(91 days)
Not Found
Unknown
The summary mentions "Automatic image analysis identifies a soft-tissue structure such as the prostate," which could potentially utilize AI/ML techniques for image recognition and tracking. However, it does not explicitly state the use of AI, ML, or related terms like deep learning or neural networks. Without further information, it's impossible to definitively confirm the presence of AI/ML.
No
The device aids in radiation therapy simulation, planning, and patient positioning, but it does not directly treat or cure a disease or condition. Its function is diagnostic imaging and guidance.
Yes
The 'Intended Use / Indications for Use' section states, "Clarity® provides 3D ultrasound and hybrid imaging of soft tissue anatomy to aid in radiation therapy simulation and planning, and to guide patient positioning prior to the delivery of treatment (Image Guided Radiation Therapy)." The 'Device Description' also notes it "integrates medical diagnostic ultrasound." These phrases explicitly indicate its use for diagnostic imaging purposes to inform subsequent therapeutic actions.
No
The device description clearly states that the Clarity® system integrates medical diagnostic ultrasound, real-time optical position tracking, and proprietary software. It includes hardware components such as an integrated ultrasound scanner, touch screen, computer system, optical tracking system, patient/couch position tracking tools, and ultrasound probes. Therefore, it is not a software-only medical device.
Based on the provided information, the Clarity® device is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs are used to examine specimens derived from the human body. The intended use and device description clearly state that Clarity® is used for imaging and tracking soft tissue anatomy within the patient's body using ultrasound. It does not analyze samples like blood, urine, or tissue that have been removed from the body.
- The purpose of Clarity® is for radiation therapy simulation, planning, and guidance. This is a therapeutic and imaging application, not a diagnostic test performed on a specimen.
- The device description focuses on imaging technology, optical tracking, and software for image analysis and guidance. These are not the typical components of an IVD.
In summary, Clarity® is a medical imaging and guidance system used in radiation therapy, not a device for performing diagnostic tests on in vitro specimens.
N/A
Intended Use / Indications for Use
Clarity® is indicated for use in external beam radiation therapy. It provides 3D ultrasound and hybrid imaging of soft tissue anatomy to aid in radiation therapy simulation and planning, and to guide patient positioning prior to the delivery of treatment (Image Guided Radiation Therapy).
When configured with an autoscan probe kit for transperineal ultrasound (TPUS) imaging, Clarity® may be used to continuously track and monitor the prostate and to accurately and precisely guide patient positioning during the delivery of treatment (Intrafractional Position Tracking and Monitoring).
When configured with a gating option, Clarity® may also interface with radiation delivery systems equipped with a compatible external gating control device. With this option, while in tracking and monitoring mode, Clarity® can signal the radiation delivery system to automatically impose a beam-hold when the tracked anatomy has exceeded pre-defined monitoring (tracking) limits, and signal again to release the tracked anatomy returns to a position within those limits (Exception gating has been shown to be compatible with radiation delivery systems equipped with Elekta's Response™ Gating Control System.
Product codes
IYE, IWB, KPQ
Device Description
The Clarity® system integrates medical diagnostic ultrasound, real-time optical position tracking and proprietary software to acquire and reconstruct 3D images of soft-tissue anatomy for use in external beam radiation therapy. Clarity® offers a non-invasive, non-ionizing means for accurate and precise localization of anatomical structures and patient positioning relative to the treatment isocenter.
The Clarity® system (Model 310C00) is configured around a mobile image acquisition station with an integrated ultrasound scanner, high-resolution touch screen, and high-performance computer system running the Clarity® software. It may be used at the patient's side in the CT-Sim room (Clarity® Sim) and the treatment room (Clarity® Guide) when equipped with a ceiling-mounted optical tracking system, patient/couch position tracking tools and, optionally, remote control and treatment monitoring equipment. With the gating option, the Clarity® Guide acquisition station may interface with radiation delivery systems equipped with a compatible external gating control device.
Each acquisition station is configured with up to three optically-tracked ultrasound probes: one or two hand-held probes for manual scanning and a motorized (autoscan) probe for automated scanning. The user can select the probe and scanning method that is most appropriate for the target anatomy and the patient's clinical presentation. The autoscan probe remains in contact with the patient for continuous imaging of the prostate and surrounding anatomy using specifically designed positioning apparatus for transperineal ultrasound (TPUS); it is operated from the acquisition station's remote control and monitoring equipment interface (touch-screen identical to that on the mobile acquisition station).
A multimodality imaging phantom is used to calibrate Clarity® to the room coordinate system and to verify system integrity for sub-millimeter target localization accuracy and precision within each room (daily and monthly QC).
A dedicated high-performance server and workstation computer system running the Clarity® software is connected to Clarity® acquisition stations through the hospital's local area network. The server houses the central database and web server, and provides for interoperability with other imaging and treatment planning/simulation systems via the DICOM 3/RT protocol. The workstation is used for multimodality image fusion and review, soft-tissue structure definition, approval of patient positioning references, setup of monitoring parameters, and review of treatment and QC data. Optionally, additional Clarity® workstations may be connected to the central Clarity® server.
The Clarity® software is designed to step the user through a radiation therapy workflow or "course" and QC procedures. Different courses are defined to help classify patients in the database and to present the user with reminders, default choices and configuration settings tailored to the target anatomy (e.g., prostate, bladder, liver, uterus & cervix, breast, head & neck). Such configurations include probe type, imaging (scan) presets, contouring and assisted segmentation tools, alert values for target misalignment, and prostate monitoring (tracking) parameters.
The typical use of the system for a radiation therapy course begins with the acquisition of a baseline 3D ultrasound (3DUS) scan with the patient in the planning position. The planning CT is imported, registered and fused with the 3DUS on the Clarity® workstation to verify the alignment of the target anatomy. The structures of interest are then defined and a baseline positioning reference including, if applicable, monitoring (prostate tracking) parameters are approved. Optionally, the 3DUS and related contours may be exported via DICOM to a third-party virtual simulator or treatment planning system.
Mentions image processing
Yes
Mentions AI, DNN, or ML
No
Input Imaging Modality
3D ultrasound, CT
Anatomical Site
Soft tissue anatomy, prostate, bladder, liver, uterus & cervix, breast, head & neck
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Clarity® has been developed and tested in compliance with regulatory guidance and recognized consensus safety standards. Software and system verification testing was conducted under typical and reasonably foreseeable use error and boundary conditions. Localization accuracy and precision specifications were verified with multimodality phantoms. Exception gating was validated with Elekta's Response™ Gating Control System under simulated treatment conditions. Formative evaluations and simulated use of the modified device with representative end-users were conducted in accordance with FDA guidance on human factors and usability engineering to assure the safe and effective performance of critical tasks. The test results from verification and validation activities demonstrate that Clarity® fulfills its design and risk management requirements, and is as safe and effective for its intended use as the predicate device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.5050 Medical charged-particle radiation therapy system.
(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002
October 8, 2014
Elekta Ltd. % Mr. George Papagiannis Regulatory Affairs 2050 de Bleury, Suite 200 Montreal, Quebec, H3A 2J5 CANADA
Re: K141855
Trade/Device Name: Clarity® Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE, IWB, KPQ Dated: July 10, 2014 Received: July 11, 2014
Dear Mr. Papagiannis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
Page 2—Ms. Yusko
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetw/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K141855
Device Name
Clarity®
Indications for Use (Describe)
Clarity® is indicated for use in external beam radiation therapy. It provides 3D ultrasound and hybrid imaging of soft tissue anatomy to aid in radiation therapy simulation and planning, and to guide patient positioning prior to the delivery of treatment (Image Guided Radiation Therapy).
When configured with an autoscan probe kit for transperineal ultrasound (TPUS) imaging, Clarity® may be used to continuously track and monitor the prostate and to accurately and precisely guide patient positioning during the delivery of treatment (Intrafractional Position Tracking and Monitoring).
When configured with a gating option, Clarity® may also interface with radiation delivery systems equipped with a compatible external gating control device. With this option, while in tracking and monitoring mode, Clarity® can signal the radiation delivery system to automatically impose a beam-hold when the tracked anatomy has exceeded pre-defined monitoring (tracking) limits, and signal again to release the tracked anatomy returns to a position within those limits (Exception gating has been shown to be compatible with radiation delivery systems equipped with Elekta's Response™ Gating Control System.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image shows the word "ELEKTA" in a bold, sans-serif font. To the left of the word is a circle with a dashed line going through it. The dashed line consists of five dots, starting from the upper left of the circle and going down to the center.
510(k) Summary
October 07, 2014
Trade/Device Name: | Clarity® |
---|---|
Common Name (GMDN): | Patient positioning system, ultrasound |
Regulation/Classification: | Medical charged-particle radiation therapy system |
(21 CFR 892.5050, Product Code IYE) | |
Radionuclide radiation therapy system | |
(21 CFR 892.5750, Product Code IWB) | |
Radiation therapy simulation system | |
(21 CFR 892.5840, Product Code KPQ) | |
Regulatory Class: | Class II |
Review Panel: | Radiology |
Submitter/Manufacturer: | Elekta Ltd. |
Establishment Registration No: | 3004747535 |
Address: | 2050 Bleury, Suite 200 |
Montreal, Quebec H3A 2J5, Canada | |
Tel: (+1) 514-840-9600 Fax: (+1) 514-840-9666 | |
Contact: | Tony Falco, PhD |
Introduction
This 510(k) Summary has been prepared in accordance with 21 CFR 807.92. It summarizes device safety and effectiveness information to provide an understanding of the basis for a determination of substantial equivalence.
Predicate Device Information
Clarity® (K121663, Dec 05, 2012; Product Codes: IYE, IWB, KPQ)
Intended Use / Indications for Use
Clarity® is indicated for use in external beam radiation therapy. It provides 3D ultrasound and hybrid imaging of soft tissue anatomy to aid in radiation therapy simulation and planning, and to guide patient positioning prior to the delivery of treatment (Image Guided Radiation Therapy).
When configured with an autoscan probe kit for transperineal ultrasound (TPUS) imaging, Clarity® may be used to continuously track and monitor the motion of the prostate and to accurately and precisely guide patient positioning during the delivery of treatment (Intrafractional Position Tracking and Monitoring).
When configured with a gating option, Clarity® may also interface with radiation delivery systems equipped with a compatible external gating control device. With this option, while in tracking and monitoring mode, Clarity® can signal the radiation delivery system to automatically impose a beamhold when the position of the tracked anatomy has exceeded pre-defined monitoring (tracking) limits, and signal again to release the beam-hold when the tracked anatomy returns to a position within those limits (Exception Gating). Exception gating has been shown to be compatible with radiation delivery systems equipped with Elekta's Response™ Gating Control System.
4
Device Description
The Clarity® system integrates medical diagnostic ultrasound, real-time optical position tracking and proprietary software to acquire and reconstruct 3D images of soft-tissue anatomy for use in external beam radiation therapy. Clarity® offers a non-invasive, non-ionizing means for accurate and precise localization of anatomical structures and patient positioning relative to the treatment isocenter.
The Clarity® system (Model 310C00) is configured around a mobile image acquisition station with an integrated ultrasound scanner, high-resolution touch screen, and high-performance computer system running the Clarity® software. It may be used at the patient's side in the CT-Sim room (Clarity® Sim) and the treatment room (Clarity® Guide) when equipped with a ceiling-mounted optical tracking system, patient/couch position tracking tools and, optionally, remote control and treatment monitoring equipment. With the gating option, the Clarity® Guide acquisition station may interface with radiation delivery systems equipped with a compatible external gating control device.
Each acquisition station is configured with up to three optically-tracked ultrasound probes: one or two hand-held probes for manual scanning and a motorized (autoscan) probe for automated scanning. The user can select the probe and scanning method that is most appropriate for the target anatomy and the patient's clinical presentation. The autoscan probe remains in contact with the patient for continuous imaging of the prostate and surrounding anatomy using specifically designed positioning apparatus for transperineal ultrasound (TPUS); it is operated from the acquisition station's remote control and monitoring equipment interface (touch-screen identical to that on the mobile acquisition station).
A multimodality imaging phantom is used to calibrate Clarity® to the room coordinate system and to verify system integrity for sub-millimeter target localization accuracy and precision within each room (daily and monthly QC).
A dedicated high-performance server and workstation computer system running the Clarity® software is connected to Clarity® acquisition stations through the hospital's local area network. The server houses the central database and web server, and provides for interoperability with other imaging and treatment planning/simulation systems via the DICOM 3/RT protocol. The workstation is used for multimodality image fusion and review, soft-tissue structure definition, approval of patient positioning references, setup of monitoring parameters, and review of treatment and QC data. Optionally, additional Clarity® workstations may be connected to the central Clarity® server.
The Clarity® software is designed to step the user through a radiation therapy workflow or "course" and QC procedures. Different courses are defined to help classify patients in the database and to present the user with reminders, default choices and configuration settings tailored to the target anatomy (e.g., prostate, bladder, liver, uterus & cervix, breast, head & neck). Such configurations include probe type, imaging (scan) presets, contouring and assisted segmentation tools, alert values for target misalignment, and prostate monitoring (tracking) parameters.
The typical use of the system for a radiation therapy course begins with the acquisition of a baseline 3D ultrasound (3DUS) scan with the patient in the planning position. The planning CT is imported, registered and fused with the 3DUS on the Clarity® workstation to verify the alignment of the target anatomy. The structures of interest are then defined and a baseline positioning reference including, if applicable, monitoring (prostate tracking) parameters are approved. Optionally, the 3DUS and related contours may be exported via DICOM to a third-party virtual simulator or treatment planning system.
5
To assist with patient positioning prior to each treatment session, a new 3DUS scan is acquired and used to determine target displacement relative to the baseline planning-day position. Optical tracking of couch position allows for accurate and precise patient repositioning relative to the treatment isocenter (Image Guided Radiation Therapy).
Automatic image analysis identifies a soft-tissue structure such as the prostate in successive transperineal 3DUS images, which are acquired continuously during treatment, and allows Clarity® to track its motion and assist with patient repositioning (Intrafractional Position Tracking and Monitoring). When configured with the gating option, while in tracking and monitoring mode, Clarity® can signal the radiation delivery system to automatically impose a beam-hold when the tracked structure position has exceeded pre-defined monitoring (tracking) limits, and signal again to release the beam-hold when the structure returns to a position within those limits (Exception Gating).
Clarity® may optionally be configured to send calculated couch shifts for patient repositioning to the operator at the couch control user interface using the MOSAIQ® Workflow Manager.
A web-based interface is available for remote review and approval of positioning references and other treatment parameters, and review of completed treatment session and QC procedure data.
Comparison with Predicate Device
The current release of Clarity®, which includes usability improvements and expanded indications for exception gating, is substantially equivalent to the predicate device. The differences in labeling and technological characteristics do not raise new questions of safety and effectiveness for the intended use. A side-by-side comparison of key device characteristics is presented in the following tables:
Device Characteristic | Predicate Device (K121663) | Current Device |
---|---|---|
Equipment | ||
Medical imaging modality | ||
(acquisition station) | Mobile console with integrated | |
diagnostic ultrasound scanner; | ||
Broadband curved (convex), linear, | ||
and autoscan probes | No Significant Change | |
(Upgraded computer system | ||
specifications and improved | ||
mobile console aesthetics) | ||
Optical tracking system | Passive infrared position sensor | |
powered from acquisition station | No Significant Change | |
(Independent power supply) | ||
Patient/couch localization | Rigid tools fitted with infrared light | |
reflecting markers | No Significant Change | |
System calibration and | ||
Quality Control | Proprietary multimodality phantom | No Significant Change |
Remote control interface | Touch-screen duplicating main | |
console via KVM device for image | ||
acquisition with autoscan probe | No Significant Change | |
(Upgraded KVM device | ||
specifications for gating option) | ||
Basic safety and essential | ||
performance standards | IEC 60601-1:2005 | |
IEC 60601-2-37:2007 (Ultrasound) | ||
IEC 60601-1-2:2007 (EMC) | ||
IEC 60601-1-6:2010 (Usability) | AAMI/ANSI ES60601-1:2012 | |
IEC 60601-2-37:2007 | ||
IEC 60601-1-2:2007 | ||
IEC 60601-1-6:2010 |
6
Device Characteristic | Predicate Device (K121663) | Current Device |
---|---|---|
Software / Functionality | ||
System calibration to room | ||
coordinate system and | ||
Quality Control | Manual characterization of phantom | |
from high-resolution CT scan; | ||
Software-assisted calibration of | ||
optical sensor tilt detector, room's | ||
coordinate system and probe; | ||
Software-assisted daily QC process; | ||
Software-configurable mobile | ||
acquisition station for CT/Tx room | No Significant Change | |
(Simplified system calibration | ||
and daily QC process; auto- | ||
configuration of mobile | ||
acquisition station based on | ||
optical tracking system ID) | ||
Ultrasound image | ||
acquisition for daily | ||
target localization | Ultrasound imaging (scan) presets; | |
Image adjustment controls; | ||
Display of previous scans for | ||
consistent image acquisition; | ||
Assisted soft-tissue segmentation; | ||
3DUS-to-3DUS comparison | No Significant Change | |
(Added "Live Guidance" for | ||
consistent probe positioning in | ||
daily prostate scans) | ||
Couch localization and | ||
control interface | Option to send couch shifts for patient | |
repositioning to MOSAIQ® | No Significant Change | |
Intrafractional prostate | ||
motion management | Tracking & monitoring with autoscan | |
probe kit for transperineal ultrasound | No Significant Change | |
(Improved user interface and | ||
tracking indicators) | ||
Exception gating during | ||
prostate monitoring | No | Optional with compatible |
radiation delivery systems | ||
Workstation & Server | ||
functionality | System-wide database; Automated | |
multimodality image registration and | ||
fusion; workspaces for contouring and | ||
positioning reference definition; | ||
DICOM RT import/export, including | ||
beam & MLC data; positioning and | ||
monitoring parameters definition | No Significant Change | |
(Compatible with Windows 7) | ||
Web-based interface | Remote review of patient data and | |
positioning reference approval | No Significant Change | |
(Improved monitoring reports and | ||
added review of QA/QC data) |
Summary of Clinical & Non-Clinical Testing
Clarity® has been developed and tested in compliance with regulatory guidance and recognized consensus safety standards. Software and system verification testing was conducted under typical and reasonably foreseeable use error and boundary conditions. Localization accuracy and precision specifications were verified with multimodality phantoms. Exception gating was validated with Elekta's Response™ Gating Control System under simulated treatment conditions. Formative evaluations and simulated use of the modified device with representative end-users were conducted in accordance with FDA guidance on human factors and usability engineering to assure the safe and effective performance of critical tasks. The test results from verification and validation activities demonstrate that Clarity® fulfills its design and risk management requirements, and is as safe and effective for its intended use as the predicate device.