K Number
K141303
Date Cleared
2014-12-12

(207 days)

Product Code
Regulation Number
880.6250
Panel
HO
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Nitrile films form a barrier to body fluids and bloodborne Pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

AI/ML Overview

The provided document describes the acceptance criteria and performance of the Nitrile Powder Free Patient Examination Gloves, Blue Color based on non-clinical performance data, as clinical data was not deemed necessary for this device.

Here's the breakdown of the information requested:

  • 1. A table of acceptance criteria and the reported device performance
CharacteristicsStandard / Acceptance CriteriaDevice Performance
DimensionASTM standard D 6319-10 (For Length, Width, Thickness)Meets ASTM standard D 6319-10 (Length: ≥230mm min for all sizes; Width: Small 84-86mm, Medium 94-97mm, Large 105-109mm, X large 114-119mm; Thickness: Finger 0.10mm min, Palm 0.06mm min)
Physical PropertiesASTM standard D 6319-10 (Before aging/after aging: Elongation ≥500%, Tensile Strength ≥ 14MPa)Meets ASTM standard D 6319-10 (Before aging/after aging: Elongation: 520-580%, Tensile Strength: 22-34 MPa)
Freedom from pinholes21 CFR 800.20; ASTM D6319-10; ASTM D 5151-06 (Reapproved 2011) (Specifically, for waterleak test on pinhole AQL 2.5 at Inspection Level I)Meets ASTM D 5151-06 (Reapproved 2011) - Holes at Inspection Level I, AQL 2.5
Powder ResidualASTM standard D 6319-10 and D6124-06 (Reapproved 2011) (

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.