AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GE200 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE200 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The GE200 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE200 Blood Glucose Monitoring System should not be used for the diagnosis of , or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). The GE200 Blood Glucose Test Strips are for use with the GE200 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

The GE300 Talking Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE300 Talking Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The GE300 Talking Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by diabetic individuals at home as an aid to monitor the effectiveness of diabetes control. Only limited audible information is available, not intended as an aid for the visually impaired. The GE300 Talking Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The GE300 Talking Blood Glucose Test Strips are for use with the GE300 Talking Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE300 Series Control Solution are for use with the GE300 Talking Blood Glucose Meter and the GE300 Talking Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The Rightest Blood Glucose Monitoring System GM650 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Monitoring System GM650 is intended to be used by a single person and should not be shared. The Rightest Blood Glucose Monitoring System GM650 is intended for self testing outside the body (in vitro diagnostic use) by diabetic individuals at home as an aid to monitor the effectiveness of diabetes control. Only limited and ble, is available, not intended as an aid for the visually impaired. The Rightest Blood Glucose Monitoring System GM650 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The Rightest Blood Glucose Test Strips GS650 are for use with the Rightest Blood Glucose Meter GM650 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Control Solutions GC650 are for use with the Rightest Blood Glucose Meter GM650 and the Rightest Blood Glucose Test Strip GS650 to check that the meter and test strips are working together properly and that the test is performing correctly.

The Rightest Blood Glucose Monitoring System GM700 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Monitoring System GM700 is intended to be used by a single person and should not be shared. The Rightest Blood Glucose Monitoring System GM700 is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System GM700 should not be used for the diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The Rightest Blood Glucose Test Strips GS700 are for use with the Rightest Blood Glucose Meter GM700 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

Device Description

The GE200 Blood Glucose Monitoring System consists of the GE200 Blood Glucose Meter, GE200 Blood Glucose Test Strip, and GE200 Control Solution. The GE200 Blood Glucose Meter, when used with the GE200 Blood Glucose Test Strips, quantitatively measures glucose in fresh capillary whole blood. The performance is verified by the GE200 Control Solution.

The GE300 Talking Blood Glucose Monitoring System consists of the GE300 Talking Blood Glucose Meter, GE300 Talking Blood Glucose Test Strip, and GE300 Series Control Solution. The GE300 Talking Blood Glucose Meter, when used with the GE300 Talking Blood Glucose Test Strips, quantitatively measures glucose in fresh capillary whole blood. The performance is verified by the GE300 Series Control Solution.

The Rightest Blood Glucose Monitoring System GM650 consists of the Rightest Blood Glucose Meter GM650, Rightest Blood Glucose Test Strip GS650, and Rightest Control Solution GC650. The Rightest Blood Glucose Meter GM650, when used with the Rightest Blood Glucose Test Strips GS650, quantitatively measures glucose in fresh capillary whole blood. The performance is verified by the Rightest Control Solution GC650.

The Rightest Blood Glucose Monitoring System GM700 consists of the Rightest Blood Glucose Meter GM700, Rightest Blood Glucose Test Strip GS700, and Rightest Control Solution GC700. The Rightest Blood Glucose Meter GM700, when used with the Rightest Blood Glucose Test Strips GS700, quantitatively measures glucose in fresh capillary whole blood. The performance is verified by the Rightest Control Solution GC700.

AI/ML Overview

The provided document describes the non-clinical and clinical studies performed for the BIONIME Righhtest Blood Glucose Monitoring System (GM700, GM650) and GE Blood Glucose Monitoring System (GE200, GE300) to demonstrate their substantial equivalence. The primary change for these systems is a revised hematocrit (HCT) range for the test strips (GS700/GS650/GS750 for Rightest brand and identical strips for GE200/GE300).

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document refers to adherence to ISO 15197:2003 for precision and references other CLSI guidelines for interference studies, but it doesn't explicitly state specific numerical acceptance criteria for each test (e.g., specific bias percentages for linearity or accuracy). However, it consistently states that "the results fall within the acceptance criteria" or "meet the acceptance criteria."

However, for accuracy, the general standard for blood glucose meters within ISO 15197:2003 (which was referenced) is typically:

  • For glucose concentrations < 75 mg/dL (4.2 mmol/L): 95% of results must be within ±15 mg/dL (±0.83 mmol/L) of the reference method.
  • For glucose concentrations ≥ 75 mg/dL (4.2 mmol/L): 95% of results must be within ±20% of the reference method.

Given the document mentions "no significant difference in slope, intercept, and correlation coefficient" for accuracy and "falls within the acceptance criteria" for precision, it implies these ISO standards were met.

Study TypeAcceptance Criteria (Implied/General Standard)Reported Device Performance
Accuracy (System Accuracy)95% of results within specific ranges compared to reference (e.g., ±15 mg/dL for < 75 mg/dL, ±20% for ≥ 75 mg/dL, per ISO 15197:2003)Met acceptance criteria; no significant difference in slope, intercept, and correlation coefficient.
LinearityLinear regression between device and reference method with mean slopes above 1.00 and R² values above 0.99 (as stated in the document for the predicate).Mean slopes above 1.00 and R² values above 0.99. Met acceptance criteria.
PrecisionRepeatability and intermediate precision within specified limits (as per ISO 15197:2003).Repeatability of within-day falls within the acceptance criteria. Met acceptance criteria.
Specimen VolumeMaintained existing claimed minimum blood specimen volume (0.75 µL).Existing claimed minimum blood specimen volumes of 0.75µL unaffected.
InterferenceBiases within specified percentages (e.g., less than ±10% bias for non-interfering substances).Most substances within criteria. Xylose, Ascorbic Acid, Dopamine, Uric Acid showed > ±10% bias at certain concentrations (noted as limitations).
Altitude EffectStability of bias compared to reference during various operated altitudes within criteria.Stability of bias compared to REF during various operated altitude were all within the criteria.
Operating Temperature/HumidityStability of bias compared to reference during various operated temperature and humidity within criteria.Stability of bias compared to REF during various operated temperature and humidity were all within the criteria.
Hematocrit RangeStability compared to reference during various HCT ranges within criteria.Stability compared to REF during various HCT ranges were all within the criteria. Acceptable HCT range is 20-65%.
Lay User PerformanceLay users could obtain accurate results (implies meeting accuracy standards).Showed substantial equivalence to predicate device in finger, palm, and forearm positions.

2. Sample sizes used for the test set and the data provenance:

  • Linearity Study Data Provenance: Not explicitly stated, but given it evaluates venous blood samples, it's likely a controlled lab study, probably prospective.
  • Precision Study Sample Size:
    • Within-day: 500 test strips (across 3 lots)
    • Between-day: 300 test strips (across 3 lots)
    • Data Provenance: Not explicitly stated, implied controlled lab study, probably prospective.
  • Specimen Volume Study Data Provenance: Not explicitly stated, likely a controlled lab study, probably prospective.
  • Interference Study Data Provenance: Not explicitly stated, likely a controlled lab study, probably prospective.
  • Altitude Study Data Provenance: Not explicitly stated, likely a controlled lab study, probably prospective.
  • Operating Temperature and Humidity Study Data Provenance: Not explicitly stated, likely a controlled lab study, probably prospective.
  • Hematocrit Study Data Provenance: Not explicitly stated, likely a controlled lab study, probably prospective.
  • System Accuracy Study (Whole Blood) Sample Size: A total of 120 patients participated.
    • Data Provenance: Not explicitly stated (e.g., country of origin), but it's a clinical study on patients, which would typically be prospective for this type of accuracy assessment.
  • Lay User Study Sample Size: A total of 100 laypersons.
    • Data Provenance: Not explicitly stated (e.g., country of origin), but it's a clinical study on lay users, which would typically be prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • For all the non-clinical studies (Linearity, Precision, Specimen Volume, Interference, Altitude, Operating Temperature/Humidity, Hematocrit), the ground truth was established by a reference method, specifically the YSI 2300 Analyzer. This is an automated laboratory instrument, not human experts.
  • For the clinical System Accuracy Study, the comparison was made against "plasma glucose values on reference lab instrument," which is the YSI 2300 Analyzer. No human experts are explicitly mentioned as establishing this ground truth for the test.
  • For the Lay User Study, the evaluation method is comparing the lay users' results to a "predicate device." The ground truth for direct accuracy is again, by implication from other studies, the YSI 2300 Analyzer. No explicit mention of human experts establishing ground truth for the test sets in this context.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • No adjudication method by human experts is described in the document for any of the test sets. The ground truth for all quantitative measurements relies on comparison to a reference laboratory instrument (YSI 2300 Analyzer).

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This is not applicable as the device is a blood glucose monitoring system, not an AI diagnostic tool that human readers would interpret. There are no "human readers" in the context of interpreting results from this device; patients or healthcare professionals read the numerical glucose value displayed by the meter.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes, the performance of the device (meter and test strip) is inherently standalone as it provides a direct numerical result. The non-clinical studies (Linearity, Precision, Interference, etc.) and the System Accuracy study report on the algorithm's (device's) performance compared to a reference standard without human interpretation as an intermediate step. The "Lay User Study" evaluates the user's ability to operate the device and obtain accurate results, not their interpretation of an algorithm's output.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • The ground truth used for all quantitative performance analyses (linearity, precision, interference, system accuracy, etc.) was a reference laboratory instrument, specifically the YSI 2300 Analyzer, which provides plasma glucose values. This is an objective, instrumental reference method.

8. The sample size for the training set:

  • This document is a 510(k) summary for a medical device (blood glucose monitoring system), not an AI/machine learning model. Therefore, the concept of a "training set" in the context of AI is not applicable. The device's operational parameters and calibration are established through engineering and analytical studies, not typically by training a model on a large dataset in the way AI systems are trained.

9. How the ground truth for the training set was established:

  • As mentioned above, the concept of a "training set" for an AI model is not applicable to this device. For the calibration and internal algorithms of the blood glucose meter, the ground truth would be established through a rigorous process involving samples with known glucose concentrations, validated by a reference method like the YSI 2300 Analyzer during the device's development and manufacturing.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a family of three people.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 9, 2015

BIONIME COPORATION C/O FENG-YU LEE REGULATORY CONSULTANT 29222 RANCHO VIEJO ROAD, SUITE 218 SAN JUN CAPISTRANO CA 92675

Re: K141292

Trade/Device Name: Rightest Blood Glucose Monitoring System GM700, Rightest Blood Glucose Monitoring System GM650, GE200 Blood Glucose Monitoring System GE 300 Talking Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345

Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, LFR Dated: April 28, 2015 Received: May 1, 2015

Dear Feng-Yu Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For: Courtney H. Lias, Ph.D.

Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

GE200 Blood Glucose Monitoring System

Indications for Use (Describe)

The GE200 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE200 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The GE200 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE200 Blood Glucose Monitoring System should not be used for the diagnosis of , or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly).

The GE200 Blood Glucose Test Strips are for use with the GE200 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

Device Name

GE300 Talking Blood Glucose Monitoring System

Indications for Use (Describe)

The GE300 Talking Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE300 Talking Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The GE300 Talking Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by diabetic individuals at home as an aid to monitor the effectiveness of diabetes control. Only limited audible information is available, not intended as an aid for the visually impaired. The GE300 Talking Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The GE300 Talking Blood Glucose Test Strips are for use with the GE300 Talking Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

The GE300 Series Control Solution are for use with the GE300 Talking Blood Glucose Meter and the GE300 Talking Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|× Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

Rightest Blood Glucose Monitoring System GM650

Indications for Use (Describe)

The Rightest Blood Glucose Monitoring System GM650 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Monitoring System GM650 is intended to be used by a single person and should not be shared.

The Rightest Blood Glucose Monitoring System GM650 is intended for self testing outside the body (in vitro diagnostic use) by diabetic individuals at home as an aid to monitor the effectiveness of diabetes control. Only limited and ble, is available, not intended as an aid for the visually impaired. The Rightest Blood Glucose Monitoring System GM650 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The Rightest Blood Glucose Test Strips GS650 are for use with the Rightest Blood Glucose Meter GM650 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

The Rightest Control Solutions GC650 are for use with the Rightest Blood Glucose Meter GM650 and the Rightest Blood Glucose Test Strip GS650 to check that the meter and test strips are working together properly and that the test is performing correctly.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{8}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

Rightest Blood Glucose Monitoring System GM700

Indications for Use (Describe)

The Rightest Blood Glucose Monitoring System GM700 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Monitoring System GM700 is intended to be used by a single person and should not be shared.

The Rightest Blood Glucose Monitoring System GM700 is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System GM700 should not be used for the diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The Rightest Blood Glucose Test Strips GS700 are for use with the Rightest Blood Glucose Meter GM700 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{9}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{10}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: K141292

1. Submitter's Identification:

BIONIME CORPORRATION NO 100. Sec. 2, Daging St., South Dist., 40242 Taichung City, Taiwan Phone Number: 886-4-23692388 FAX Number: 886-4-22617568

c/o IVDD Regulatory Consultant 29222 Rancho Viejo Road, Suite 218 San Juan Capistrano, CA 92675 Contact Person: Feng-Yu Lee Phone Number: 1-949-218-0929 Fax Number: 1-218-0928

Date Summary Prepared: April 28, 2015

2. Name of the Device:

Rightest Blood Glucose Monitoring System GM700 Rightest Blood Glucose Monitoring System GM650 GE200 Blood Glucose Monitoring System GE300 Talking Blood Glucose Monitoring System

3. Common or Usual Name: Blood Glucose Monitoring System

Product CodeClassificationRegulation SectionPanel
NBW; System, Test, BloodGlucose, Over-the-CounterClass II21 CFR 862.1345Clinical Chemistry 75
LFR; GlucoseDehydrogenase, GlucoseClass II21 CFR 862.1345Clinical Chemistry 75

{11}------------------------------------------------

4. Device Description:

4.1 Bionime Rightest Blood Glucose Monitoring System GM700 consists of the following devices: Rightest Blood Glucose Meter GM700, Rightest Blood Glucose Test Strip GS700, and Rightest Control Solution GC700. The Rightest Blood Glucose Meter GM700, Rightest Blood Glucose Test Strips GS700 are manufactured by BIONIME Corporation. The Rightest Blood Glucose Meter GM700, when used with the Rightest Blood Glucose Test Strips GS700, quantitatively measures glucose in fresh capillary whole blood. The performance of the Rightest Blood Glucose Monitoring System GM700 is verified by the Rightest Control Solution GC700.

4.2 Bionime Rightest Blood Glucose Monitoring System GM650 consists of the following devices: Rightest Blood Glucose Meter GM650, Rightest Blood Glucose Test Strip GS650, and Rightest Control Solution GC650. The Rightest Blood Glucose Meter GM650, Rightest Blood Glucose Test Strips GS650, and Lancing Device are manufactured by BIONIME Corporation. The Rightest Blood Glucose Meter GM650, when used with the Rightest Blood Glucose Test Strips GS650, quantitatively measures glucose in fresh capillary whole blood. The performance of the Rightest Blood Glucose Monitoring System GM650 is verified by the Rightest Control Solution GC650.

4.3 Bionime GE200 Blood Glucose Monitoring System consists of the following devices: GE200 Blood Glucose Meter, GE200 Blood Glucose Test Strip, and GE200 Control Solution. The GE200 Blood Glucose Meter, GE200 Blood Glucose Test Strips are manufactured by BIONIME Corporation. The GE200 Blood Glucose Meter, when used with the GE200 Blood Glucose Test Strips, quantitatively measures glucose in fresh capillary whole blood. The performance of the GE200 Blood Glucose Monitoring System is verified by the GE200 Control Solution.

4.4 Bionime GE300 Talking Blood Glucose Monitoring System consists of the following devices: GE300 Talking Blood Glucose Meter, GE300 Talking Blood Glucose Test Strip, and GE300 Series Control Solution. The GE300 Talking Blood Glucose Meter, GE300 Talking Blood Glucose Test Strips are manufactured by BIONIME Corporation. The GE300 Talking Blood Glucose Meter, when used with the GE300 Talking Blood Glucose Test Strips, quantitatively measures glucose in fresh capillary whole blood. The performance of the GE300 Talking Blood Glucose Monitoring System is verified by the GE300 Series Control Solution.

Intended Use (identical to K123008): 5.

Rightest BGMS GM700/GM650

5.1

The Rightest Blood Glucose Monitoring System GM700 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700 is intended to be used by a single person and should not be shared.

{12}------------------------------------------------

The Rightest Blood Glucose Monitoring System GM700 is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System GM700 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The Rightest Blood Glucose Test Strips GS700 are for use with the Rightest Blood Glucose Meter GM700 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

5.2

The Rightest Blood Glucose Monitoring System GM650 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM650 is intended to be used by a single person and should not be shared.

The Rightest Blood Glucose Monitoring System GM650 is intended for self-testing outside the body (in vitro diagnostic use) by diabetic individuals at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System GM650 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The Rightest Blood Glucose Test Strips GS650 are for use with the Rightest Blood Glucose Meter GM650 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

GE GE200/GE300 BGMS

5.3

The GE200 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE200 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The GE200 Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE200 Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The GE200 Blood Glucose Test Strips are for use with the GE200 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

{13}------------------------------------------------

The GE300 Talking Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE300 Talking Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The GE300 Talking Blood Glucose Monitoring System is intended for self -testing outside the body (in vitro diagnostic use) by diabetic individuals at home as an aid to monitor the effectiveness of diabetes control. The GE300 Talking Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The GE300 Talking Blood Glucose Test Strips are for use with the GE300 Talking Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

6. Predicate Device Information:

The Rightest and GE series Blood Glucose Monitoring Systems are substantially equivalent to the predicate devices below:

Name:Rightest BGMS GM700
Device Company:Bionime Corporation
510(K) Number:K110737/K123008
Name:Rightest BGMS GM650
Device Company:Bionime Corporation
510(K) Number:K120423/K123008
Name:GE200 BGMS
Device Company:Bionime Corporation
510(K) Number:K123008
Name:GE300 BGMS
Device Company:Bionime Corporation
510(K) Number:K120423/K123008

5.4

8-4

{14}------------------------------------------------

Comparison to Predicate Devices: Specification Comparison 7.

Subject DevicePredicate Device (K123008)
ItemRightest BGMS GM700 (modified teststrip)Rightest BGMS GM700
Similarities and Differences
Intended UseIt is intended to be used for quantitativemeasurement of glucose in freshcapillary whole blood as an aid tomonitor the effectiveness of diabetescontrol in people with diabetessame
Detection MethodDehydrogenase ElectrochemicalBiosensorsame
Measuring Time5 secsame
Measurement Unitmg/dLsame
Sample SiteCapillary whole blood from fingertips,forearm or palmsame
Voice FunctionNosame
Meter Dimensions99 mm x 46 mm x 17.5 mmsame
Meter Weight57 ± 5 g with batterysame
Power SupplyOne CR2032 batterysame
Battery LifeApprox. 1000 measurementssame
BacklightNosame
InterfaceLCD displaysame
LCD Display Area52.6 mm x 32.0 mmsame
Memory Capacity500 blood glucose test resultssame
InterferenceUric acid ≥ 16 mg/dLXylose ≥ 18 mg/dLAscorbic acid ≥ 3 mg/dLDopamine HCl ≥ 1.25 mg/dLL-Dopa ≥ 2 mg/dLUric acid ≥ 16 mg/dLXylose ≥10 mg/dLAscorbic acid ≥3 mg/dLDopamine HCl ≥ 1.25 mg/dLL-Dopa ≥ 2 mg/dL
CodingAuto Codingsame
OperatingTemperature Range43-111°F (6-44°C)same
Operating RelativeHumidity Range10-90%same
Test Strip StorageConditions39-86 °F (4-30°C ), 10-90% relativehumiditysame
Test Strip Shelf Life(after opened vial)4 monthssame
Meter StorageConditions14-140 °F (-10-60°C )same
Control Solution3 levels (Level 1, 2, and 4)same
Strip Reagent EnzymeFAD-Glucose Dehydrogenasesame
Hematocrit Range20-65%30-55%
Sample Volume0.75 µLsame
Measurement Range20-600 mg/dLsame

A. Rightest Blood Glucose Monitoring Systems, GM700, GM650

{15}------------------------------------------------

Subject DevicePredicate Device (K123008)
ItemRightest BGMS GM650 (modified teststrip)Rightest BGMS GM650
Similarities and Differences
Intended UseIt is intended to be used for quantitativemeasurement of glucose in freshcapillary whole blood as an aid tomonitor the effectiveness of diabetescontrol in people with diabetessame
Detection MethodDehydrogenase ElectrochemicalBiosensorsame
Measuring Time5 secsame
Measurement Unitmg/dLsame
Sample SiteCapillary whole blood from fingertips,forearm or palmsame
Voice FunctionYessame
Meter Dimensions105 mm x 50 mm x 18 mmsame
Meter Weight80.0 ± 5 g with batteriessame
Power SupplyTwo 1.5V(AAA) batteriessame
Battery LifeApprox. 1000 measurementssame
BacklightNosame
InterfaceLCD displaysame
LCD Display Area48mm x 34mmsame
Memory Capacity500 blood glucose test resultssame
InterferenceUric acid ≧ 16 mg/dLXylose ≥18 mg/dLAscorbic acid ≥3 mg/dLDopamine HCl ≥ 1.25 mg/dLL-Dopa ≥ 2 mg/dLUric acid = 16 mg/dLXylose ≧10 mg/dLAscorbic acid ≥3 mg/dLDopamine HCl = 1.25 mg/dLL-Dopa ≥ 2 mg/dL

.

.

Rightest/GE BGMSs, Revised Test Strip K141292 04/28/2015

.

{16}------------------------------------------------

.

CodingAuto Codingsame
OperatingTemperature Range43-111°F (6-44C)same
Operating RelativeHumidity Range10-90%same
Test Strip StorageConditions39-86 °F (4-30°C ), 10-90% relativehumiditysame
Test Strip Shelf Life(after opened vial)4 monthssame
Meter StorageConditions14-140 °F (-10-60°C )same
Control Solution3 levels (Level 1, 2, and 4)same
Strip Reagent EnzymeFAD-Glucose Dehydrogenasesame
Hematocrit Range20-65%30-55%
Sample Volume0.75 µLsame
Measurement Range20-600 mg/dLsame

B. GE Blood Glucose Monitoring Systems, GE200, GE300 .

ItemSubject DevicePredicate Device (K123008)
GE200 BGMS (modified test strip)GE BGMS GE200
Similarities and Differences
Intended UseIt is intended to be used for quantitativemeasurement of glucose in freshcapillary whole blood as an aid tomonitor the effectiveness of diabetescontrol in people with diabetessame
Detection MethodDehydrogenase ElectrochemicalBiosensorsame
Measuring Time5 secsame
Measurement Unitmg/dLsame
Sample SiteCapillary whole blood from fingertips,forearm or palmsame
Memory Capacity1000 blood glucose test resultssame
Voice FunctionNosame
Meter Dimensions96 mm x 46 mm x 17.5 mmsame
Meter Weight$65 \pm 5$ g with batteriessame
Power SupplyTwo CR2032 batteriessame

{17}------------------------------------------------

Bionime

:

Battery LifeApprox. 1000 measurementssame
BacklightYessame
InterfaceLCD displaysame
LCD Display Area55mm x 37mmsame
InterferenceUric acid ≥ 16 mg/dLXylose ≥18 mg/dLAscorbic acid ≥ 3 mg/dLDopamine HCl ≥ 1.25 mg/dLL-Dopa ≥ 2 mg/dLUric acid ≥ 16 mg/dLXylose ≥10 mg/dLAscorbic acid ≥3 mg/dLDopamine HCl ≥ 1.25 mg/dLL-Dopa ≥ 2 mg/dL
CodingAuto Codingsame
OperatingTemperature Range43-111°F (6-44C)same
Operating RelativeHumidity Range10-90%same
Test Strip StorageConditions39-86 °F (4-30°C), 10-90% relativehumiditysame
Test Strip Shelf Life(after opened vial)4 monthssame
Meter StorageConditions14-140 °F (-10-60°C )same
Control Solution3 levels (Level 1, 2, and 4)same
Strip Reagent EnzymeFAD-Glucose Dehydrogenasesame
Hematocrit Range20-65%30-55%
Sample Volume0.75 µLsame
Measurement Range20-600 mg/dLsame
ItemSubject DevicePredicate Device (K123008)
GE300 BGMS (modified test strip)GE BGMS GE300
Similarities and Differences
Intended UseIt is intended to be used for quantitativemeasurement of glucose in freshcapillary whole blood as an aid tomonitor the effectiveness of diabetescontrol in people with diabetessame
Detection MethodDehydrogenase ElectrochemicalBiosensorsame
Measuring Time5 secsame
Measurement Unitmg/dLsame
Sample SiteCapillary whole blood from fingertips,forearm or palmsame
Voice FunctionYessame
Meter Dimensions105 mm x 50 mm x 18 mmsame
Meter Weight80.0 ± 5 g with batteriessame
Power SupplyTwo 1.5V(AAA) batteriessame
Battery LifeApprox. 1000 measurementssame
BacklightNosame
InterfaceLCD displaysame
LCD Display Area48mm x 34mmsame
Memory Capacity500 blood glucose test resultssame
InterferenceUric acid ≥ 16 mg/dLXylose ≥ 18 mg/dLAscorbic acid ≥ 3 mg/dLDopamine HCl ≥ 1.25 mg/dLL-Dopa ≥ 2 mg/dLUric acid ≥ 16 mg/dLXylose ≥10 mg/dLAscorbic acid ≥3 mg/dLDopamine HCl ≥ 1.25 mg/dLL-Dopa ≥ 2 mg/dL
CodingAuto Codingsame
OperatingTemperature Range43-111°F (6-44C)same
Operating RelativeHumidity Range10-90%same
Test Strip StorageConditions39-86 °F (4-30°C ), 10-90% relativehumiditysame
Test Strip Shelf Life(after opened vial)4 monthssame
Meter StorageConditions14-140 °F (-10-60°C )same
Control Solution3 levels (Level 1, 2, and 4)same
Strip Reagent EnzymeFAD-Glucose Dehydrogenasesame
Hematocrit Range20-65%30-55%
Sample Volume0.75 µLsame
Measurement Range20-600 mg/dLsame

.

:

8-8

:

:

:

{18}------------------------------------------------

8. Technology Characteristics:

Bromine's Rightest™ GM700/GM650 and GE200/GE300 GDH Strip with revised HCT range, is paired with electronic device that utilizes the electrical characteristic technology for measuring the glucose level in human blood. A relatively small drop of blood is placed on a disposable test strip coated with FAD-glucose Dehydrogenase (FAD-GDH) which interacts with the software driven digital talking meter. Within five seconds, the level of blood glucose will be shown on the digital display screen.

{19}------------------------------------------------

Rightest™ GM700/GM650 and GE200/GE300 GDH BGMS Test Strip, with revised HCT range, requires only minimum of 0.75 microliter of blood for the testing, therefore it reduces the time and effort required for testing and improves the compliance of diabetic people to their testing regimens.

9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

Three meter models were used for testing: GM650/GE300, GM700, and GE200. GDH (GS700/GS650/GS750) test strip was used for testing, GS700, GS650, and GS750 test strips are identical to each other for Rightest brand, GE200 and GE300 test strips are identical to GS750 and GS650 with the exception of trade name.

Non-clinical tests were evaluated to establish the performance, functionality, safety and reliability of the Rightest™ GM700/GM650 and GE200/GE300 GDH Strip with revised HCT range. The performed evaluations include:

Linearity and sensitivity, readability assessment, safety test, precimen volume, interference, hematocrit, and test strip stability (close vial and open vial), are summarized below:

Linearity

To reevaluate linearity of HCT revised test strip, 3 lots of revised blood glucose test strips (GS650/GS700/GS750, HCT 20-65%) were used to evaluate detection range. Venous blood samples spiked to 14 levels of glucose concentrations were used, ranging from low to high level (0 to 640 mg/dL), and evaluated compared to reference method YSI 2300.

The tested glucose concentration ranged from 0 to 640 mg/dL and hematocrit value from 20-65%. Analysis of results indicated linear regression between device and reference method shows mean slopes above 1.00 and R2 values above 0.99. The results support the claim that glucose assay is linear from 20-600 mg/dL - identical to prior 510k cleared claims.

Precision:

Based on the requirements and standards of ISO 1597:2003 to reevaluate the precision of GE series the Rightest and Blood -Glucose -Monitoring Systems (GM700/GM650/GE200/GE300), the precision study evaluates repeatability of betweenday testing, conducted over 10 days at 5 glucose concentrations. 10 meters (3 models) were evaluated for repeatability as well as intermediate precision. 3 lots of revised test strips (GS650/GS700/GS750, HCT range 20-65%) were used; 500 test strips were used for within-day and 300 test strips were used for between-day testing. The venous blood sample results were compared against reference method YSI 2300 Analyzer.

The repeatability evaluation and intermediate precision results of Rightest and GE series Blood Glucose Monitoring Systems (GM700/GM650/GE200/GE300) demonstrate that repeatability of within-day is evaluated at 5 glucose concentrations and falls within the acceptance criteria.

{20}------------------------------------------------

Specimen Volume

The Specimen Volume Study reevaluated the specimen volume of Rightest and GE series Blood Glucose Monitoring Systems (GM700/GM650/GE200/GE300), using 3 lots of revised test strips (GS650/GS700/GS750, HCT range 20-65%), testing venous blood at 3 glucose concentration levels and at 9 specimen volumes ranging from 0.60 to 3.0 µL. The obtained study results support the existing claimed minimum blood specimen volumes of 0.75uL, which is unaffected with the recent design change.

Interference Study

The Interference Study was performed under CLSI: EP7-A2, using 2 glucose levels of venous blood samples and 19 interference substances to reevaluate the interference the GE series -Blood Rightest and and Systems (GM700/GM650/GE200/GE300). 3 lots of revised test strips (GS650/GS700/GS750, HCT range 20-65%) were used.

According to study results, xylose (20 and 40 mg/dL), ascorbic acid (6 mg/dL), dopamine (2.5 mg/dL), and uric acid (20 and 24 mg/dL) indicate percentage of interference over ±10% bias in normal glucose concentration, and xylose (40 mg/dL) over ±10% bias in high glucose concentration. This may lead to inaccurate test results, and is stated in the limitations section of the test strip package insert. The remaining 15 interference substances are within the criteria for the 2 glucose concentrations.

Altitude

The Altitude Effect Study was designed to evaluate the performance of Rightest and GE series Blood Glucose Monitoring Systems (GM700/GM650/GE200/GE300) under environment factors at 3 different altitudes: 0, 1000, 2000 and 3000 meters (0, 3280, 6561, and 9842 ft.) above sea level. Test procedures were performed in 3 replicates using 3 lots of revised test strips (GS650/ GS700/GS750, HCT range 20-65%) with glucose control solutions at 3 levels - low, normal and high concentration. The results were compared against reference method YSI 2300 Analyzer.

The obtained study results indicate that the stability of bias compared to the REF during various operated altitude were all within the criteria.

Operation Temperature and Humidity Study

The Operating Temperature and Humidity Study was designed to evaluate the performance of revised GDH test strip under environment factors of operated temperature and humidity. 6 various temperature and humidity conditions using three lots of revised test strips (GS650/GS700/GS750, HCT 20-65%) with glucose control solutions at 3 levels - low, normal and high concentration, and each sample was tested in 5 replicates. The results were compared against reference method YSI 2300 Analyzer.

The obtained study results indicate that the stability of bias compared to the REF during various operated temperature and humidity were all within the criteria. The results demonstrated operation conditions at 43-111°F and 10-90% humidity remain unaffected with the HCT specification change.

{21}------------------------------------------------

Hematocrit Study

The Hematocrit Study was designed to evaluate the Hematocrit levels for revised GS650/GS700/GS750 GDH test strip with HCT range from 20-65%, 6 glucose concentration levels of venous blood samples were evaluated. Three lots of revised test strips (GS650/GS700/GS750) were used, and each sample was tested in 5 replicates. The results were compared against reference method YSI 2300 Analyzer.

The obtained study results indicate that the stability compared to the REF during various HCT ranges were all within the criteria. The acceptable HCT range of the revised test strips is 20-65%.

These studies and evaluations were either performed internally by professional personnel in Bionime or by third party, performed by qualified personnel, with proper calibrated/maintained equipment and under properly-controlled environmental conditions. All of the evaluated performances passed and meet the acceptance criteria set in the study protocol.

9. Discussion of Clinical Tests Performed:

Three meter models were used for testing: GM650/GE300, GM700, and GE200. GDH (GS700/GS650/GS750) test strip was used for testing. GS700. GS650, and GS750 test strips are identical to each other for Rightest brand, GE200 and GE300 test strips are identical to GS750 and GS650 with the exception of trade name.

System Accuracy Study (Whole Blood):

The accuracy studies of Rightest™ GS650/GS700/GS750 Strip with revised HCT range with previous FDA cleared Rightest Blood Glucose Monitoring Systems: GM700, GM650/GE200 and GE300 meters, were performed by comparing whole blood (plasma equivalent) glucose values on the Rightest™ GM700/GM650 and GE200/GE300 GDH Strip with revised HCT range with plasma glucose values on reference lab instrument.

A total of 120 patients were participated. The study result demonstrates that the accuracy of Rightest™ GM700/GM650/GE200/GE300 GDH Strip with revised HCT range met the acceptance criteria, there was no significant difference in slope, intercept, and correlation coefficient.

Lay User Study

A Lay User Performance Evaluation Study was performed to demonstrate that lay users could obtain accurate results using the subject devices. The study was performed using capillary whole blood from fingertip, palm and forearm sample sites by evaluating total of 100 laypersons. The study result shows substantial equivalence to predicate device used in finger, palm and forearm position.

{22}------------------------------------------------

10. Conclusions:

Results of performance evaluation of the Rightest™ GM700/GM650 and GE200/GE300 GDH Strip with revised HCT range demonstrates that:

Subject devices Rightest BGMS GM700/GM650 and GE200/GE300 BGMS with revised HCT specification are substantially equivalent to the predicate devices Rightest Blood Glucose Monitoring System GM700/GM650 and GE200/GE300 BGMS (K123008) in commercial distribution.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.