K Number
K140883

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-07-30

(114 days)

Product Code
Regulation Number
888.3565
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
  2. Correction of varus, valgus, or posttraumatic deformity.
  3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.
    The Regenerex femoral augments are indicated for use with the Vanguard Total Knee System.
    The Regenerex tibial augments are indicated for use with standard and offset Biomet® Tibial Trays.
    Femoral components and tibial tray components with porous coatings are indicated for cemented and uncemented biological fixation application. Non-coated (Interlok) femoral components, tibial tray components and all polyethylene patellar components are indicated for cemented application only. Regenerex and OsseoTi components are intended only for uncemented biologic fixation application.
    The Vanguard DA 360 components are not intended for use with the Vanguard PS Open Box Porous Femoral components. The Vanguard DA 360 components are not approved for sale in the United States.
Device Description

The Vanguard 360 OsseoTi Tibial Sleeve Augments are porous, modular augment sleeves that have been designed to address a wide range of defects in cases of severe bone loss in both primary and revision cases. The subject augments are constructed from a previously cleared materials/additive manufacturing process known as OsseoTi. The OsseoTi process results in a fully integrated part with solid and porous regions. This structure leads to a highly interconnected volume of porosity and promotes biological fixation.
The subject OsseoTi Tibial Sleeve Augments mount directly onto the previously cleared Vanguard 360 Tibial Tray boss via Morse tapers. The subject augments are available in half and full sleeve configurations and multiple sizes.

AI/ML Overview

The medical device in question is the Vanguard® 360 OsseoTi™ Tibial Sleeve Augments.

Here's an analysis of the acceptance criteria and study information provided:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Tests ConductedAcceptance Criteria (Implicit)Reported Device Performance
Construct TestingStatic Axial (Pull Off) ForceMet established criteriaMet or exceeded criteria
Torque ResistanceMet established criteriaMet or exceeded criteria
FatigueMet established criteriaMet or exceeded criteria
FrettingMet established criteriaMet or exceeded criteria
MRI Compatibility JustificationMet established criteriaMet or exceeded criteria
OsseoTi Material CharacterizationInterconnecting PorosityMet established criteriaMet or exceeded criteria
MicrographsMet established criteriaMet or exceeded criteria
Porosity and Pore SizeMet established criteriaMet or exceeded criteria
RoughnessMet established criteriaMet or exceeded criteria
Abrasion ResistanceMet established criteriaMet or exceeded criteria
Mechanical StrengthMet established criteriaMet or exceeded criteria
Chemical CompositionMet established criteriaMet or exceeded criteria
Shear FatigueMet established criteriaMet or exceeded criteria
Static ShearMet established criteriaMet or exceeded criteria
Static TensileMet established criteriaMet or exceeded criteria
Animal Data(Not explicitly named, but conducted)Met established criteriaMet or exceeded criteria

Note: The document states "All testing met or exceeded the established acceptance criteria," but does not explicitly list the quantitative values of these criteria. The implicit acceptance criteria for each test are that the device performed within the intended use and did not raise new safety or efficacy issues, aligning with the performance of predicate devices.


2. Sample Size Used for the Test Set and Data Provenance

The provided document describes non-clinical tests (material characterization and construct testing) and "animal data." It explicitly states:

  • Clinical Performance Data/Information: None provided as a basis for substantial equivalence.

Therefore, there is no human clinical test set for this submission. The tests performed were laboratory-based and animal studies. The sample sizes for these tests are not specified in the document.

  • Data Provenance: The tests are described as non-clinical (laboratory-based material and mechanical testing) and animal data. No country of origin is specified for these studies, nor is it explicitly stated if they were retrospective or prospective, though non-clinical and animal studies are typically prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

As there was no human clinical test set for a performance evaluation, the concept of "ground truth" derived from expert consensus on patient data (e.g., radiology reports) is not applicable here. The "ground truth" for the non-clinical tests would be defined by engineering specifications, material science standards, and established biomechanical testing protocols. The qualifications of the engineers/scientists conducting these tests are not provided.


4. Adjudication Method for the Test Set

Not applicable, as there was no human clinical test set requiring adjudication of findings (e.g., medical image interpretations).


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical Performance Data/Information: None provided as a basis for substantial equivalence." This submission relies on non-clinical and animal data to establish substantial equivalence.


6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable to the device described. The device is a physical knee implant component, not an algorithm or AI-driven system. Therefore, standalone algorithm performance is not relevant.


7. The Type of Ground Truth Used

For the non-clinical testing and animal data, the "ground truth" would be established by:

  • Engineering specifications and standards: For mechanical properties, fatigue, fretting, static axial force, torque resistance.
  • Material science standards: For interconnecting porosity, micrographs, porosity and pore size, roughness, abrasion resistance, mechanical strength, chemical composition, shear fatigue, static shear, static tensile.
  • Animal study endpoints: For biocompatibility, bone ingrowth, or other relevant biological responses (details not provided).

There was no "expert consensus," "pathology," or "outcomes data" from human subjects used to establish ground truth for this 510(k) submission.


8. The Sample Size for the Training Set

This question is not applicable. The device is a physical medical implant, not an AI or machine learning model that requires a training set.


9. How the Ground Truth for the Training Set was Established

This question is not applicable, as there is no training set for a physical medical implant.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for Biomet Manufacturing Corp. The logo consists of the word "BIOMET" in a stylized font, with the words "MANUFACTURING CORP." printed below. The text is black and the background is white. The logo is simple and professional.

510(k) SUMMARY

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92

Submitter Information
NameBiomet Manufacturing Corp.
Address56 East Bell DriveP.O. Box 587Warsaw, IN 46581-0857
EstablishmentRegistration Number1825034
Name of contact personJason HeckamanRegulatory Affairs Project Manager
Phone number(574) 371-3707
Fax number(574) 372-1683
Date preparedJuly 30, 2014
Name of device
Trade or proprietarynameVanguard® 360 OsseoTi™ Tibial Sleeve Augments
Common or usualnameKnee Prosthesis
Classification nameKnee joint patellofemorotibial metal/polymer porous-coateduncemented prosthesis (21CFR §888.3565)
Classification panelOrthopedic
Regulation21 CFR §888.3565
Product Code(s)MBH, JWH
Legally marketed device(s)to which equivalence isclaimedK093293 Vanguard 360 Revision Knee SystemK121149 Vanguard SSK 360 Revision Knee SystemK072336 Regenerex Porous Titanium Sleeve AugmentsK102896 Zimmer Trabecular Metal Tibial Cone Augments
Reason for 510(k)submissionThe Vanguard 360 OsseoTi Tibial Sleeve Augments are newproducts that are an addition to the Biomet® Vanguard® 360Knee Joint System (K093293 and K121149).
Device descriptionThe Vanguard 360 OsseoTi Tibial Sleeve Augments are porous,modular augment sleeves that have been designed to address awide range of defects in cases of severe bone loss in bothprimary and revision cases. The subject augments areconstructed from a previously cleared materials/additivemanufacturing process known as OsseoTi. The OsseoTi processresults in a fully integrated part with solid and porous regions.This structure leads to a highly interconnected volume of porosityand promotes biological fixation.
The subject OsseoTi Tibial Sleeve Augments mount directly ontothe previously cleared Vanguard 360 Tibial Tray boss via Morsetapers. The subject augments are available in half and full sleeveconfigurations and multiple sizes.
Intended use of the deviceThe Vanguard 360 OsseoTi Revision Sleeve Tibial Augments areoptional components intended to be used with the previouslycleared Vanguard 360 Revision Knee System as part of a totalknee construct. The subject augments are intended foruncemented biologic fixation.
Indications for use1. Painful and disabled knee joint resulting from osteoarthritis,rheumatoid arthritis, traumatic arthritis where one or morecompartments are involved.2. Correction of varus, valgus, or posttraumatic deformity.3. Correction or revision of unsuccessful osteotomy, arthrodesis,or failure of previous joint replacement procedure.The Regenerex femoral augments are indicated for use with theVanguard Total Knee System.
The Regenerex tibial augments are indicated for use withstandard and offset Biomet® Tibial Trays.
Femoral components and tibial tray components with porouscoatings are indicated for cemented and uncemented biologicalfixation application. Non-coated (Interlok) femoral components,tibial tray components and all polyethylene patellar componentsare indicated for cemented application only. Regenerex andOsseoTi components are intended only for uncemented biologicfixation application.
The Vanguard DA 360 components are not intended for use withthe Vanguard PS Open Box Porous Femoral components. TheVanguard DA 360 components are not approved for salein the United States.Summary of the technological characteristics of the device compared to the predicate

P.O. Bax 557 V/201908 RV 4/5501-0567
Total Froor 800-340-9500 Offer 57-1 267 1633 Meto Fex: 574 267 5137 1. 2017

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for Biomet Manufacturing Corp. The logo consists of the word "BIOMET" in a stylized font, with the words "MANUFACTURING CORP." printed below. The font is bold and sans-serif, and the letters are all capitalized. The logo is black and white.

The technological characteristics of the Vanguard 360 OsseoTi Augments are similar to those of predicate systems (K093293/K121149, K072336, K102896) in terms of geometry, design and principle of operation. The subject OsseoTi augments use the same Morse taper for attachment to the previously cleared Vanquard 360 tibial boss as the predicate Vanguard 360 Cruciate Wing augments but with varying taper engagement. The cone shape and sizing of the subject device is similar to the previously cleared Regenerex (K072336) and Zimmer Trabecular Metal Tibial Cone

PO. Box 557 are, in 46681-0587 Tot Free: 200 348 952 - 534 287 6532

{2}------------------------------------------------

TURING CORP. MANUFA

Augments (K102896). The subject OsseoTi augments are constructed from a previously cleared material/manufacturing process.

CE DE LEGION CONTROLLER PERFORMANCE DATA CONTACT CONTROLLER CONTROLLER CONTROLLER

SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE

OsseoTi material characterization tests and construct testing used in support of the substantial equivalence included the following:

Construct testing / justifications:

  • Static Axial (Pull Off) Force .
  • Torque Resistance .
  • Fatique ●
  • Fretting .
  • MRI Compatibility Justification

OsseoTi characterization:

  • Interconnecting Porosity ●
  • Micrographs
  • Porosity and Pore Size
  • Roughness
  • Abrasion Resistance
  • Mechanical Strength
  • Chemical Composition
  • Shear Fatique
  • Static Shear
  • Static Tensile

Animal Data

All testing met or exceeded the established acceptance criteria.

SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION

Clinical Performance Data/Information: None provided as a basis for substantial equivalence.

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA

The culmination of the results of the mechanical construct testing, OsseoTi characterization and animal data indicated the devices performed within the intended use, did not raise any new safety and efficacy issues, and are substantially equivalent to the predicate devices.

P.O. Box 557

Warm, IN 46581-6587
Toll Free: 800.348.5500
Office: 574 287.5533
Mech Fax: 574.267.8137
www.boxoft.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 30, 2014

Biomet, Incorporated Mr. Jason Heckaman Regulatory Affairs Project Manager 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0857

Re: K140883

Trade/Device Name: Vanguard® 360 OsseoTi™ Tibial Sleeve Augments Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH, JWH Dated: July 1, 2014 Received: July 2, 2014

Dear Mr. Heckaman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{4}------------------------------------------------

Page 2 - Mr. Jason Heckaman

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Lori A. Wiggins

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

K140883 510(k) Number (if known):

Device Name: - Vanguard® 360 Osseo™ Ti Tibial Sleeve Augments

    1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
    1. Correction of varus, valgus, or posttraumatic deformity.
    1. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

The Regenerex femoral augments are indicated for use with the Vanguard Total Knee System.

The Regenerex tibial augments are indicated for use with standard and offset Biomet® Tibial Trays.

Femoral components and tibial tray components with porous coatings are indicated for cemented and uncemented biological fixation application. Non-coated (Interlok) femoral components, tibial tray components and all polyethylene patellar components are indicated for cemented application only. Regenerex and OsseoTi components are intended only for uncemented biologic fixation application.

The Vanguard DA 360 components are not intended for use with the Vanguard PS Open Box Porous Femoral components. The Vanguard DA 360 components are not approved for sale in the United States.

Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/5/Picture/15 description: The image shows the text "Elizabeth L. Frank -S" at the top. Below that is the date and time "2014 07.30 16:50:12". Finally, at the bottom is the text "-04'00'".

Page 1 of 1

Division of Orthopedic Devices

§ 888.3565 Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis.

(a)
Identification. A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.(b)
Classification. Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.