(236 days)
The InMode WMface device is intended for use in dermatologic procedures, for noninvasive treatment of mild to moderate facial wrinkles and rhytides.
The InMode WMface device is designed to deliver radio frequency (RF) energy which is emitted into the skin in a uniform manner, via two pairs of bi-polar electrodes. RF energy heats the tissue to trigger collagen remodeling for the treatment of wrinkles and rhytides. The InMode WMface device is designed to deliver RF energy to the skin. The device provides individual adjustment of RF power to achieve maximum efficiency and safety for each patient. The ergonomic hand piece allows efficient treatment of the skin. The InMode WMface device consists of an AC/DC power supply unit, RF generator, controller and user interface including a LCD screen and functional buttons. The hand piece is connected to the console via a cable and a foot switch activates the energy delivery to the hand piece. The hand piece comprises a handle and two pairs of electrodes.
The InMode WMface Device is an electrosurgical cutting and coagulation device intended for noninvasive treatment of mild to moderate facial wrinkles and rhytides.
Here's an analysis of its acceptance criteria and the study proving its performance:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Output Parameters | RF Output Power | 10-65 Watt (Matches predicate device) |
| RF Output Frequency | 1 MHz ± 2% (Matches predicate device) | |
| Skin Temperature Maintenance | Consistently maintained below specified temperature limit during recommended treatment. | |
| Thermal Energy Distribution | Identical energy distribution pattern | Demonstrated identical pattern to predicate device in ex-vivo tests. |
| Penetration depth | Demonstrated identical penetration depth to predicate device in ex-vivo tests. | |
| Maximal temperature | Demonstrated identical maximal temperature to predicate device in ex-vivo tests. | |
| Safety | Compliance with IEC 60601-1, IEC 60601-1-2 | Complies with these standards. |
| Compliance with IEC 60601-2-2 | Complies with this standard. | |
| Biocompatibility | Materials in contact with skin | All device materials in contact with skin surface are biocompatible. |
| Cleaning | Device is intended for multiple use and must be cleaned | Cleaning instructions are provided in the Instruction For Use. |
2. Sample Size and Data Provenance:
- Test Set Sample Size: Not applicable. The primary studies described are bench tests and ex-vivo tests, not clinical studies with human participants.
- Data Provenance: The bench tests and ex-vivo tests were likely conducted in a controlled laboratory environment. The document does not specify a country of origin for the data but the applicant is based in Israel. The studies are not described as retrospective or prospective clinical trials.
3. Number of Experts and Qualifications for Ground Truth:
Not applicable. The reported studies are non-clinical (bench and ex-vivo) and do not involve expert assessment of clinical outcomes or images for establishing ground truth. Ground truth for these tests would be based on instrument readings and physical measurements.
4. Adjudication Method for Test Set:
Not applicable, as no human-reviewed test set is mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No, an MRMC comparative effectiveness study was not done. The submission relies on non-clinical performance data (bench and ex-vivo tests) to demonstrate substantial equivalence to a predicate device, rather than comparative clinical efficacy against human readers or other devices.
6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop):
Yes, the studies conducted were standalone. The device's performance was evaluated independently in bench tests and ex-vivo tests. There is no mention of a human-in-the-loop component in these studies.
7. Type of Ground Truth Used:
The ground truth for the non-clinical studies was based on:
- Instrumental Measurements: For RF output power, frequency, and skin temperature.
- Physical Observations/Measurements: For thermal energy distribution, penetration depth, and maximal temperature in the ex-vivo porcine tissue model.
8. Sample Size for Training Set:
Not applicable, as this is a medical device and not an AI/algorithm-based diagnostic or imaging device that would typically have a training set for machine learning. The "training" of the device is through its engineering design and manufacturing to meet specified parameters.
9. How Ground Truth for Training Set Was Established:
Not applicable for the reasons stated in point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 3, 2014
InMode MD LTD. % Ms. Ahava Stein A. Stein - Regulatory Affairs Consulting LTD. 20 Hata'as Street, Suite 102 Kfar Saba 44425 Israel
Re: K140926
Trade/Device Name: InMode WMface Device Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 21, 2014 Received: October 29, 2014
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
INDICATIONS FOR USE
510(k) Number (if known): K_140926
InMode WMface Device Device Name:
Intended Use Statement:
The InMode WMface device is intended for use in dermatologic procedures, for noninvasive treatment of mild to moderate facial wrinkles and rhytides.
√ Prescription Use _ (Per 21 C.F.R. 801 Subpart D) C)
OR
Over-The-Counter Use (Optional Format Subpart
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
510(K) SUMMARY INMODE WMFACE DEVICE
510(k) Number K140926
Applicant Name:
| Company Name: | InMode MD Ltd. |
|---|---|
| Address: | Tabor Building, Shaar YokneamPOB 44Yokneam 20692Israel |
| Tel: +972-4-9096307 | |
| Fax: +972-4-9096310 | |
| E-mail: ahava@asteinrac.com |
Contact Person:
| Official Correspondent: | Ahava Stein |
|---|---|
| Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. |
| Address: | 20 Hata'as Str., Suite 102Kfar Saba 44425Israel |
| Tel: | +972-9-7670002 |
| Fax: | +972-9-7668534 |
| E-mail: | ahava@asteinrac.com |
| Date Prepared: | 25 November 2014 |
| Trade Name: | InMode WMface Device |
| Classification Name: | CFR Classification section 878.4400; (Product code GEI) |
| Classification: | Class II Medical Device |
Predicate Device:
The InMode WMface device is substantially equivalent to the following predicate devices.
| Manufacturer | Device | 510(k) No. |
|---|---|---|
| EndyMed Ltd. Imagine TC Skin Treatment System K083461 |
{4}------------------------------------------------
Device Description:
The InMode WMface device is designed to deliver radio frequency (RF) energy which is emitted into the skin in a uniform manner, via two pairs of bi-polar electrodes. RF energy heats the tissue to trigger collagen remodeling for the treatment of wrinkles and rhytides.
The InMode WMface device is designed to deliver RF energy to the skin. The device provides individual adjustment of RF power to achieve maximum efficiency and safety for each patient. The ergonomic hand piece allows efficient treatment of the skin
The InMode WMface device consists of an AC/DC power supply unit, RF generator, controller and user interface including a LCD screen and functional buttons. The hand piece is connected to the console via a cable and a foot switch activates the energy delivery to the hand piece. The hand piece comprises a handle and two pairs of electrodes.
Following are the InMode WMface device specifications: RF Output Power: 10-65 Watt RF Output Frequency: 1[MHz] ± 2% Dimension: 46 x 46 x 100 cm (18.2 x 18.2 x 40 in) Weight: 30 Kg (66 lbs) Main Line Frequency (nominal): 50-60 Hz Input Voltage (nominal): 100-240 VAC
Intended Use/Indication for Use:
The InMode WMface device is intended for use in dermatologic procedures for noninvasive treatment of mild to moderate facial wrinkles and rhytides.
Performance Standards:
The InMode WMface Device has been tested and complies with the following voluntary recognized standards:
- IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for . Safety, 2005; + CORR. 1 (2006) + CORR. 2 (2007)/ EN 60601-1:2006
- IEC 60601-1-2 (2007), Medical Electrical Equipment Part 1-2: General . Requirements for Safety - Collateral Standard: Electromagnetic Compatibility --Requirements and Tests: and IEC 60601-2-2 (2009): Medical Electrical Equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment (section 202.6.1 (Emission) and section 202.6.2 (Immunity).
- IEC 60601-2-2 (2009): Medical Electrical Equipment Part 2: Particular . requirements for the safety of high frequency surgical equipment; for use in conjunction with IEC 60601-1:2005
{5}------------------------------------------------
Sterilization, Cleaning and Biocompatibility:
The InMode WMface device is not supplied sterilized by the user. The device is intended for multiple use and therefore, it must be cleaned according to the instructions provided in the Instruction For Use. The InMode WMface hand piece comes in contact with the skin surface. All device materials in contact with the skin surface are biocompatible.
Non-Clinical (Bench) Performance Data:
Several performance tests were conducted to evaluate the effectiveness and safety of the InMode WMface device and its equivalence to the predicate device. A bench test was performed to measure the accuracy of the RF output parameters in the InMode WMface device and compare them to the RF output measurements in the predicate device (EndyMed Imagine TC Skin Treatment System). The results of the bench test demonstrated that the InMode WMface device has the same RF output specifications as the predicate device and therefore, is substantially equivalent to the predicate device.
An additional bench test was performed to demonstrate that the InMode WMface device and the predicate device consistently maintain the skin temperature below the specified temperature limit during the recommended treatment period and using different RF output power settings. The results demonstrated that both devices tested at the same RF energy outputs maintain the skin temperature below the specified temperature limit.
An additional ex-vivo test was performed to evaluate and compare the thermal energy distribution generated by the WMface device and the EndyMed Imagine System. The testing performed on a porcine tissue model using different RF output power settings and time intervals demonstrated an identical energy distribution pattern, penetration depth and maximal temperature in both devices.
Clinical Performance Data:
Not Applicable
Substantial Equivalence:
The indications for use and technological characteristics of the InMode WMface device are substantially equivalent to the indications for use and technological characteristics of the EndyMed Imagine TC Skin Treatment System.
The design and components in the InMode WMface device, including the console (with power supply, RF generator, controller and display panel) and the hand piece applicator (with cable and connector to console) are similar to the design and components found in the predicate EndyMed Imagine TC Skin Treatment System. The minor differences are insignificant and do not influence the safety or efficacy of the device as shown in the performance tests provided in section 18. The performance specifications (including frequency, RF output power and pulse duration) of the InMode WMface device are
{6}------------------------------------------------
substantially equivalent to those in the EndyMed Imagine TC Skin Treatment System. Both devices demonstrated similar temperature and thermal profiles behaviors as was shown in the performance tests. The safety features and compliance with safety standards in the InMode WMface device are similar to the safety features and compliance with safety standards found in the predicate device. Patient contact materials are also similar. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns. Furthermore, the new InMode WMface device underwent performance testing, including software validation testing (provided in Section 16) and electrical and mechanical safety testing according to IEC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2 (provided in Section 17) and bench tests (provided in Section 18). These performance tests demonstrated that the minor differences in the device specifications meet the system requirements and do not raise new safety or effectiveness concerns.
Consequently, it can be concluded that the InMode WMface device is substantially equivalent to the predicate EndyMed Imagine TC Skin Treatment System, cleared under 510(k) K083461, and therefore, may be legally marketed in the USA.
Conclusions:
Based on the performance testing and comparison to predicate devices, the InMode WMface device is substantially equivalent to the EndyMed Imagine TC Skin Treatment predicate device.
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.