(29 days)
RANGE /DENALI/MESA and SMALL STATURE and ARI are cleared for the following indications: Non-cervical fixation as an adjunct to fusion for the following indications: degenerative disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stcnosis: curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor: pscudarthrosis: and/or failed previous fusion.
Except for hooks, when used as an anterolatcral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.
Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatric patients. The Range Spinal System for pediatric use is intended to be used with autograft. Pediatric pediatric pediale screw fixation is limited to a posterior approach.
The Range Spinal System is a top-loading, multiple component, posterior (thoracic-lumbar) spinal fixation system which consist of pedicle screws, rods, locking set screws, hooks. rod connectors and transverse connectors.
Materials: The devices are manufactured from Titanium Alloy and Cobalt Chrome per ASTM and ISO standards.
Function: The system functions as an adjunct to fusion to provide immobilization and stabilization of the posterior thoracic and lumbar spine.
The purpose of this submission is to add 6.35mm implants to the system.
This FDA submission describes a spinal fixation system, not an AI/ML powered device. As such, the typical acceptance criteria and study designs for AI/ML devices (like accuracy, sensitivity, specificity, and MRMC studies) are not applicable.
The submission focuses on establishing substantial equivalence to previously marketed devices based on design features and material properties. The primary way this device meets its "acceptance criteria" is by demonstrating it is as safe and effective as existing, legally marketed spinal systems.
Here's an analysis based on the provided document, addressing the original prompt's categories where applicable for a non-AI device:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category (for this type of device) | Reported Device Performance (from the document) |
|---|---|
| Material Composition Conformity | Manufactured from Titanium Alloy and Cobalt Chrome per ASTM and ISO standards. |
| Mechanical Performance (Static Compression Bending) | Worst case components previously tested in static compression bending in accordance with ASTM F1717. Proposed implants determined not to represent a new worst case. |
| Mechanical Performance (Static Torsion) | Worst case components previously tested in static torsion in accordance with ASTM F1717. Proposed implants determined not to represent a new worst case. |
| Mechanical Performance (Dynamic Compression) | Worst case components previously tested in dynamic compression in accordance with ASTM F1717. Proposed implants determined not to represent a new worst case. |
| Design Feature Substantial Equivalence | Design features of components were compared to predicate devices and found to be substantially the same. |
| Intended Use Substantial Equivalence | Indications for use are the same as predicate devices; the submission specifically adds 6.35mm implants to the existing system. |
2. Sample size used for the test set and the data provenance
- Sample Size: Not applicable in the context of clinical "test sets" for AI/ML performance. For mechanical testing, the "worst case components" were selected for testing, implying a representative sample of components or configurations were tested to cover the range of mechanical properties. The specific number of components or implants tested is not provided in this summary.
- Data Provenance: Not applicable in the context of clinical data provenance. The mechanical test data would be generated in a laboratory setting (likely within the company or a certified testing facility).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable as this is a mechanical device, not an AI/ML diagnostic or prognostic tool requiring expert interpretation of outputs to establish ground truth. Substantial equivalence for this device is based on technical comparisons and mechanical testing against recognized standards.
4. Adjudication method for the test set
- Not applicable. There is no "adjudication" in the sense of reconciling clinical interpretations for this type of device. The determination of "worst case" for mechanical testing would be an engineering assessment.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is specifically for evaluating the impact of AI on human reader performance, which is not relevant for a spinal implant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No, a standalone algorithm performance evaluation was not done. This device is a physical medical implant, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For mechanical testing, the "ground truth" is defined by established engineering standards (ASTM F1717) and successful performance within those standards, demonstrating adequate strength and durability.
- For substantial equivalence, the "ground truth" is the established safety and effectiveness profile of the predicate devices.
8. The sample size for the training set
- Not applicable. This device does not use an AI/ML "training set."
9. How the ground truth for the training set was established
- Not applicable. As there is no AI/ML component, there is no "training set" or ground truth for it.
Summary regarding the device's "acceptance":
The Range Spinal System gained FDA clearance (K141147) by demonstrating substantial equivalence to predicate devices already on the market. This means the FDA concluded that the new device is as safe and effective as the existing devices. The key elements for this determination were:
- Design and Material Comparison: The new components (6.35mm implants) were found to be "substantially the same" in design features and materials (Titanium Alloy and Cobalt Chrome per ASTM/ISO standards) as predicate devices.
- Mechanical Performance: The "worst case components" of the system were previously tested against established standards (ASTM F1717) for static compression bending, static torsion, and dynamic compression. The proposed new implants were determined by engineering analysis not to represent a "new worst case," implying they perform comparably or better than previously tested components and meet the required mechanical integrity.
- Intended Use: The indications for use are consistent with those of the predicate devices.
Essentially, the "study" for this device was a combination of engineering analysis and mechanical testing against industry standards, alongside a direct comparison of its technical characteristics and intended use to already cleared predicate devices.
{0}------------------------------------------------
JUN 0 3 2014
510(k) Summary for the Range (Mesa/Denali) Spinal System
This 510(k) summary for the Range Spinal System is provided as required per Section 513(i)(3) of the Food, Drug and Cosmetic Act.
1. Submitter :
:
ે.
K2M, Inc. 751 Miller Drive SE. Suite F1 Leesburg, VA 20175 Contact Person : Nancy Giezen K2M, Inc. Telephone: 703-777-3155
Date Prepared: 06//03/14
| 2. Tradename: | Range Spinal System, |
|---|---|
| Common Name: | Spinal Fixation System |
| Classification Name: | Pedicle Screw Spinal System (21CFR 888.3070)Spinal Interlaminal Fixation Orthosis (21CFR 888.3050)Orthosis, Spondylolisthesis Spinal Fixation (21CFR 888.3070 |
| Device Product Code: | MNI, KWP, MNH, OSH, NKB |
| Regulatory Class: | Class II, III |
3. Predicate or legally marketed devices which are substantially equivalent :
- K2M Range K052404, K070229, K080792, K121016, K121630, K133944 .
- Medtronic CD Horizon K961633, K090390, K132328 .
- � LDR EasySpine K082592
4. Description of the device:
The Range Spinal System is a top-loading, multiple component, posterior (thoracic-lumbar) spinal fixation system which consist of pedicle screws, rods, locking set screws, hooks. rod connectors and transverse connectors.
Materials: The devices are manufactured from Titanium Alloy and Cobalt Chrome per ASTM and ISO standards.
Function: The system functions as an adjunct to fusion to provide immobilization and stabilization of the posterior thoracic and lumbar spine.
The purpose of this submission is to add 6.35mm implants to the system.
5. Intended Use:
RANGE /DENALI/MESA and SMALL STATURE and ARI are cleared for the following indications:
Non-cervical fixation as an adjunct to fusion for the following indications: degenerative disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stcnosis: curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor: pscudarthrosis: and/or failed previous fusion.
Except for hooks, when used as an anterolatcral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.
Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatric patients. The Range Spinal
{1}------------------------------------------------
- System for pediatric use is intended to be used with autograft. Pediatric pediatric pediale screw fixation is limited to a posterior approach.
-
- Comparison of the technological characteristics of the device to predicate and legally marketed devices : The design features of the components were compared to predicate devices and were found to be substantially the same as these systems.
7. Comparison of the performance data of the device to predicate and legally marketed devices :
Worst case components of the Range Spinal System were previously tested in static compression bending . static torsion and dynamic compression in accordance with ASTM F1717. The proposed implants were determined not to represent a new worst case (engineering analysis) and are considered substantially equivalent to the predicate devices.
8. Conclusion:
There are no significant differences between the Range Spinal Systems currently being marketed which would adversely affect the use of the product. They are substantially equivalent to these other devices in design, function, material and intended use.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3, 2014
K2M, Incorporated Ms. Nancy Giezen Manager, Regulatory Affairs 751 Miller Drive SE, Suite F1 Leesburg, Virginia 20175
Re: K141147
Trade/Device Name: Range Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI, MNH, KWP Dated: May 2, 2014 Received: May 5, 2014
Dear Ms. Giezen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Nancy Giezen
[f you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K141147
Device Name Range Spinal System
Indications for Use (Describe)
RANGE /DENALI/MESA and SMALL STATURE and ARI are cleared for the following indications: Non-cervical fixation as an adjunct to the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor: pscudarthrosis: and/or failed previous fusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.
Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatic patients. The Range Spinal System for pediatric use is intended to be used with autograft. Pediatic pedicle screw fixation is limited to a posterior approach.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
Zane W-Wyatt
Division of Orthopedic Devices
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
N/A