K Number
K140910
Date Cleared
2014-11-19

(224 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CenterCross Catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices.

Device Description

The CenterCross Catheter is a sterile, single-use, single lumen over-the-wire support catheter to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature to facilitate exchange of guidewires and other interventional devices.

The CenterCross Catheter consists of an inner shaft, outer sheath, and a proximal handle that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily expandable and retractable nitinol structure, which when deployed by the physician expands to the width of the artery to aid interventionalists in establishing greater support near the treatment site.

Subsequent to conventional guidewire placement, atherectomy devices, PTCA catheters, and/or stents may be used to provide therapeutic benefit. The CenterCross Catheter in and of itself does not provide therapeutic benefit beyond simple facilitation of guidewire support. The CenterCross Catheter is similar in its design and it achieves its intended use by means of the same mechanisms as the predicate devices.

AI/ML Overview

The provided text is a 510(k) Premarket Notification for the CenterCross Catheter, which is a medical device. This type of document is a submission to the FDA demonstrating that the device is substantially equivalent to a legally marketed predicate device. As such, it focuses on comparing the new device to existing ones rather than establishing novel acceptance criteria or conducting studies to prove independent performance against such criteria as one might find for a truly novel device or an AI/software as a medical device (SaMD).

Therefore, much of the requested information regarding acceptance criteria, specific performance metrics, sample sizes for test sets, expert ground truth establishment, MRMC studies, or standalone algorithm performance simply does not apply or is not detailed in this type of regulatory submission.

However, I can extract the information that is present:

1. A table of acceptance criteria and the reported device performance:

The document does not present a formal table of "acceptance criteria" and "reported device performance" in the way one might expect for a diagnostic or algorithmic device. Instead, it lists general "Nonclinical Testing Summary" categories. The "acceptance criteria" are implied to be "meeting established specifications" which allow for a substantial equivalence determination, rather than specific quantitative thresholds. The "reported device performance" is summarized as the device "meets the established specifications necessary for consistent performance for its intended use" and "functions to its specifications and intended use."

Acceptance Criteria (Implied)Reported Device Performance
Device functions as intended for specified uses.Confirmed to function to specifications and intended use.
Consistent performance across various aspects (tensile, torque, etc.)Meets established specifications for consistent performance.
Safe and effective for its intended use.No new issues of safety or effectiveness are raised.
Comparable to predicate devices in technological characteristics.Substantially equivalent to predicate devices.
Biocompatible.Demonstrated biocompatibility.
Sterilizable and maintains shelf-life.Sterilization adopted into validated process; packaging and shelf-life evaluated.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

The document does not specify sample sizes for non-clinical bench testing. The data provenance is "bench testing" and an "in vivo animal validation study." No human data or specific country of origin for data is mentioned, which is typical for pre-market submissions of physical devices where initial verification and validation are often done in-house or through contract labs. These are typically prospective tests performed for the purpose of the submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

This information is not applicable and not provided. The testing described is primarily mechanical and material characterization, and an animal study. "Ground truth" in the context of expert medical interpretation (e.g., for imaging or diagnostic algorithms) is not relevant here.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable and not provided. Adjudication methods are typically used for establishing ground truth in clinical studies involving human interpretation or uncertain outcomes, not for bench testing of physical device performance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No MRMC study was conducted or is mentioned. This type of study is relevant for AI/SaMD devices that assist human readers (e.g., radiologists interpreting images), which is not the nature of the CenterCross Catheter.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This is not applicable as the CenterCross Catheter is a physical medical device, not an algorithm or software. Standalone performance for an algorithm is not relevant here.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the non-clinical testing is based on established engineering specifications, recognized standards, and the expected functional performance of catheters. For the "in vivo animal validation study," the "ground truth" would be the observed physiological response and device performance within the animal model, assessed by researchers or veterinary professionals, but specifics are not detailed.

8. The sample size for the training set:

This is not applicable. "Training set" refers to data used to train machine learning models. The CenterCross Catheter is a physical device and does not involve a training set in this context.

9. How the ground truth for the training set was established:

This is not applicable since there is no training set for a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 19, 2014

Roxwood Medical, Inc. Mehrdad Farhangnia President & CEO 400 Seaport Ct. Suite #103 Redwood City, California 94063

Re: K140910 Trade/Device Name: CenterCross Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: November 7, 2014 Received: November 10, 2014

Dear Mr. Farhangnia,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

ROXWOOD MEDICAL, INC.

CenterCross Catheter 510(k) PREMARKET NOTIFICATION

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K140910

Device Name: CenterCross Catheter

Indications For Use:

The CenterCross Catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices.

Prescription Use X And/Or (21 CFR Part 801 Subpart D)

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

510(k) SUMMARY

510(k) Notification K

GENERAL INFORMATION

Applicant:

Roxwood Medical, Inc. 400 Seaport Ct., Suite #103 Redwood City, CA 94063 U.S.A. Phone: 408.373.4751 FAX: 650.779.4554

Contact Person:

Mehrdad Farhangnia Roxwood Medical, Inc. 400 Seaport Ct., Suite #103 Redwood City, CA 94063 U.S.A. Phone: 408.373.4751 FAX: 650.779.4554

Date Prepared: April 28, 2014

DEVICE INFORMATION

The CenterCross Catheter is a percutaneous catheter for use in the coronary and peripheral vasculature.

Trade Name: CenterCross Catheter

Generic/Common Name: Percutaneous Catheter

Classification: 21 CFR§870.1250, Class II

Product Code: DQY

{4}------------------------------------------------

510(k) SUMMARY (CONT.)

PREDICATE DEVICE(S)

  • Roxwood Medical, MultiCross Support Catheter (K121763)
  • Radius Medical, Prodigy Support Catheter (K130714) ●

INDICATIONS FOR USE

The CenterCross Catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices.

PRODUCT DESCRIPTION

The CenterCross Catheter is a sterile, single-use, single lumen over-the-wire support catheter to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature to facilitate exchange of guidewires and other interventional devices.

The CenterCross Catheter consists of an inner shaft, outer sheath, and a proximal handle that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily expandable and retractable nitinol structure, which when deployed by the physician expands to the width of the artery to aid interventionalists in establishing greater support near the treatment site.

Subsequent to conventional guidewire placement, atherectomy devices, PTCA catheters, and/or stents may be used to provide therapeutic benefit. The CenterCross Catheter in and of itself does not provide therapeutic benefit beyond simple facilitation of guidewire support. The CenterCross Catheter is similar in its design and it achieves its intended use by means of the same mechanisms as the predicate devices.

TECHNOLOGICAL CHARACTERISTICS

The technological characteristics of the CenterCross Catheter are similar to the predicate devices. Performance data is provided to support the determination of substantial equivalence.

SUBSTANTIAL EQUIVALENCE

The CenterCross Catheter is substantially equivalent to the Roxwood Medical MultiCross Support Catheter and the Radius Medical Prodigy Support Catheter. The subject device and the predicate devices are percutaneous catheters. The proposed indications for use for the CenterCross Catheter are substantially equivalent to the indications for use for the predicate devices. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the CenterCross Catheter is substantially equivalent to the predicate devices.

{5}------------------------------------------------

510(k) SUMMARY (CONT.)

TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

Extensive bench testing was conducted on the CenterCross Catheter to evaluate the performance of the device and to support a determination of substantial equivalence to the predicate devices. Non-clinical testing assessed the following aspects of the device:

Nonclinical Testing Summary:

  • Tensile Strength ●
  • Torque Strength .
  • . Simulated Use
    • · Catheter Compatibility
    • · Guidewire Compatibility
    • · Delivery, Deployment and Retraction
    • · Durability
    • Contrast Injection
  • Flexibility & Kink Resistance ●
  • Leak ●
  • Particulate ●
  • Expansion Force .
  • In vivo animal validation study ●
  • Biocompatibility ●
  • Sterilization Adoption into a validated sterilization process ●
  • Packaging and shelf-life .

All testing was performed in accordance with recognized standards. The collective results of the non-clinical testing demonstrate that the CenterCross Catheter meets the established specifications necessary for consistent performance for its intended use and is substantially equivalent to the predicate devices.

CONCLUSION

The CenterCross Catheter is a percutaneous vascular catheter and shares its design and mechanism of action with the identified predicate devices. The results of the performance testing confirm that the CenterCross Catheter functions to its specifications and intended use and exhibit the appropriate characteristics of a percutaneous vascular support catheter. The CenterCross Catheter is substantially equivalent to the predicate devices in terms of technological characteristics, intended use and performance. No new issues of safety or effectiveness are raised by the CenterCross Catheter.

SUMMARY

The CenterCross Catheter is substantially equivalent to the predicate devices.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).