K Number
K140723

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-08-19

(151 days)

Product Code
Regulation Number
892.1720
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NOMAD MD is a handheld and portable general purpose X-ray system. The device uses a fixed tube current and voltage (kVp) and, therefore, is limited to taking diagnostic X-rays of extremities.

It is intended to be used by a qualified and trained clinician on both adult and pediatric patients. It is not intended to replace a radiographic system with variable tube current and voltage (kVp) which may be required for full optimization of image quality and radiation exposure for different exam types.

Device Description

The NOMAD MD is a handheld, general purpose X-ray system that operates on 14.4 V DC supplied by a rechargeable NiCd battery pack. The X-ray tubehead, X-ray controls, and power source are assembled into a single hand-held enclosure. The package includes a spare battery, a Charging Cradle, and an AC-to-DC Power Supply.

AI/ML Overview

The provided text describes the safety and effectiveness testing for the NOMAD MD Handheld X-ray System, as part of its 510(k) premarket notification. Here's a breakdown of the requested information:

1. Table of acceptance criteria and the reported device performance:

The document doesn't explicitly state "acceptance criteria" for diagnostic image quality in a quantitative, pass/fail table format. Instead, it states the objective was to demonstrate substantial equivalence to the predicate device (MinXray HF 100H+) for diagnostic image quality.

Feature / CriterionAcceptance Criteria (Implied)Reported Device Performance
Image QualityDiagnostic quality images of adult and pediatric extremities, equivalent to predicate device."NOMAD MD can produce diagnostic images of adult and pediatric extremities and they are of equivalent diagnostic quality as the MinXray device."
Operator ExposureBelow the requirements of 21 CFR 1020.30(k) leakage radiation limits."Test results indicate that the level of exposure is well below the requirements of 21 CFR 1020.30(k)."
Extremity ImagingAbility to capture diagnostic quality images of extremities from fingers (1.5cm) to thighs (21cm)."NOMAD MD has been validated... to capture diagnostic quality images of body part extremities, such as fingers (from 1.5cm) and thighs (up to 21cm)."
Compliance with StandardsMeet requirements of relevant IEC and CFR standards (e.g., IEC 60601-1, 21 CFR 1020.30/31)."EMC and Electrical Safety testing... found to meet all the requirements in standards IEC 60601-1: 2005, IEC 60601-1-3:2008, IEC 60601-1-6:2010, IEC 60601-2-28:2010, IEC 60601-2-54:2009, IEC 60825:2007, IEC 62366:2007, and IEC 60601-1-2:2007." (Also 21 CFR 1020.30, 1020.31).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Sample Size: The document mentions "radiographic images of foot, knee, elbow, and hand phantoms" were used. It does not specify the exact number of phantoms or the number of images taken.
  • Data Provenance: Not specified, but likely obtained in a controlled laboratory or testing environment by the manufacturer (Aribex, Inc., located in Orem, UT, USA). The study appears to be prospective as it involved taking new images for comparison.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: "A board-certified radiologist" (singular) was used.
  • Qualifications of Experts: "Board-certified radiologist." No further details on years of experience are provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Based on the description "reviewed by a board-certified radiologist who determined that images from both devices were of diagnostic quality and, therefore, substantially equivalent," there was no formal adjudication method described involving multiple readers. It was a single-reader assessment.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC study was done. The described study was a comparison of image quality between two X-ray devices, not an assessment of human reader performance with or without AI assistance.
  • No AI component is mentioned in this device or its evaluation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. The NOMAD MD is an X-ray imaging device, not an AI algorithm. Its performance was evaluated based on the diagnostic quality of the images it produced, assessed by a human expert.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth for image quality was expert assessment/consensus by a single board-certified radiologist. The images were of phantoms, simplifying the "ground truth" to the radiologist's determination of "diagnostic quality."

8. The sample size for the training set:

  • Not applicable. This device is an X-ray emitter, not an AI or imaging processing algorithm that typically requires a training set. The descriptions relate to hardware performance, radiation safety, and image generation capabilities.

9. How the ground truth for the training set was established:

  • Not applicable, as there was no training set for an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a wing-like shape above them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 19, 2014

Aribex, Inc. % Mr. Sanjay Ahuja Director, Regulatory Affairs 744 South 400 East OREM UT 84097

Re: K140723 Trade/Device Name: Nomad MD Handheld X-ray System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: June 5, 2014 Received: July 21, 2014

Dear Mr. Ahuja:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Smh. 7)

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140723

Device Name NOMAD MD Handheld X-ray System

Indications for Use (Describe)

The NOMAD MD is a handheld and portable general purpose X-ray system. The device uses a fixed tube current and voltage (LVp) and, therefore, is limited to taking diagnostic X-rays of extremities.

It is intended to be used by a qualified and trained clinician on both adult and pediatric patients. It is not intended to replace a radiographic system with variable tube current and voltage (kVp) which may be required for full optimization of image quality and radiation exposure for different exam types.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Form Approved: OMB No. 0910-0120

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

NOMAD MD Handheld X-ray System

Submitter:

Aribex, Inc. 744 South 400 East Orem, UT 84097 Phone No: 801-226-5522 FAX No:

Date of Summary:

March 14, 2014

Contact Person:

Christopher Dodge, Senior Regulatory Affairs Specialist or Sanjay Ahuja, Ph.D., Director, Regulatory Affairs

Device Name:

  • Trade Name: NOMAD MD Handheld X-ray System ●
  • . Common Name:
  • Nomad MD
  • Classification Name: Mobile X-ray system 21CFR 892.1720
  • Regulation:
  • Product Code: . IZL

Predicate Device for Which Substantial Equivalence is Claimed:

MinXray HF 100H+ Portable X-ray Unit (K052721)

Device Description:

The NOMAD MD is a handheld, general purpose X-ray system that operates on 14.4 V DC supplied by a rechargeable NiCd battery pack. The X-ray tubehead, X-ray controls, and power source are assembled into a single hand-held enclosure. The package includes a spare battery, a Charging Cradle, and an AC-to-DC Power Supply.

Consideration was given to three FDA guidance documents, "Pediatric Information for X-ray Imaging Device Premarket Notifications" issued on May 10,

{5}------------------------------------------------

2012, "Radiation Safety Considerations for X-Ray Equipment for Hand-Held Use," issued on December 24, 2008, and "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices," dated August 6,1999. Aribex determined that the solid state quidance did not apply to the NOMAD MD because it is an X-ray emitter and the quidance document is written for X-ray image recording systems, not for X-ray emitters.

Indications for Use/Intended Use:

The NOMAD MD is a handheld and portable general purpose X-ray system. The device uses a fixed tube current and voltage (kVp) and, therefore, is limited to taking diagnostic X-rays of extremities.

It is intended to be used by a qualified and trained clinician on both adult and pediatric patients. It is not intended to replace a radiographic system with variable tube current and voltage (kVp) which may be required for full optimization of image quality and radiation exposure for different exam types.

Summary of Technological Characteristics:

The NOMAD MD has been demonstrated to be substantially equivalent to the predicate device, the MinXray HF 100H+ Portable X-ray Unit (cleared under K052721). The proposed device has been verified and validated to satisfy the requirements derived from the indications for use and the substantial equivalence table.

NOMAD MD shares many of the same components and features as the predicate device including:

  • Portable X-ray source
  • . Collimation of X-ray
  • Two stage triggering mechanism .
  • Source to skin guard to limit distance between X-ray source and patient .
  • Constant potential waveform
  • Similar user interface with tactile membrane pad with up-down push ● buttons for making selections
  • . Similar alerts and alarms that indicate equipment state of readiness to use

The MinXray operator instructions require that the operator "stand as far as possible from the X-ray unit while holding the exposure switch and pressing the button" in order to avoid being exposed to excess leakage and backscatter radiation.

The following is a comparison between Nomad MD (proposed) and the predicate device, the MinXray HF 100H+ Portable X-ray Unit (cleared under K052721).

{6}------------------------------------------------

MinXray HF100H+(K052721)Nomad MDX-ray System (K140723)
Feature
Regulation No.21 CFR 892.172021 CFR 892.1720
Regulation ClassIIII
Product CodeIZLIZL
Indications forUse/Intended Use:Intended for use by aqualified/trained physician ortechnician on both adult andpediatric subjects for takingdiagnostic x-rays.The NOMAD MD is ahandheld and portablegeneral purpose X-raysystem. The device uses afixed tube current andvoltage (kVp) and,therefore, is limited totaking diagnostic X-rays ofextremities.
It is intended to be used bya qualified and trainedclinician on both adult andpediatric patients. It is notintended to replace aradiographic system withvariable tube current andvoltage (kVp) which may berequired for full optimizationof image quality andradiation exposure fordifferent exam types.
Principle of OperationGeneral Purpose Diagnostic X-RayGeneral PurposeDiagnostic X-Ray
TECHNOLOGICAL:
Size: Body8.75"Hx9.5"Wx15.35"Lexcluding skin guards9.5"H x 5.25"Wx10"L(excluding Source Skinguard)
Weight40.9 lbs.11.0 lbs.
Source to skin distance30 cm30 cm
Focal Spot1.2 mm0.4 mm
CollimatorFour manually and steplesslyadjustable shutters with lightbeam type central x-rayindicator (Advantech R72)Four manually andsteplessly adjustableshutters with LED LightField Center Indicator
Triggering MechanismTwo stage triggeringTwo stage triggering
User InterfaceUp-down push buttons for kVpselections and exposure timeselections with LED indicatorsand mAs indicators.Up-down buttons forexposure time selection,with timer display.
Energy Source100-140 VAC 50/60 HzRechargeable 14.4 V DCNiCd battery pack
Exposure Time0.03-0.2 sec (in 0.01 sec Steps)0.2-0.4 sec (in 0.02 sec Steps)0.4-1.0 sec (in 0.05 sec Steps)0.02 - 0.99 seconds in 0.01increments
FeatureMinXray HF100H+(K052721)Nomad MDX-ray System (K140723)
1.0-4.0 sec (in 0.1 sec Steps)
mA30 mA @ 40-60 kV25 mA @ 62-80 kV20 mA @ 82-100 kV2.0 mA fixed
kVp40-100kVp75 kVp fixed
WaveformConstant PotentialConstant Potential
Alerts and AlarmsAs least two specific alerts andalarms indicate equipment stateof readiness to useSeven specific alerts andalarms indicatingequipment's state ofreadiness to use
Class II LaserNot a feature on this unitUsed to indicate thedistance from the x-raysource to the target to bediagnosed.
User Interface EaseTactile membrane pad withaudible signals indicatessettings.Large illuminated digitalexposure setting screen.Tactile membrane pad withaudible signals indicatessettings.Large illuminated digitalexposure setting screen.
SafetyExposure switch buttonilluminates if unit misfires. Theunit will lock up and needs tobe restarted from the offpositionDuring the exposure an audiblesignal and the X-ray indicatorilluminates.Locked upon powering onwith LC indicator, along withelongated audible signal.Unit will power down after2.5 minutes of inactivity.After 10 seconds of beingenabled, the unit will lock ifexposure does not occur.Continuous audible signal isemitted along with indicatorlights indicating thatexposure is completed.
Low Battery Indicatorand Audible SignalNot a feature on this unitbecause it is AC/DC powered.LED illuminates and flasheswhen battery charge isneeds or needs rechargingAudible signal indicates lowbattery signal
Electrical SafetyStandardsUL 2601, IEC60601IEC60601-1, EN60601-2
Performance Standard21 CFR 1020.3021 CFR 1020.30, 1020.31IEC60601-3

{7}------------------------------------------------

Per the recommendations of the FDA Guidance document on "Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use," Aribex has incorporated internal shielding into the design of the NOMAD MD. This shielding, made from a proprietary bismuth blend, encapsulates the X-ray source assembly and prevents the operator of the device from being exposed to any appreciable amounts of leakage radiation. Test results indicate that the level of exposure is

{8}------------------------------------------------

well below the requirements of 21 CFR 1020.30(k). Evidence of the effectiveness of this internal shielding can be found in Section 19 NOMAD MD Performance Testing - Bench. Specifically, TP-0223, Test Case Scatter Radiation NOMAD MD, identifies the operator exposure dose within the significant zone of occupancy. See also QR-0200 Risk Analysis. NOMAD MD - Risk ID R009. R017.

Since the NOMAD MD, unlike its dental predecessor, is held on average 28 inches from the patient, it was not considered practical to incorporate an external backscatter shield into the design of the device. The effective size of the backscatter shield would need to be about three feet in diameter. For this reason, Aribex recommends in its operator manual (MP-0221, Section 3.3, Section 4.7[6], and Section 7.5) the use of a lead apron and thyroid collar when operating the NOMAD MD. Also see QR-0200 Risk Analysis, NOMAD MD - Risk ID R061.

As evidenced through Design Verification and Validation, NOMAD MD can produce diagnostic images of adult and pediatric extremities and they are of equivalent diagnostic quality as the MinXray device.

An extremity is generally defined as the distal or terminal portion of any bodily limb, or any portion of bodily limbs such as arms and legs. The NOMAD MD has been validated (see Section 19, TP-0302, Extremity Thickness Performance, NOMAD MD) to capture diagnostic quality images of body part extremities, such as fingers (from 1.5cm) and thighs (up to 21cm).

Based upon an analysis of the technological differences, and the substantiation through design verification and validation. Aribex determines that the safety and effectiveness of the proposed NOMAD MD is substantially equivalent to the MinXray predicate device and does not raise any new concerns.

Non-Clinical Test Data:

Performance bench testing was conducted as part of design control to ensure the safety and effectiveness of the NOMAD MD is equivalent to the predicate device. The safety and effectiveness of NOMAD MD for the proposed indications for use is substantiated through verification and validation testing on diagnostic images.

An image comparison study was also conducted by taking radiographic images of foot, knee, elbow, and hand phantoms using both the NOMAD MD and the predicate device. The images were then reviewed by a board-certified radiologist who determined that images from both devices were of diagnostic quality and, therefore, substantially equivalent.

Software documentation has been provided in Section 17 according to the FDA's requirement for software that meets the moderate level of concern.

{9}------------------------------------------------

EMC and Electrical Safety testing on the NOMAD MD was performed and found to meet all the requirements in standards IEC 60601-1: 2005, IEC 60601-1-3:2008, IEC 60601-1-6:2010, IEC 60601-2-28:2010, IEC 60601-2-54:2009, IEC 60825:2007, IEC 62366:2007, and IEC 60601-1-2:2007.

Clinical Test Data:

Clinical testing was deemed not required to support the indications for use and therefore, it has not been conducted on the Nomad MD proposed product.

Conclusion:

Based on the comparison of the indications for use/intended use, the technological characteristics, and the operation of the device, Aribex concludes that NOMAD MD is substantially equivalent to the predicate device. In addition, based upon the review of the performance, standards analysis, and labeling, the differences between NOMAD MD device and the MinXray predicate device do not raise new concerns on safety and effectiveness for the proposed indications of use.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.