LIQUICHEK TUMOR MARKER CONTROL
K141073 · Bio-Rad Laboratories · JJY · Jun 23, 2014 · Clinical Chemistry
Device Facts
| Record ID | K141073 |
| Device Name | LIQUICHEK TUMOR MARKER CONTROL |
| Applicant | Bio-Rad Laboratories |
| Product Code | JJY · Clinical Chemistry |
| Decision Date | Jun 23, 2014 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Intended Use
Liguichek Tumor Marker Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.
Device Story
Liquichek Tumor Marker Control is a trilevel liquid quality control serum derived from human source material with added animal-derived constituents, stabilizers, and preservatives. It is used in clinical laboratories to monitor the precision of testing procedures for various tumor marker analytes. The control is provided in a liquid form and is designed for use with specific instrumentation, such as the Siemens Dimension Vista Systems. Laboratory personnel use the control by running it alongside patient samples; the resulting values are compared against the provided mean values and ±3SD ranges to verify the performance and precision of the testing system. By identifying potential variations caused by reagents, instrumentation, or technique, the device helps ensure the reliability of clinical test results, thereby supporting accurate clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only. Stability studies (real-time and accelerated) were conducted to establish thawed opened, thawed unopened, and shelf-life stability claims. Acceptance criteria were met for all stability parameters.
Technological Characteristics
Trilevel liquid control; human source material with animal-derived constituents, stabilizers, and preservatives. Formulated for specific use with Siemens Dimension Vista Systems. Storage requirements: -20 to -50 °C (unopened). Stability: 10 days thawed/opened, 15 days thawed/unopened at 2-8 °C. No software or electronic components.
Indications for Use
Indicated for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for specific tumor marker analytes (AFP, B2-M, CA 15-3, CA 19-9, CA 125, CEA, Ferritin, hCG, Prolactin, Total PSA, Free PSA) in clinical laboratory settings.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
- Liquichek Tumor Marker Control (K071675)
Related Devices
- K163015 — Liquichek Tumor Marker Control-Level 1; Liquichek Tumor Marker Control-Level 2; Liquichek Tumor Marker Control-Level 3; Liquichek Tumor Marker Control- Trilevel MiniPak · Bio-Rad Laboratories · Jan 10, 2017
- K071675 — LIQUICHEK TUMOR MARKER CONTROL · Bio-Rad Laboratories · Jul 3, 2007
- K030705 — CLINIQA LIQUID QC TUMOR MARKER CONTROL LEVEL 1, LEVEL 2, LEVEL 3, AND TRI LEVEL · Cliniqa Corporation · Mar 26, 2003
- K972235 — PRECICONTROL TUMOR MARKER 1 AND 2 · Boehringer Mannheim Corp. · Aug 12, 1997
- K983807 — LYPHOCHEK TUMOR MARKER CONTROL, MODEL #580 · Bio-Rad · Dec 16, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary Liquichek Tumor Marker Control K141073
#### 1.0 Submitter
Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, California 92618-2017 Telephone: (949) 598-1200 (949) 598-1557 Fax:
### Contact Person
Suzanne Parsons Regulatory Affairs Manager Telephone: (949) 598-1467
### Date of Summary Preparation
June 20, 2014
#### 2.0 Device Identification
| Product Trade Name: | Liquichek Tumor Marker Control |
|---------------------|----------------------------------------------------|
| Common Name: | Multi-Analyte Controls, All Kinds (Assayed) |
| Review Panel: | Clinical Chemistry and Clinical Toxicology Devices |
| Classifications: | Class I, Reserved |
| Product Code: | JJY |
| Regulation Number | 21 CFR 862.1660 |
#### 3.0 Device to Which Substantial Equivalence is Claimed
Liquichek Tumor Marker Control Bio-Rad Laboratories Irvine, California
510 (k) Number: K071675
#### 4.0 Description of Device
Liquichek Tumor Marker Control, a trilevel liquid control is prepared from human source material with added constituents of human and animal origin, stabilizers and preservatives.
The human source material used to manufacture this control was tested by FDA accepted methods and found non-reactive for Hepatitis B Surface Antigen (HBsAg), antibody to Hepatitis C (HCV) and antibody to HIV-1/HIV-2.
JUN 2 3 2014
{1}------------------------------------------------
#### 5.0 Value Assignment
The mean values and the corresponding ±3SD ranges printed in this insert were derived from replicate analyses and are specific for this lot of product. The tests listed were performed by the manufacturer and/or independent laboratories using manufacturer supported reagents and a representative sampling of this lot of product. It is recommended that each laboratory establish its own acceptable ranges and use those provided only as guides. Laboratory established ranges may vary from those listed during the life of this control. Variations over time and between laboratories may be caused by differences in laboratory technique, instrumentation and reagents, or by manufacturer test method modifications.
#### 6.0 Intended use
Liguichek Tumor Marker Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.
#### 7.0 Comparison of the new device with the Predicate Device
The new Liquichek Tumor Marker Control claims substantial equivalence to the Liguichek Tumor Marker Control currently in commercial distribution (K071675). Table 1 (below) contains comparison information of similarities and differences between the new and predicate device to which substantial equivalence is claimed.
(This portion of the page is intentionally left blank.)
{2}------------------------------------------------
### Table 1: Similarities and Differences between new and predicate device
| Characteristics | Liquichek Tumor Marker Control<br>(New Device) | Liquichek Tumor Marker Control<br>(Predicate Device, K071675) |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Similarities | |
| Intended Use | This product is intended for use as an assayed<br>quality control serum to monitor the precision of<br>laboratory testing procedures for the analytes<br>listed in the package insert. | This product is intended for use as an assayed<br>quality control serum to monitor the precision of<br>laboratory testing procedures listed in the<br>package insert. |
| Form | Liquid | Liquid |
| Matrix | Human source material and constituents of<br>animal origin | Human source material and constituents of<br>animal origin |
| | Differences | |
| Test System | Siemens Dimension Vista Systems | Multiple Analyzers |
| Vial Type | Siemens Dimension Vista Vials | General control vials |
| Thawed and Unopened<br>(On-Board Stability) | 15 days at 2 - 8 °C | 60 days at 2 - 8 °C except for:<br>IGF-1, PAP: 35 days<br>Free PSA: 30 days<br>CA 125: 14 days |
| Thawed & Opened | 10 days at 2 - 8 °C | 30 days at 2 - 8 °C except for:<br>IGF-1: 15 days<br>CA 125: 12 days |
| Storage Unopened<br>(Shelf life) | At -20 to -50 °C until the expiration date | At -20 to -70 °C until the expiration date |
| Analytes | Contains:<br>Alpha Fetoprotein (AFP)<br>Beta-2-Microglobulin (B2-M)<br>CA 15-3<br>CA 19-9<br>CA 125<br>Carcinoembryonic Antigen (CEA)<br>Ferritin<br>Human Chorionic Gonadotropin (hCG)<br>Prolactin<br>Prostate Specific Antigen, Total (Total PSA)<br>Prostate Specific Antigen, Free (Free PSA)<br>Does not contain:<br>CA 27.29<br>Prostatic Acid Phosphatase (PAP)<br>Insulin-like Growth Factor-I (IGF-1)<br>Thyroglobulin (Tg) | Contains:<br>Alpha Fetoprotein (AFP)<br>Beta-2-Microglobulin (B2-M)<br>CA 15-3<br>CA 19-9<br>CA 125<br>Carcinoembryonic Antigen (CEA)<br>Ferritin<br>Prolactin<br>Prostate Specific Antigen, Total (Total PSA)<br>Prostate Specific Antigen, Free (Free PSA)<br>CA 27.29<br>Prostatic Acid Phosphatase (PAP)<br>Insulin-like Growth Factor-I (IGF-1)<br>Human Chorionic Gonadotropin (hCG)<br>Thyroglobulin (Tg) |
{3}------------------------------------------------
#### 8.0 Statement of Supporting Data
Real-time stability studies were conducted to establish the thawed opened and thawed unopened stability claims. Accelerated stability studies were conducted to establish the shelf-life stability claim. Acceptance Criteria were met to support the product claims as follows:
| Thawed and Opened Stability: | 10 days at 2 to 8 °C |
|--------------------------------|-----------------------------|
| Thawed and Unopened Stability: | 15 days at 2 to 8 °C |
| Shelf Life Stability: | 28 months at -20°C to -50°C |
#### 9.0 Conclusion
Based on the performance characteristics indicated above, the Bio-Rad Liquichek Tumor Marker Control is substantially equivalent to the predicate device K071675.
All supporting data is retained on file at Bio-Rad Laboratories.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2014
BIO-RAD LABORATORIES SUZANNE PARSONS REGULATORY AFFAIRS MANAGER 9500 JERONIMO RD. IRVINE CA 92618-2017
Re: K141073
Trade/Device Name: Liquichek Tumor Marker Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I Product Code: JJY Dated: April 23, 2014 Received: April 25, 2014
Dear Ms. Parsons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Ms. Parsons
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
## Elizabeth A. Stafford -S
for Maria M Chan, Ph.D.
Director Division of Immunology and Hematology Devices
Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) k141073
Device Name
Liquichek Tumor Marker Control
### Indications for Use (Describe)
Liquichek Tumor Marker Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.
| Analytes: |
|----------------------------------------------|
| Alpha Fetoprotein (AFP) |
| Beta-2-Microglobulin (B2-M) |
| CA 15-3 |
| CA 19-9 |
| CA 125 |
| Carcinoembryonic Antigen (CEA) |
| Ferritin |
| Human Chorionic Gonadotropin (hCG) |
| Prolactin |
| Prostate Specific Antigen, Total (Total PSA) |
| Prostate Snerific Antigen Free (Free PSA) |
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Elizabeth A. Stafford -S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review Instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
> EP PSC Publishing Services (301) J43-6740