K Number
K141073

Validate with FDA (Live)

Date Cleared
2014-06-23

(59 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Liquichek Tumor Marker Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.

Device Description

Liquichek Tumor Marker Control, a trilevel liquid control is prepared from human source material with added constituents of human and animal origin, stabilizers and preservatives.

AI/ML Overview

Here's a summary of the acceptance criteria and the study details for the Liquichek Tumor Marker Control, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Thawed and Opened Stability10 days at 2 to 8 °C (Met acceptance criteria)
Thawed and Unopened Stability15 days at 2 to 8 °C (Met acceptance criteria)
Shelf Life Stability28 months at -20°C to -50°C (Met acceptance criteria)

Study Details

  1. Sample size used for the test set and the data provenance: Not explicitly stated. The document mentions "replicate analyses" and "a representative sampling of this lot of product" for value assignment, but no specific number of samples or details on provenance beyond the manufacturer and/or independent laboratories. The document indicates the human source material used to manufacture the control was tested by FDA accepted methods and found non-reactive for specific viruses.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for this type of device. The study is for stability claims of a quality control material, not for diagnostic accuracy requiring expert consensus on clinical cases. Ground truth for control materials is typically established through precise analytical methods and value assignment processes.

  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This study does not involve adjudication of expert readings or clinical interpretations.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a stability study for a quality control material, not an AI-powered diagnostic device or a comparative effectiveness study involving human readers.

  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is a control material, not an algorithm.

  6. The type of ground truth used:

    • Value Assignment: The mean values and ±3SD ranges printed in the product insert were derived from replicate analyses using manufacturer supported reagents and a representative sampling of the product. These values serve as the established "truth" for monitoring precision.
    • Stability Studies: For stability, the ground truth is the initial established value of the control material, against which subsequent measurements over time (under various conditions) are compared to ensure they remain within acceptable ranges. The acceptance criteria themselves define the acceptable deviation from this established truth.
  7. The sample size for the training set: Not applicable. This device is a quality control material and does not involve artificial intelligence or machine learning that requires a training set.

  8. How the ground truth for the training set was established: Not applicable, as there is no training set for this device.

{0}------------------------------------------------

510(k) Summary Liquichek Tumor Marker Control K141073

1.0 Submitter

Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, California 92618-2017 Telephone: (949) 598-1200 (949) 598-1557 Fax:

Contact Person

Suzanne Parsons Regulatory Affairs Manager Telephone: (949) 598-1467

Date of Summary Preparation

June 20, 2014

2.0 Device Identification

Product Trade Name:Liquichek Tumor Marker Control
Common Name:Multi-Analyte Controls, All Kinds (Assayed)
Review Panel:Clinical Chemistry and Clinical Toxicology Devices
Classifications:Class I, Reserved
Product Code:JJY
Regulation Number21 CFR 862.1660

3.0 Device to Which Substantial Equivalence is Claimed

Liquichek Tumor Marker Control Bio-Rad Laboratories Irvine, California

510 (k) Number: K071675

4.0 Description of Device

Liquichek Tumor Marker Control, a trilevel liquid control is prepared from human source material with added constituents of human and animal origin, stabilizers and preservatives.

The human source material used to manufacture this control was tested by FDA accepted methods and found non-reactive for Hepatitis B Surface Antigen (HBsAg), antibody to Hepatitis C (HCV) and antibody to HIV-1/HIV-2.

JUN 2 3 2014

{1}------------------------------------------------

5.0 Value Assignment

The mean values and the corresponding ±3SD ranges printed in this insert were derived from replicate analyses and are specific for this lot of product. The tests listed were performed by the manufacturer and/or independent laboratories using manufacturer supported reagents and a representative sampling of this lot of product. It is recommended that each laboratory establish its own acceptable ranges and use those provided only as guides. Laboratory established ranges may vary from those listed during the life of this control. Variations over time and between laboratories may be caused by differences in laboratory technique, instrumentation and reagents, or by manufacturer test method modifications.

6.0 Intended use

Liguichek Tumor Marker Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.

7.0 Comparison of the new device with the Predicate Device

The new Liquichek Tumor Marker Control claims substantial equivalence to the Liguichek Tumor Marker Control currently in commercial distribution (K071675). Table 1 (below) contains comparison information of similarities and differences between the new and predicate device to which substantial equivalence is claimed.

(This portion of the page is intentionally left blank.)

{2}------------------------------------------------

Table 1: Similarities and Differences between new and predicate device

CharacteristicsLiquichek Tumor Marker Control(New Device)Liquichek Tumor Marker Control(Predicate Device, K071675)
Similarities
Intended UseThis product is intended for use as an assayedquality control serum to monitor the precision oflaboratory testing procedures for the analyteslisted in the package insert.This product is intended for use as an assayedquality control serum to monitor the precision oflaboratory testing procedures listed in thepackage insert.
FormLiquidLiquid
MatrixHuman source material and constituents ofanimal originHuman source material and constituents ofanimal origin
Differences
Test SystemSiemens Dimension Vista SystemsMultiple Analyzers
Vial TypeSiemens Dimension Vista VialsGeneral control vials
Thawed and Unopened(On-Board Stability)15 days at 2 - 8 °C60 days at 2 - 8 °C except for:IGF-1, PAP: 35 daysFree PSA: 30 daysCA 125: 14 days
Thawed & Opened10 days at 2 - 8 °C30 days at 2 - 8 °C except for:IGF-1: 15 daysCA 125: 12 days
Storage Unopened(Shelf life)At -20 to -50 °C until the expiration dateAt -20 to -70 °C until the expiration date
AnalytesContains:Alpha Fetoprotein (AFP)Beta-2-Microglobulin (B2-M)CA 15-3CA 19-9CA 125Carcinoembryonic Antigen (CEA)FerritinHuman Chorionic Gonadotropin (hCG)ProlactinProstate Specific Antigen, Total (Total PSA)Prostate Specific Antigen, Free (Free PSA)Does not contain:CA 27.29Prostatic Acid Phosphatase (PAP)Insulin-like Growth Factor-I (IGF-1)Thyroglobulin (Tg)Contains:Alpha Fetoprotein (AFP)Beta-2-Microglobulin (B2-M)CA 15-3CA 19-9CA 125Carcinoembryonic Antigen (CEA)FerritinProlactinProstate Specific Antigen, Total (Total PSA)Prostate Specific Antigen, Free (Free PSA)CA 27.29Prostatic Acid Phosphatase (PAP)Insulin-like Growth Factor-I (IGF-1)Human Chorionic Gonadotropin (hCG)Thyroglobulin (Tg)

{3}------------------------------------------------

8.0 Statement of Supporting Data

Real-time stability studies were conducted to establish the thawed opened and thawed unopened stability claims. Accelerated stability studies were conducted to establish the shelf-life stability claim. Acceptance Criteria were met to support the product claims as follows:

Thawed and Opened Stability:10 days at 2 to 8 °C
Thawed and Unopened Stability:15 days at 2 to 8 °C
Shelf Life Stability:28 months at -20°C to -50°C

9.0 Conclusion

Based on the performance characteristics indicated above, the Bio-Rad Liquichek Tumor Marker Control is substantially equivalent to the predicate device K071675.

All supporting data is retained on file at Bio-Rad Laboratories.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 23, 2014

BIO-RAD LABORATORIES SUZANNE PARSONS REGULATORY AFFAIRS MANAGER 9500 JERONIMO RD. IRVINE CA 92618-2017

Re: K141073

Trade/Device Name: Liquichek Tumor Marker Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I Product Code: JJY Dated: April 23, 2014 Received: April 25, 2014

Dear Ms. Parsons:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2-Ms. Parsons

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Elizabeth A. Stafford -S

for Maria M Chan, Ph.D.

Director Division of Immunology and Hematology Devices

Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K141073

Device Name

Liquichek Tumor Marker Control

Indications for Use (Describe)

Liquichek Tumor Marker Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.

Analytes:
Alpha Fetoprotein (AFP)
Beta-2-Microglobulin (B2-M)
CA 15-3
CA 19-9
CA 125
Carcinoembryonic Antigen (CEA)
Ferritin
Human Chorionic Gonadotropin (hCG)
Prolactin
Prostate Specific Antigen, Total (Total PSA)
Prostate Snerific Antigen Free (Free PSA)

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Elizabeth A. Stafford -S

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review Instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

EP PSC Publishing Services (301) J43-6740

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.