(276 days)
Not Found, Not Found
No
The summary describes a standard electronically controlled, pneumatically powered ventilator with display and alarm functions. There is no mention of AI or ML in the intended use, device description, or performance studies.
Yes
The device is an emergency and transport ventilator designed to provide ventilatory support to patients in respiratory and/or cardiac arrest or respiratory distress, directly treating these conditions.
No
The device is a ventilator, which provides respiratory support, and displays ventilation parameters; it does not diagnose medical conditions.
No
The device description clearly states it is an electronically controlled, pneumatically powered ventilator with hardware components like a screen, power adapter, and battery pack. It is not solely software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states that the device is a ventilator designed to provide ventilatory support to patients. This is a life-support device that interacts directly with the patient's respiratory system.
- Device Description: The description details a mechanical ventilator that delivers air to the patient's lungs. It monitors physiological parameters like pressure and volume, but it does not perform any tests on biological samples (like blood, urine, or tissue) outside of the body.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples, detecting analytes, or providing diagnostic information based on in vitro testing.
IVD devices are specifically designed to perform tests on samples taken from the human body to provide information for diagnosis, monitoring, or screening. This device does not fit that description.
N/A
Intended Use / Indications for Use
o two e700, e600 and e500 are a time-cycled, volume-constant and pressure-controlled (only e700) emergency and transport ventilator designed for use in the pre-hospital, inter-hospital and transport settings. It is intended for use with adult, child, infant patients with a tidal volume from 50 ml (100 ml for e500) upwards who are in respiratory and/or cardiac arrest or respiratory distress and who require ventilatory support.
Product codes
BTL
Device Description
The proposed o_two e700, o_two e600 and o_two e500 are time-cycled, volume-constant and pressure controlled (o_two e700 only) emergency and transport ventilators.
They are electronically controlled, pneumatically powered ventilators which can be run using AC/DC power adapter or an internal rechargeable Lithium Ion battery pack.
The devices use a 4.3" TFT screen to display live ventilation parameters (Tidal and minute volumes, proximal airway pressure and breathing rate) as well as ventilation pressure & volume wave forms, ventilation modes, settings, alarm limits and battery status. The wide range of both visual and audible alarms provides the healthcare professional with warnings of any changes in patient or device parameters.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adult, child and infant patients
Intended User / Care Setting
pre-hospital, inter-hospital and transport settings
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
A comparative side-by-side bench testing was performed on o_two e700 and Oxylog 3000 to demonstrate substantial equivalence of the proposed to the predicate Oxylog 3000.
The summary of the test results follows:
- -Both ventilators delivered equivalent volume-time, pressure-time and flow- time wave forms under the same ventilation mode;
- Both ventilators were responsive to spontaneous breathing trigger; -
- -Ventilation Vt, frequency and I:E ratio or Ti of both units were close to target or preset parameters;
- -Both units delivered similar pressure supports
We have also performed the following safety/essential performance bench testing as per to IEC60601-1, IEC60601-1-2 and other applicable standards with respect to mechanical, electrical, software, usability and biocompatibility.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 868.5925 Powered emergency ventilator.
(a)
Identification. A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 16, 2015
O-Two Medical Technologies Inc. Mr. David Zhang Ouality Assurance Manager 7575 Kimbel Street Mississauga, Ontario CANADA
Re: K141595
Trade/Device Name: o two e700, e600 and e500 Regulation Number: 21 CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: II Product Code: BTL Dated: February 12, 2015 Received: February 18, 2015
Dear Mr. Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K141595
Device Name
o two e700, e600, e500 Electronic Transport Ventilators
Indications for Use (Describe)
o two e700, e600 and e500 are a time-cycled, volume-constant and pressure-controlled (only e700) emergency and transport ventilator designed for use in the pre-hospital, inter-hospital and transport settings. It is intended for use with adult, child, infant patients with a tidal volume from 50 ml (100 ml for e500) upwards who are in respiratory and/or cardiac arrest or respiratory distress and who require ventilatory support.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
---|---|
------------------------------------------------------------------------------------------------- | ------------------------------------------------------------------------------------------------ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
| Submitter's Name & Address: | O-Two Medical Technologies
7575 Kimbel St.
Mississauga, Ontario L5S1C8
Canada
Tel - 905-677-9410 |
|-----------------------------|--------------------------------------------------------------------------------------------------------------|
| Official Contact: | David Zhang |
| Application Date: | 2014-04-25 |
| Proprietary or Trade Name: | o_two e700, o_two e600, o_two e500 |
| Common/Usual Name: | Electronic Transport Ventilators |
| Classification Name: | Ventilator, emergency, powered (resuscitator)
(21 CFR 868.5925, product code: BTL) |
| Device Class: | Class II |
| Classification Panel: | Anesthesiology |
Oxylog 3000
Carevent PAR
510k Summary
Device Description:
Predicate Devices:
The proposed o_two e700, o_two e600 and o_two e500 are time-cycled, volume-constant and pressure controlled (o_two e700 only) emergency and transport ventilators.
• 510(k) number K062267
• 510(k) number K081330
· Manufactured Draeger medical GmbH
· Manufactured by O-Two Medical Technologies Inc.
They are electronically controlled, pneumatically powered ventilators which can be run using AC/DC power adapter or an internal rechargeable Lithium Ion battery pack.
The devices use a 4.3" TFT screen to display live ventilation parameters (Tidal and minute volumes, proximal airway pressure and breathing rate) as well as ventilation pressure & volume wave forms, ventilation modes, settings, alarm limits and battery status. The wide range of both visual and audible alarms provides the healthcare professional with warnings of any changes in patient or device parameters.
4
Indications for Use:
o_two e700, e600 and e500 are a time-cycled, volume-constant and pressure-controlled (only e700) emergency and transport ventilator designed for use in the pre-hospital, intra-hospital, inter-hospital and transport settings. It is intended for use with adult, child and infant patients with a tidal volume from 50 ml (100 ml for e500) upwards who are in respiratory and/or cardiac arrest or respiratory distress who require ventilatory support.
Patient Population:
Adult, child and infant patients with a tidal volume from 50 ml (100 ml for o_two e500) upwards.
Contraindications: NA
Environment of Use:
They are intended for use in the prehospital, intrahospital and ground transport settings.
| Intended
use | Proposed
o_two e700, o_two e600, o_two e500 | Predicate
K062267 Oxylog 3000® | Substantial
Equivalence |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Operating
principle | time-cycled, volume-constant and
pressure-controlled (only o_two
e700) emergency and transport
ventilators | time-cycled, volume-constant and
pressure controlled
emergency and transport ventilator | Yes |
| Patient
population | intended for adult, child and infant
patients with a tidal volume from 50
ml* upwards who are in respiratory
and/or cardiac arrest or respiratory
distress and who require ventilatory
support.
- o_two e500: from 100 ml upwards | intended for patients with a tidal volume
from 50 ml upwards. | Yes |
| Environment
of use | pre-hospital, intra-hospital, inter-
hospital and ground transport settings | Mobile use for emergency medical care or
primary care of emergency patients:
- During transport in emergency rescue
vehicles or aircrafts including helicopters
(pre-hospital, transport); - In accident and emergency departments,
in the recovery room. (intra-hospital)
Mobile use for secondary transfers: - During transfer by road or air(inter-
hospital, transport) - When moving ventilated patients around
in the hospital. (intra-hospital) | Yes (air
transport
excluded) |
| Operating
principle | Proposed
o_two e700, o_two e600, o_two e500 | Predicate
K062267 Oxylog 3000® | Substantial
Equivalence |
| Control logic | time-cycled, volume-constant and
pressure-controlled emergency and
transport ventilators | time-cycled, volume-constant and
pressure controlled emergency and
transport ventilator | Yes |
| Flow & frequency
control | Solenoid Valves activated by the
microprocessor at the controlled
intervals to deliver the desired flow Gas blender controlled by the
microprocessor to deliver the
variable flow rates as per the
ventilation mode Yes | | |
| | During inhalation, the pressure control
valve (VCA) controls the breathing
valve (Anti-Lock up valve) blocking
the exhaust path resulting in all gas
delivered to patient | During inspiration, the pressure
control (V6) controls the Breathing
Valve (V10) to seal off against
atmosphere air | Yes |
| Expiration | During expiration , the VCA controls
the breathing valve (Anti-Lock up
valve) to open the patient airway
exhaust path and releasing all exhaled
gas to ambient. | During expiration, the V6 controls
Breathing Valve (V10) to adjust the
required patient pressure by
controlling the pressure in the
inspiration hose. | Yes |
| Pressure Control | the pressure control valve (VCA)
provide pressure ventilation and
support by controlling Airway
pressure to desired values | the pressure control (V6) reduces the
pressure in the inspiration hose to
control pressure support or Pmax
when the target values are reached | Yes |
| Safety | In the event of a fault, the pressure
control valve (VCA) opens to
atmosphere to vent excess pressure;
A safety valve limits patient airway
pressure to 80 cm H2O in the presence
of an excesses pressure | In the event of a fault, the pressure
control (V6) opens to atmosphere to
vent excess pressure;
The relieve SV (set to 80 cmH2O)
opens in the presence of an excesses
pressure | Yes |
| Monitoring | The flow and airway pressure signals
measured on the patient side are
transmitted to the pressure sensors for
flow and airway pressure curve
display as well as the measured tidal
volume and Pmax, Pmean. | The flow and airway pressure signals
measured on the patient side are
transmitted to the pressure sensors for
flow and airway pressure curve
display as well as the measured tidal
volume and Pmax, Pmean. | Yes |
| Characteristic | Proposed
o_two e700, o_two e600 and
o_two e500 | Predicate
K062267 Oxylog 3000 or
K081330 CAREvent PAR | Substantial
Equivalence |
| Product code | BTL | BTL | Equivalent to
CAREvent PAR |
| Ventilation modes | CMV, ACV,
SIMV, SIMV / PS
BiLVL, BiLVL /PS,
CPAP, CPAP /PS | CMV, CMVassist (ACV)
SIMV, SIMV /PS
BIPAP(BiLVL), BIPAP (BiLVL)/PS
CPAP, CPAP /PS | Equivalent to
Oxylog 3000 |
| CPR mode | CPR mode synchronized with
audible prompts and visual
animated display | CPR mode synchronized with
audible prompts and visual
animated display | Equivalent to
CAREvent PAR |
| | Chest
compression | 30 compression within 18 sec
audible prompts and visual
animation | 30 compression within 18 sec
audible prompts and visual
animation |
| | Ventilation | Two 1 sec mandatory breaths ;
Breath interval: 2 sec
Compression/ ventilation rate: 30:2
visual animation | Two 1 sec mandatory breaths;
Breath interval: 2 sec
Compression/ ventilation rate: 30:2
visual animation |
| | | Pmax:
60 cmH2O-adult,
40 cmH2Ol-child & infant | Pmax:
60 cmH2O-adult only |
| Waveforms | volume-time, pressure-time and
flow- time | volume-time, pressure-time and
flow- time | Equivalent to
Oxylog 3000 |
| Trigger Sensitivity | 1 to 15 L/min | 3 to 15 L/min | Equivalent to
Oxylog 3000 |
| Input pressure | Compressed O2
45 PSI to 87 PSI | Compressed O2
43.5 PSI to 87 PSI | Equivalent to
Oxylog 3000 |
| PSV (pressure
support ventilation) | 0, 4- 35 cmH2O (e700) | 0- 35 cmH2O | Equivalent to
Oxylog 3000 |
| Ventilation
Frequency | 5 to 60 breath/min | 2 to 60 breath/min (SIMV, BIPAP)
5 to 60 breath/min (CMV, ACV) | Equivalent to
Oxylog 3000 |
| Tidal Volume (L) | 50 ml to 2.0 L | 50 ml to 2.0 L | Equivalent to
Oxylog 3000 |
| Manual ventilation/
Inspiration hold | Yes | Yes | Equivalent to
Oxylog 3000 |
| Inspiration time to
expiration time ratio | 3: 1 to 1: 4 | 3:1 to 1: 4 | Equivalent to
Oxylog 3000 |
| Inspiration time Ti
(sec.) | 0.3 - 9 sec | 0.2 to 10 sec | Equivalent to
Oxylog 3000 |
| PEEP/ CPAP
(cm H2O) | 0 - 20 cm H2O | 0 - 20 cm H2O | Equivalent to
Oxylog 3000 |
| F;O2 (%) | 60% & 100% | 60% & 100% | Equivalent to
Oxylog 3000 |
| | | | |
| Pmax | 10- 80 cmH2O | 20- 60 cmH2O | Equivalent to
Oxylog 3000 |
| Safety relief valve | Opens at 80 cmH2O | Opens at 80 cmH2O | Equivalent to
Oxylog 3000 |
| Inhalation pressure
(cm H2O) | 4 - 50 | 3 - 55 | Equivalent to
Oxylog 3000 |
| Apnea back up time | 10 to 60 sec | 15 to 60 sec | Equivalent to
Oxylog 3000 |
| Battery Operating
time | 18 hrs | 4 hrs | Better than
Oxylog 3000 |
| Live monitoring | Mve, Vte,
Paw(AV), Paw(Peak),
Fbpm | Mve, Vte,
Paw(AV), Paw(Peak), Pplat, PEEP
Fbpm, FiO2 (%) | Equivalent to
Oxylog 3000 |
| Wave form
displayed | Pressure and flow | Pressure and flow | Equivalent to
Oxylog 3000 |
| Alarms
Audible/Visual &
indications | Pmax, Pmin,
MVHigh, MVLow,
Low Battery (20% increments)
BCI,
Supply pressure Low or No,
APNEA | Pmax, Pmin,
MVHigh, MVLow,
Low Battery (25% increments)
Leakage,
Supply pressure,
APNEA | Equivalent to
Oxylog 3000 |
| Accessories | - AC/DC power supply - Lithium ion battery pack
- Patient ventilation circuit
- Oxygen supply hose | - AC/DC power supply Input:
- Lithium ion battery pack
- Patient ventilation circuit
- Oxygen supply hose | Equivalent to
Oxylog 3000 |
Comparative table- Intended use
5
Comparative table- Operating principle
6
Comparative table- Technological characteristics/specifications of performance
7
Substantially equivalence to the predicate devices:
The proposed devices (o two e700, o two e600 and o two e500) have the equivalent intended use, Patient populations, environment of use, contra-indications and intended user to predicate devices.
The proposed devices and the Oxylog 3000/K062267 provide the similar ventilation modes including CMV, SIMV, ACV, BiLevel, CPAP and Pressure Control and Pressure Support. Furthermore, o two e700, o two e600 and o two e500 use the similar operating principle as the Oxylog 3000. Specific features and range of performance specifications of the o two e700, o two e600 and o two e500 are also found equivalent to those on the Oxylog 3000/K062267.
As there are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices regarding the intended patient population, use environment/setting, contraindications, end users, as well as the technological characteristics (operating principles, waveforms, start-up mode, breath triggering and range of performance specifications), the proposed o_two e600 and o_two e500 are viewed as substantially equivalent to the predicate devices - Oxylog 3000/K062267 with the exception of CPR mode.
8
In addition, o_two e700, o_two e600 and o_two e500 incorporate an extra CPR mode which use the same CPR function as the predicate CARevent PAR/K081330. As a result, they are considered as substantially equivalent regarding their intended use, use environment/setting, contraindications, end users and technological characteristics.
Summary of Performance Testing:
A comparative side-by-side bench testing was performed on o_two e700 and Oxylog 3000 to demonstrate substantial equivalence of the proposed to the predicate Oxylog 3000.
The summary of the test results follows:
- -Both ventilators delivered equivalent volume-time, pressure-time and flow- time wave forms under the same ventilation mode;
- Both ventilators were responsive to spontaneous breathing trigger; -
- -Ventilation Vt, frequency and I:E ratio or Ti of both units were close to target or preset parameters;
- -Both units delivered similar pressure supports
We have also performed the following safety/essential performance bench testing as per to IEC60601-1, IEC60601-1-2 and other applicable standards with respect to mechanical, electrical, software, usability and biocompatibility.
| Safety & Essential
performance testing | Testing standards/ Comparative testing | Test
Result |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Safety & Essential
performance | IEC 60601-1:2005 | Comply |
| | EN794-3:1998/A2 :2009, ISO10651-3:1997, Product specifications,
ISO 80601-2-12 :2011 | Comply |
| EMC | IEC 60601-1-2:2007 w/ increased levels:
ESD: ±8kV contact & ±15kV Air
Radiated Immunity: 30V/m
Power Freq Magnetic: 30A/m | Comply |
| Vibration/ Bump | Vibration (sinusoidal) per IEC60068-2-6 Fc
Random vibration per IEC60068-2-36 Fdb, IEC60068-2-64 Fh
Bump per IEC60068-2-29 Eb, IEC60068-2-27 Type 1
Crash (10 g) per EN 1789 :2007 Clause 4.5.9 6.2, 6.3.5 | Comply |
| Environmental | ISO10651-3:1997 Clause 10.2.1, EN 794-3:1998/ A1: 2005 Sec.10.2.1 R | Comply |
| Altitude | EN794-3:1998/A2 :2009 10.2.1 c) | Comply |
| Software | IEC 60601-1:2005 Sec.14, ANSI/AMMI/IEC 62304:2006 | Comply |
| Usability | IEC 60601-1-6 :2010, IEC62366 :2007, Human Factor/Usability Validation | Comply |
| Safety - Battery | IEC62133:2002, EN60950-1:2006/ A11:2009/ A1:2010/A12:2011 | Comply |
| Transportation- battery | UN 38.3 test | Comply |
| Bio-compatibility | ISO 10993-1: 2009, ISO 10993-5:2009, ISO 10993-10: 2010 | Comply |
| Function/ Validation test | O-Two Validation Protocol-e700/e600 and Validation Protocol-e500 | Comply |
9
Conclusion:
The results of the above comparative performance and specification as well as bench testing demonstrate that the proposed o_two e700, o_two e600 and o_two e500 are as safe, as effective and perform as well as the legally marketed predicate devices - Oxylog 3000 (K062267) and CARevent PAR (K081330).