(169 days)
The Sterilucent Sterilization Container System is a device intended to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. This container system is intended to be used in the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer Lumen Cycles. The sterility maintenance period has been validated for a period of up to 180 days.
Reusable baskets and accessory items (clips, posts, pins, dividers, bars and mats) are intended to organize and secure enclosed medical devices during sterilization, transport, and storage of the container.
Consumable accessory items (filter media, data card and tamper evident arrows) provide a range of functions and are indicated as single-use devices. Filter media allows ingress of sterilant while providing a microbial barrier. Data cards are used to record information regarding a specific sterilization process load. Tamper evident arrows provide a visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains an external process indicator that serves as a visual indication that the system has been exposed to the Sterilization process.
The Sterilucent Sterilization Container System is an assortment of rigid, reusable, stackable containers that are used to sterilize other medical devices and maintain sterility of those devices until used. The container system is comprised of a lid, bottom, filter, tamper evident arrows, and data cards.
The container system houses reusable baskets of varying depths and organizing accessory items that are used to organize and secure surgical instrumentation and/or other medical devices.
The provided text describes the acceptance criteria and performance of the Sterilucent Sterilization Container System (K142109) through non-clinical studies.
Here's the breakdown of the information you requested:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria for Sterilization Performance (Sterilization Container System):
- Demonstrate a 12 log reduction in microbial load.
- Achieve a Sterility Assurance Level (SAL) of 10^-6.
- Demonstrate sterility maintenance for a period of up to 180 days.
- Maintain appropriate microbial barrier properties following exposure to hydrogen peroxide sterilization processes under worst-case conditions.
Reported Device Performance:
- Sterilization efficacy testing demonstrated a 12 log reduction and a sterility assurance level (SAL) of 10^-6.
- Real-time event-related shelf-life studies demonstrated sterility maintenance for a 180-day time period.
- Whole package microbial challenge testing demonstrated appropriate microbial barrier properties following exposure to the Sterilucent hydrogen peroxide sterilization processes under worst-case conditions.
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific "sample size" in terms of number of devices or iterations for the efficacy and sterility maintenance tests. It refers to "studies" and "testing" without quantifying the number of units tested.
- Sample size for test set: Not explicitly stated (e.g., "three containers were tested" or "n=X runs"). The document refers to "studies" and "testing."
- Data provenance: Prospective non-clinical studies. The country of origin of the data is not specified, but the device is submitted for FDA clearance in the USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not provided in the document. Non-clinical studies for sterilization effectiveness typically rely on laboratory assays and established microbiological standards, rather than expert human interpretation of results in the way, for example, a radiology AI model would.
4. Adjudication Method for the Test Set
This information is not applicable as the studies are non-clinical (laboratory-based microbiological and physical testing), not involving human readers or interpretation requiring adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of images or other data, not for sterilization container systems.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This question is not applicable as the device is a physical sterilization container system, not an algorithm or AI model. The performance described is that of the physical device itself.
7. The Type of Ground Truth Used
The ground truth for the non-clinical studies was established using microbiological assays, specifically:
- Biological indicator (BI) overkill method: For sterilization efficacy (demonstrating a 12 log reduction and SAL of 10^-6).
- Microbial challenge testing: For evaluating microbial barrier properties, using a specified quantity of Bacillus atrophaeus colony forming units (CFU).
- Real-time event-related shelf-life studies: To determine sterility maintenance over time.
8. The Sample Size for the Training Set
This information is not applicable. The device is a physical product (a sterilization container system), not a software or AI model that requires a training set. The studies described are validation and performance tests, not training data for an algorithm.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the reasons stated in point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a way that they appear to be interconnected. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 20, 2015
Sterilucent, Inc. Peter Kalkbrenner Director of Engineering 1400 Marshall Street NE Minneapolis, MN 55413
Re: K142109
Trade/Device Name: Sterilucent Sterilization Container System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Container Regulatory Class: II Product Code: KCT Dated: December 19, 2014 Received: December 22, 2014
Dear Mr. Kalkbrenner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K142109
Device Name Sterilucent Sterilization Container System
Indications for Use (Describe)
The Sterilucent Sterilization Container System is a device intended to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. This container system is intended to be used in the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer Lumen Cycles. The sterility maintenance period has been validated for a period of up to 180 days.
Reusable baskets and accessory items (clips, posts, pins, dividers, bars and mats) are intended to organize and secure enclosed medical devices during sterilization, transport, and storage of the container.
Consumable accessory items (filter media, data card and tamper evident arrows) provide a range of functions and are indicated as single-use devices. Filter media allows ingress of sterilant while providing a microbial barrier. Data cards are used to record information regarding a specific sterilization process load. Tamper evident arrows provide a visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains an external process indicator that serves as a visual indication that the system has been exposed to the Sterilization process.
(Continued on separate page.)
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
Indications for Use Continuation page (Form FDA 3881 (1/14)
| Container | Basket | Contents/Configuration | Sterilization Cycle | |
|---|---|---|---|---|
| Catalog # | Catalog # | LumenCycle | Non-LumenCycle | |
| SL065 | SL011 | Baskets | YES | YES |
| SL066 | SL012 | |||
| SL067 | SL013 | |||
| SL068 | SL014 | Stacking Baskets | YES | YES |
| SL069 | SL015 | Lumen: 1mm (ID) or larger x 60mm (L) orshorter | YESQty. 10 | NO |
| SL070 | SL016 | |||
| SL071 | SL017 | |||
| SL072 | SL018 | |||
| SL073 | SL019 | Lumen: 2mm (ID) or larger x 250mm (L) orshorter | YESQty. 10 | NO |
| SL074 | SL020 | |||
| SL075 | SL021 | |||
| SL076 | SL022 | |||
| SL076 | SL023 | Lumen: 3mm (ID) or larger x 350mm (L) orshorter | YESQty. 10 | NO |
| SL076 | SL023 | |||
| SL077 | SL024 | Occluded/Mated Challenge | YES | YES |
| SL078 | SL025 | |||
| SL098 | SL026 | |||
| SL098 | SL027 | Silicone Support Bars | YES | YES |
| SL098 | SL028 | Dividers | YES | YES |
| SL098 | SL029 | |||
| SL098 | SL030 | |||
| SL098 | SL031 | Brackets | YES | YES |
| SL098 | SL032 | |||
| SL098 | SL033 | |||
| SL098 | SL034 | Silicone Mat | YES | YES |
| SL098 | SL035 | |||
| SL098 | SL036 | |||
| SL098 | SL037 | Filter | SL082SL083SL084 | SL082SL083SL084 |
| SL098 | SL038 | |||
| SL098 | SL039 | |||
| SL098 | SL040 | Data Card | SL081 | SL081 |
| SL098 | SL041 | |||
| SL098 | SL042 | |||
| SL098 | SL043 | Tamper-Evident Arrow | SL085 | SL085 |
| SL098 | SL044 | |||
| SL098 | SL045 | |||
| SL098 | SL046 | Stack Height | NO STACKING | NO STACKING |
| SL098 | SL047 | |||
| SL098 | SL048 | |||
| SL098 | SL049 | Max. Total Container System Weight | 10 lbs(4.5 kg) | 25 lbs(11.3 kg) |
| SL098 | SL050 |
Container and Accessory Device Challenges
{4}------------------------------------------------
| CatalogCode | Container Description | External Dimensions | Total Container Plus LoadWeight | |
|---|---|---|---|---|
| Lumen Cycle¹ | Non-Lumen² | |||
| SL065 | Mini | 10.2" x 7.2" x 3.2" | 5 lbs(2.3 kg) | 5 lbs(2.3 kg) |
| SL066 | Quarter Length | 9.5" x 12.4" x 3.8" | 8 lbs(3.3 kg) | 8 lbs(3.3 kg) |
| SL067 | Half Length, 4" Deep | 11.8" x 12.4" x 4.5" | 10 lbs(4.5 kg) | 12 lbs³(5.4 kg) |
| SL068 | Half Length, 5" Deep | 11.8" x 12.4" x 5.3" | 10 lbs(4.5 kg) | 12 lbs³(5.4 kg) |
| SL069 | Half Length, 6" Deep | 11.8" x 12.4" x 6.1" | 10 lbs(4.5 kg) | 12 lbs³(5.4 kg) |
| SL070 | Mid Length, 4" Deep | 19.2" x 12.4" x 4.5" | 10 lbs(4.5 kg) | 20 lbs(9.1 kg) |
| SL071 | Mid Length, 5" Deep | 19.2" x 12.4" x 5.3" | 10 lbs(4.5 kg) | 20 lbs(9.1 kg) |
| SL098 | Mid Length, 6" Deep | 19.2" x 12.4" x 6.1" | 10 lbs(4.5 kg) | 20 lbs(9.1 kg) |
| SL072 | Mid Length, 8" Deep | 19.2" x 12.4" x 7.8" | 10 lbs(4.5 kg) | 20 lbs(9.1 kg) |
| SL073 | Full Length, 4" Deep | 23.1" x 12.4" x 4.5" | 10 lbs(4.5 kg) | 25 lbs(11.3kg) |
| SL074 | Full Length, 5" Deep | 23.1" x 12.4" x 5.3" | 10 lbs(4.5 kg) | 25 lbs(11.3 kg) |
| SL075 | Full Length, 6" Deep | 23.1" x 12.4" x 6.1" | 10 lbs(4.5 kg) | 25 lbs(11.3kg) |
| SL076 | Full Length, 7" Deep | 23.1" x 12.4" x 7.0" | 10 lbs(4.5 kg) | 25 lbs(11.3kg) |
| SL077 | Small Narrow, 3" Deep | 20.8" x 7.3" x 3.9" | 10 lbs(4.5 kg) | 10 lbs(4.5 kg) |
| SL078 | Small Narrow, 5" Deep | 20.8" x 7.3" x 5.2" | 10 lbs(4.5 kg) | 10 lbs(4.5 kg) |
Maximum Container Loads
1 The Lumen Cycle validation testing was conducted using a maximum of ten (10) lumens per load. The validation studies were performed using the following device models and weights:
| Model | SL065 | SL066 | SL069 | SL072 | SL076 | SL078 |
|---|---|---|---|---|---|---|
| Validation Load Weight (lbs.) | 5.1 | 8.0 | 10.3 | 14.0 | 17.1 | 10.1 |
2 The Non-Lumen Cycle validation studies were performed using the following device models and weights:
| Model | SL065 | SL066 | SL069 | SL072 | SL076 | SL078 | |
|---|---|---|---|---|---|---|---|
| Validation Load Weight (lbs.) | 5.0 | 8.0 | 12.1 | 20.1 | 25.1 | 10.1 |
3 Two (2) containers, each containing 12 lbs. (5.4 kg), for a total chamber load weight of 24 lbs. (10.8 kg), has also been validated for use in the PSD-85 Non-Lumen Cycle.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Sterilucent. The logo consists of a blue abstract symbol to the left of the word "sterilucent" in blue, sans-serif font. Below the word "sterilucent" is the phrase "scientific sterilization solutions" in a smaller, blue, sans-serif font. The background of the image is light blue.
510(k) Summary for the Sterilucent Sterilization Container System K142109
| Owner: | |
|---|---|
| Address: |
Telephone:
Sterilucent, Inc. 1400 Marshall Street NE Minneapolis, MN 55413 612-767-3260 612-767-3261
Contact:
Fax:
Peter R. Kalkbrenner Director of Engineering
612-767-3253
612-767-3261
Telephone: Fax:
Summary Date: 16 January 2015
{7}------------------------------------------------
1. Device Name and Classification
| Trade Name: | Sterilucent Sterilization Container System |
|---|---|
| Common/Usual Name: | Sterilization Container |
| Classification Name: | Sterilization Wrap, Containers, Trays, Cassettes & OtherAccessories |
| Product Code: | KCT (21 CFR 880.6850) |
| Class: | II |
2. Predicate Device
Genesis™ Reusable Rigid Container System (K112535)
3. Device Description
The Sterilucent Sterilization Container System is an assortment of rigid, reusable, stackable containers that are used to sterilize other medical devices and maintain sterility of those devices until used. The container system is comprised of a lid, bottom, filter, tamper evident arrows, and data cards.
The container system houses reusable baskets of varying depths and organizing accessory items that are used to organize and secure surgical instrumentation and/or other medical devices.
4. Statement of Intended Use:
The Sterilucent Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The container system is intended to be used in the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer Lumen and Non-Lumen Cycles. The sterility maintenance period has been validated for a period of up to 180 days.
Reusable baskets and accessory items (clips, posts, pins, dividers, brackets, bars and mats) are intended to organize and secure enclosed medical devices during sterilization, transport, and storage of the container.
Consumable accessory items (filter media, data card and tamper evident arrows) provide a range of functions and are indicated as single-use devices. Filter media allows ingress and egress of sterilant while providing a microbial barrier. Data cards are used to record information regarding a specific sterilization process load. Tamper evident arrows provide a
{8}------------------------------------------------
visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains an external process indicator that serves as a visual indication that the system has been exposed to the Sterilucent VHP sterilization process.
| Container | Basket | Sterilization Cycle | |||
|---|---|---|---|---|---|
| Catalog # | Catalog # | Contents/Configuration | Lumen | Non-Lumen | |
| Cycle | Cycle | ||||
| ട് സ്കാരംSLOE6 | SL011SL012 | Baskets | YES | YES | |
| SL067SL068 | SL013SL014 | Stacking Baskets | YES | YES | |
| SLO69SL070SL071 | SL015SL016SL017 | Lumen: 1mm (ID) or larger x 60mm (L) orshorter | YESQty. 10 | NO | |
| SL072SL073 | SL018SL019SL020 | Lumen: 2mm (ID) or larger x 250mm (L) orshorter | YESQty. 10 | NO | |
| SL074SL075SL076 | SL021SL022SL023 | Lumen: 3mm (ID) or larger x 350mm (L) orshorter | YESQty. 10 | NO | |
| SL077SL078 | SL024SL025 | Occluded/Mated Challenge | YES | YES | |
| SL098 | SL026SL027 | Silicone Support Bars | YES | YES | |
| SL028SL029 | Dividers | YES | YES | ||
| SL030SL031 | Brackets | YES | YES | ||
| SL032SL033 | Silicone Mat | YES | YES | ||
| SL034SL035SL036SL037 | Filter | SL082SL083SL084 | SL082SL083SL084 | ||
| SL038SL039 | Data Card | SL081 | SL081 | ||
| SL040SL041SL042SL043SL044SL045SL046SL047SL048SL049SL050 | Tamper-Evident Arrow | ട് സ്റ്റാറ്റ് ട്രോഗ്യമായി സ്റ്റാമിക്കുന്നു. അവലംബം കോട്ടിക്കുന്നു വിവ്യാവിക്കുന്നു. അവലംബം കേരള | ട് സ്റ്റാറ്റ് ട്രോഗ്യമായി സ്റ്റാമിക്കുന്നു. അവലംബം കോട്ടിക്കുന്നു കേരള | ||
| Stack Height | NO STACKING | NO STACKING | |||
| Max. Total Container System Weight | 10 lbs(4.5 kg) | 25 lbs(11.3 kg) |
Container and Accessory Device Challenges
{9}------------------------------------------------
| CatalogCode | Container Description | External Dimensions | Total Container Plus LoadWeight | |
|---|---|---|---|---|
| Lumen Cycle¹ | Non-Lumen² | |||
| SL065 | Mini | 10.2" x 7.2" x 3.2" | 5 lbs(2.3 kg) | 5 lbs(2.3 kg) |
| SL066 | Quarter Length | 9.5" x 12.4" x 3.8" | 8 lbs(3.3 kg) | 8 lbs(3.3 kg) |
| SL067 | Half Length, 4" Deep | 11.8" x 12.4" x 4.5" | 10 lbs(4.5 kg) | 12 lbs³(5.4 kg) |
| SL068 | Half Length, 5" Deep | 11.8" x 12.4" x 5.3" | 10 lbs(4.5 kg) | 12 lbs³(5.4 kg) |
| SL069 | Half Length, 6" Deep | 11.8" x 12.4" x 6.1" | 10 lbs(4.5 kg) | 12 lbs³(5.4 kg) |
| SL070 | Mid Length, 4" Deep | 19.2" x 12.4" x 4.5" | 10 lbs(4.5 kg) | 20 lbs(9.1 kg) |
| SL071 | Mid Length, 5" Deep | 19.2" x 12.4" x 5.3" | 10 lbs(4.5 kg) | 20 lbs(9.1 kg) |
| SL098 | Mid Length, 6" Deep | 19.2" x 12.4" x 6.1" | 10 lbs(4.5 kg) | 20 lbs(9.1 kg) |
| SL072 | Mid Length, 8" Deep | 19.2" x 12.4" x 7.8" | 10 lbs(4.5 kg) | 20 lbs(9.1 kg) |
| SL073 | Full Length, 4" Deep | 23.1" x 12.4" x 4.5" | 10 lbs(4.5 kg) | 25 lbs(11.3 kg) |
| SL074 | Full Length, 5" Deep | 23.1" x 12.4" x 5.3" | 10 lbs(4.5 kg) | 25 lbs(11.3 kg) |
| SL075 | Full Length, 6" Deep | 23.1" x 12.4" x 6.1" | 10 lbs(4.5 kg) | 25 lbs(11.3 kg) |
| SL076 | Full Length, 7" Deep | 23.1" x 12.4" x 7.0" | 10 lbs(4.5 kg) | 25 lbs(11.3 kg) |
| SL077 | Small Narrow, 3" Deep | 20.8" x 7.3" x 3.9" | 10 lbs(4.5 kg) | 10 lbs(4.5 kg) |
| SL078 | Small Narrow, 5" Deep | 20.8" x 7.3" x 5.2" | 10 lbs(4.5 kg) | 10 lbs(4.5 kg) |
Maximum Container Loads
1 The Lumen Cycle validation testing was conducted using a maximum of ten (10) lumens per load. The validation studies were performed using the following device models and weights:
| Model | SL065 | SL066 | SL069 | SL072 | SL076 | SL078 |
|---|---|---|---|---|---|---|
| Validation Load Weight (lbs.) | 5.1 | 8.0 | 10.3 | 14.0 | 17.1 | 10.1 |
2 The Non-Lumen Cycle validation studies were performed using the following device models and weights:
| Model | SL065 | SL066 | SL069 | SL072 | SL076 | SL078 |
|---|---|---|---|---|---|---|
| Validation Load Weight (lbs.) | 5.0 | 8.0 | 12.1 | 20.1 | 25.1 | 10.1 |
3 Two (2) containers, each containing 12 lbs. (5.4 kg), for a total chamber load weight of 24 lbs. (10.8 kg), has also been validated for use in the PSD-85 Non-Lumen Cycle.
{10}------------------------------------------------
5. Technological Characteristics Summary Comparison
The Sterilucent Sterilization Container System has similar technological characteristics as the predicate device:
| Summary of Technological Characteristics of the Device Compared to the | ||
|---|---|---|
| Predicate Device | ||
| Characteristic | New DeviceSterilucent SterilizationContainer SystemK142109 | Predicate DeviceGenesis™ Reusable RigidContainer SystemK112535 |
| Container | 5000 and 1100 Series AnodizedAluminum;300 Series Stainless Steel | Same |
| Container Gasket | Closed Cell Silicone Foam | Same |
| Reusable Baskets | 304 Electropolished Stainless Steel | Same |
| Clips, Posts, Pins | 300 and 400 Series Stainless Steel | Same |
| Dividers, Brackets | 5000 Series Aluminum | Same |
| Bars, Mats | Silicone Elastomer | Same |
| Filter Material | SMS Polypropylene | Same |
| Data Card | High-Density Polyethylene (Tyvek®) | Same |
| Tamper Evident Arrow | Polypropylene | Same |
| Arrow Process Indicator | Synthetic Substrate printed with ReactiveInk | Same |
| Sterilization Modality | Vaporized Hydrogen Peroxide (VHP) | Same (some models alsoapproved for steam andETO) |
| Indications for Use | The Sterilucent Sterilization ContainerSystem is a device intended to be used toenclose another medical device that is tobe sterilized by a healthcare provider. Itis intended to allow sterilization of theenclosed medical device and also tomaintain sterility of the enclosed deviceuntil used. The container system isintended to be used in the SterilucentPSD-85 Hydrogen Peroxide SterilizerLumen and Non-Lumen Cycles.Reusable baskets and accessory items(clips, posts, pins, dividers, brackets, barsand mats) are intended to organize andsecure enclosed medical devices duringsterilization, transport, and storage of thecontainer. | The Genesis ReusableRigid SterilizationContainer System is adevice intended to be usedto enclose another medicaldevice that is to besterilized by a healthcareprovider. It allowssterilization of theenclosed medical deviceand maintains sterility ofthe enclosed device untilused for a maximum of180 days.Containers are suitable fordynamic air removal (pre-vacuum) steam |
| Consumable accessory items (e.g. filtermedia, data card and tamper evidentarrows) provide a range of functions andare indicated as single-use devices. Filtermedia allows ingress and egress ofsterilant while providing a microbialbarrier. Data cards are used to recordinformation regarding a specificsterilization process load. Tamper evidentarrows provide a visual indication thatthe container system has not beeninadvertently opened prior to use. Eacharrow contains an external processindicator that serves as a visual indicationthat the system has been exposed to theSterilucent VHP sterilization process. | use pre-vacuum steamsterilization and 100%ethylene oxide sterilizationwhen used as described inthe instructions for use.Reusable baskets andaccessory items (pins,dividers, mats, etc) areintended to organize andsecure enclosed medicaldevices duringsterilization, transport, andstorage of the container.Data cards are used torecord informationregarding a specificsterilization process load.Filter media allows ingressand egress of sterilantwhile providing amicrobial barrier. Tamperevident arrows provide avisual indication that thecontainer system has notbeen inadvertently openedprior to use. Each arrowcontains a modality-specific external processindicator that serves as a | |
| visual indication that thesystem has been exposedto a specific sterilizationparameter. Data cards,filters and tamper evident | ||
| arrows are single use only. |
{11}------------------------------------------------
6. Summary of Non-Clinical Performance Data
Sterilization performance studies were conducted for the Sterilization Container System and all acceptance criteria were met. Sterilization efficacy testing demonstrated a 12 log reduction and a sterility assurance level (SAL) of 10 ° using the biological indicator (BI) overkill method and half-cycle validation under worst case conditions. Real time event related shelf life studies demonstrated sterility maintenance for a 180 day time period. Whole
{12}------------------------------------------------
package microbial challenge testing, exposing a container to a minimum of 1 x 106 Bacillus atrophaeus colony forming units (CFU) via an aerosol challenge, demonstrated appropriate microbial barrier properties following exposure to the Sterilucent hydrogen peroxide sterilization processes under worst case conditions.
7. Summary of Clinical Performance Data
N/A - No clinical tests were conducted for this submission.
8. Overall Performance Conclusion Statement
The non-clinical studies demonstrated that the Sterilization Container System is as safe, as effective, and performs as well as the predicate device for the sterilization of the enclosed medical devices and maintains sterility of those devices for a period of up to 180 days. The proposed device and the predicate device are composed of the same designs, materials, and manufacturing characteristics. The proposed device is substantially equivalent to the predicate device.
§ 880.6850 Sterilization wrap.
(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).