K Number
K140579
Date Cleared
2015-02-06

(337 days)

Product Code
Regulation Number
884.2660
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SD3 Series Ultrasonic Pocket Dopplers (hereinafter called "the Doppler") are intended to be used by health care professionals including registered nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices.

The 2 MHz and/or 3 MHz waterproof probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. The 3 MHz waterproof probe is used for more than 9-week gestation and the 2 MHz is used for 12-week gestation. They can also be used to verify fetal heart viability.

The 4 MHz, 5 MHz and/or 8 MHz waterproof vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease.

Device Description

The SD3 series Ultrasonic Pocket Doppler is a hand-held device for non-invasive measurement and display of fetal heart rate and blood flow velocity utilizing the principle of Doppler shift of an ultrasound. The ultrasound is transmitted from the probe to patient body (maternal abdominal wall), and moves through biophysical objects. The acoustic ultrasound is reflected by blood and moving objects such as the fetal heart. The reflected ultrasound is received by the probe and is converted into electric signals. The waveform data are applied to the CPU for all the digital processing on OLED Display, operation keys. The audio signal is taken off for the routing to the speaker to generate the analogue signals before digital processing. The following probes are supplied with the SD3 series Ultrasonic Pocket Doppler: 1. 2 MHz for fetal heart rate. 2. 3 MHz for fetal heart rate 3. 4 MHz for detections of arterial and venous blood flow velocity. 4. 5 MHz for detections of arterial and venous blood flow velocity. 5. 8 MHz for detections of arterial and venous blood flow velocity.

AI/ML Overview

The provided document describes the Edan Instruments, Inc. SD3 Series Ultrasonic Pocket Doppler (K140579). Here's an analysis of the acceptance criteria and supporting studies:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/TestAcceptance CriteriaReported Device Performance
FHR Measuring Range50 bpm ~ 240 bpm50 bpm ~ 240 bpm (Met)
FHR Resolution1 bpm1 bpm (Met)
FHR Accuracy±2 bpm (for 2MHz and 3Mhz probes)±2 bpm (Met, based on FHR simulator testing)
FHR Sensitivity (3 MHz probe)9 weeks gestation9 weeks gestation (Met, based on clinical testing)
FHR Sensitivity (2 MHz probe)12 weeks gestation11-week to 13-week gestation (Met, based on clinical testing)
Clinical Detection Rate (2 MHz, 11-13 weeks)95% "pass" rate157 out of 162 cases (96.9%) were "pass" (Met)
Clinical Detection Rate (3 MHz, 9-10(+) weeks)95% "pass" rate95 out of 99 cases (95.9%) were "pass" (Met)
Repeatability of FHR measurements95% confidenceRepeatable to a 95% confidence (Met, based on Students T-test)

2. Sample Sizes Used for the Test Set and Data Provenance

  • FHR Accuracy and Repeatability: The document doesn't specify a sample size for the FHR simulator testing (non-clinical). It only states that a simulator was used.
  • Clinical Detection Rate:
    • 2 MHz probe: 162 tests for 11-week to 13-week pregnant women.
    • 3 MHz probe: 99 tests for 9-week to 10(+)-week pregnant women.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical data. It mentions "Physicians at 6 hospitals," but the location of these hospitals is not specified. However, the submitter Edan Instruments, Inc. is based in Shenzhen, P.R. China, suggesting the data may be from China, but this is not confirmed. The studies are prospective in nature, as they involve actively taking measurements.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • FHR Accuracy and Repeatability (Non-clinical): For the FHR simulator testing, the "ground truth" was established by the FHR simulator itself, which provides a known, controlled FHR. No human experts were involved in establishing this ground truth.
  • Clinical Detection Rate: The "ground truth" for the clinical tests was likely the presence of a fetal heart rate (FHR) that could be detected. The determination of a "pass" or "fail" (whether FHR could be detected by the device) was made by "physicians at 6 hospitals." Their specific qualifications beyond "physicians" are not detailed, but the intended users are described as "health care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians." It's implied that these professionals are the ones making the assessment during the clinical tests.

4. Adjudication Method for the Test Set

The document does not describe an explicit adjudication method (like 2+1 or 3+1 consensus) for the clinical tests. The "pass" or "fail" determination for FHR detection appears to be a direct result of whether the device could detect an FHR and was likely determined by the operating healthcare professional.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not conducted. This device is an ultrasonic pocket doppler, not an AI-powered diagnostic tool requiring human-in-the-loop performance measurement. The evaluation focuses on the device's ability to detect FHR and blood flow.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, a standalone performance evaluation was done for the device's core functionality.

  • FHR Accuracy and Repeatability: These were tested using an FHR simulator, which is an assessment of the device's performance in a controlled environment without human interpretation influencing the measurement itself.
  • Clinical Detection Rate: While humans (healthcare professionals) operated the device, the "pass" or "fail" criterion was whether the device itself could detect the FHR. This is a measure of the device's standalone capability in a clinical setting.

7. The Type of Ground Truth Used

  • FHR Accuracy and Repeatability: Instrumental ground truth (established by an FHR simulator providing known values).
  • Clinical Detection Rate: Clinical observation/detection by healthcare professionals, confirming the presence or absence of a detectable fetal heart rate. It's not explicitly stated if this was confirmed by another gold standard (e.g., a full ultrasound), but the primary goal was to show the Doppler's ability to detect.

8. The Sample Size for the Training Set

The document does not mention any training set. This is a typical regulatory submission for a hardware medical device with embedded algorithms, not a machine learning or AI algorithm that requires a separate training phase with a distinct training dataset. The device operates based on physics principles (Doppler shift) and signal processing, not on learned patterns from a training set in the AI sense.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set for a machine learning algorithm was mentioned or used.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 9, 2015

Edan Instruments, Inc Queena Chen Certification Engineer 3/F - B, Nanshan Medical Equipments Park Nanhai Road 1019# Shekou. Nanshan Shenzhen 518067 P.R. China

Re: K140579 Trade/Device Name: SD3 Series Ultrasonic Pocket Doppler- SD3 LITE, SD3, SD3 PLUS, SD3 PRO, SD3 VASCULAR Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG, DPW Dated: January 4, 2015 Received: January 7, 2015

Dear Queena Chen,

This letter corrects our substantially equivalent letter of February 6, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joyce M. Whang -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K140579

Device Name: SD3 Series Ultrasonic Pocket Doppler SD3 LITE, SD3, SD3 PLUS, SD3 PRO, SD3 VASCULAR

Intended Use:

The SD3 Series Ultrasonic Pocket Dopplers (hereinafter called "the Doppler") are intended to be used by health care professionals including registered nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices.

The 2 MHz and/or 3 MHz waterproof probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. The 3 MHz waterproof probe is used for more than 9-week gestation and the 2 MHz is used for 12-week gestation. They can also be used to verify fetal heart viability.

The 4 MHz, 5 MHz and/or 8 MHz waterproof vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease.

n Prescription Use Or Over the Counter Use

(21 CFR Part 801 Subpart D)

Over the Counter Use _

(21 CFR Part 801 Subpart C)

IPI FASE DO NOT WRITE RFI OW THIS I INF-C.ONTINUIF ON ANOTHER PAGE IFNFFDFD)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

{3}------------------------------------------------

510(k) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

1. Submitter:Edan Instruments, Inc.
3/F-B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, Shekou, Nanshan Shenzhen, 518067 P.R. China
Tel.: (0755) 26856469
Fax: (0755) 26882223
Contact Person:Queena Chen
Prepare date:2015-02-04
2. Device name and classification:Device Name: SD3 series Ultrasonic Pocket Doppler
Model: SD3 LITE, SD3, SD3 PLUS, SD3 PRO, SD3 VASCULAR, hereinafter called SD3 series
Classification Name:
21 CFR 884.2660 Fetal ultrasonic monitor and accessories
21 CFR 870.2100 Cardiovascular blood flowmeter
Product code: KNG, DPW
Regulatory Class: Class II
3.Premarket Notification Class III Certification and SummaryNot applicable, the subject device is Class II.
4. Predicate Device(s):Sonotrax Series Ultrasonic Pocket Doppler / K101960 / Edan Instruments, Inc.
5. Device Description:The SD3 series Ultrasonic Pocket Doppler is a hand-held device for non-invasive measurement and display of fetal heart rate and blood flow velocity utilizing the principle of Doppler shift of an ultrasound. The ultrasound is transmitted from the probe to patient body (maternal abdominal wall), and moves through biophysical objects. The acoustic ultrasound is reflected by blood and moving objects such as the fetal heart. The reflected ultrasound is received by the probe and is converted into electric signals. The waveform data are applied to the CPU for all the digital processing on OLED Display, operation keys. The audio signal is taken off for the routing to the speaker to generate the analogue signals before digital processing. The following probes are supplied

Page 1 / 5

{4}------------------------------------------------

with the SD3 series Ultrasonic Pocket Doppler: 1. 2 MHz for fetal heart rate. 2. 3 MHz for fetal heart rate 3. 4 MHz for detections of arterial and venous blood flow velocity. 4. 5 MHz for detections of arterial and venous blood flow velocity. 5. 8 MHz for detections of arterial and venous blood flow velocity. The subject device is not a sterilized product, so there is no detailed information about its sterilization status and the shelf life included in the submission. FHR Performance: Sensitivity: 9 weeks gestation (3 MHz) FHR Measuring Range: 50 bpm ~ 240 bpm Resolution: 1 bpm ±2 bpm Accuracy: The SD3 Series Ultrasonic Pocket Dopplers (hereinafter called "the 6. Indications for Doppler") are intended to be used by health care professionals including Use: registered nurses, practical nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices. The 2 MHz and/or 3 MHz waterproof probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. The 3 MHz waterproof probe is used for more than 9-week gestation and the 2 MHz is used for 12-week gestation. They can also be used to verify fetal heart viability.

The 4 MHz, 5 MHz and/or 8 MHz waterproof vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease.

7. Predicate Device Comparison

Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device.

The differences between the subject device and predicate device include physical specifications, display type. Please refer to following table. All above differences do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.

ItemSONOTRAXSD3
Manufacturer/K#EDAN Instruments/ K101960EDAN Instruments/ K140579
Intended UseThe Sonotrax series of Ultrasonic PocketDoppler are intended for use by health careprofessionals including registered nurses,practical nurses, midwives, ultrasoundtechnicians, and physicians assistants, byprescription from licensed physicians inThe SD3 Series Ultrasonic PocketDopplers (hereinafter called “theDoppler”) are intended to be used byhealth care professionals includingregistered nurses, practical nurses,midwives, ultrasound technicians, and
hospitals, clinics and private offices.physician assistants, by prescription from
licensed physicians in hospitals, clinics and
private offices.
The 2 MHz and/ or 3 MHz obstetricalprobes are indicated for the detection of fetalheart rate from early gestation thru deliveryand as a general indication of fetal wellbeing. They can also be used to verify fetalheart viability following patient trauma.The 2 MHz and/or 3 MHz waterproofprobes are indicated for the detection offetal heart rate from early gestation thrudelivery and as a general indication of fetalwell being. The 3 MHz waterproof probe isused for more than 9-week gestation andthe 2 MHz is used for 12-week gestation.They can also be used to verify fetal heartviability.
The 4 MHz, 5 MHz and/or 8 MHz vascularprobes are indicated for the detection ofblood flow in veins and arteries for assistingin the detection of peripheral vasculardisease.The 4 MHz, 5 MHz and/or 8 MHzwaterproof vascular probes are indicatedfor the detection of blood flow in veins andarteries for assisting in the detection ofperipheral vascular disease.
Installationandusehand-heldhand-held
Mode of operationContinuous operationContinuous operation
Dimensions138mm (W) × 85 mm (L) × 32 mm (H)168mm (W)×67 mm (L) × 31 mm (H)
Weight290g<350 g (including the battery)
Operating:5°C ~40°C5°C ~40°C
Transport/Storage:-20°C~55°C-20~55°C
25%~85% RH25% ~ 80%RH
Operating:(no condensation)(no condensation)
Transport/25%~93% RH25% ~ 93%RH
Storage:(no condensation)(no condensation)
Operating:860hPa~1060hPa860hPa ~1060hPa
Transport/Storage:700hPa~1060hPa700hPa ~1060hPa
Type of protectionagainstelectricshockInternally powered equipmentInternally powered equipment
Thedegreeofprotection againstelectric shockType BType B
ThedegreeofThe main unit : IPX0,The main unit : IPX0,
protection againstharmful ingress ofprobes : IPX8probes : IPX8
liquidThedegreeof
safetyofapplication inthepresenceofaflammable gasEquipment not suitable for use in presenceof flammable gasesEquipment not suitable for use in presenceof flammable gases
The degree of RFGroup 1, Class BGroup 1, Class B
Theelectrical,Complies with the standard:Complies with the standard:
mechanicalandIEC 60601-1:2005IEC 60601-1:2005
thermalsafety
EvaluationIEC 60601-2-37:2007IEC 60601-2-37:2007
EMC EvaluationComplies with the standard:IEC60601-1-2:2007Complies with the standard:IEC60601-1-2:2007
SoftwarelifeComplies with the standard:IECComplies with the standard:IEC
cycle processes62304:200662304:2006
Acoustic outputEvaluationComplies with the standard:Complies with the standard:
IEC 61157:2007IEC 61157:2007
IEC 60601-2-37:2007IEC 60601-2-37:2007
NEMA UD 2-2004NEMA UD 2-2004
BiocompatibilityEvaluationComplies with the standard: ISO 109931-1.Complies with the standard: ISO109931-1.
Clean the equipment case, probe, etc. asabove, and then wipe the probe with anClean the equipment case, probe, etc. asabove, and then wipe the probe with an
Disinfectionalcoholalcohol
impregnated wipe (ethanol 75%,isopropanol alcohol ≤ 70% orglutaraldehyde ≤3.6%).impregnated wipe (ethanol 75%,isopropanol alcohol ≤ 70% orglutaraldehyde ≤3.6%).
Wipe the probe with a clean, dry cloth toremove any remaining moisture.Wipe the probe with a clean, dry cloth toremove any remaining moisture.
Display45mm*25mm LCD display0.96'' OELD double color screen
Battery Supply2*LR6AA3*LR6AA
18650 Lithium Battery
Auto Shut down1 minute after no signal or operation, autoshut down1 minute after no signal or operation, autoshut down
Probe replacement, auto shut down
Resolution1 bpm1 bpm
Accuracy±3bpm±2bpm
Sensitivity10 weeks gestation (3MHz)/105.4dB in 200mm distance9 weeks gestation (3MHz)/130dB in 200mm distance
FHR MeasuringRange50bpm ~ 240bpm50bpm ~ 240bpm
Recording Length240s240s

Page 2 / 5

{5}------------------------------------------------

{6}------------------------------------------------

8. Performance Testing:

Non-clinical test:

The following safety standards are conducted on the subject device:

(1) IEC 60601-1:2005 Safety requirements for medical electrical systems

(2) IEC 60601-1-2:2007 Electromagnetic compatibility requirements and tests

(3) AIUM/NEMA UD 2-2004 Acoustic output measurement standard for diagnostic ultrasound equipment

(4) IEC 60601-2-37:2007 Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment

(5) ISO 10993-1:2009, ISO 10993-5:2009 and ISO 10993-10:2010 Biological evaluation of medical devices

(6) Y Y0449:2009 Ultrasonic Doppler fetal monitor (a Chinese standard)-Performance requirements and methods of measurement and reporting

(7) IEC61266:1994 Ultrasonics Hand-held probe Doppler fetal heartbeat

detectors-Performance requirements and methods of measurement and reporting

Accuracy of FHR for the SD3 devices was tested with the 2 MHz and 3 MHz probes using an FHR simulator for FHR between 50 and 240 BPM. Testing showed the accuracy met the pre-specified criteria of ±2 BPM.

Page 4 / 5

{7}------------------------------------------------

Repeatability of FHR measurements for the SD3 devices was shown by taking multiple measurements and comparing with set values on the FHR simulator. A students T test analysis showed that measurements were repeatable to a 95% confidence.

Clinical test:

Physicians at 6 hospitals used the 2 MHz or the 3 MHz probe depending on the gestational age. There were 162 tests for 11-week to 13-week pregnant women and 99 tests for 9-week to 10(+)-week pregnant women. The test was considered a "pass" if the FHR could be detected and a "fail" if no FHR was detected. A pass rate of 95% was required to support use at a particular gestational age.

Of the 162 cases for the 2 MHz probe from 11-week to 13-week , 157 were considered a "pass." This met the requirements of the test.

Of the 99 cases for the 3 MHz probe for 9-week to 10(+)-week pregnant women, 95 were considered a "pass." This met the requirements of the test.

The subject device passed all testing. The tests were selected to show substantial equivalence between the subject device and the predicate, and since the testing passed, substantial equivalence is shown.

9. Conclusion:

Verification and validation testing was conducted on the SD3 series Ultrasonic Pocket Doppler SD3 series and all testing passed prespectied criteria. This premarket notification submission demonstrates that the Diagnostic Ultrasonic System is substantially equivalent to the predicate devices.

§ 884.2660 Fetal ultrasonic monitor and accessories.

(a)
Identification. A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.(b)
Classification. Class II (performance standards).