    • a. Method comparison with predicate device:

The 25-Hydroxy Vitamin DS EIA Assay was compared against the 25-Hydroxy Vitamin D EIA assay (K021163) for the quantitative determination of 25-Hydroxy Vitamin D (and other hydroxylated metabolites), following CLSI EP-9A2, "Method Comparison and Bias Estimation Using Patient Samples".

Correlation to 25-Hydroxy Vitamin D EIA, AC57 (K021163)

Passing Bablok$y = 0.88x + 3.23$ ng/mL$y = 0.88x + 8.05$ nmol/L
Slope, 95% Confidence Interval0.86 to 0.910.86 to 0.91
Intercept, 95% Confidence Interval2.71 to 3.83 ng/mL6.85 to 9.45 nmol/L
Linear Regression$y = 0.82x + 4.86$ ng/mL$y = 0.82x + 12.35$ nmol/L
Slope, 95% Confidence Interval0.80 to 0.840.80 to 0.84
Intercept, 95% Confidence Interval4.14 to 5.57 ng/mL10.35 to 13.92 nmol/L
Correlation Coefficient, r0.970.97
n195195
Difference Plot-0.61-1.53

Additional method comparison study was performed against the reference measurement procedure using 109 in dependent native serum samples which has been value assigned by the Ghent reference method procedure. Regression analysis was summarized below:

ng/mLnmol/L
Passing Bablok$y = 0.96x - 0.11$ ng/mL$y = 0.96x - 0.34$ nmol/L
Slope, 95% Confidence Interval0.88 to 1.020.88 to 1.02
Intercept, 95% Confidence Interval-1.62 to 1.88 ng/mL-4.15 to 4.68 nmol/L
Deming Regression$y = 0.97x - 0.84$ ng/mL$y = 0.97x - 2.07$ nmol/L
Slope, 95% Confidence Interval0.90 to 1.040.90 to 1.04
Intercept, 95% Confidence Interval-2.75 to 1.08 ng/mL-6.85 to 2.71 nmol/L
Correlation Coefficient, r0.9470.947
n109109
Diffrence Plot-1.77-4.42

{13}------------------------------------------------

b. Matrix comparison:

Summary of the statistical methods used for alternative blood tube types to serum blood tube using a paired tube comparison is shown below.

Passing-Bablok analysis of the test and comparator assays was performed, taking the slope, intercept and correlation coefficient (with 95 % confidence intervals) and difference bias plot.

SSTEDTASodium Heparin
n283828
Passing bablok$y=0.99x+0.18 g/mL$$y=0.97x+0.60ng/mL$$y=1.07x-0.47ng/mL$
Slope,95% confidence level0.95 to 1.031.00 to 1.011.00 to 1.10
Intercept, 95% confidencelevel-0.17 to 0.65ng/mL-0.15 to 1.04ng/mL-1.21 to 0.68ng/mL
Correlation Coefficient r0.991.000.99
Mean Bias0.410.071.11
Lithium HeparinCitrate
n2828
Passing BablokY=1.04x -0.22 ng/mLY=1.03x -0.70 ng/mL
Slope, 95% confidence level1.01 to 1.080.98 to 1.06
Intercept, 95% Confidence level-0.73 to 0.60 ng/mL-1.36 to 0.49 ng/mL
Correlation Coefficient, r1.000.99
Mean Bias1.500.43

The design criterion was that the slope must be 0.85 to 1.15 and the intercept +/- 7 ng/mL and r ≥0.90.

Serum separator tubes (SST), EDTA plasma, Sodium Heparin plasma, Lithium Heparin plasma and citrate plasma tube data do not present any new issues of safety or effectiveness for the 25-Hydroxy Vitamin De EIA assay.

3. Expected values/Reference range:

An expected values study performed according to the non-parametric method in CLSI protocol C28-A2.

Samples from 280 apparently light skin and dark skin healthy male adults (71.1%) and female adults (28.9%) aged 21-77 years living in geographical diverse regions of the United States to represent a broad spectrum of UV light exposure in the intended population were assayed in the 25-Hydroxy Vitamin Do Assay. Samples were from individuals with normal values for intact PTH, calcium, phosphate, and TSH, and not taking any interfering medications. The following ranges were determined using the 25-Hydroxy Vitamin De assay and are provided for information only. The 95 % reference interval for apparently healthy adults, were calculated by a non-parametric

{14}------------------------------------------------

method following guidance from CLSI C28-A3 " Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory".

Obtained normal adults range:12.3 to 49.0 ng/mL (30.7 to 122.5 nmol/L) (n=280)

n%Observed Sample Range(ng/mL)Median(ng/mL)
Non-supplemented21978.2%11.2 to 45.925.1
Northern US11641.4%10.3 to 35.122.6
Southern US10336.8%14.6 to 48.029.5
Supplemented6121.8%15.4 to 86.830.2
Overall280100%12.3 to 49.026.0

Observed sample ranges were:

n%Observed Sample Range(nmol/L)Median(nmol/L)
Non-supplemented21978.2%28.0 to 114.662.8
Northern US11641.4%25.6 to 87.656.5
Southern US10336.8%36.6 to 120.073.9
Supplemented6121.8%38.5 to 217.175.5
Overall280100%30.7 to 122.565.0

The above ranges should be considered as guidelines only; it is recommended that each laboratory establish its own expected range based for its own patient population.

The 95% reference interval was calculated by a non-parametric method following C28-A2. The following range was obtained:

12.3 to 49.0 ng/mL (n = 280) Normal Adults 30.7 to 122.5 nmol/L (n = 280)

The package insert states that there is no universal agreement on the optimal concentration of 250HD. Ranges should be based on clinical decision values that apply to both sexes of all ages rather than population based reference ranges for 250HD.

E. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 862.1825 Vitamin D test system.

(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.