K Number
K141774
Device Name
ZOLL X SERIES
Date Cleared
2014-11-19

(140 days)

Product Code
Regulation Number
870.5310
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The X Series is intended for use by trained medical personnel who are familiar with basic monitoring, vital sign assessment, emergency cardiac care, and the use of the X Series, The X Series is also intended for use by (or on the order of) physicians at the scene of an emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other similar areas of a hospital. The usage may be in an ambulance or at the scene of an emergency. It is also intended to be used during the transport of patients. The X Series will be used primarily on patients experiencing symptoms of cardiac arrest or in post trauma situation. It may also be used whenever it is required to monitor any of those functions that are included (as options) in the device. The X Series unit can be used on pediatric patients (as described in the following table) and on adult patients (21 years of age or older) with and without heart dysfunction.

Device Description

The predicate X Series is a lightweight, portable device designed to be used by trained medical personnel who are familiar with vital signs monitoring and emergency cardiac care. As in its previously cleared configuration, the modified X Series combines the functions of a Semi-automatic (AED)/ manual defibrillator, external transcutaneous pacer and patient monitors (including ECG, SpO2, SpCO, SpMet, CO2, NIBP, IBP, temperature and respiration monitoring).

With the current application, we are proposing to revise the X Series communication processor software to create a web page with a graphical trend format and countdown timer for ventilations that consolidates the existing SBP (systolic blood pressure from the NIBP monitoring), SpO2, EtCO2 data and the timer on the display of the X Series. The access to this web page will be marketed as the Traumatic Brain Injury (TBI) Dashboard. This TBI Dashboard assists the operator who is ventilating a patient to follow recommendations in the TBI guidelines. Through an encrypted and password protected communications protocol, a suitable separate device with touchscreen, web browser capability and Bluetooth communications can display the TBI Dashboard.

As described above, the proposed software change impacts only the communication aspects of the X Series device. Therapeutic and monitoring capabilities of the X Series device as cleared under K133269 have remained unchanged and are not impacted by the proposed change. The X Series therapeutic and monitoring functionality is entirely controlled by access directly through the X Series device, itself, and not through the TBI Dashboard. TBI Dashboard is intended to supplement and not replace any part of the device monitoring.

AI/ML Overview

The provided text describes the ZOLL X Series, a multi-function medical device that includes capabilities such as defibrillation, ECG monitoring, CPR monitoring, pacing, and various physiological monitoring functions (NIBP, temperature, SpO2, respiration, CO2, invasive pressure). The 510(k) submission (K141774) is specifically for a revision to the X Series communication processor software to add a "TBI Dashboard" functionality.

However, the document does not explicitly state specific acceptance criteria in terms of numerical performance targets (e.g., sensitivity, specificity, accuracy, error rates) for the device's various functions, nor does it provide detailed descriptions of studies proving these criteria are met. Instead, the submission relies on demonstrating substantial equivalence to predicate devices.

The "Performance Testing" section states: "Extensive performance testing in the form of the software verification and system level validation ensures that the ZOLL X Series performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications. Safety testing assures that the device complies with applicable sections of recognized industry and safety standards." This is a general statement about the type of testing performed, not a detailed report of the results against specific criteria.

The 510(k) summary focuses on the TBI Dashboard functionality, which is a software revision. For this new feature, the document asserts substantial equivalence to two predicate devices:

  • ZOLL RescueNet Link (K111296) for collecting and displaying data in a web page format.
  • L770-CPR Resuscitation Timer (K062080) for ventilation rate timer functionality.

Given this, I will infer what the "acceptance criteria" likely are based on the substantial equivalence claims and the general nature of medical device clearances. The "reported device performance" will be based on the general statements made about meeting requirements.

Here's the breakdown of the requested information:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Inferred)Reported Device Performance
For all existing X Series functions (e.g., Defibrillation, ECG, NIBP, SpO2, etc.): Maintains performance as previously cleared predicate (ZOLL X Series K133269) and meets functional requirements and performance specifications."Features and functions cleared with the ZOLL X Series (K133269) device has remained unchanged ... ensures that the ZOLL X Series performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications."
For TBI Dashboard (Data Display/Web Console): Functionality is equivalent to ZOLL RescueNet Link (K111296) for displaying physiological data and user-entered data via web page."The proposed TBI Dashboard functionality and RescueNet Link (K111296) both collect and display data... in the form of a web page that can be accessed by the authorized users through standard web browser technology." "TBI Dashboard is intended to supplement and not replace any part of the device monitoring."
For TBI Dashboard (Ventilation Rate Timer): Functionality is equivalent to L770-CPR Resuscitation Timer (K062080) for providing a visual aid for ventilation rate."Ventilation rate timer functionality in the TBI Dashboard is substantially equivalent to the the L770-CPR Resuscitation Timer (K062080). Both the devices are visual aids for a rescuer trained in the protocol they support..."
Safety: Device complies with applicable recognized industry and safety standards."Safety testing assures that the device complies with applicable sections of recognized industry and safety standards."
Software Verification & Validation: Software operates as intended and meets design specifications."Extensive performance testing in the form of the software verification and system level validation ensures that the ZOLL X Series performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications."

2. Sample size used for the test set and the data provenance

The document does not provide details on specific sample sizes for test sets or data provenance (e.g., country of origin, retrospective/prospective) for the software verification and system-level validation. The nature of the submission (a software update for a display feature) suggests that these tests would primarily involve engineering and software validation rather than clinical data from a "test set" in the traditional sense of patient data. For the original X Series functions, such data would have been part of the K133269 submission, but not detailed here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable/provided. The submission relies on demonstrating substantial equivalence to predicate devices and general software verification/system validation, rather than a clinical study requiring expert-established ground truth for a novel diagnostic or treatment outcome.

4. Adjudication method for the test set

Not applicable/provided for the reasons stated above.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done or reported. This submission is for a device update adding a dashboard, not a novel diagnostic AI algorithm. The TBI Dashboard is described as a "supplement and not replace any part of the device monitoring," and the ventilation timer as a "visual aid," implying it supports human users rather than being an AI system that improves human "reading" performance in a diagnostic context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a patient monitor/defibrillator with software features, not a standalone algorithm. The TBI Dashboard is a display feature for existing data.

7. The type of ground truth used

For the software revision adding the TBI Dashboard, the "ground truth" would likely be the correct functioning of the software according to its design specifications, accuracy in displaying data from the device sensors, and adherence to accepted communication protocols for web page display. This would be established through software unit testing, integration testing, and system validation against known correct values or predicate device behavior. No external expert consensus, pathology, or outcomes data is mentioned for this particular update.

8. The sample size for the training set

Not applicable. The X Series and its TBI Dashboard are not described as machine learning or AI-based systems that require a "training set" for model development.

9. How the ground truth for the training set was established

Not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 19, 2014

Zoll Medical Corporation Tanmay Shukla Sr. Regulatory Affairs Specialist 269 Mill Road Chelmsford, Massachusetts 01824-4105

Re: K141774 Trade/Device Name: Zoll X series Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ, LIX, LDD, DRT, DRO, DXN, DSK, CCK, DQA Dated: October 20, 2014 Received: October 21, 2014

Dear Tanmay Shukla,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Section 4 - Indications for Use

510(k) Number (if known):

Device Name: X Series

Intended Use:

The X Series is intended for use by trained medical personnel who are familiar with basic monitoring, vital sign assessment, emergency cardiac care, and the use of the X Series, The X Series is also intended for use by (or on the order of) physicians at the scene of an emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other similar areas of a hospital. The usage may be in an ambulance or at the scene of an emergency. It is also intended to be used during the transport of patients. The X Series will be used primarily on patients experiencing symptoms of cardiac arrest or in post trauma situation. It may also be used whenever it is required to monitor any of those functions that are included (as options) in the device. The X Series unit can be used on pediatric patients (as described in the following table) and on adult patients (21 years of age or older) with and without heart dysfunction.

Pediatric Patient SubpopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age.
Infant1 month to 2 years of age.
Child2 to 12 years of age.
Adolescent12 to 21 years of age.

When the pediatric patient is less than 8 years of age or weighs less than 55 lbs. (25 kg.), use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine the patient's exact age or weight.

Manual Defibrillation

Use of the X Series in the manual mode for external and internal defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • · Unconsciousness.
  • · Absence of breathing.
  • · Absence of pulse.

Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.

The unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. Qualified medical personnel must decide when synchronized cardioversion is appropriate.

The patient population will range from newborn (neonate) to adult.

Semiautomatic Operation (AED)

X Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.

They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.

Use of the X Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • · Unconsciousness.
  • · Absence of breathing.
  • Absence of pulse.

When the patient is less than 8 years of age or weighs less that 55 lbs. (25 Kg), you must use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine patient's exact age or weight.

ECG Monitoring

The X Series is intended for use to monitor and/or record 3-. 5-. or 12-lead ECG waveform and heart rate, and to alarm when heart rate is above or below limits set by the operator. The patient population will range from newborn (neonate) to adult, with and without heart dysfunction.

CPR Monitoring

The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a minimum compression depth of at least 1.5 (3.8 cm) or 2.0 inches (5.0 cm), depending on the configuration, for adult patients.

{4}------------------------------------------------

External Transcutaneous Pacing

This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include:

  • · Resuscitation from standstill or bradycardia of any etiology:
  • · As a standby when standstill or bradycardia might be expected:
  • · Suppression of tachycardia.
  • · Pediatric pacing.

Non-Invasive Blood Pressure Monitoring

The X Series is intended for use to make non-invasive measurements of arterial pressure and heart rate, and to alarm if either parameter is outside of the limits set by the user. Measurements are made using an inflatable cuff on the patient's arm or leg. The patient population will range from newborn (neonate) to adult.

Temperature Monitoring

The X Series is intended for use to make continuous temperature measurements of rectal, esophageal, or surface temperatures, and to alarm if the temperature is outside of the limits set by the user. The patient population will range from newborn (neonate) to adult.

SpO2 Monitoring

The X Series pulse CO-oximeter, with Masimo Rainbow SET technology and the Rainbow series of sensors, is intended for use for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and/or methemoglobin saturation (SpMet). The pulse CO-oximeter and accessories are indicated for use on adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused, in hospitals, hospital-tvpe facilities, or in mobile environments.

Respiration Monitoring

The X Series is intended for use to continuously monitor respiration rate and to alarm if the rate falls outside of the range set by the operator. Because the measurement method actually measures respiratory effort, apnea episodes with continued respiratory effort (such as obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The patient population will range from newborn (neonate) to adult.

CO2 Monitoring

The X Series is intended for use to make continuous noninvasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath rate. The patient population will range from newborn (neonate) to adult.

{5}------------------------------------------------

Invasive Pressure Monitoring

The X Series is intended for use to display and make continuous invasive pressure measurements from any compatible pressure transducer. The primary intended uses are arterial blood pressure, central venous pressure and intracranial pressure monitoring. Any contraindications of the particular transducer selected by the user shall apply. The patient population will range from newborn (neonate) to adult.

12-Lead Analysis

The 12-lead ECG Analysis is intended for use in acquiring, analyzing and reporting ECG data, and to provide interpretation of the data for consideration by caregivers. The interpretations of ECG data offered by the device are only significant when used in conjunction with caregiver overread as well as consideration of all other relevant patient data. The 12-lead ECG Analysis is intended for use on adults (> 18 years of age).

Web Console

X Series Web Console functionality allows medical personnel to view physiological data available on a connected X Series device. Physiological data may include ECG, noninvasive blood pressure, SpO2/SpCO/SpMet, end-tidal carbon dioxide, respiration rate, invasive blood pressure, temperature, and Heart Rate. For each patient connected to an X Series, the X Series collects patient data, consolidates the data on an embedded web page that can be requested, through an Ethernet connection, by a web browser on a PC for display.

TBI Dashboard

X Series TBI Dashboard functionality creates an embedded web page that consolidates concurrently displayed data on the X Series for SBP (systolic blood pressure), SpO2, EtCO2 data and timer into graphical trend formats and a countdown timer for ventilations. The web page can be requested, through an Bluetooth connection, using a standard web browser on a viewing device for display.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "ZOLL" in a sans-serif font. The letters are a light blue color. A small registered trademark symbol is located to the right of the second "L" in the word.

ZOLL Medical Corporation Worldwide Headquarters 269 Mill Road Chelmsford, MA 01824 U.S.A.

978-421-9655 978-421-0025 Main Fax

510(k) Summarv:

Applicant's Name and Address:ZOLL Medical Corporation269 Mill RoadChelmsford, MA 01824
Application Correspondent:Tanmay Shukla978-421-9171

June 30, 2014 Date Summary Prepared:

Classification: Class III

Device Name

Product Code

Automated External Defibrillators (MKJ) Cardiopulmonary Resuscitation Aid (LIX) Low-Energy - Defibrillators (LDD) Cardiac Monitors - including Cardiotachometer and Rate Alarms (DRT) External Transcutaneous Cardiac Non-Invasive Pacemaker (DRO) Noninvasive Blood Pressure Measurement System (DXN) Blood Pressure Computer (DSK) Carbon Dioxide Gas Analyzer (CCK) Oximeter (DQA)

Predicate DevicesZOLL X Series (K133269)
ZOLL RescueNet Link (K111296)
L770-CPR Resuscitation Timer (K062080)

ZOLL X Series

{7}------------------------------------------------

Description:

The predicate X Series is a lightweight, portable device designed to be used by trained medical personnel who are familiar with vital signs monitoring and emergency cardiac care. As in its previously cleared configuration, the modified X Series combines the functions of a Semi-automatic (AED)/ manual defibrillator, external transcutaneous pacer and patient monitors (including ECG, SpO2, SpCO, SpMet, CO2, NIBP, IBP, temperature and respiration monitoring).

With the current application, we are proposing to revise the X Series communication processor software to create a web page with a graphical trend format and countdown timer for ventilations that consolidates the existing SBP (systolic blood pressure from the NIBP monitoring), SpO2, EtCO2 data and the timer on the display of the X Series. The access to this web page will be marketed as the Traumatic Brain Injury (TBI) Dashboard. This TBI Dashboard assists the operator who is ventilating a patient to follow recommendations in the TBI guidelines. Through an encrypted and password protected communications protocol, a suitable separate device with touchscreen, web browser capability and Bluetooth communications can display the TBI Dashboard.

As described above, the proposed software change impacts only the communication aspects of the X Series device. Therapeutic and monitoring capabilities of the X Series device as cleared under K133269 have remained unchanged and are not impacted by the proposed change. The X Series therapeutic and monitoring functionality is entirely controlled by access directly through the X Series device, itself, and not through the TBI Dashboard. TBI Dashboard is intended to supplement and not replace any part of the device monitoring.

Indications for Use:

The X Series is intended for use by trained medical personnel who are familiar with basic monitoring, vital sign assessment, emergency cardiac care, and the use of the X Series. The X Series is also intended for use by (or on the order of) physicians at the scene of an emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other similar areas of a hospital. The usage may be in an ambulance or at the scene of an emergency. It is also intended to be used during the transport of patients. The X Series will be used primarily on patients experiencing symptoms of cardiac arrest or in post trauma situation. It may also be used whenever it is required to monitor any of those functions that are included (as options) in the device. The X Series unit can be used on pediatric patients (as described in the following table) and on adult patients (21 years of age or older) with and without heart dysfunction.

Pediatric Patient SubpopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age.
Infant1 month to 2 years of age.
Child2 to 12 years of age.
Adolescent12 to 21 years of age.

{8}------------------------------------------------

Page 3 of 6 When the pediatric patient is less than 8 years of age or weighs less than 55 lbs. (25 kg.), use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine the patient's exact age or weight.

Manual Defibrillation

Use of the X Series in the manual mode for external and internal defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • · Unconsciousness.
  • · Absence of breathing.
  • Absence of pulse.

This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.

The unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. Qualified medical personnel must decide when synchronized cardioversion is appropriate.

The patient population will range from newborn (neonate) to adult.

Semiautomatic Operation (AED)

X Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.

They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.

Use of the X Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • · Unconsciousness.
  • · Absence of breathing.
  • · Absence of pulse.

When the patient is less than 8 years of age or weighs less that 55 lbs. (25 Kg), you must use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine patient's exact age or weight.

K141774

{9}------------------------------------------------

ECG Monitoring

The X Series is intended for use to monitor and/or record 3-, 5-, or 12-lead ECG waveform and heart rate, and to alarm when heart rate is above or below limits set by the operator. The patient population will range from newborn (neonate) to adult, with and without heart dysfunction.

CPR Monitoring

The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a minimum compression depth of at least 1.5 (3.8 cm) or 2.0 inches (5.0 cm), depending on the configuration, for adult patients.

External Transcutaneous Pacing

This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include:

  • . Resuscitation from standstill or bradycardia of any etiology:
  • As a standby when standstill or bradycardia might be expected: ●
  • Suppression of tachycardia.
  • Pediatric pacing. ●

Non-Invasive Blood Pressure Monitoring

The X Series is intended for use to make non-invasive measurements of arterial pressure and heart rate, and to alarm if either parameter is outside of the limits set by the user. Measurements are made using an inflatable cuff on the patient's arm or leg. The patient population will range from newborn (neonate) to adult.

Temperature Monitoring

The X Series is intended for use to make continuous temperature measurements of rectal, esophageal, or surface temperatures, and to alarm if the temperature is outside of the limits set by the user. The patient population will range from newborn (neonate) to adult.

SpO2 Monitoring

The X Series pulse CO-oximeter, with Masimo Rainbow SET technology and the Rainbow series of sensors, is intended for use for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and/or methemoglobin saturation (SpMet). The pulse CO-oximeter and accessories are indicated for use on adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused, in hospitals, hospital-type facilities, or in mobile environments.

{10}------------------------------------------------

Respiration Monitoring

The X Series is intended for use to continuously monitor respiration rate and to alarm if the rate falls outside of the range set by the operator. Because the measurement method actually measures respiratory effort, apnea episodes with continued respiratory effort (such as obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The patient population will range from newborn (neonate) to adult.

CO2 Monitoring

The X Series is intended for use to make continuous noninvasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and breath rate. The patient population will range from newborn (neonate) to adult.

Invasive Pressure Monitoring

The X Series is intended for use to display and make continuous invasive pressure measurements from any compatible pressure transducer. The primary intended uses are arterial blood pressure, central venous pressure and intracranial pressure monitoring. Any contraindications of the particular transducer selected by the user shall apply. The patient population will range from newborn (neonate) to adult.

12-Lead Analysis

The 12-lead ECG Analysis is intended for use in acquiring, analyzing and reporting ECG data, and to provide interpretation of the data for consideration by caregivers. The interpretations of ECG data offered by the device are only significant when used in conjunction with caregiver overread as well as consideration of all other relevant patient data. The 12-lead ECG Analysis is intended for use on adults (> 18 years of age).

Web Console

X Series Web Console functionality allows medical personnel to view physiological data available on a connected X Series device. Physiological data may include ECG, noninvasive blood pressure, SpO2/SpCO/SpMet, end-tidal carbon dioxide, respiration rate, invasive blood pressure, temperature, and Heart Rate. For each patient connected to an X Series, the X Series collects patient data, consolidates the data on an embedded web page that can be requested, through an Ethernet connection, by a web browser on a PC for display.

TBI Dashboard

X Series TBI Dashboard functionality creates an embedded web page that consolidates concurrently displayed data on the X Series for SBP (systolic blood pressure). SpO2, EtCO2 data and timer into graphical trend formats and a countdown timer for ventilations. The web page can be requested, through an Bluetooth connection, using a standard web browser on a viewing device for display.

{11}------------------------------------------------

Substantial Equivalence - Non-Clinical Evidence:

Features and functions cleared with the ZOLL X Series (K133269) device has remained unchanged in the proposed version of the device. TBI Dashboard functionality to the X Series device, added through a software revision, is substantially equivalent to the ZOLL RescueNet Link (K111296) and L770-CPR Resuscitation Timer (K062080). Safety, efficacy and substantial equivalence was shown through software verification and system level validation.

Substantial Equivalence - Clinical Evidence:

N/A - Clinical evidence was not necessary to show substantial equivalence

Comparison of Technological Characteristics

Features and functions cleared with the ZOLL X Series (K133269) have remained unchanged in the proposed version of the device. The proposed TBI Dashboard functionality and RescueNet Link (K111296) both collect and display data, that is entered by the user (caregiver) or captured from the device(s) to which they are connected, in the form of a web page that can be accessed by the authorized users through standard web browser technology. As with the predicate RescueNet link, TBI Dashboard is intended to supplement and not replace any part of the device monitoring.

Ventilation rate timer functionality in the TBI Dashboard is substantially equivalent to the the L770-CPR Resuscitation Timer (K062080). Both the devices are visual aids for a rescuer trained in the protocol they support; which can automatically set the target ventilation rate to the guideline-recommended rate based upon the selections made at initiation.

Performance Testing:

Extensive performance testing in the form of the software verification and system level validation ensures that the ZOLL X Series performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications. Safety testing assures that the device complies with applicable sections of recognized industry and safety standards.

Conclusion

The information provided in this 510(k) demonstrates that the ZOLL X Series' features and functions are substantially equivalent to those of the indicated commercially distributed devices with regard to performance, safety and effectiveness.

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.