(206 days)
The Illuminated Mediastinal Access Port is intended to aid the surgeon in direct visualization of the mediastinum and facilitate the introduction and removal of surgical instruments during surgical procedures.
The Illuminated Mediastinal Access Port is specifically indicated for use in the anterior, superior and middle mediastinum.
The Illuminated Mediastinal Access Port is a trocar/cannula system used to provide mediastinal surgical access. The device includes a blunt trocar for dissecting tissue planes. The open cannula provides for direct visualization of the surgical space and includes small notches at the rim to anchor sutures.
The base of the cannula has fiber optic bundles incorporated that may be connected to a fiber optic light cable and an endoscopic light source. This enhances lighting and visualization within the access device. The device does not include a light source.
The trocar/cannula access device also includes a handle for connection to a support arm. This allows the mediastinoscope to be held in position so that the surgeon can operate using both hands, without holding or re-positioning the mediastinoscope, and to provide stability for the mediastinoscope during operation.
The entire device is disposable, single use and is provided pre-sterilized.
This document is a 510(k) summary for the Illuminated Mediastinal Access Port, which is a medical device intended to aid surgeons in visualizing the mediastinum and facilitating surgical instrument introduction and removal. The document focuses on demonstrating substantial equivalence to a predicate device, not on presenting a detailed study proving the device meets an explicit set of acceptance criteria.
Therefore, many of the requested details about acceptance criteria, study design, and ground truth are not present in the provided text. The document primarily describes various performance tests conducted to ensure the device's safety and functionality in comparison to a predicate device.
Here's an attempt to extract relevant information and note what is missing based on your request:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state a table of acceptance criteria with corresponding performance metrics. Instead, it describes general performance tests and states that the device was ensured to meet certain conditions or functions comparably to the predicate.
| Acceptance Criteria (Inferred from testing) | Reported Device Performance (Summary) |
|---|---|
| No major deformation under mechanical load | No major deformation under worst-case mechanical load conditions |
| No breakage of device parts | No breakage of other parts of the device (specifically fiber optic bundles) |
| Fiber optic bundles don't pull apart | Fiber optic bundles don't pull apart under normal conditions of use |
| Illumination quality verified | Illumination quality verified following mechanical tests and simulated use testing |
| Biocompatibility | Studies performed in accordance with ISO 10993-1:2009 for a limited contact duration device |
| Sterility Assurance Level | Sterilized to an SAL of 1x10^-6 using ethylene oxide (validated EO cycle) |
| Sterile packaging integrity | Maintained a sterile barrier over time in shelf life testing |
| Product functionality maintained | Functionality maintained in shelf life testing |
| Withstand distribution conditions | Adequately withstood ISTA 2A distribution standards for packaging |
| Appropriate incision size | Established appropriate incision size through cadaver testing |
| Forces on tissue during insertion/extraction | Evaluated forces during insertion/extraction in cadaver testing, demonstrated device does not break or is not damaged |
| Safe light output | Spectral irradiance was "considerably less" than standard light source, indicating safe use |
2. Sample size used for the test set and the data provenance
- Sample Size: Not explicitly stated for any of the individual tests. The document refers to "worst case mechanical load conditions," "normal conditions of use," and "cadaver models," but without specific numbers.
- Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be laboratory-based ("bench" testing), cadaver testing, and potentially internal studies by the manufacturer or a contracted lab. This would typically be considered prospective for the device being tested, but the setting (lab/cadaver) is not a clinical one.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This information is not provided. The term "ground truth" as typically used in AI/diagnostic device studies (e.g., confirmed disease status) is not directly applicable here as this is a surgical access device, not a diagnostic one.
- For assessments like "appropriate incision size" and ability of surgeons to grip the handle, the document mentions "surgeons" in the cadaver models, but does not specify their number, qualifications, or how their observations were formalized into a "ground truth" or acceptance criterion.
4. Adjudication method for the test set
- Not applicable as this is not a diagnostic study requiring adjudication of findings. Performance tests were conducted against technical specifications or functional requirements.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not a diagnostic device involving human readers or AI.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable as this is not an AI/algorithm-based device.
7. The type of ground truth used
"Ground truth" in the AI/diagnostic sense is not directly applicable. For this device, the "truth" is based on:
- Engineering specifications/standards: Mechanical load conditions, durability, sterilization standards (e.g., ISO 10993-1, ANSI/AAMI/ISO 11135-2007).
- Functional performance: "No major deformation," "illumination quality verified," "maintain sterile barrier."
- Clinical observation (cadaver): "Appropriate incision size," "forces applied to tissue," surgeon feedback on handle grip. These are observational data points rather than a pre-defined "ground truth."
8. The sample size for the training set
- Not applicable. This is not an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established
- Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized symbol featuring three overlapping human profiles facing to the right, with flowing lines suggesting movement or connection.
January 9, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Aegis Surgical Ltd % Ms. Tammy Carrea Regulatory Consultant RegMatters, LLC 113 Grantwood Drive Holly Springs, NC 27540
Re: K141622
Trade/Device Name: Illuminated Mediastinal Access Port Regulation Number: 21 CFR 874.4720 Regulation Name: Mediastinoscope and Accessories Regulatory Class: Class II Product Code: EWY Dated: December 7, 2014 Received: December 11, 2014
Dear Ms. Carrea:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K141622
Device Name Illuminated Mediastinal Access Port
Indications for Use (Describe)
The Illuminated Mediastinal Access Port is intended to aid the surgeon in direct visualization of the mediastinum and facilitate the introduction and removal of surgical instruments during surgical procedures.
The Illuminated Mediastinal Access Port is specifically indicated for use in the anterior, superior and middle mediastinum.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K141622
510(k) Summary [In accordance with 21CFR 807.92]
1. Submitter / 510(k) Holder
Aegis Surgical, Ltd. 4020 Stirrup Creek Drive - Suite 115 Durham, NC 27703 USA
| Contact Person: | Tammy B. CarreaRegMatters, LLC |
|---|---|
| ----------------- | ------------------------------------ |
Date Prepared: December 3, 2014
2. Device Name
| Proprietary Name: | Illuminated Mediastinal Access Port |
|---|---|
| Common/Usual Name: | Mediastinoscope, Surgical |
| Classification Name: | Mediastinoscope and accessories |
| Regulation: | 874.4720 |
| Classification: | Class II |
| Product Code: | EWY |
3. Predicate Device(s)
| CLASS | MANUFACTURER | 510(k)NUMBER | DEVICE NAME/DESCRIPTION |
|---|---|---|---|
| II | Richard Wolf | K971166 | Richard Wolf OpticalMediastinoscope |
4. Device Description
The Illuminated Mediastinal Access Port is a trocar/cannula system used to provide mediastinal surgical access. The device includes a blunt trocar for dissecting tissue planes. The open cannula provides for direct visualization of the surgical space and includes small notches at the rim to anchor sutures.
The base of the cannula has fiber optic bundles incorporated that may be connected to a fiber optic light cable and an endoscopic light source. This enhances lighting and
{4}------------------------------------------------
visualization within the access device. The device does not include a light source.
The trocar/cannula access device also includes a handle for connection to a support arm. This allows the mediastinoscope to be held in position so that the surgeon can operate using both hands, without holding or re-positioning the mediastinoscope, and to provide stability for the mediastinoscope during operation.
The entire device is disposable, single use and is provided pre-sterilized.
5. Intended Use
The Illuminated Mediastinal Access Port is intended to aid the surgeon in direct visualization of the mediastinum and facilitate the introduction and removal of surgical instruments during surgical procedures.
The Illuminated Mediastinal Access Port is specifically indicated for use in the anterior, superior and middle mediastinum.
Comparison of Technological Characteristics with Predicate 6.
The intended use of the Aegis Surgical Illuminated Mediastinal Access Port is the same as the Richard Wolf Optical Mediastinoscope cleared in K971166.
The principles of operation and operational characteristics are the same and include mediastinal access, surgical instrument passage, visualization and illumination, the ability to spread tissue planes and increase surgical access. Both devices are used in a sterile state.
The devices are different in that the Richard Wolf Optical Mediastinoscope is reusable and able to be re-sterilized between uses. The Aegis Surgical Illuminated Mediastinal Access Port is pre-sterilized and single use. Sterilization to an SAL of 1x10-6 using ethylene oxide was demonstrated via a validated EO cycle.
The two devices are different in that the Richard Wolf Optical Mediastinoscope includes the use of an endoscope while the Aegis Illuminated Mediastinal Access Port provides direct mediastinal visualization but similarly includes illumination by a standard endoscopic light source.
The Illuminated Mediastinal Access Port device and predicate device have differences in the size and shape of the cannula. While both devices provide an open channel for access to the mediastinum and the ability to introduce and utilize surgical instruments through the cannula channel, the predicate device (Richard Wolf Optical Mediastinoscope, K971166) has a narrower channel and instruments are introduced singularly. The Illuminated Mediastinal Access Port device has a wider, trapezoid shaped cannula that allows the clinician to pass multiple instruments through the channel simultaneously. This larger diameter has the ability to improve visualization of the work space and the ability to internally suture as needed.
{5}------------------------------------------------
Another technical difference is the length of the working channel such that the Illuminated Mediastinal Access Device is for use in the anterior, superior and middle mediastinum. Although the two devices have differences in size, the differences do not introduce new or different issues of safety and effectiveness and any possible questions regarding use have been tested using bench or cadaver models to evaluate use and safety.
The handle grip of the Illuminated Mediastinal Access Port device is shorter than that of the predicate device, however, in evaluations of the device in cadaver models, surgeons were able to adequately grip the handle and control the insertion and positioning of the device within the incision. Once positioned, the Illuminated Mediastinal Access Port may be attached to a surgical support arm to provide hands free support and stability during the surgical procedure. More recently marketed versions of the predicate device also include a connection point for connecting to a surgical support arm as displayed in marketing literature.
7. Performance Testing Summary
Like the Richard Wolf Optical Mediastinoscope, the Aegis Surgical Illuminated Mediastinal Access Port was tested to ensure that there was no major deformation of the device when tested under worst case mechanical load conditions.
In addition durability tests were performed to demonstrate that there was no breakage of other parts of the device, specifically the fiber optic bundles were tested and compared to similar medical devices to assure that they do not pull apart under normal conditions of use. The illumination quality was also verified following mechanical tests and simulated use testing.
Biocompatibility studies were performed in accordance with ISO 10993-1:2009 for a limited contact duration device.
Sterilization studies were also performed in accordance with ANSI/AAMI/ISO 11135-2007 "Medical Devices-Validation and routine control of ethylene oxide sterilization" and AAMI TIR 28:2009, "Product adoption and process equivalence for ethylene oxide sterilization" to demonstrate that the Aegis Surgical Illuminated Mediastinal Access Port is able to be ethylene oxide (EO) sterilized to a sterility assurance level of 1x10 6. Shelf life testing was performed to demonstrate that the sterile packaging maintained a sterile barrier over time and to demonstrate that the product functionality was maintained.
Transit testing was performed to demonstrate that the sterile packaging, corrugate box, and outer shipping container were able to adequately withstand the test conditions required for ISTA 2A distribution standards.
Cadaver testing was performed to establish an appropriate incision size and to evaluate the forces applied to tissue during insertion and extraction. Forces in the lateral and axial directions were also evaluated under worst case conditions to demonstrate that the device does not break or is not damaged.
Lastly the Illuminated Mediastinal Access Port was tested with a compatible light source for spectral irradiance under normal and worst case light output conditions. The testing
{6}------------------------------------------------
compared the output of the light source when connected to a standard fiber optic light cable and when connected to a standard fiber optic light cable in tandem with the Illuminated Mediastinal Access Port. The spectral irradiance of the Illuminated Mediastinal Access Port was considerably less than that of the standard light source configuration when tested under both typical clinical and worst case light output conditions, indicating that the Illuminated Mediastinal Access Port is safe for use with standard light sources and light cables.
Conclusion 8.
The Illuminated Mediastinal Access Port has been shown to be equivalent to the predicate in terms of intended use and operation and the risks of any technological differences have been mitigated through testing.
§ 874.4720 Mediastinoscope and accessories.
(a)
Identification. A mediastinoscope and accessories is a tubular tapered electrical endoscopic device with any of a group of accessory devices which attach to the mediastinoscope and is intended to examine or treat tissue in the area separating the lungs. The device is inserted transthoracicly and is used in diagnosis of tumors and lesions and to determine whether excision of certain organs or tissues is indicated. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel. This generic type of device includes the flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, and flexible biopsy curette, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.