K Number
K140745
Date Cleared
2014-04-17

(23 days)

Product Code
Regulation Number
870.1650
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Encore 26 Advantage Kit is a kit of sterile disposable devices intended of ruse as accessories for percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.
Encore™ 26 Inflation Device: indicated for use with balloon dilatation catheters to create and monitor pressure in the balloon, and to deflate the balloon.
Gateway™ Plus Y-Adaptor: used for providing hemostasis around balloon dilatation catheters, guidewires and other therapeutic devices during general intravascular procedures.
Torque Device: used for guidewire manipulation during general intravascular procedures.
Insertion Tool: used to facilitate the introduction of a quidewire during general intravascular procedures.

Device Description

The Encore™ 26 Advantage Kit is a Kit of sterile disposable devices intended for use as accessories and percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (Encore™ 26 Advantage Kit), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a clinical study with acceptance criteria and performance metrics in the way one might for a novel AI algorithm or diagnostic tool.

Therefore, many of the requested sections are not applicable to the provided document. I will fill in the information that is present or directly inferable from the document.

1. A table of acceptance criteria and the reported device performance

The submission states, "Design Verification and Design Validation Testing was performed to verify that the performance and usability of the modified Insertion Tool remains substantially equivalent to the predicate device via K140673. In addition Sterilization, Packaging and Biocompatibility testing verifies the overall substantial equivalence to the kit predicate."

While specific numerical acceptance criteria and performance values for each test are not listed in this summary document, the overall acceptance criterion is substantial equivalence to the predicate device(s). The reported performance is that the device met this criterion.

Acceptance Criteria (Implied)Reported Device Performance
Substantial equivalence in performance to predicate device (K123214 and K922410) for the modified Insertion ToolDesign Verification and Design Validation Testing showed the modified Insertion Tool remains substantially equivalent to the predicate device.
Substantial equivalence in sterilization, packaging, and biocompatibility to the kit predicateSterilization, Packaging and Biocompatibility testing verifies overall substantial equivalence to the kit predicate.
No new safety or performance issues raisedNo new safety or performance issues were raised during the device testing.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the summary. The document refers to "Design Verification and Design Validation Testing" but does not detail the specific sample sizes used for these engineering/performance tests. Data provenance (country of origin, retrospective/prospective) is not applicable or provided for this type of device submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable to a device like a balloon inflation kit. Ground truth, in the context of diagnostic algorithms, involves expert labeling of data. For a hardware medical device, "ground truth" would relate to engineering specifications and performance benchmarks, not expert consensus on medical images or diagnoses.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are typically used in clinical studies involving human interpretation or subjective assessments, not for performance testing of a mechanical device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device, and no MRMC study was conducted or mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm-based device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For this type of device, the "ground truth" for performance testing would be established by:

  • Engineering specifications and standards: Performance parameters (e.g., pressure capabilities, sealing integrity, material strength, dimensional accuracy) are measured against predefined engineering requirements and industry standards.
  • Predicate device performance: The performance of the predicate device serves as the benchmark for substantial equivalence.

8. The sample size for the training set

Not applicable. This is not a machine learning or AI device that uses a "training set."

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

K140745

Special 510(k) Submission

i

APR 1 7 2014

510(k) Summary

per 21 CFR §807.92

Submitter's AddressContact Name andInformationBoston Scientific CorporationOne Scimed PlaceMaple Grove, MN 55311Lisa MeeSenior Regulatory Affairs SpecialistPhone: 763-494-1185Fax:763-494-2222e-mail: lisa.mee@bsci.com
Date Prepared24 Mar 2014
Proprietary NameEncore™ 26 Advantage Kit
Common NameBalloon Inflation KitComon name of the Kit components:Inflation device, Insertion Tool, Y-Adaptor and Torque Device
Product CodeMAV - Syringe, Balloon Inflation
ClassificationClass II, 21 CFR Part 870.1650 - Cardiovascular
Predicate Device(s)Encore™ 26 Advantage KitK123214Nov 13, 2012
SCIMED AVENUE CoronaryK922410Jul 23, 1992
Guidewire Insertion Tool
Device DescriptionThe Encore™ 26 Advantage Kit is a Kit of sterile disposabledevices intended for use as accessories and percutaneouscoronary angiography (PTCA) procedures. They allow forballoon inflation and wire control.
Intended UseThe Encore™ 26 Advantage Kits are intended for use asaccessories for percutaneous coronary angiography (PTCA)procedures. They create and monitor balloon inflation andfacilitate wire introduction and control.Individual Device Intended Use:Encore™ 26 Inflation Device: used with balloon●dilatation catheters to create and monitor pressure in theballoon, and to deflate the balloon.Gateway™ Plus Y-Adaptor: used for providinghemostasis around guidewires, balloon dilatationcatheters, and other therapeutic devices.Torque Device: used for guidewire manipulation.Insertion Tool: used to facilitate the introduction of aguidewire.

{1}------------------------------------------------

Indications for UseIndividual Device Indications for Use:
Comparison ofTechnologicalCharacteristicsThe Encore™ 26 Inflation devices are recommended foruse with balloon dilatation catheters to create andmonitor pressure in the balloon, and to deflate theballoon. The GateWay™ PLUS Y-Adapter is recommended forproviding homeostasis around balloon dilatationcatheters, guidewires, and other therapeutic devicesduring general intravascular procedures. The Torque Device is used for guidewire manipulationduring general intravascular procedures. The Insertion Tool is used to facilitate the introduction ofa guidewire during general intravascular procedures.
The proposed Encore™ 26 Advantage Kit Inflation device, Y-Adaptor and Torque device components incorporatesubstantially equivalent design, fundamental technology,manufacturing processes, packaging, sterilization and intendeduse as the Inflation device, Y-Adaptor and Torque devicecomponents featured in the Boston Scientific predicate Encore™26 Advantage Kit (K123214).

The proposed Encore " 26 Advantage Kit Insertion Tool ce component incorporates substantially equivalent design, fundamental technology, manufacturing processes, packaging, sterilization and intended use as the SCIMED AVENUE Coronary Guidewire Insertion Tool device cleared for marketing under K922410.

Comparison to Predicates:

.

CharacteristicProposed compared toPredicates
Mechanism of ActionSame mechanism of action.
ComponentsSame components,configuration, design andfunction.
MaterialsModified Insertion Tool hasequivalent materials (sheathand hub colorant).
PackagingSame packaging materials andpackaging configuration.
Sterilization Method/SALSame method and level ofsterility assurance.
Device CompatibilitySame compatibility.
Device DimensionsModified Insertion Tool hasequivalent dimensions.
BiocompatibilitySame biocompatibility.

{2}------------------------------------------------

Encore™ 26 Advantage Kit

Special 510(k) Submission

Performance Data

Design Verification and Design Validation Testing was performed to verify that the performance and usability of the modified Insertion Tool remains substantially equivalent to the predicate device via K140673. In addition Sterilization, Packaging and Biocompatibility testing verifies the overall substantial equivalence to the kit predicate. No additional testing was required for kit inclusion.

No new safety or performance issues were raised during the device testing. Therefore, these devices may be considered substantially equivalent to the predicate device.

Conclusion

Based on the indications for use, technological characteristics, safety and performance testing, the proposed Encore™ 26 Advantage Kit has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the Encore™ 26 Advantage Kit as submitted in K123214 and the SCIMED AVENUE Coronary Guidewire Insertion Tool as submitted in K922410.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Public Health Scrvice

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 17, 2014

Boston Scientific Corporation c/o Lisa Mee Senior Regulatory Affairs Specialist One Scimed Place Maple Grove, MN 55311

Re: K140745

Trade/Device Name: Encore™ 26 Advantage Kit Regulation Number: 21 CFR 870.1650 Regulation Name: Balloon Inflation Kit Regulatory Class: II Product Code: MAV Dated: March 24. 2014 Received: March 25, 2014

Dear Ms. Mee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{4}------------------------------------------------

Page 2 - Ms. Lisa Mee

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Bram D. Zuckerman -S

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K140745

Device Name: Encore™ 26 Advantage Kit

Indications for Use:

The Encore 26 Advantage Kit is a kit of sterile disposable devices intended of ruse as accessories for percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.

  • Encore™ 26 Inflation Device: indicated for use with balloon dilatation . catheters to create and monitor pressure in the balloon, and to deflate the balloon.
  • Gateway™ Plus Y-Adaptor: used for providing hemostasis around balloon . dilatation catheters, guidewires and other therapeutic devices during general intravascular procedures.
  • Torque Device: used for guidewire manipulation during general ◆ intravascular procedures.
  • Insertion Tool: used to facilitate the introduction of a quidewire during . general intravascular procedures.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR .

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bram D. Zuckerman -S 2014.04.17 17:31:34 -04'00'

Boston Scientific Corporation

. Attachment 1

§ 870.1650 Angiographic injector and syringe.

(a)
Identification. An angiographic injector and syringe is a device that consists of a syringe and a high-pressure injector which are used to inject contrast material into the heart, great vessels, and coronary arteries to study the heart and vessels by x-ray photography.(b)
Classification. Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.