(102 days)
The Nitrile Powder Free Examination Gloves -Violet Blue (VBLU) is a non-sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
The Nitrile Powder Free Examination Gloves - Dak Blue (DKBU) is a non-sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
The Nittle Powder Free Examination Gloves - Aqua Blue (ABLU) is a non-sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Nitrile Powder Free Examination Gloves- Aqua Blue (ABLU) Nitrile Powder Free Examination Gloves- Violet Blue (VBLU) Nitrile Powder Free Examination Gloves- Dark Blue (DKBU)
This document is a 510(k) premarket notification for Nitrile Powder Free Examination Gloves. It details the indications for use of the gloves but does not contain information regarding detailed acceptance criteria or the specific studies used to demonstrate device performance against such criteria.
The document primarily focuses on:
- Regulatory Clearance: Acknowledging the review of the 510(k) application and determining substantial equivalence to legally marketed predicate devices.
- Device Identification: Trade names (Aqua Blue, Violet Blue, Dark Blue), regulation number, regulation name, regulatory class, and product code.
- Indications for Use: Stating that the gloves are "a non-sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner."
- Regulatory Compliance: Reminders about general controls provisions, potential additional controls, and contact information for further advice.
Therefore, based on the provided text, I cannot extract the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment.
The document is a clearance letter and "Indications for Use" statement, not a detailed technical report or study summary.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
June 20, 2014
Hartalega NGC SDN BHD Ms. Nurul Aisyah Kong Senior Manager- Quality Assurance No. 1 Persiaran Tanjung Lot PT4330 Kawasan Perindustrian Tanjung Jalan B20, 64000 Sepang Selangor Darul Ehsan, Malaysia
Re: K140610
Trade/Device Name: Nitrile Powder Free Examination Gloves- Aqua Blue (ABLU) Nitrile Powder Free Examination Gloves- Violet Blue (VBLU) Nitrile Powder Free Examination Gloves- Dark Blue (DKBU) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: May 19, 2014 Received: May 22, 2014
Dear Ms. Kong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 - Ms. Nurul Aisyah Kong
found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
MarysBonner-S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K140610
Device Name
Nitrile Powder Free Examination Gloves - Violet Blue (VBLU)
Indications for Use (Describe)
The Nitrile Powder Free Examination Gloves -Violet Blue (VBLU) is a non-sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
Do Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| A can and proportion and a comprehensive to the country of a. | Book and a production a comprehens a comparent of the compares of the controlled・・・・・・・・・・・・ | ||
|---|---|---|---|
| Concurrence of Center for Davices and Radiological Health (CDRH) /Signature) |
Sreekanth Gutala -S
Digitally signed by Sreekanth Gutala -5 o=U.S. Government, ou=HHS People, 00300.100.1.1=2000540490, peranti Gutala - S te: 2014.06.19 14:57:44 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1996.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, Including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Offica of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Acre (105) 148 145 140
... .. . . . . . . . . . . . . . . . . . . . . . . . . .
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K140610
Device Name
Nitrile Powder Free Examination Gloves - Dark Blue (DKBU)
Indications for Use (Describe)
The Nitrile Powder Free Examination Gloves - Dak Blue (DKBU) is a non-sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
| And Annual Annual Annual Annual Annual Andrew Andrew And | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | For a book for a finance for any and the many of the collection of the production of the difference the desires | |||
|---|---|---|---|---|---|
| Type of Use (Select one or both, as applicable) | |||||
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED,
A CONSULTION PORTER FOR FOR FOR FOR FDA USE ONLY CAN CONTROLLER CONTROLLER CONSTITUTION OF Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) igned by Sreekanth Gutala -S
=U.S. Government ou=HHS. ou=FDA 0.9.2342.19200300.100.1.1=2000540490. reekanti Gutala -S 2014.06.19 14:58:24 -04'00
This section applies only to requirements of the Papenwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data scurces, gather and maintain the deta needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or soonsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
4
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K140610
Device Name
Nitrile Powder Free Examination Gloves - Aqua Blue (ABLU)
Indications for Use (Describe)
The Nittle Powder Free Examination Gloves - Aqua Blue (ABLU) is a non-sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
이 참석 기술 FOR FOR FOR FOR USE ONEY - 1 - 1 - 1 - 2 - 2 - 2 Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Digitally sloned by Sreekanth Gutata -S o=U.S. Government, ou=HHS, ou=FDA 00300.100.1.1=2000540490, 42.19 eekanti Gutala -S Date: 2014.06.19 14:54:27 -04'00 This section applies only to requirements of the Paperwork Reduction Act of 1995.
'DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF ENAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete. and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.