K Number
K140607
Manufacturer
Date Cleared
2014-05-08

(59 days)

Product Code
Regulation Number
888.3030
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The aap LOQTEQ® Olecranon Plates are indicated for fixation of fractures, osteotomies and nonunions of the olecranon, particularly in osteopenic bone.

Device Description

The aap LOQTEQ® Olecranon Plate consists of bone plates and bone screws, to be implanted by a surgeon in order to achieve an internal fixation of bone fragments typically after fractures. If the plates are used in conjunction with locking screws, a so called internal fixator will be realized (internal fixation).
The aap LOQTEQ® Olecranon Plate System consists of:
• LOQTEQ® Olecranon Plate (left and right)
to be used with
• LOQTEQ® Cortical Screw 2.7, small head T8, self-tapping
• Cortical Screw 2.5, small head T8, self-tapping
• LOQTEQ® Cortical Screw 3.5, T15, self-tapping
• Cortical Screw 3.5, self-tapping
• Set of Instruments aap LOQTEQ® Olecranon Plate

AI/ML Overview

The provided documentation describes the aap LOQTEQ® Olecranon Plate, a metallic bone fixation appliance. The study conducted to demonstrate its acceptance criteria focuses on non-clinical mechanical testing to establish substantial equivalence to predicate devices.

Here's an analysis based on your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Mechanical performance of the aap system to be substantially equivalent to predicate devices.Customized 4-point bending tests (static and dynamic) based on ASTM F382-99 demonstrated substantial equivalence with respect to mechanical performance.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state the specific number of aap LOQTEQ® Olecranon Plates or predicate devices tested for the mechanical evaluation. It refers to "tests" and "test results" in the plural, implying multiple tests were conducted, but the exact sample dimension is not provided.
  • Data Provenance: The data is from non-clinical testing conducted by the manufacturer, aap Implantate AG, in Germany. This is a prospective evaluation of the device's mechanical properties.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • This question is not applicable as the study described is a non-clinical mechanical testing study, not a study involving human diagnosis or interpretation where expert ground truth would be established. The "ground truth" here is the physical performance data against established ASTM standards.

4. Adjudication Method for the Test Set

  • This question is not applicable for a non-clinical mechanical testing study. Adjudication methods are typically relevant for studies involving human interpretation or clinical outcomes where discrepancies need to be resolved.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • This question is not applicable. The provided document describes a medical device (bone plate), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study comparing human readers with and without AI assistance was not conducted.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • This question is not applicable. As mentioned above, the device is a physical bone plate, not an algorithm or AI system.

7. The Type of Ground Truth Used

  • The "ground truth" for this non-clinical study is defined by established mechanical testing standards, specifically ASTM F382-99, and the performance of the legally marketed predicate devices. The acceptance criterion for the new device was to demonstrate "substantial equivalence" to the mechanical performance of these predicate devices.

8. The Sample Size for the Training Set

  • This question is not applicable. There is no "training set" in the context of this non-clinical mechanical testing study. Training sets are relevant for machine learning or AI models.

9. How the Ground Truth for the Training Set Was Established

  • This question is not applicable for the same reasons as #8.

{0}------------------------------------------------

Summary of Safety and Effectiveness

MAY 0 8 2014

Sponsor:aap Implantate AGLorenzweg 5D-12099 Berlin Germany
Company Contact:Dr. Christian ZietschPhone:+49-30-750-19-193Fax: +49-30-750-19-111
DateFebruary 26 / 2013
Trade Name:aap LOQTEQ® Olecranon Plate
Common Name:Olecranon Plate
Classification Name andReference:21 CFR 888.3030 Single/multiple component metallicbone fixation appliances and accessories - Class II
Product CodeHRS
Predicate device:Synthes (USA) 3.5mm LCP Olecranon Plates, as part ofthe 3.5 / 4.5 mm LCP System under the premarket notifica-tion K082807 (Jan 13, 2009) and Synthes (USA) SmallFragment Dynamic Compression Locking (DCL) Systemunder the premarket notification K000684 (Apr 28, 2000)
Device Description:The aap LOQTEQ® Olecranon Plate consists of boneplates and bone screws, to be implanted by a surgeon inorder to achieve an internal fixation of bone fragments typi-cally after fractures. If the plates are used in conjunctionwith locking screws, a so called internal fixator will be real-ized (internal fixation).
The aap LOQTEQ® Olecranon Plate System consists of:• LOQTEQ® Olecranon Plate (left and right)to be used with• LOQTEQ® Cortical Screw 2.7, small head T8,self-tapping• Cortical Screw 2.5, small head T8, self-tapping• LOQTEQ® Cortical Screw 3.5, T15, self-tapping• Cortical Screw 3.5, self-tapping• Set of Instruments aap LOQTEQ® Olecranon Plate
Material:Implants are made of Ti6Al4V (ASTM F136 or ISO 5832-3)
Indications:The aap LOQTEQ® Olecranon Plates are indicated forfixation of fractures, osteotomies and nonunions of theolecranon, particularly in osteopenic bone.
Substantial EquivalenceThe Substantial Equivalence of the new device and thepredicate device is based on similar intended use, design,functionality, components and materials in use.
Documentation including mechanical testing to show thesubstantial equivalence has been provided with this sub-mission.
Performance Data(Non-Clinical and /or Clinical):Non-clinical tests have been performed and showthe substantial equivalence of the device.

{1}------------------------------------------------

Summary of Non-clinical tests:

Type of test:

Customized 4-point bending tests of bone plates, based on ASTM F382-99 (static and dynamic)

Assessment of test results:

Substantial equivalence with respect to the mechanical performance of the aap system could be stated due to the test results gained. The subject device is substantial equivalent, and whose performance meets the requirements of its pre-defined acceptance criteria and intended uses.

Documentation regarding the mechanical testing to show the substantial equivalence has been provided with this submission.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image contains the words "Public Health Service" in bold, black font. The text is stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The background of the image is white.

ood and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 8, 2014

aap Implantate AG Dr. Christian Zietsch Manager Regulatory Affairs Lorenzweg 5 12099 Berlin Germany

Re: K140607

Trade/Device Name: aap LOQTEQ® Olecranon Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: February 28, 2014 Received: March 10, 2014

Dear Dr. Zietsch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Dr. Christian Zietsch

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDeyices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.

Sincerely yours.

Lori A. Wiggins

  • for
    Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

Indications for Use Statement

510(k) Number (if known): K140607

Device Name: aap LOQTEQ® Olecranon Plate

Indications for Use:

The aap LOQTEQ® Olecranon Plates are indicated for fixation of fractures, osteotomies and nonunions of the olecranon, particularly in osteopenic bone.

Prescription Use × (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth=D&Frank -S

Division of Orthopedic Devices

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.