K Number
K121763

Validate with FDA (Live)

Date Cleared
2012-11-27

(165 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MultiCross Support Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

Device Description

The MultiCross Support Catheter is a sterile, single-use, triple lumen, over-the-wire, disposable percutaneous support catheter designed for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature for guidewire exchange. The MultiCross Support Catheter consists of an inner shaft, outer sheath, and a proximal handle that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and retractable distal Nitinol scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide an anchoring to aid the user in establishing greater support near the treatment site. Subsequent to conventional guidewire placement, therapeutic devices such as atherectomy devices, PTCA catheters, and/or stents may be used to provide therapeutic benefit. The MultiCross Support Catheter by itself does not provide therapeutic benefit beyond simple facilitation of guidewire support.

AI/ML Overview

The provided 510(k) summary for the MultiCross Support Catheter (K121763) describes a medical device, not an AI/ML-driven software or diagnostic tool. Therefore, many of the requested categories in the prompt, which are typically relevant for AI/ML device evaluations (e.g., sample size for test/training sets, ground truth methodology, expert qualifications, MRMC studies, standalone performance, etc.), are not applicable or not explicitly detailed in this type of submission.

This submission focuses on demonstrating "substantial equivalence" to predicate devices through engineering bench testing and an animal study, which are standard for physical medical devices of this type.

Here’s an attempt to address the prompt based only on the provided document, acknowledging that most AI/ML-specific criteria will not be found:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/TestAcceptance Criteria (Implied)Reported Device Performance
Bench TestingDevice meets established specifications for consistent performance for its intended use and is substantially equivalent to predicate devices.All testing performed in accordance with recognized standards. Results demonstrate the device meets established specifications for consistent performance for its intended use and is substantially equivalent.
Catheter Inspection (Visual/Physical Dimensions)Not explicitly stated, but implies meeting design specifications.Performed (results not detailed but deemed acceptable for equivalence).
Particulate TestingNot explicitly stated, but implies meeting acceptable limits.Performed (results not detailed but deemed acceptable for equivalence).
Radial Outward ForceNot explicitly stated, but implies meeting functional requirements.Performed (results not detailed but deemed acceptable for equivalence).
Contrast Flush InjectionNot explicitly stated, but implies meeting functional requirements.Performed (results not detailed but deemed acceptable for equivalence).
Simulated UseDevice performs as intended during simulated clinical scenarios.Performed (results not detailed but deemed acceptable for equivalence).
Torque StrengthNot explicitly stated, but implies meeting mechanical strength requirements.Performed (results not detailed but deemed acceptable for equivalence).
Flexibility and Kink ResistanceNot explicitly stated, but implies meeting mechanical and functional requirements.Performed (results not detailed but deemed acceptable for equivalence).
Leak: Liquid & AirNot explicitly stated, but implies no unacceptable leaks.Performed (results not detailed but deemed acceptable for equivalence).
Tensile StrengthNot explicitly stated, but implies meeting mechanical strength requirements.Performed (results not detailed but deemed acceptable for equivalence).
Corrosion Resistance TestNot explicitly stated, but implies no unacceptable corrosion.Performed (results not detailed but deemed acceptable for equivalence).
Coating IntegrityNot explicitly stated, but implies coating remains intact and functional.Performed (results not detailed but deemed acceptable for equivalence).
BiocompatibilityMeets ISO 10993 or equivalent standards for patient contact.Performed (results not detailed but deemed acceptable for equivalence).
Sterilization ValidationDevice can be effectively sterilized without compromising integrity.Performed (results not detailed but deemed acceptable for equivalence).
Packaging and Shelf-lifePackaging maintains sterility and device integrity over specified shelf-life.Performed (results not detailed but deemed acceptable for equivalence).
In vivo Animal Validation StudySafety acceptance criteria for the study are met, and performance/handling are acceptable.Based on pathology and histopathology results, the safety acceptance criteria were met. No untoward observations were found by the clinician regarding performance and handling.

2. Sample Size for Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated for bench tests (which would typically involve multiple units tested). For the animal study, the sample size is not numerically provided, but it is referred to as "a GLP animal study."
  • Data Provenance: The studies were conducted by Roxwood Medical, Inc. (applicant). The location of the animal study (country of origin) is not specified. The studies are prospective in the sense that they were designed and executed to test the device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • The document refers to a "clinician" who made performance and handling observations in the animal study.
  • The qualifications of this clinician are not specified (e.g., radiologist with X years of experience).
  • For pathology and histopathology results in the animal study, it's implied that experts (e.g., pathologists) evaluated the tissues, but their number and specific qualifications are not detailed.

4. Adjudication Method for the Test Set

  • Not applicable as this is a physical device submission. Ground truth regarding safety (e.g., pathology) is established by standard medical/scientific evaluation, not multi-reader adjudication in the context of diagnostic performance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. This type of study is typically for evaluating the diagnostic performance of software or imaging systems with human readers.

6. If a Standalone (algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical medical device; there is no "algorithm only" performance to evaluate.

7. The Type of Ground Truth Used

  • For the animal study:
    • Pathology and Histopathology results: Used to establish safety ground truth.
    • Clinician observations: Used to establish performance and handling ground truth.
  • For bench testing: The "ground truth" is adherence to engineering specifications and recognized standards for physical performance parameters.

8. The Sample Size for the Training Set

  • Not applicable. This is a physical device, and there is no "training set" in the AI/ML sense. Design and development would involve iterative testing, but not a formally defined "training set" for an algorithm.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for an AI/ML algorithm.

{0}------------------------------------------------

510(k) SUMMARY K121763

NOV 2 7 2012

This summary of the 510(k) premarket notification for the MultiCross Support Catheter is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR8807.92.

GENERAL INFORMATION

Applicant:

Roxwood Medical, Inc. 150 Executive Park Boulevard, Suite 3700 San Francisco, CA 94134 U.S.A. Phone: 408-373-4751 Fax: 415-467-1253

Contact Person:

Mehrdad Farhangnia Chief Executive Officer Roxwood Medical, Inc. 150 Executive Park Boulevard, Suite 3700 San Francisco, CA 94134 U.S.A. Phone: 408-373-4751 Fax: 415-467-1253

Date Prepared: June 13, 2012

DEVICE INFORMATION

The MultiCross Support Catheter is a percutaneous catheter for use in the coronary and peripheral vasculature.

Trade Name: MultiCross Support Catheter

Generic/Common Name: Percutaneous Catheter

Classification: 21 CFR 870.1250, Class II

Product Code: DQY

{1}------------------------------------------------

MULTICROSS SUPPORT CAT 510(k) PREMARKET NOTIFICATION

PREDICATE DEVICE(S)

  • Vascular Solutions Twin-Pass™ OTW Dual Access Catheter (K072618) �
  • EndoCross ENABLER-P Catheter System (K092174) �
  • BridgePoint Medical Stingray Orienting Balloon Catheter (K080987) .

INDICATIONS FOR USE

The MultiCross Support Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

PRODUCT DESCRIPTION

The MultiCross Support Catheter is a sterile, single-use, triple lumen, over-the-wire, disposable percutaneous support catheter designed for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature for guidewire exchange.

The MultiCross Support Catheter consists of an inner shaft, outer sheath, and a proximal handle that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and retractable distal Nitinol scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide an anchoring to aid the user in establishing greater support near the treatment site.

Subsequent to conventional guidewire placement, therapeutic devices such as atherectomy devices, PTCA catheters, and/or stents may be used to provide therapeutic benefit. The MultiCross Support Catheter by itself does not provide therapeutic benefit beyond simple facilitation of guidewire support. The MultiCross Support Catheter is similar in its design and it achieves its intended use by means of the same mechanisms as the predicate devices.

TECHNOLOGICAL CHARACTERISTICS

The technological characteristics of the MultiCross Support Catheter are similar to the predicate devices. Performance data were provided to support the determination of substantial equivalence.

FeatureMultiCross SupportCatheter(Subject Device)Vascular SolutionsTwin-PassTM OTW DualAccess CatheterEndoCrossENABLER-PCatheter SystemBridgePoint MedicalStingray OrientingBalloon Catheter
510(k) NumberK_____K052257K092174K080987
Indications forUseThe MultiCrossSupport Catheter isintended to be used inconjunction with asteerable guidewire toaccess discrete regionsof the coronary andperipheral vasculatureand for guidewireThe Twin-Pass DualAccess Catheter is to beused in conjunction withsteerable guidewires inorder to access discreteregions of the coronaryand peripheral arterialvasculature, to facilitateplacement of guidewiresand other interventionalThe ENABLER-PCatheter System isintended to be usedin conjunction with asteerable guidewireto access discreteregions of theperipheralvasculature and forThe BridgePointMedical StingrayTMOrienting BalloonCatheters are indicatedfor directing, steering,controlling, andsupporting aguidewire in order toaccess discrete regions

{2}------------------------------------------------

ROXWOOD MEDICAL, INC.

MULTICROSS SUPPORT CATHETER 510(k) PREMARKET NOTIFICATION

FeatureMultiCross SupportCatheter(Subject Device)exchange.Vascular SolutionsTwin-Pass™ OTW DualAccess Catheterdevices, and for useduring two guidewireprocedures.EndoCrossENABLER-PCatheter Systemguidewire exchange.BridgePoint MedicalStingray OrientingBalloon Catheterof the coronary andperipheral vasculature.
Guide CatheterCompatibility7F5F7F6F
WorkingLength135 cm140 cm107 cm135cm
GuidewireCompatibility0.014" Guidewirecompatible;Up to three guidewiresmay be exchanged0.014" Guidewirecompatible;Up to two guidewiresmay be exchanged0.035" GuidewireCompatible0.014" Guidewirecompatible;
LumenTripleDualDualTriple
Distal ShaftCapabilityThe Nitinol scaffoldprovides distalanchoring andsupports theadvancement of theguidewire.Standard PercutaneousCatheter Distal ShaftStructure (no supportmechanism)The balloon providesdistal anchoring andsupports theadvancement of theguidewire.The balloon providesdistal anchoring andsupports theadvancement of theguidewire.

SUBSTANTIAL EQUIVALENCE

The MultiCross Support Catheter is substantially equivalent to Vascular Solutions Twin-Pass OTW Dual Access Catheter (K072618), the EndoCross ENABLER-P Catheter System (K092174, and the BridgePoint Medical Stingray Orienting Balloon Catheter (K080987). The subject device and the predicate devices are percutaneous catheters. The proposed indications for use for the MultiCross Support Catheter are substantially equivalent to the indications for use for the predicate devices. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the MultiCross Support Catheter is substantially equivalent to the predicate devices.

TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

Performance Testing - Bench

Extensive bench testing was conducted on the MultiCross Support Catheter to evaluate the performance of the device and to support a determination of substantial equivalence to the predicate devices. Non-clinical testing assessed the following aspects of the device:

  • Catheter Inspection (Visual Inspection/Physical Dimensions) �
  • Particulate Testing
  • Radial Outward Force .
  • Contrast Flush Injection .
  • Simulated Use �
  • Torque Strength
  • Flexibility and Kink Resistance
    • Leak: Liquid
      • Air

{3}------------------------------------------------

K121763
Page 4 of 4

MULTICROSS SUPPORT CATHETER 510(k) PREMARKET NOTIFICATION

  • Tensile Strength .
  • Corrosion Resistance Test .
  • Coating Integrity .
  • In vivo Animal Validation Study .
  • . Biocompatibility
  • Sterilization Validation .
  • Packaging and Shelf-life ♥

All testing was performed in accordance with recognized standards.

Safety and Performance Testing - Animal

A GLP animal study was performed to evaluate the safety, performance and handling of the MultiCross Support Catheter in the coronary and peripheral vasculature as compared to a control device. Based on pathology and histopathology results, the safety acceptance criteria for the study were met.

Performance and handling observations were made based on detailed characteristics of the device. No untoward observations were found by the clinician.

Conclusion of Performance Testing

The collective results of the bench testing and the GLP animal study demonstrate that the MultiCross Support Catheter meets the established specifications necessary for consistent performance for its intended use and is substantially equivalent to the predicate devices.

CONCLUSION

The MultiCross Support Catheter is a percutaneous vascular support catheter and shares its design and mechanism of action with the identified predicate devices. The results of the bench testing and GLP animal study confirm that the MultiCross Support Catheter functions to its specifications, performs as intended and exhibits the appropriate characteristics of a percutaneous support catheter. The MultiCross Support Catheter is substantially equivalent to the predicate devices in terms of technological characteristics, intended use and performance. No new issues of safety or effectiveness are raised by the MultiCross Support Catheter.

SUMMARY

The MultiCross Support Catheter is substantially equivalent to the predicate devices.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of an eagle or bird with outstretched wings, rendered in a bold, black color.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Roxwood Medical, Inc. Mr. Mehdrad Farhangnia Chief Executive Officer 150 Executive Park Blvd, Suite 3700 San Francisco, CA 94134

Re: K121763

Trade/Device Name: MultiCross Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: November 16, 2012 Received: November 19, 2012

Dear Mr. Farhangnia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

NOV 2 7 2012

{5}------------------------------------------------

Page 2 - Mr. Mehdrad Farhangnia

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic in product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours, Matthew G Hillebrenne

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

ROXWOOD MEDICAL, INC.

510(k) Number (if known): KI21 76 3

Device Name: MultiCross Support Catheter

Indications For Use:

The MultiCross Support Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

MS Kelli

Division of Cardiovascular Devices

510(k) Number K121763

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).