K Number
K140857
Date Cleared
2014-04-25

(22 days)

Product Code
Regulation Number
866.3980
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real Time PCR Instrument in conjunction with clinical and epidemiological information:

  • For determination of the genetic lineage of human influenza B viruses as B/Victoria or B/Yamagata lineage from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) from human patients with signs and symptoms of respiratory infection and/or from viral culture:
  • To provide epidemiologic information for surveillance of circulating influenza viruses.
    Performance characteristics for influenza B lineage genotyping were established during a season when influenza B/Victoria and B/Yamagata lineages were found in approximately equal proportion.
    Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
    All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
Device Description

The Influenza B Lineage Genotyping Kit contains components of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel that is used in real-time RT-PCR (rRT-PCR) assays on the ABI 7500 Fast Dx Real-Time PCR Instrument. The Influenza B Lineage Genotyping Kit consists of oligonucleotide primers and dual-labeled hydrolysis (TaqMan®) probes to be used in rRT-PCR for the in vitro qualitative detection and characterization of human influenza viruses from viral RNA in respiratory specimens from patients presenting with influenza-like illness (ILI). The Influenza B Lineage Genotyping Kit is based on technology which is used in many molecular diagnostic assays.

AI/ML Overview

This is an interesting case where the submission is a "New Special 510(k)" for a labeling change only. This means that no new performance studies were conducted, as the technological attributes of the device itself were not altered. The entire submission focuses on establishing substantial equivalence to the predicate device based on this labeling change.

Therefore, the typical acceptance criteria and study data related to device performance in detecting influenza B lineages are not present in this document. Instead, the "acceptance criteria" here implicitly refer to the FDA's acceptance of the labeling changes and the conclusion of substantial equivalence.

Given this, I will answer the questions based on the information available and clarify where specific data (like sample sizes for performance studies) are not applicable or not provided due to the nature of this submission.


Description of Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

This 510(k) submission (K140857) is a "New Special 510(k)" for the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, Influenza B Lineage Genotyping Kit. The submission states that the proposed changes are for labeling purposes only and will not alter the technological attributes of the device. Consequently, no new performance studies were conducted or are presented in this document to demonstrate the device meets new acceptance criteria.

The acceptance criteria implicitly relate to the FDA's evaluation of the labeling changes and whether the modified kit remains substantially equivalent to the predicate device (K132508) based on these changes. The study proving the device meets this "acceptance criteria" is the analysis of substantial equivalence presented.

1. Table of Acceptance Criteria and Reported Device Performance:

Since this submission is for a labeling change only, there are no new quantitative acceptance criteria or device performance metrics reported for the Influenza B Lineage Genotyping Kit that are different from its predicate device. The "performance characteristics" mentioned in the Intended Use refer to those established for the original device (K132508) during a season when both B/Victoria and B/Yamagata lineages were prevalent.

Acceptance Criteria (Implicit for Labeling Change)Reported Device Performance (as established for predicate, implied for this device)
Maintain substantial equivalence to predicate device (K132508)Declared substantially equivalent by FDA.
Labeling accurately reflects the intended use and technological characteristics of the deviceLabeling modified to create a separate Package Insert for the Influenza B Lineage Genotyping Kit, retaining all required information. No new or increased risks identified.
No alteration to device's design or technological attributesConfirmed by the submitter; no changes beyond labeling are proposed.

2. Sample Size Used for the Test Set and Data Provenance:

Not applicable. As this submission is for a labeling change only, no new test set was used, and no new performance data provenance is provided. The previous performance characteristics were established during a season when influenza B/Victoria and B/Yamagata lineages were approximately equally proportioned, but the specific sample size and geographic origin for that study are not detailed in this document.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

Not applicable. No new test set for performance evaluation was involved in this labeling-only submission.

4. Adjudication Method for the Test Set:

Not applicable. No new test set for performance evaluation was involved in this labeling-only submission.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done as part of this submission. This submission focuses solely on a labeling change, not on evaluating the device's diagnostic performance or its improvement of human readers.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

The device itself is a real-time RT-PCR diagnostic panel, which is a laboratory test, not an algorithm, and its performance is inherently standalone in the sense that the test provides a result. However, for diagnostic interpretation, the "Intended Use" explicitly states that the kit is used in conjunction with "clinical and epidemiological information," implying that the end result is interpreted by trained personnel. No new standalone performance studies were conducted for this specific 510(k) submission.

7. The Type of Ground Truth Used:

For the original performance characteristics of the predicate device (K132508), the ground truth for influenza B lineage genotyping would have been established through methods outside the scope of this document, typically through confirmatory laboratory methods such as gene sequencing or other validated reference assays. This submission does not provide new ground truth data.

8. The Sample Size for the Training Set:

Not applicable. This submission is for a labeling change only; therefore, no new training set was used or described.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. No new training set was used or described in this labeling-only submission.

{0}------------------------------------------------

K140857

New Special 510(k) CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, Influenza B Lineage Genotyping Kit

8. 510(k) Summary

APR 2 5 2014

CDC hereby submits this Special 510(k) in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter

Centers for Disease Control and Prevention 1600 Clifton Road NE Atlanta, GA 30333 Establishment Registration: 1050190

Contact Person

CDR Yon Yu, Pharm.D. Acting Associate Director for Regulatory Affairs Office of the Director National Center for Emerging and Zoonotic Infectious Diseases Centers for Disease Control and Prevention 1600 Clifton Road, NE, MS C-18 Atlanta, GA 30333 (404) 639-3046 (office) (404) 639-1275 (fax) fkb8@cdc.gov

Proprietary Name

CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, Influenza B Lineage Genotyping Kit

Common or Usual Name

Influenza B Lineage Genotyping Kit

Regulatory Information

Classification Regulation Section: 866.3980- Respiratory viral panel multiplex nucleic acid assay Classification: Class II Classification Product Code: OZE Subsequent Product Codes: NSU Panel: Microbiology

Predicate Device

CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel (K132508)

Device Description

The Influenza B Lineage Genotyping Kit contains components of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel that is used in real-time RT-PCR (rRT-PCR) assays on the ABI 7500 Fast Dx Real-Time PCR Instrument. The Influenza B Lineage Genotyping Kit consists of oligonucleotide primers and dual-labeled hydrolysis (TaqMan®) probes to be used in rRT-PCR for the in vitro qualitative detection and characterization of human influenza viruses from viral RNA in respiratory specimens from patients presenting with influenza-like illness (ILI). The Influenza B Lineage Genotyping Kit is based on technology which is used in many molecular diagnostic assays.

{1}------------------------------------------------

rRT-PCR assays are one-tube assays that first reverse-transcribe specific regions of RNA into cDNA copies. The cDNA then serves as a template for a polymerase chain reaction that utilizes a thermocyclic heating and cooling of the reaction to logarithmically amplify a specific region of DNA. The probe anneals to a specific internal target sequence located between the target loci of the forward and reverse primers. During the extension phase of the PCR cycle, the 5' nuclease activity of Tag polymerase degrades any probe molecules hybridized to amplified target sequence, causing the reporter dye to separate from the quencher dye, and generating a fluorescent signal. With each cycle, additional reporter dye molecules are cleaved from their respective probes, increasing the fluorescence intensity. Fluorescence intensity is monitored at each PCR cycle. Amplification of targets is reflected by logarithmic increase in fluorescence over time in comparison to the background signal.

Intended Use

The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information:

  • . For the determination of the genetic lineage of human influenza B viruses as B/Victoria or B/Yamagata lineage from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW], and dual nasopharyngeal/throat swabs [NPS/TS]) from human patients with signs and symptoms of respiratory infection and/or from viral culture;
  • . To provide epidemiologic information for surveillance of circulating influenza viruses.

Performance characteristics for influenza B lineage genotyping were established during a season when influenza B/Victoria and B/Yamagata lineages were found in approximately equal proportion.

Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.

All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.

Technological Characteristics

The changes proposed to the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel subject to this special 510(k) are for labeling purposes only and will not alter the technological attributes of the device.

Substantial Equivalence Comparison

The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel (K132508) will serve as the predicate for the intended change. Please see Table 1 for a detailed comparison.

{2}------------------------------------------------

New Special 510(k)
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, Influenza B Lincage Genotyping Kit

Table 1: Device ComparisonInfluenza B Lineage Genotyping Kit
CDC Human Influenza Virus Real-time PCR Diagnostic Panel (K132508)
The CDC Human Influenza Virus Real-Time PCR Diagnostic Panel is intended for usein Real-time RT-PCR assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-TimePCR Instrument in conjunction with clinical and epidemiological information:The Influenza B Lineage Genotyping Kit contains reagents and controls ofthe CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Paneland is intended for use in Real-time RT-PCR assays on an AppliedBiosystems (ABI) 7500 Fast Dx Real-Time PCR Instrument inconjunction with clinical and epidemiological information:
For qualitative detection of influenza virus type A or B from viral RNA in upperrespiratory tract clinical specimens (including nasopharyngeal swabs, nasal swabs,throat swabs, nasal aspirates, nasal washes and dual nasopharyngeal throat swabs),and lower respiratory tract specimens (including bronchoalveolar lavages,bronchial washes, tracheal aspirates, sputum, and lung tissue) from human patientswith signs and symptoms of respiratory infection and/or from viral culture:For the determination of the genetic lineage of human influenza Bviruses as B/Victoria or B/Yamagata lineage from viral RNA in upperrespiratory tract clinical specimens (including NPS, NS, TS, NA,NW, and NPS/TS) from human patients with signs and symptoms ofrespiratory infection and/or from viral culture;For the determination of the genetic lineage of human influenza Bviruses as B/Victoria or B/Yamagata lineage from viral RNA in upperrespiratory tract clinical specimens (including NPS, NS, TS, NA,NW, and NPS/TS) from human patients with signs and symptoms ofrespiratory infection and/or from viral culture;
For determination of the subtype of seasonal human influenza A virus as seasonalA/H1. A/113, and/or A/H1pdm09 from viral RNA in upper respiratory tractclinical specimens (including nasopharyngeal swabs, nasal swabs, throat swabs,nasal aspirates, nasal washes and dual nasopharyngeal/throat swabs), and lowerrespiratory tract specimens (including bronchoalveolar lavages, bronchial washes,tracheal aspirates, sputum, and lung tissue) from human patients with signs andsymptoms of respiratory infection and/or from viral culture:To provide epidemiologic information for surveillance of thecirculating influenza viruses.To provide epidemiologic information for surveillance of thecirculating influenza viruses.
For the determination of the genetic lineage of human influenza B viruses asB/Victoria or B/Yamagata lineage from viral RNA in upper respiratory tractclinical specimens (including NPS. NS. TS. NA. NW. and NPS/TS) from humanpatients with signs and symptoms of respiratory infection and/or from viral culture:Performance characteristics for influenza B lineage genotyping wereestablished during a season when influenza B/Victoria andB/Yamagata lineages were found in approximately equal proportion.Performance characteristics for influenza B lineage genotyping wereestablished during a season when influenza B/Victoria andB/Yamagata lineages were found in approximately equal proportion.
Intended Use
Testing with the influenza H5a and H5b primer and probe sets should not beperformed unless the patient meets the most current U.S. Department of Health andHuman Services (DHHS) clinical and epidemiological criteria for testing suspectA/115 specimens. The definitive identification of influenza A/15 (Asian lineage)either directly from patient specimens or from virus cultures requires additionallaboratory testing, along with clinical and epidemiological assessment inconsultation with national influenza surveillance expertsNegative results do not preclude influenza virus infection and shouldnot be used as the sole basis for treatment or other patient managementdecisions. Conversely. positive results do not rule out bacterialinfection or co-infection with other viruses. The agent detected may notbe the definite cause of disease.

Table 1: Device Compariso

8-3

.

{3}------------------------------------------------

New Special 510(k)
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, Influenza B Lineage Genotyping Kit

8-4

.

{4}------------------------------------------------

New Special 510(k)

CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, Influenza B Lineage Genotyping Kit

Risk Analysis

The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel. A risk analysis of the labeling modification to create a separate Package Insert for the Influenza B Lineage Genotyping Kit was performed to verify that the change in the package insert did not present increased or new risks to the user. No new significant risks were identified as a result of the proposed modification.

Substantial Equivalence Conclusion

The changes proposed to the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel do not alter the device's design or technological attributes. In addition, the indications for use and intended use of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel will remain the same. The modification to the package insert included with the Influenza B Lineage Genotyping Kit made available to users retains all required label information, and is determined to be substantially equivalent to the predicate.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a circular logo. The logo contains a symbol of a bird-like figure with three curved lines representing its body and wings. The symbol is surrounded by text that reads 'U.S. Department of Health & Human Services' in a circular arrangement.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20903-0002

CENTERS FOR DISEASE CONTROL AND PREVENTION April 25. 2014 CDR YON YU, Pharm. D. ACTING ASSOCIATE DIRECTOR FOR REGULATORY AFFAIRS OFFICE OF THE DIRECTOR NATIONAL CENTER FOR EMERGING AND ZOONOTIC INFECTIOUS DISEASE 1600 CLIFTON ROAD, MS-C18 ATLANTA GA 30333

Re: K140857

Trade/Device Name: CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, Influenza B Lineage Genotyping Kit Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: II Product Code: OZE. NSU Dated: April 02, 2014 Received: April 03, 2014

Dear Dr. Yu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability i warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements medical device relation (QS) regulation (21 CFR Part 820): and if applicable, the

{6}------------------------------------------------

Page 2-Dr. Yu

electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for vour device on our labeling regulations (21 CFR Parts 801 and 809). please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.lda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometries/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Uwe Scherf -S tor

Sally Hojvat. M.Sc.. Ph.D. Director. Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health

Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known): K140857

Device Name: CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panci. Influenza B Lincage Genotyping Kit

Indications For Use:

The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Intluenza Virus Real-Time RT-PCR Diagnostic Panch and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real Time PCR Instrument in conjunction with clinical and epidemiological information:

  • For determination of the genetic lineage of human influenza B viruses as B/Victoria . or B/Yamagata lineage from viral RNA in upper respiratory tract clinical specimens (including nasopharvngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) from human patients with signs and symptoms of respiratory infection and/or from viral culture:
  • To provide epidemiologic information for surveillance of circulating influenza . viruses.

Performance characteristics for influenza B lineage genotyping were established during a season when influenza B/Victoria and B/Yamagata lineages were found in approximately equal proportion.

Negative results do not prechide influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.

All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.

{8}------------------------------------------------

Prescription Use(Part 21 CFR 801 Subpart D)AND/OROver-The-Counter Use(21 CFR 801 Subpart C)
X

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Li Li -S 2014.04.24 14:44:30 -04'00'

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.