K Number
K141220
Date Cleared
2014-05-22

(10 days)

Product Code
Regulation Number
862.2570
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Applied Biosystems® 7500 Fast Dx Real-Time PCR instrument with the SDS Software is a real-time nucleic acid amplification and five color fluorescence detection system for use with FDA cleared or approved tests on human-derived specimens. The 7500 Fast Dx Real-Time PCR instrument and SDS Software are intended for use in combination with in vitro diagnostic tests labeled for use on this instrument. The 7500 Fast Dx instrument is intended for use by laboratory professionals trained in laboratory techniques, procedures, and on use of the system.

Device Description

The Applied Biosystems® 7500 Fast Dx Real-Time PCR instrument integrates a thermal cycler, a fluorimeter and application specific software. The instrument houses the thermal cycler and the fluorimeter, while the application software is run on a PC that is attached to the instrument. Samples are placed in a tube strip or 96-well low-head space plate that is moved to a Peltier-based thermal block and positioned relative to the optics using a tray loading mechanism. Excitation for all samples is provided by a halogen tungsten white source that passes through 5 switchable excitation filters prior to reaching the sample. Fluorescence emission is then detected through 5 color emissions filter wheel to a charge coupled device (CCD) camera. The instrument is designed to complete quantitative RT-PCR runs in about 40 minutes. The Sequence Detection Software (SDS) is used for instrument control, data collection and data analysis. The software provides a wizard for user-friendly set-up. The software measures cycleby-cycle real-time signals from the sample and provides a variety of tools to help the user analyze the data extracted from the samples. In addition, the software provides lamp-life monitoring and other instrument maintenance information. The software runs as an application on the Windows 7 platform.

AI/ML Overview

This document is a 510(k) summary for the Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software. It describes a Class II medical device used for real-time nucleic acid amplification and fluorescence detection.

Here's an analysis of the provided information regarding acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state acceptance criteria or report specific device performance metrics (e.g., sensitivity, specificity, accuracy) in a quantitative manner. Instead, the focus of this 510(k) is on demonstrating substantial equivalence to a predicate device due to minor modifications.

The summary states: "Based on the risk analysis evaluation results, verification testing was conducted to support the modifications in the instrument computer operating system. The verification testing report included in the submission supports substantial equivalence to the predicate device."

This implies that the "acceptance criteria" are implicitly met by demonstrating that the updated device performs equivalently to the predicate, particularly in light of the operating system change. The performance of the predicate device (K082562) would have been established previously.

A conceptual table based on the document's content would look something like this:

Acceptance Criteria (Implied)Reported Device Performance
Functional equivalence to predicate device (K082562)Verification testing supports substantial equivalence to the predicate device, especially regarding the operating system modification.
No adverse impact on safety and effectiveness due to minor changes"The changes are not substantive changes in the use of the device and do not affect the safety and effectiveness of the device when used as labeled."
Maintains real-time PCR functionality and five-color detectionDevice continues to operate as a "real-time nucleic acid amplification and five color fluorescence detection system."
Compatibility with FDA cleared/approved testsIntended for "use with FDA cleared or approved tests on human-derived specimens" and "in combination with in vitro diagnostic tests labeled for use on this instrument."

2. Sample Size Used for the Test Set and Data Provenance:

The document mentions "verification testing was conducted to support the modifications in the instrument computer operating system." However, it does not provide any specific sample sizes for this testing. It also does not specify the data provenance (e.g., country of origin, retrospective or prospective nature).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

The document does not describe the use of experts to establish ground truth for this specific verification testing. Given that the changes were primarily to the operating system and the focus is on substantial equivalence in functionality, this type of expert review is unlikely to have been a primary component of this particular submission.

4. Adjudication Method for the Test Set:

No information is provided regarding an adjudication method. This is not typically relevant for a submission focused on demonstrating functional equivalence of an instrument's operating system update.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not conducted or described in this document. This type of study is more common for diagnostic algorithms or imaging devices where human interpretation is a key component and the AI aims to assist or replace it. This document concerns a PCR instrument and its software, not an AI-assisted diagnostic interpretation tool.

6. If a Standalone Performance Study was done:

The document refers to "verification testing." While this testing would have evaluated the device's standalone performance after the operating system upgrade, it is presented in the context of demonstrating substantial equivalence to the predicate, rather than a de novo standalone performance study providing new, extensive performance metrics like sensitivity and specificity. The document states: "The verification testing report included in the submission supports substantial equivalence to the predicate device."

7. The Type of Ground Truth Used:

The document does not explicitly state the type of ground truth used for the verification testing. Given that it's a real-time PCR instrument, the ground truth for any performance validation would typically involve:

  • Reference standards/materials: Well-characterized samples with known concentrations of nucleic acid targets.
  • Comparison to a reference method: Performance compared against a gold standard or extensively validated method.
  • Known positive/negative controls: Established biological or synthetic samples with known outcomes.

The testing would primarily confirm that the upgraded system continues to accurately detect and quantify nucleic acids as expected, without being negatively affected by the new operating system.

8. The Sample Size for the Training Set:

This device is an instrument and its software, not a machine learning model that requires a "training set" in the typical sense of AI development. Therefore, there is no concept of a "training set" as described for AI studies. The software handles instrument control, data collection, and data analysis using established algorithms, not by learning from a large dataset.

9. How the Ground Truth for the Training Set was Established:

As there is no "training set" in the context of AI for this device, this question is not applicable. The software's functionality is based on pre-programmed algorithms for signal processing and quantification inherent to real-time PCR, not on learning from labelled data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Thermo Fisher Scientific. The words "Thermo Fisher" are in bold, and the word "SCIENTIFIC" is in a smaller font below. The logo is simple and professional, and it is likely used on the company's website, products, and marketing materials.

MAY 2 2 2014

Special 510(k) Summary

Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software

Submitter:Thermo Fisher ScientificLife Sciences Solutions GroupLife Technologies Holdings Pte LtdBlk 33, #07-06, Marsiling Industrial Estate, Road 3Singapore 739256
Manufacturer:Life Technologies Holdings Pte LtdBlk 33, #07-06, Marsiling Industrial Estate, Road 3Singapore 739256
Contact:Nikki Arora, Engineer, Regulatory AffairsPhone: 650-554-2268Fax: 240-379-4673Email: nikki.arora@thermofisher.com
Alternate Contact:Nicole Fisher, Manager, Regulatory AffairsPhone: 240-379-4602Fax: 240-379-4673Email: nicole.fisher@thermofisher.com

Date Prepared: May 22, 2014

Device Information:

Proprietary Trade Name: Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software

Common Name: 7500 Fast Dx

Device Class: Class II

Classification Name: Instrumentation for Clinical Multiplex Test Systems (21 CFR 862.2570)

Product Code: NSU

Predicate Device: K082562, Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software version 1.4

Device Description: The Applied Biosystems® 7500 Fast Dx Real-Time PCR instrument integrates a thermal cycler, a fluorimeter and application specific software. The instrument houses the thermal cycler and the fluorimeter, while the application software is run on a PC that is attached to the instrument. Samples are placed in a tube strip or 96-well low-head space plate that is moved to a Peltier-based thermal block and positioned relative to the optics using a tray loading mechanism.

Excitation for all samples is provided by a halogen tungsten white source that passes

{1}------------------------------------------------

through 5 switchable excitation filters prior to reaching the sample. Fluorescence emission is then detected through 5 color emissions filter wheel to a charge coupled device (CCD) camera. The instrument is designed to complete quantitative RT-PCR runs in about 40 minutes.

The Sequence Detection Software (SDS) is used for instrument control, data collection and data analysis. The software provides a wizard for user-friendly set-up. The software measures cycleby-cycle real-time signals from the sample and provides a variety of tools to help the user analyze the data extracted from the samples. In addition, the software provides lamp-life monitoring and other instrument maintenance information. The software runs as an application on the Windows 7 platform.

Intended Use/Indications of Use: The Applied Biosystems® 7500 Fast Dx Real-Time PCR instrument with the SDS Software is a real-time nucleic acid amplification and five color fluorescence detection system for use with FDA cleared or approved tests on human-derived specimens. The 7500 Fast Dx Real-Time PCR instrument and SDS Software are intended for use in combination with in vitro diagnostic tests labeled for use on this instrument. The 7500 Fast Dx instrument is intended for use by laboratory professionals trained in laboratory techniques, procedures, and on use of the system.

Minor changes were made to the intended use in order to remove the specific software version. to modernize, and to implement consistency. The changes are not substantive changes in the use of the device and do not affect the safety and effectiveness of the device when used as labeled.

Summary of technological characteristics of the device compared to the predicate device:

The Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software is substantially equivalent to the predicate device Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software version 1.4 (K082562). A comparison table is provided below:

ItemSubject Device: AppliedPredicate Device: Applied
Biosystems® 7500 Fast DxBiosystems® 7500 Fast Dx
Real-Time PCR InstrumentReal-Time PCR Instrument
with SDS Softwarewith SDS Software version 1.4
Indications for UseThe Applied Biosystems 7500Fast Dx Real-Time PCRinstrument with the SDSSoftware is a real-time nucleicacid amplification and fivecolor fluorescence detectionsystem for use with FDAcleared or approved tests onhuman-derived specimens.The 7500 Fast Dx Real-TimePCR instrument and SDSSoftware are intended for useApplied Biosystems® 7500Fast Dx Real-Time PCRinstrument with the SDSSoftware version 1.4 is a real-time nucleic acid amplificationand detection system thatmeasures nucleic acid signalsfrom reverse transcribed RNAand converts them tocomparative quantitativereadouts using fluorescentdetection of dual-labeled

Table 1: Predicate Comparison Table

{2}------------------------------------------------

in combination with in vitrodiagnostic tests labeled foruse on this instrument. The7500 Fast Dx instrument isintended for use by laboratoryprofessionals trained inlaboratory techniques,procedures, and on use of thesystem.hydrolysis probes. The 7500Fast Dx instrument is to beused only by technologiststrained in laboratorytechniques, procedures, andon use of the analyzer.
Fundamental TechnologyReal-Time PCRSame
Instrument ComputerOperating SystemWindows 7Windows XP

Non-clinical Testing Performed for Determination of Substantial Equivalence: Based on the risk analysis evaluation results, verification testing was conducted to support the modifications in the instrument computer operating system. The verification testing report included in the submission supports substantial equivalence to the predicate device.

Conclusion: The Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software has the same operating principle and technological characteristics as the previously cleared device. The changes in intended use were made only to remove the software version and to modernize the language; there is no change in the use of the device. In summary, the Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software is substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, uppercase letters. The text is centered horizontally and appears to be the heading of a document or a sign. The font is a serif typeface, and the overall impression is formal and official.

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

THERMO FISHER SCIENTIFIC (LIFE TECHNOLOGIES HOLDINGS PTE LTD.) NIKKI ARORA REGULATORY AFFAIRS ENGINEER May 22, 2014 BLK 33, MARSILING INDUSTRIAL ESTATE ROAD 3, NO. 07-06 SINGAPORE 739256

Re: K141220

Trade/Device Name: Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument with SDS Software Regulation Number: 21 CFR 862.2570 Regulation Name: Instrumentation for clinical multiplex test systems Regulatory Class: II Product Code: NSU Dated: May 13, 2014 Received: May 14, 2014

Dear Ms. Arora:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I odotal Juliates and registanents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{4}------------------------------------------------

Page 2-Ms. Arora

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

John Hobson -S for

Sally Hojvat, M.Sc., Ph.D. Director, Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

.

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510{k) Number (if known) K141220

Device Name

Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software

Indications for Use (Describe)

The Applied Biosystems 7500 Fast Dx Real-Time PCR instrument with the SDS Software is a real-time nucleic acid amplification and five color fluorescence detection system for use with FDA cleared or approved tests on human-derived specimens. The 7500 Fast Dx Real-Time PCR instrument and SDS Software are intended for use in combination with in vitro diagnostic tests labeled for use on this instrument. The 7500 Fast Dx instrument is intended for use by laboratory professionals trained in laboratory techniques, procedures, and on use of the system.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR : Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

John Hobson -S
2014.05.22 12:33:44 -04:00

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

§ 862.2570 Instrumentation for clinical multiplex test systems.

(a)
Identification. Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.