Device Specific Reports provide the user information compiled from data generated by networked devices and stored in the PharmGuard Server. This information may be used to support continuous quality improvement programs. Reports generated from the data can be used to analyze and trend various aspects of the designated infusion pump systems and therapies used.
The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 software is a medical device intended to be used in the analysis of infusion data, provide information used in forming clinical workflow decisions and improve infusion safety. The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 is an infusion pump accessory compatible with the PharmGuard® Server, which stores data exported from networked infusion pumps. The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 copies data from the PharmGuard® Server databases and uses it to create reports.
The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 can be used for transferring, displaying or storing and retrieving historical clinical data, clinical and operational alarm events, fault events, power events, maintenance events, therapy and therapy change events and telemetry events that are generated by the networked infusion pumps and stored in the PharmGuard® Server. It is also intended for near real-time monitoring of infusion pump history. For the purposes of this device "near real-time" is associated with time intervals typically measured in minutes rather than in seconds or in hours.
The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 manipulates and re-configures data to present information bevond the discrete data that is displayed on the pumps.
Reports include information on drug compliance, safety events, alarm history, drug program utilization, and pump technical status. Reports can summarize data and reveal trends in these areas by presenting results summarized by such parameters as time of day, facility, drug profile, alarm type and safety events. The information may be presented graphically, in table form and in some cases allows "drill through" to the underlying data used to produce the reports.
The Device Specific Reports Software comprises:
- . Instructions for Use and Help Files
- An executable file for launching the application .
- A Configuration Tool that provides a user interface to enter, update, and apply the . configuration parameters that operate the Data Mart associated with the Medfusion 4000® Device Specific Reports and associated SQL Server databases
- Metadata that is imported into the PharmGuard® Server System and that . determines what information within the PharmGuard® Server is available to be output via the Device Specific Report software
- Infrastructure files that interact to facilitate and coordinate the data acquisition, . storage, and processing between the PharmGuard® Repository and the Medfusion® 4000 Pump Device Specific Report data mart. The Reporting Infrastructure performs the following major tasks:
- o Extract, Transform, and Load (ETL) data from the PharmGuard Repository into the Reporting Infrastructure data mart.
- o Store device event data in the data mart
- Process requests for data retrieval for device specific reports 0
I am sorry, but based on the provided text, I cannot provide a table of acceptance criteria and reported device performance or other detailed study information. The document is a 510(k) summary for the Smiths Medical Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0.
Here's what the document does state regarding performance and studies:
- Performance Testing: "Smiths Medical performed software validation testing on the Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0."
- Testing Conclusion: "All testing met pre-established specifications, and successfully demonstrated that the device performed as intended."
- Clinical Studies: "Human clinical studies were deemed unnecessary to evaluate the safety or effectiveness of the Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0."
- Equivalence: The device was found substantially equivalent to a predicate device (K111386, Medfusion® Model 4000 Series Syringe Infusion Pump, PharmGuard® Toolbox 2.0 Medication Safety Software).
However, the document does not contain any specific details regarding:
- The actual "pre-established specifications" that served as acceptance criteria.
- The quantitative results of the "software validation testing" to show how the device performed against those criteria.
- Sample sizes for test sets or training sets.
- Data provenance, expert qualifications, adjudication methods, or ground truth types.
- Any multi-reader multi-case studies or standalone algorithm performance.
Therefore, I cannot fulfill your request for a table of acceptance criteria and reported performance, nor any of the other detailed study information, as these specifics are not present in the provided 510(k) summary.
{0}------------------------------------------------
smiths medical
bringing technology to life
Smiths Medical 1265 Grey Fox Road St. Paul, MN 55112 T: 651 633 2556 F: 651 628 7459 www.smiths-medical.com
510(k) Summary
510(k) 141193
JUL 2 5 2014
DATE PREPARED: 16-JULY-2014
I. GENERAL INFORMATION
| Applicant's Nameand Address: | Smiths Medical ASD, Inc.1265 Grey Fox RoadSt. Paul, MN 55112 |
|---|---|
| Contact Person: | Tom BlissSr. Specialist, Regulatory Affairs |
| Common/UsualName: | Infusion Safety Management Software |
| Product Code: | PHC |
| Proprietary Name: | Medfusion® 4000 Pump Device Specific Reports Software forPharmGuard® Server, Version 2.0 |
| Predicate Device/510(k): | K111386, Medfusion® Model 4000 Series Syringe Infusion Pump,PharmGuard® Toolbox 2.0 Medication Safety Software |
II. DEVICE DESCRIPTION
The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 software is a medical device intended to be used in the analysis of infusion data, provide information used in forming clinical workflow decisions and improve infusion safety. The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 is an infusion pump accessory compatible with the PharmGuard® Server, which stores data exported from networked infusion pumps. The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 copies data from the PharmGuard® Server databases and uses it to create reports.
{1}------------------------------------------------
The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 can be used for transferring, displaying or storing and retrieving historical clinical data, clinical and operational alarm events, fault events, power events, maintenance events, therapy and therapy change events and telemetry events that are generated by the networked infusion pumps and stored in the PharmGuard® Server. It is also intended for near real-time monitoring of infusion pump history. For the purposes of this device "near real-time" is associated with time intervals typically measured in minutes rather than in seconds or in hours.
The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 manipulates and re-configures data to present information bevond the discrete data that is displayed on the pumps.
Reports include information on drug compliance, safety events, alarm history, drug program utilization, and pump technical status. Reports can summarize data and reveal trends in these areas by presenting results summarized by such parameters as time of day, facility, drug profile, alarm type and safety events. The information may be presented graphically, in table form and in some cases allows "drill through" to the underlying data used to produce the reports.
The Device Specific Reports Software comprises:
- . Instructions for Use and Help Files
- An executable file for launching the application .
- A Configuration Tool that provides a user interface to enter, update, and apply the . configuration parameters that operate the Data Mart associated with the Medfusion 4000® Device Specific Reports and associated SQL Server databases
- Metadata that is imported into the PharmGuard® Server System and that . determines what information within the PharmGuard® Server is available to be output via the Device Specific Report software
- Infrastructure files that interact to facilitate and coordinate the data acquisition, . storage, and processing between the PharmGuard® Repository and the Medfusion® 4000 Pump Device Specific Report data mart. The Reporting Infrastructure performs the following major tasks:
- o Extract, Transform, and Load (ETL) data from the PharmGuard Repository into the Reporting Infrastructure data mart.
- o Store device event data in the data mart
- Process requests for data retrieval for device specific reports 0
III. DEVICE INTENDED USE
Device Specific Reports provide the user information compiled from data generated by networked devices and stored in the PharmGuard Server. This information may be used to support continuous quality improvement programs. Reports generated from the data can be used to analyze and trend various aspects of the designated infusion pump systems and therapies used.
{2}------------------------------------------------
IV. SUMMARY OF STUDIES
Performance Testing
Smiths Medical performed software validation testing on the Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0
Clinical Studies
Human clinical studies were deemed unnecessary to evaluate the safety or effectiveness of the Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0
Testing Conclusion
All testing met pre-established specifications, and successfully demonstrated that the device performed as intended. The testing results allowed for a conclusion to be made that the Medfusion® Model 4000 Series Syringe Infusion Pump with the Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 was as safe and effective as the predicate device.
VII. STATEMENT OF EQUIVALENCE
The Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 is an accessory to the Medfusion® Model 4000 Series Syringe Infusion Pump. The design and use of the Medfusion® 4000 Pump Device Specific Reports Software. Version 2.0 does not affect the device classification, intended use, technological characteristics, or risks associated with the pump. Validation tests confirmed the suitability of the Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 device for its intended use. The test results did not raise new safety or performance questions, and confirmed that the Medfusion® Model 4000 Series Syringe Infusion Pump with the Medfusion® 4000 Pump Device Specific Reports Software, Version 2.0 accessory is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes extending from its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Public I lealth Service
Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 25, 2014
Thomas Bliss Senior Regulatory Affairs Specialist Smiths Medical ASD. Inc. 1265 Grey Fox Road St. Paul, Minnesota 55112
Re: K141193
Trade/Device Name: Medfusion 4000 Pump Device Specific Reports Software for PharmGuard Server Version 2.0 Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: PHC Dated: May 7, 2014 Received: May 8, 2014
Dear Mr. Bliss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Bliss
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
K141193 510(k) Number (if known):
Device Name:
Medfusion® 4000 Pump Device Specific Reports Software for PharmGuard® Server, Version 2.0
Indications for Use:
Device Specific Reports provide the user information compiled from data generated by networked devices and stored in the PharmGuard Server. This information may be used to support continuous quality improvement programs. Reports generated from the data can be used to analyze and trend various aspects of the designated infusion pump systems and therapies used.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary E. Brooks -A 2014.07.17 14:21:48 -04'00'
§ 880.5725 Infusion pump.
(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).