(81 days)
The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Multix Fusion system is not meant for mammography.
The Multix Fusion uses a mobile (wired) or portable (wireless) digital detector for generating diagnostic images by converting x-rays into electronic signals. The Multix Fusion is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
The Multix Fusion is a modification of Siemens' Ysio X-ray system cleared under Premarket Notification K081722 on 08/22/2008. The Multix Fusion offers the following system configurations:
- An analog radiography system or; .
- A digital radiography system with a mobile (wired) and/or a portable . (wireless) flat panel detector or;
- An optional Multix DR-Upgrade for the analog version of the Multix Fusion . and other tables within the Multix and Vertix analog family.
This modification does not affect the general intended use of the device nor does it alter its fundamental scientific technology.
The Multix Fusion system consists of a radiologic table, x-ray generator, x-ray tube, flat panel detector (wired or wireless) and Bucky-wall stand. It will be marketed as an analog system, a digital radiographic system with a mobile (wired) and/or portable (wireless) flat panel detector or an optional Multix DR- Upgrade for the analog version of the Multix Fusion and other tables within the Multix and Vertix analog family.
The key components are a patient table, a ceiling mounted support and a Bucky wall stand which are available in different configurations. A semi-motorized movement in the ceiling-mounted X-ray tube is available. The optional Multix DR-Upgrade for the analog version of the Multix Fusion consists of a mobile (wired) and/or portable (wireless) flat panel detector and an imaging system. Components used to upgrade (Multix DR-Upgrade) the analog version of the Multix Fusion and other tables within the Multix and Vertix analog family are the same or similar components cleared in the Ysio 510(k) K0817227 on August 22, 2008.
The Multix Fusion may be combined into different configurations to provide specialized customer requirements.
Similar to the cleared Ysio stationary x-ray system, the Multix Fusion and the Mutix DR-Upgrade have the same or similar comparable components. The subject device the Multix Fusion does not affect the indication for use nor the intended use of the device nor does it alter its fundamental scientific technology.
The provided text is a 510(k) summary for the Multix Fusion X-ray System. It describes the device, its intended use, and its substantial equivalence to a predicate device (Ysio).
However, the document does not contain information about formal acceptance criteria for device performance, a study to prove those criteria are met, specific performance metrics, sample sizes for test/training sets, data provenance, expert qualifications, ground truth establishment, or any MRMC (Multi-Reader Multi-Case) study results.
The document states:
- "Non-clinical testing was conducted for the Multix Fusion during product development. This includes verification and validation testing as well as phantom testing. The risk analysis was combation and risk controls implemented to mitigate identified hazards. The testing results supports that all the software specificating have met the acceptance criteria." (Page 3, Section 9)
- "Testing for verification and validation of the device was found acceptable to support the claims of substantial eqivalence." (Page 3, Section 9)
Therefore, based on the provided text, I cannot complete the requested tables and information because the details of the acceptance criteria, the study design, specific performance metrics, and data characteristics are not included.
Here's an explanation of why each requested point cannot be fulfilled:
- Table of acceptance criteria and reported device performance: The document generally states that "software specifications have met the acceptance criteria" and "testing...was found acceptable to support the claims of substantial equivalence," but it does not specify what those criteria are or what the actual performance values were.
- Sample size used for the test set and data provenance: No information on test sets, their size, or data provenance (e.g., country of origin, retrospective/prospective) is provided. The testing mentioned refers to "phantom testing" and "verification and validation," which are typically done internally and not necessarily on patient data as part of a formal clinical reader study for performance claims.
- Number of experts used to establish ground truth and qualifications: No information about experts or ground truth establishment is given.
- Adjudication method: Not mentioned.
- MRMC comparative effectiveness study: The document does not mention any MRMC study or any comparison of human readers with vs. without AI assistance. This device is a diagnostic imaging system, not an AI-based interpretation tool.
- Standalone (algorithm only) performance: Not mentioned. The device itself is an X-ray system, not a standalone AI algorithm.
- Type of ground truth used: Not mentioned.
- Sample size for training set: No information on training sets is provided, as this is a traditional medical device, not an AI model requiring a training set in that context.
- How ground truth for training set was established: Not applicable based on the available information.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows handwritten text on a white background. The text at the top reads "K121513", which appears to be a code or identifier. Below this, the text "Page 1 of 4" indicates that this is the first page of a four-page document. The handwriting is clear and legible, with a simple, unadorned style.
AUG 1 0 2012
510(k) Summary: Multix Fusion X-ray System
Company:
: - -
Siemens Medical Systems, Inc. 1 Valley Stream Parkway Malvern, PA 19355
July 26, 2012 Date Prepared:
Date Frepared. In our offectiveness information is being submitted in This order) - Sammants of SMDA 1990 and 21 CFR § 807.92.
General Information: 1.
Importer / Distributor: Siemens Medical Solutions, Inc. 51 Valley Stream Parkway, E-50 Malvern, PA 19355 Establishment Registration Number: 2240869
Location of Manufacturing Site:
Siemens AG Medical Solutions X-Ray Products Henkestrasse 127 DE-91052 Erlangen Establishment Registration Number: 3002808157
Manufacturer:
Siemens Shanghai Medical Equipment Ltd. 278 Zhou Zhu Road, Shanghai 201318, China Headquarters: Siemens AG Wittelsbacherplatz 2 D-80333 Munich 2, Germany Establishment Registration Number: 3003202425
2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway D-02 Malvern, PA 19355 Fax: (610) 448-1787 Phone: (610) 448 -3536
Email: patricia.d.jones@siemens.com
- Device Name and Classification: 3. Trade Name: Classification Name: Classification Panel:
Multix Fusion Stationary X-Ray System Radiology
{1}------------------------------------------------
Classification Regulation: Device Class: Product Code:
21 CFR §892.1680 Class II 90 KPR
-
- Legally Marketed Predicate Device Trade Name: Ysio 510(k) #: K0817227 Clearance Date: August 22, 2008 Classification Name: Solid State X-ray Imager (SSXI) Classification Panel: Radiology CFR Section: 21 CFR §892.1680 Device Class: Class II Product Code: 90 MQB
KIZISI3
- Legally Marketed Predicate Device Trade Name: Ysio 510(k) #: K0817227 Clearance Date: August 22, 2008 Classification Name: Solid State X-ray Imager (SSXI) Classification Panel: Radiology CFR Section: 21 CFR §892.1680 Device Class: Class II Product Code: 90 MQB
age 2 of H
5. Device Description:
The Multix Fusion is a modification of Siemens' Ysio X-ray system cleared under Premarket Notification K081722 on 08/22/2008. The Multix Fusion offers the following system configurations:
- An analog radiography system or; .
- A digital radiography system with a mobile (wired) and/or a portable . (wireless) flat panel detector or;
- An optional Multix DR-Upgrade for the analog version of the Multix Fusion . and other tables within the Multix and Vertix analog family.
This modification does not affect the general intended use of the device nor does it alter its fundamental scientific technology.
The Multix Fusion system consists of a radiologic table, x-ray generator, x-ray tube, flat panel detector (wired or wireless) and Bucky-wall stand. It will be marketed as an analog system, a digital radiographic system with a mobile (wired) and/or portable (wireless) flat panel detector or an optional Multix DR- Upgrade for the analog version of the Multix Fusion and other tables within the Multix and Vertix analog family.
The key components are a patient table, a ceiling mounted support and a Bucky wall stand which are available in different configurations. A semi-motorized movement in the ceiling-mounted X-ray tube is available. The optional Multix DR-Upgrade for the analog version of the Multix Fusion consists of a mobile (wired) and/or portable (wireless) flat panel detector and an imaging system. Components used to upgrade (Multix DR-Upgrade) the analog version of the Multix Fusion and other tables within the Multix and Vertix analog family are the same or similar components cleared in the Ysio 510(k) K0817227 on August 22, 2008.
The Multix Fusion may be combined into different configurations to provide specialized customer requirements.
Similar to the cleared Ysio stationary x-ray system, the Multix Fusion and the Mutix DR-Upgrade have the same or similar comparable components. The subject device the Multix Fusion does not affect the indication for use nor the intended use of the device nor does it alter its fundamental scientific technology.
{2}------------------------------------------------
K121513
Page 3 of 4
6. Indication for Use:
The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Multix Fusion system is not meant for mammography.
The Multix Fusion uses a mobile (wired) or portable (wireless) digital detector for generating diagnostic images by converting x-rays into electronic signals. The Multix Fusion is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
7. Substantial Equivalence:
The Multix Fusion is substantially equivalent to the commercially available Siemens Ysio radiographic x-ray system. The Ysio was described in premarket notification K081722 which received FDA Clearance on August 22, 2008. Ysio is indicated for tomographic procedures, The Multix Fusion is not indicated for tomographic procedures. The technical differences between the Multix Fusion and the Ysio are listed below:
| ComparableProperties | Subject DeviceMultix Fusion | Predicate DeviceYsio (K081722) | ComparisonResults |
|---|---|---|---|
| X-ray tube | 80 kW Two-focus | 80 kW Two-focus | Same |
| Flat panel detectorportable | 43cm x 36 cmType: 4336X | 43cm x 36 cmType: 3543pR | Same imagingtechnology |
| Digital Imaging system | Fluorospot CompactHigh Res DigitalImaging | Fluorospot CompactHigh Res DigitalImaging | SimilarFunctionality |
| Collimator | ACSS collimator(Automatic CassetteSize Sensing) | ACSS collimator(Automatic CassetteSize Sensing) | SimilarFunctionality |
| X-ray Generator | 80 kW65 kW55 kW | 80 kW65 kW | Similar |
| Conventionalfilm/screen systemsor CR cassettes | Film/Screen or CRCassettes | Film/Screen or CRCassettes | Same |
Table 1: Subject and Predicate Device Comparable Properties
8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The Multix Fusion does not have significant changes in materials, energy source, or technological characteristics when compared to the predicate device. The intended use and fundamental scientific technology are similar to the predicate device. The Multix Fusion is designed based on the Ysio radiography x-ray system. It uses the same or similar components cleared in the Ysio (e.g. tube, generator, ceilingmounted tube support, table, Bucky wall stand and imaging system). The Multix Fusion is similar to the predicated system with respect to technical characteristics,
{3}------------------------------------------------
K121513
Page 4 of 4
performance and intended use. The subject device the Multix Fusion does not affect the indication for use nor the intended use of the device nor does it alter its fundamental scientific technology
9. Non-clinical Testing:
Non-clinical testing was conducted for the Multix Fusion during product development. This includes verification and validation testing as well as phantom testing. The risk analysis was combation and risk controls implemented to mitigate identified hazards. The testing results supports that all the software specificating have met the acceptance criteria.
Testing for verification and validation of the device was found acceptable to support the claims of substantial eqivalence.
10. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Multix Fusion is continually monitored, and if an error occurs , the system the mank I asion is continually montored, and If an
displayed displayed.
Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processionals familiar will and performed. To mininize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
11. Conclusion as to Substantial Equivalence:
The Multix Fusion is intended for the same use as Ysio (with the exception of tomography). It uses the same or similar components cleared for the Ysio (e.g. tube, generator, ceiling-mounted tube support, table, Bucky wall stand and imaging system). It is Siemens opinion that the Multix Fusion is substantially equivalent to the cleared predicate device the Ysio radiographic x-ray system.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle symbol on the right and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular fashion around the left side. The eagle symbol is composed of three stylized lines that suggest the wings and body of an eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service .
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
AUG 10 2012
Ms. Patricia D. Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway, E-50 MALVERN PA 19355
Re: K121513
Trade/Device Name: Multix Fusion Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: July 26, 2012 Received: July 27, 2012
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice mource no real forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirements as bet form in the quar device as described in your Section 510(k) premarket witi anow you to begal finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you desire specifice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Special 510(k) Submission: Multix Fusion
Indications for Use Statement
510(k) Number (if known):
Device Name: Multix Fusion
Indications for Use:
The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Multix Fusion system is not meant for mammography.
The Multix Fusion uses a mobile (wired) or portable (wireless) digital detector for generating diagnostic images by converting x-rays into electronic signals. The Multix Fusion is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
hnd D'O'hm
Division Sian-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K121513
Page 1 of
Special 510(k) Multix Fusion
Siemens Medical Systems, Inc.
§ 892.1680 Stationary x-ray system.
(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.