K Number
K140775
Date Cleared
2014-07-23

(117 days)

Product Code
Regulation Number
870.4875
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Turbo Elite devices are indicated for use in the treatment of infrainguinal stenosis and occlusions. When used in conjunction with the Turbo Booster and/or as an accessory to the Turbo Tandem System, the devices are indicated for atherectomy of infrainguinal arteries.

The 0.014" and 0.018" Over-the-wire (OTW) Turbo Elite laser catheters are also indicated for use as an accessory to the use of the Turbo Tandem System in the treatment of femoropoplited artery in-stent restenosis (ISR) in bare nitinol stents, when used in conjunction with Percutaneous Transluminal Angioplasty (PTA).

Device Description

Spectranetics Turbo Elite Laser Ablation Catheters are percutaneous intravascular devices constructed of multiple optical fibers arranged around a guidewire lumen for Over-the-Wire (OTW) configurations, and bundled together for Rapid-Exchange (Rx) versions. Turbo Elite Laser Ablation Catheters are available in an Over the Wire (OTW) configuration and a Rapid Exchange (RX) configuration. The Turbo Elite laser catheters in the OTW configuration are available in six different catheter tip sizes (0.9mm, 1.4mm, 1.7mm, 2.0mm, 2.3mm, and 2.5mm) and three different guide-wire compatibilities (0.014", 0.018", and 0.035"). New indication clearance is only being requested for the OTW configurations of Turbo Elite catheters compatible with 0.014" and 0.018" quidewires.

The Turbo-Tandem Laser Guide Catheter with Laser Atherectorny Catheter (Turbo-Tandem System) is a laser atherectomy catheter constrained within a guiding catheter to facilitate the offset (biased position) of the laser atherectomy catheter. The Turbo-Tandem System is designed to be used to directionally ablate infrainguinal concentric lesions in vessels 5mm or greater at or above the knee. The Turbo Tandem Laser Guide Catheter with Laser Atherectorny Catheter is available in two sizes; a 7F and 8F with a 2.0mm equivalent laser catheter embedded in the system. New indication clearance is only being requested for the Turbo-Tandem 7F catheters.

The multifiber laser catheters transmit ultraviolet energy from the Spectranetics CVX-300® Excimer Laser System to an obstruction in the patient's artery. The ultraviolet energy is delivered to the tip of the laser catheter to photoablate fibrous, calcific, and atheromatous lesions, thus recanalizing diseased vessels. Photoablation is the process by which energy photons cause molecular bond disruption at the cellular level without thermal damage to surrounding tissue.

AI/ML Overview

Acceptance Criteria and Study Details for Spectranetics Turbo-Tandem System and Turbo Elite Laser Catheters

1. Table of Acceptance Criteria and Reported Device Performance

The provided document focuses on seeking a new Indication for Use (IFU) based on clinical evidence, rather than presenting specific numerical acceptance criteria for a device performance test where the device's output is measured against a predefined standard. The clinical study (EXCITE trial) compared the device combination (ELA+PTA) against another treatment (PTA alone) to demonstrate non-inferiority for safety and superiority for effectiveness for the new indication.

Device Effectiveness (Primary Effectiveness Hypothesis):

Acceptance CriteriaReported Device Performance (EXCITE Trial)
Freedom from Target Lesion Revascularization (TLR) through 6 months with ELA+PTA superior to PTA alone.The primary effectiveness hypothesis was met.

Device Safety (Primary Safety Hypothesis):

Acceptance CriteriaReported Device Performance (EXCITE Trial)
Freedom from a Major Adverse Event (MAE) through 30 days with ELA+PTA non-inferior to PTA alone. MAE defined as all-cause death, major amputation in the target limb, or target lesion revascularization (TLR).The primary safety hypothesis was met.
(Additional analysis) Superiority for the safety endpoint.Met (implicitly supported by the statement "Analysis of the data has shown superior safety and effectiveness results compared to PTA.").
No statistical difference in major amputation rates, mortality, serious adverse events, or adverse events between groups.Met.

Note: The document does not provide specific numerical thresholds or percentages for "superiority" or "non-inferiority" for these clinical endpoints. It only states that the hypotheses were met.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set (Clinical Data): The EXCITE trial
    • Sample Size: Not explicitly stated as a number of patients, but referenced as "a prospective randomized controlled trial performed respectively in a 2:1 randomization scheme."
    • Data Provenance: Prospective, randomized controlled trial. The country of origin of the data is not specified in the provided text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This type of information is generally not applicable to clinical trials comparing medical device efficacy and safety directly in patients. The "ground truth" in this context is based on clinical outcomes (e.g., freedom from TLR, occurrence of MAE) observed in patients following treatment, as assessed by clinical investigators and defined trial endpoints, rather than expert consensus on a diagnostic image or interpretation. The trial would have been monitored by a Data Safety Monitoring Board (DSMB) and events adjudicated by independent committees, but the roles and qualifications of such individuals are not detailed here.

4. Adjudication Method for the Test Set

The document mentions that the EXCITE trial involved a "primary safety hypothesis" and "primary effectiveness hypothesis" and that "freedom from a major adverse event (MAE)" was defined. Clinical trials typically have independent clinical events committees (CECs) to adjudicate endpoints like MAE and TLR to ensure consistency and minimize bias. However, the specific adjudication method (e.g., 2+1, 3+1, none) is not detailed in this summary.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. MRMC studies are typically used to assess the performance of diagnostic devices or imaging interpretation by different human readers. The EXCITE trial was a clinical trial comparing the effectiveness and safety of a treatment device (laser atherectomy with PTA) against another treatment (PTA alone), not a study of human readers' performance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No, a standalone study (algorithm only) was not done. The "device" in question is a physical medical device (laser ablation catheter and system) used by a physician during a procedure, not a diagnostic algorithm or AI system. The EXCITE trial evaluated the performance of the device in conjunction with human intervention (the procedure).

7. Type of Ground Truth Used for the Test Set

The ground truth for the clinical effectiveness and safety endpoints in the EXCITE trial was based on clinical outcomes data from patients, specifically:

  • Freedom from Target Lesion Revascularization (TLR): A clinical outcome representing the need for a repeat revascularization procedure of the treated lesion.
  • Freedom from Major Adverse Events (MAE): A composite clinical outcome including all-cause death, major amputation in the target limb, or target lesion revascularization (TLR).

These are direct patient-level clinical events, not expert consensus on images or pathology results in the typical sense of AI/image analysis studies.

8. Sample Size for the Training Set

The concept of a "training set" is not applicable here, as the device is a physical medical instrument, not a machine learning algorithm that underwent a training phase. The EXCITE trial was a clinical validation study to support a new indication for use.

9. How the Ground Truth for the Training Set Was Established

As noted above, there was no "training set" in the context of developing an algorithm. Therefore, the establishment of ground truth for a training set is not relevant to this submission.

{0}------------------------------------------------

Spectranetics Corporation

JUL 2 3 2014

K140775 p. 1 of 4

.

1 510(k) Summary

This 510(k) summary was prepared in accordance with 21 CFR 807.92

Prepared on July 9, 2014

· 1.1 Subject Device

Device Trade NameSpectranetics Turbo-Tandem System andTurbo Elite Laser Catheters
Classification Name21 CFR 870.4875, Intraluminal Artery Stripper
Device Common NameTurbo-Tandem: Percutaneous Laser Ablation CatheterTurbo Elite: Laser Atherectomy Catheters
Device ClassClass II
Classification PanelCardiovascular
Product CodeMCW
510(k) TypeTraditional
Model NumbersTurbo-Tandem: 472-110Turbo Elite: 410-152, 414-151, 417-152, 420-006, 423-001,425-011
SponsorSpectranetics9965 Federal DriveColorado Springs, CO 80921
Establishment Registration No3007284006
Primary ContactAmanda M. JohnsonVice President of Regulatory and Medical AffairsOffice: (719) 447-2452Fax: (719) 447-2040Email: Amanda.Johnson@SPNC.com
Secondary ContactMs. Christine GodleskiSenior Regulatory Affairs SpecialistOffice: (719) 447-2432Fax: (719) 447-2040Email: Christine.Godleski@SPNC.com

1.2 Predicate Device

The Turbo-Tandem system is being compared to the following legally marketed predicate device:

510(k) NumberK094036, K091299
ManufacturerThe Spectranetics Corporation
Trade NameSpectranetics Turbo-Tandem System
Device Common NamePercutaneous Laser Ablation Catheter

:

{1}------------------------------------------------

Spectranetics Corporation

The Turbo Elite Laser Catheters are being compared to the following legally marketed predicate device:

510(k) NumberK060012, K052514, K052296
ManufacturerThe Spectranetics Corporation
Trade NameSpectranetics Turbo Elite System
Device Common NamePercutaneous Laser Ablation Catheter

1.3 Indications for Use

The devices are currently cleared with the following IFUs:

Turbo Elite:For use in the treatment of infrainguinal stenoses and occlusions.
Turbo-Tandem:Indicated for atherectomy of infrainguinal arteries.

Spectranetics is proposing the following new IFU statements based on clinical evidence:

The Turbo Elite devices are indicated for use in the treatment of infrainguinal Turbo Elite: stenosis and occlusions. When used in conjunction with the Turbo Booster and/or as an accessory to the Turbo Tandem System, the devices are indicated for atherectomy of infrainquinal arteries. The 0.014" and 0.018" Over-the-wire (OTW) Turbo Elite laser catheters are also indicated for use as an accessory to the use of the Turbo-Tandem System in the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, when used in conjunction with Percutaneous Transluminal Angioplasty (PTA).

Turbo-Tandem: The 7 and 8 French Turbo-Tandem systems are indicated for atherectorny of infrainguinal arteries.

The 7 French Turbo-Tandem System is indicated for laser atherectomy of de novo or restenotic lesions in native infrainquinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) only in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA). A > 2.0mm pilot channel must be present for treatment using the Turbo-Tandem.

1.4 Device Description

Spectranetics Turbo Elite Laser Ablation Catheters are percutaneous intravascular devices constructed of multiple optical fibers arranged around a guidewire lumen for Over-the-Wire (OTW) configurations, and bundled together for Rapid-Exchange (Rx) versions. Turbo Elite Laser Ablation Catheters are available in an Over the Wire (OTW) configuration and a Rapid Exchange (RX) configuration. The Turbo Elite laser catheters in the OTW configuration are available in six different catheter tip sizes (0.9mm, 1.4mm, 1.7mm, 2.0mm, 2.3mm, and 2.5mm) and three different guide-wire compatibilities (0.014", 0.018", and 0.035"). New indication clearance is only being requested for the OTW configurations of Turbo Elite catheters compatible with 0.014" and 0.018" quidewires.

The Turbo-Tandem Laser Guide Catheter with Laser Atherectorny Catheter (Turbo-Tandem System) is a laser atherectomy catheter constrained within a guiding catheter to facilitate the offset (biased position) of the laser atherectomy catheter. The Turbo-Tandem System is designed to be used to directionally ablate infrainguinal concentric lesions in vessels 5mm or greater at or above the knee. The Turbo Tandem Laser Guide Catheter with Laser Atherectorny Catheter is available in two sizes; a 7F and 8F with a 2.0mm equivalent laser catheter embedded in the system. New indication clearance is only being requested for the Turbo-Tandem 7F catheters.

{2}------------------------------------------------

The multifiber laser catheters transmit ultraviolet energy from the Spectranetics CVX-300® Excimer Laser System to an obstruction in the patient's artery. The ultraviolet energy is delivered to the tip of the laser catheter to photoablate fibrous, calcific, and atheromatous lesions, thus recanalizing diseased vessels. Photoablation is the process by which energy photons cause molecular bond disruption at the cellular level without thermal damage to surrounding tissue.

1.5 Technological Characteristics

There have been no changes to the design or function of the predicate devices. The only change being proposed is to the indication of use. The safety and effectiveness of the existing technological characteristics are being supported by clinical, animal, and bench testing ..

Performance Data 1.6

The following testing was conducted to validate and verify that the subject device met all specifications and was substantially equivalent to the predicate device.

1.6.1 Design Verification and Validation Testing submitted in support of prior 510(k)s

  • Solarization .
  • Tissue ablation efficiency .
  • Physical testing .
  • Trackability ●
  • Torquability ●
  • Tensile .
  • Laser Lifetime .
  • . Packaging stability

Bench testing submitted in support of IDE G110039 and this 510(k) 1.6.2

  • . Stent Fatigue testing (post lasing and visual inspection )
  • . Stent Axial, Bending, Torsion, and Fatigue Study (post - lasing and visual inspection)
  • . Stent corrosion testing (post lasing, inspection and fatigue testing)

1.6.3 Sterilization

  • . There have been no changes to the sterilization process for either the Spectranetics Turbo Elite or the Turbo-Tandem catheters; therefore the sterilization validation previously submitted under the existing 510(k) files is unchanged.

Biocompatibility: 1.6.4

  • . There have been no changes to the materials of construction, or manufacturing process for either the Spectranetics Turbo Elite or the Turbo-Tandem catheters, therefore the biocompatibility testing previously submitted under the existing 510(k) files is unchanged.

{3}------------------------------------------------

Preclinical and Clinical Data: 1.6.5

  • . Pre-clinical Data: Pre-clinical testing was completed to demonstrate the safety of a 2.0mm Turbo Elite Laser Ablation Catheter for ablation treatment of in-stent restenosis via the porcine, stented peripheral artery injury model.
  • Clinical Data: The EXCITE trial evaluated the safety and effectiveness of Excimer Laser . Atherectomy (ELA) using the Spectranetics Turbo Elite Laser Ablation Catheter and the Spectranetics Turbo-Tandem Laser Guide Catheter with Laser Atherectomy Catheter with adjunctive percutaneous transluminal angioplasty (PTA) in comparison with PTA alone in the treatment of femoropopliteal bare nitinol in-stent restenosis in vessels ≥5mm. This trial was a prospective randomized controlled trial performed respectively in a 2:1 randomization scheme. The primary safety hypothesis was that freedom from a major adverse event (MAE) through 30 days with ELA+PTA, which included all-cause death, major amputation in the target limb, or target lesion revascularization (TLR), would be non-inferior to PTA. An additional analysis was also conducted to evaluate superiority for the safety endpoint. The primary effectiveness hypothesis was that freedom from TLR through 6 months with ELA+PTA would be superior to PTA. Both primary safety and effectiveness hypothesis were met. There was no statistical difference in major amputation rates, mortality, serious adverse events, or adverse events between groups,

Substantial Equivalence 1.7

The clinical data presented in this 510(k) was developed and collected through IDE G110039. Analysis of the data has shown superior safety and effectiveness results compared to PTA. which supports clearance of the device for the new indication. The currently marketed Turbo Elite and Turbo-Tandem product designs (materials, construction and specifications) are unchanged from their original design and performance specifications and are substantially equivalent.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 23, 2014

Spectranetics Corporation Ms. Amanda Johnson Vice President of Regulatory & Medical Affairs 9965 Federal Drive Colorado Springs, Colorado 80921-3617

Re: K140775

Trade/Device Name: Turbo-Tandem System, Turbo-Elite Atherectomy Catheters Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: June 20, 2014 Received: June 23, 2014

Dear Ms. Johnson,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{5}------------------------------------------------

Page 2 - Ms. Amanda Johnson

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Nicole (a Jorahim -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140775

Device Name Turbo Elite Laser Catheter

Indications for Use (Describe)

The Turbo Elite devices are indicated for use in the treatment of infrainguinal stenosis and occlusions. When used in conjunction with the Turbo Booster and/or as an accessory to the Turbo Tandem System, the devices are indicated for atherectomy of infrainguinal arteries.

The 0.014" and 0.018" Over-the-wire (OTW) Turbo Elite laser catheters are also indicated for use as an accessory to the use of the Turbo Tandem System in the treatment of femoropoplited artery in-stent restenosis (ISR) in bare nitinol stents, when used in conjunction with Percutaneous Transluminal Angioplasty (PTA).

Type of Use (Select one or both, as applicable)

J Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

THE THE THE THE FOR FOR FOR FOR FOR FOA USE ONLY 1991-1995 1991

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

*An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 870.4875 Intraluminal artery stripper.

(a)
Identification. An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)(b)
Classification. Class II (performance standards).