K Number
K140546
Date Cleared
2014-06-06

(94 days)

Product Code
Regulation Number
862.3870
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Healgen THC One Step Marijuana Test is an immunochromatographic assay for the qualitative determination of 11-nor-A9-THC-9-COOH in human urine at a Cut-Off concentration of 50 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Healgen mAMP One Step Methamphetamine Test is an immunochromatographic assay for the qualitative determination of methamphetamine in human urine at a Cut-Off concentration of 1000 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Device Description

Immunochromatographic assays for Marijuana and Methamphetamine Urine Tests use a lateral flow, one step system for the qualitative detection of 11-nor-Δ9-THC-9-COOH and Methamphetamine (target analyte) in human urine. Each assay uses a monoclonal antibody-dye conjugate against drugs with gold chloride and fixed drug-protein conjugates and anti-mouse IgG polyclonal antibody in membranes.

AI/ML Overview

The provided document describes the performance characteristics of the Healgen THC One Step Marijuana Test and the Healgen mAMP One Step Methamphetamine Test, and concludes that they are substantially equivalent to a predicate device.

Here's an analysis of the acceptance criteria and the studies that support it:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for these devices are implicitly defined by the results of the performance studies presented, primarily focused on their ability to correctly identify drug presence/absence at and around the specified cutoff concentrations. The studies demonstrate that for samples at or above +25% of the cutoff, the devices consistently show positive results, and for samples at or below -25% of the cutoff, they consistently show negative results. At the cutoff concentration, there is some variability, which is expected for qualitative tests.

Healgen THC One Step Marijuana Test (Cut-off: 50 ng/mL)

Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
PrecisionConsistent results at defined concentrations (100% agreement for -100%/-75%/-50%, +25%/+50%/+75%/+100% cut off)THC Strip Format: -100%, -75%, -50% cut off: 50-/0+ (100% negative)+25%, +50%, +75%, +100% cut off: 50+/0- (100% positive)Cut off: 12-/38+THC Cassette Format: Similar results (e.g., 18-/32+ at cut off) THC Cup Format: Similar results (e.g., 14-/36+ at cut off) THC Dipcard Format: Similar results (e.g., 20-/30+ at cut off)
Cut-off VerificationAll positive at and above +25% Cut-off; all negative at and below -25% Cut-off.THC: All positive at and above +25% Cut-off; all negative at and below -25% Cut-off.
InterferenceNo interference from common physiological/pathological substances at specified concentrations.No interference from a long list of compounds at 100 µg/mL.
SpecificityPredictable cross-reactivity with related compounds.THC: 11-nor-Δ9-THC-9-COOH (100% at 50 ng/mL), Delta-9-THC (0.1% at 50,000 ng/mL), 11-nor-delta-9-THC-carboxyglucuronide (67% at 75 ng/mL), 11-Hydroxy-Δ9-THC (1% at 5,000 ng/mL), etc.
Effect of Urine Specific Gravity and pHNo impact on results at defined range.All positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off within pH 4-9 and specific gravity 1.000-1.035.
Method Comparison (Clinical Samples)Agreement with GC/MS (concordance)THC Strip, Cassette, Cup, Dip Card: High concordance for "Negative" and "High Positive" samples. Some discordant results (false negatives) for samples "Near Cutoff Positive" (52-53 ng/mL, i.e., slightly above the 50 ng/mL cutoff).
Lay User StudyHigh percentage of correct results, especially for samples further from the cutoff.THC: 100% correct for -100%, -75%, -50%, +50%, +75% Cutoff samples. 95% correct for -25% Cutoff (1 false positive) and +25% Cutoff (1 false negative).

Healgen mAMP One Step Methamphetamine Test (Cut-off: 1000 ng/mL)

Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
PrecisionConsistent results at defined concentrations (100% agreement for -100%/-75%/-50%, +25%/+50%/+75%/+100% cut off)MET Strip Format: -100%, -75%, -50% cut off: 50-/0+ (100% negative)+25%, +50%, +75%, +100% cut off: 50+/0- (100% positive)Cut off: 14-/36+MET Cassette, Dip Card, Cup Formats: Similar results (e.g., 20-/30+ at cut off, 24-/26+ at cut off, 22-/28+ at cut off respectively)
Cut-off VerificationAll positive at and above +25% Cut-off; all negative at and below -25% Cut-off.MET: All positive at and above +25% Cut-off; all negative at and below -25% Cut-off.
InterferenceNo interference from common physiological/pathological substances at specified concentrations.No interference from a long list of compounds at 100 µg/mL.
SpecificityPredictable cross-reactivity with related compounds.MET: D-Methamphetamine (100% at 1000 ng/mL), MDEA (5% at 20,000 ng/mL), Procaine (1.7% at 60,000 ng/mL), MDMA (40% at 2500 ng/mL), Ephedrine (1% at 100,000 ng/mL), etc.
Effect of Urine Specific Gravity and pHNo impact on results at defined range.All positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off within pH 4-9 and specific gravity 1.000-1.035.
Method Comparison (Clinical Samples)Agreement with GC/MS (concordance)MET Strip, Cassette, Dip Card, Cup: High concordance for "Negative" and "High Positive" samples. Some discordant results (false negatives) for samples "Near Cutoff Positive" (1003-1008 ng/mL, i.e., slightly above the 1000 ng/mL cutoff).
Lay User StudyHigh percentage of correct results, especially for samples further from the cutoff.MET: 100% correct for -100%, -75%, -50%, +50%, +75% Cutoff samples. 90% correct for -25% Cutoff (2 false positives) and 95% correct for +25% Cutoff (1 false negative).

2. Sample Sizes and Data Provenance

  • Precision Study: For each drug (THC and MET) and each format (Strip, Cassette, Cup, Dip Card), 8 concentrations (-100% cut off to +100% cut off) were tested. For each concentration, tests were performed two runs per day for 25 days. This implies 50 individual tests per concentration per lot, and 3 lots were used for each format.
    • Example for THC Strip: 8 concentrations x 50 tests/concentration x 3 lots = 1200 tests.
    • Data Provenance: The document does not explicitly state the country of origin but implies an in-house laboratory setting (referred to as "in-house" for comparison studies, and "prepared by spiking drug in negative samples" for precision, cut-off, and interference studies). The data is retrospective in the sense that samples were prepared and then tested. The urine samples were "negative samples" (for spiking) or "drug-free urine" (for interference), suggesting they were sourced specifically for testing purposes.
  • Cut-off Verification Study: 150 samples were used, equally distributed at concentrations of -50%, -25%, Cut-Off, +25%, +50% Cut-Off. So, 30 samples per concentration.
    • Data Provenance: Similar to precision, prepared by spiking.
  • Interference Study: Urine samples (drug-free and spiked with target drugs +/-25% Cut-Off) were used for testing various interfering substances. The number of samples per substance is not specified, but it was tested using three batches of each device for all formats.
    • Data Provenance: Assumed to be artificially created/spiked samples.
  • Method Comparison Study: For each drug and each format, 80 unaltered clinical urine samples were used (40 negative, 40 positive).
    • Data Provenance: The samples are described as "clinical samples," suggesting real-world patient samples. The country of origin is not specified, but the study was performed "in-house." This is retrospective data.
  • Lay-user Study: 140 lay persons participated. Urine samples were prepared at 7 different concentrations (negative, +/-75%, +/-50%, +25% of cutoff). For each concentration, 20 samples were prepared.
    • Data Provenance: Artificially prepared by spiking into drug-free pooled urine specimens.

3. Number of Experts and Qualifications for Ground Truth

  • Precision, Cut-off, Interference, Specificity, Effect of Urine Specific Gravity and pH:
    • The ground truth for these analytical performance studies was established by GC/MS (Gas Chromatography/Mass Spectrometry). This is a highly accurate and widely accepted gold standard method for quantifying drug concentrations in urine.
    • The document states that "Each drug concentration was confirmed by GC/MS" and that the "concentrations of the samples were confirmed by GC/MS." No human experts are explicitly mentioned for establishing ground truth from GC/MS results, as GC/MS is an objective analytical method.
  • Method Comparison Study:
    • The ground truth for the 80 clinical samples was established by GC/MS results.
    • No specific number or qualifications of human experts are mentioned for interpreting the GC/MS results or establishing the ground truth from them.
  • Lay-user Study:
    • Ground truth was established by GC/MS for the spiked urine samples.

4. Adjudication Method for the Test Set

  • Precision, Cut-off, Interference, Specificity, Effect of Urine Specific Gravity and pH: For these analytical studies, GC/MS was the objective reference. The testing was done by unspecified personnel; "sample aliquots were blinded labeled by the person who prepared the samples and won't take part in the sample testing" (precision study). This suggests a blinding mechanism but no explicit multi-reader adjudication process for the actual device results.
  • Method Comparison Study: Three laboratory assistants were the "operators" who ran the devices and presumably interpreted the results. The comparison was against GC/MS. The individual results of each viewer are reported, and then discordant results are listed. There is no explicit mention of an adjudication method (e.g., 2+1, 3+1 consensus among the three viewers) to arrive at a single device result per sample if their interpretations differed. Each viewer's interpretation is compared independently to the GC/MS ground truth.
  • Lay-user Study: The study involved "140 lay persons" each testing "1 blind labeled sample and a device." The "Lay person results" table shows aggregate counts of positive/negative results. It does not describe an adjudication process for discordant interpretations among lay users on a single sample, as each lay user tested a unique sample (or set of samples for the overall study) and their individual interpretation was the data point.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was described. The studies focused on the performance of the device itself, observed by laboratory assistants (method comparison) or lay users (lay-user study), against an objective ground truth (GC/MS). There is no comparison of "human readers improve with AI vs without AI assistance" because the device is a simple, qualitative immunochromatographic assay, not an AI-powered diagnostic system requiring human interpretation of complex AI outputs.

6. Standalone Performance Study (Algorithm Only)

The device itself is a standalone, qualitative diagnostic test (immunochromatographic assay). All the performance studies (precision, cut-off, interference, specificity, method comparison, lay-user study) assess the standalone performance of the device without human interpretation being the primary variable. The "human-in-the-loop" component is essentially the reading and interpretation of the colored lines on the test strip as positive or negative, which is intrinsic to this type of device. There isn't an "algorithm" in the sense of a sophisticated computational model that could be evaluated separately from the physical test.

7. Type of Ground Truth Used

The primary and consistently used ground truth for all analytical and comparative studies was GC/MS (Gas Chromatography/Mass Spectrometry). This is an objective, highly accurate analytical method for drug concentration determination.

8. Sample Size for the Training Set

The document does not describe the development of an "algorithm" or "AI" system that would typically require a training set. This is an immunochromatographic assay, which relies on chemical and biological reactions rather than machine learning algorithms. Therefore, there is no "training set" in the context of AI. The product validation data described (precision, cut-off, interference, specificity, method comparison) serves to demonstrate the device's performance, not to "train" it.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" in the context of AI/algorithm development for this device, this question is not applicable. The device's inherent design and manufacturing processes are validated by the performance studies using GC/MS as the ground truth.

{0}------------------------------------------------

.K140546

JUN 0 6 2014

510(k) SUMMARY

    1. Date:
      May 29, 2014

Submitter: 2.

HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX77401

    1. Contact person: Jianqiu Fang HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX77401 Telephone: 713-733-8088 Fax: 713-733-8088 Email: bryan@healgen.com
    1. Device Name:

Healgen mAMP One Step Methamphetamine Test

Healgen THC One Step Marijuana Test

Classification:

Product CodeCFR #Panel
LDJ21 CFR, 862.3870 Cannabinoid Test SystemToxicology
LAF21 CFR, 862.3610 Methamphetamine Test SystemToxicology

FIRST CHECK DIAGNOSTICS LLC

6. Intended Use

Healgen THC One Step Marijuana Test is an immunochromatographic assay for the qualitative determination of 11-nor-A9-THC-9-COOH in human urine at a Cut-Off concentration of 50 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Healgen mAMP One Step Methamphetamine Test is an immunochromatographic assay for the qualitative determination of methamphetamine in human urine at a Cut-Off concentration of 1000 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred

{1}------------------------------------------------

confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

    1. Device Description
      Immunochromatographic assays for Marijuana and Methamphetamine Urine Tests use a lateral flow, one step system for the qualitative detection of 11-nor-Δ9-THC-9-COOH and Methamphetamine (target analyte) in human urine. Each assay uses a monoclonal antibody-dye conjugate against drugs with gold chloride and fixed drug-protein conjugates and anti-mouse IgG polyclonal antibody in membranes.

8. Substantial Equivalence Information

A summary comparison of features of the Healgen THC One Step Marijuana Test and Healgen mAMP One Step Methamphetamine Test and the predicate devices is provided in Table 1 & Table 2.

ItemDevicePredicate -K052115
Indication(s) forUseFor the qualitative determination of11-nor-Δ9-THC-9-COOH in human urine.Same (but thenumber of drugsdetected is different)
Calibrator11-nor-Δ9-THC-9-COOHSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitative to indicate positive or negativeresultSame
Specimen TypeHuman UrineSame
Cut-Off Values50 ng/mLSame
Intended UseFor over-the-counter and prescription uses.For over-the-counteruse.
ConfigurationsStrip, Cassette, Cup, Dip CardCup

Table 1: Features Comparison of Healgen THC One Step Marijuana Test and the Predicate Devices

Table 2: Features Comparison of Healgen mAMP One Step Methamphetamine Test and the Predicate Devices

ltemDevicePredicate - K052115
Indication(s) forUseFor the qualitative determination ofMethamphetamine in human urine.Same(but the numberof drugs detected isdifferent)
CalibratorMethamphetamineSame

{2}------------------------------------------------

ItemDevicePredicate - K052115
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Type of TestQualitative to indicate positive or negativeresultSame
Specimen TypeHuman UrineSame
Cut-Off Values1000 ng/mLSame
Intended UseFor over-the-counter and prescription uses.For over-the-counteruse.
ConfigurationsStrip, Cassette, Cup, Dip CardCup
    1. Test Principle
      It is a rapid test for the qualitative detection of 11-nor-A9-THC-9-COOH and Methamphetamine in urine samples. It is a lateral flow chromatographic immunoassay. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen below its cut-off concentration, it will not saturate the binding sites of its specific antibody coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off , +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blinded labeled by the person who prepared the samples and won't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days. The results obtained are summarized in the following table.

DrugResult-100%cutoff-75%cutoff-50%cutoff-25%cutoffcut off+25%cutoff+50%cutoff+75%cut+100%cutt off
Lot:THC111000150-/0+50-/0+50-/0+50-/0+12-/38+50+/0-50+/0-50+/0-50+/0-
Lot:THC111000250-/0+50-/0+50-/0+50-/0+12-/38+50+/0-50+/0-50+/0-50+/0-
Lot:THC111000350-/0+50-/0+50-/0+50-/0+12-/38+50+/0-50+/0-50+/0-50+/0-
Result-100% cutoff-75%cutoff-50%cutoff-25%cutoffcut off+25%cutoff+50%cutoff+75%cutoff+100%cutt off
DrugLot:THC111000450-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-
Lot:THC111000550-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-
Lot:THC111000650-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-
THC Cup FormatResult-100%cutoff-75%cutoff-50%cutoff-25%cutoffcut off+25%cutoff+50%cutoff+75%cutoff+100%cutt off
DrugLot:THC111000750-/0+50-/0+50-/0+50-/0+14-/36+50+/0-50+/0-50+/0-50+/0-
Lot:THC111000850-/0+50-/0+50-/0+50-/0+14-/36+50+/0-50+/0-50+/0-50+/0-
Lot:THC111000950-/0+50-/0+50-/0+50-/0+14-/36+50+/0-50+/0-50+/0-50+/0-
THC Dipcard FormatResult-100% cutoff-75%cutoff-50%cutoff-25%cutoffcut off+25%cutoff+50%cutoff+75%cutoff+100%cutt off
DrugLot:THC111001050-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot:THC111001150-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot:THC111001250-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
MET Strip FormatResult-100% cutoff-75%cutoff-50%cutoff-25%cutoffcut off+25%cutoff+50%cutoff+75%cutoff+100%cutt off
DrugLot:MET111000150-/0+50-/0+50-/0+50-/0+14-/36+50+/0-50+/0-50+/0-50+/0-
Lot:MET111000250-/0+50-/0+50-/0+50-/0+14-/36+50+/0-50+/0-50+/0-50+/0-
Lot:MET111000350-/0+50-/0+50-/0+50-/0+14-/36+50+/0-50+/0-50+/0-50+/0-
MET Cassette FormatResult-100% cutoff-75%cutoff-50% cutoff-25%cutoffcut off+25%cutoff+50%cutoff+75%cutoff+100%cutt off
DrugLot:MET111000450-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot:MET111000550-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot:MET111000650-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
MET Dip Card FormatResult-100% cutoff-75%cutoff-50%cutoff-25%cutoffcut off+25%cutoff+50%cutoff+75%cutoff+100%cutt off
DrugLot:MET111000750-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot:MET111000850-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot:MET111000950-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
MET CUP FormatResult-100% cutoff-75%cutoff-50% cutoff-25%cutoffcut off+25%cutoff+50%cutoff+75%cutoff+100%cutt off
DrugLot:MET111001050-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot:MET111001150-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

THC Strip Format

THC Cassette Format

{3}------------------------------------------------

:

රි

{4}------------------------------------------------

b.Linearity

Not applicable, this is a visually read device

  • c. Stability
    It is stable at 4-30 ℃ for 24 months based on the accelerated stability study at 45 ℃ and real time stability determination at both 4 ℃ and 30 ℃.

d.Cut-off

Total 150 samples equally distributed at concentrations of -50% Cut-Off; -25% Cut-Off; Cut-Off; +25% Cut-Off; +50% Cut-Off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% Cut-off and all negative at and below -25% Cut-off for both Marijuana and Methamphetamine. The following cut-off values for the test devices have been verified.

TestCalibratorCut-off (ng/mL)
One Step Marijuana Test11-nor-Δ9-THC-9-COOH50
One Step Methamphetamine TestMethamphetamine1000

e. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentration at 25% below and 25% above Cut-Off level respectively. These urine samples were tested using three batches of each device for all formats.

Compounds that showed no interference at a concentration of 100ug/mL are summarized in the following tables. There were no differences observed for different formats.

4-AcetamidophenolEstrone-3-sulfatePenicillin-G
AcetophenetidinEthyl-p-aminobenzoatePentazocine
N-AcetylprocainamideFenoprofenPentobarbital
Acetylsalicylic acidFurosemidePerphenazine
AminopyrineGentisic acidPhencyclidine
AmitryptylineHemoglobinPhenelzine
AmobarbitalHydralazinePhenobarbital
AmoxicillinHydrochlorothiazidePhentermine
AmpicillinHydrocodoneL-Phenylephrine
Ascorbic acidHydrocortisoneß-Phenylethlamine
D,L-AmphetamineO-Hydroxyhippuric acidß-Phenyllethylamine
L-Amphetamine3-HydroxytyraminePhenylpropanolamine
ApomorphineIbuprofenPrednisolone
AspartameImipraminePrednisone
AtropineIproniazidProcaine
Benzilic acid(-) IsoproterenolPromazine
Benzoic acidIsoxsuprinePromethazine
BenzoylecgonineKetamineD,L-Propanolol
BenzphetamineKetoprofenD-Propoxyphene
BilirubinLabetalolD-Pseudoephedrine

THC:

{5}------------------------------------------------

BrompheniramineLevorphanolQuinidine
CaffeineLoperamideQuinine
Chloralhydrate.MaprotilineRanitidine
ChloramphenicolMeprobamateSalicylic acid
ChlordiazepoxideMethadoneSecobarbital
ChlorothiazideMethoxyphenamineSerotonin (5-Hydroxytyramine)
(±) Chlorpheniramine(+) 3,4-MethylenedioxyamphetamineSulfamethazine
Chlorpromazine(+)3,4-MethylenedioxymethamphetamineSulindac
ChlorquineMethylphenidateTemazepam
CholesterolMethyprylonTetracycline
ClomipramineMorphine-3-β-DglucuronideTetrahydrocortisone, 3 Acetate
ClonidineNalorphineTetrahydrocortisone3(5-Dglucuronide)
Cocaine hydrochlorideNaloxoneTetrahydrozoline
CodeineNalidixic acidThebaine
CortisoneNaltrexoneThiamine
(-) CotinineNaproxenThioridazine
CreatinineNiacinamideD, L-Thyroxine
DeoxycorticosteroneNifedipine ·Tolbutamine
DextromethorphanNorcodeinTriamterene
DiazepamNorethindroneTrifluoperazine
DiclofenacD-NorpropoxypheneTrimethoprim
DiflunisalNoscapineTrimipramine
DigoxinD, L-OctopamineTryptamine
DiphenhydramineOxalic acidD, L-Tryptophan
DoxylamineOxazepamTyramine
Ecgonine hydrochlorideOxolinic acidPrD, L-Tyrosine
Ecgonine methylesterOxycodoneUric acid
(-) Y EphedrineOxymetazolineVerapamil
Erythromycinp-HydroxymethamphetamineZomepirac
β-EstradiolPapaverine

МЕТ

:

4-Acetamidophenol(-) Y EphedrinePenicillin-G
AcetophenetidinErythromycinPentazocaine
N-Acetylprocainamideß-EstradiolPentobarbital
Acetylsalicylic acidEstrone-3-sulfatePerphenazine
AminopyrineEthyl-p-aminobenzoatePhencyclidine
AmitryptylineFenfluraminePhenelzine
AmobarbitalFenoprofenPhendimetrazine
AmoxicillinFurosemidePhenobarbital
AmpicillinGentisic acidPhetoin
Ascorbic acidHemoglobinL-Phenylephrine

{6}------------------------------------------------

ApomorphineHydralazineβ-Phenylethlamine
AspartameHydrochlorothiazidePhenylpropanolamine
AtropineHydrocodonePrednisolone
Benzilic acidHydrocortisonePrednisone
Benzoic acidO-Hydroxyhippuric acidProcaine
Benzoylecgonine3-HydroxytyraminePromazine
BilirubinIbuprofenPromethazine
BrompheniramineImipramineD,L-Propanolol
Caffeine(-) IsoproterenolPropiomazine
CannabidiolIsoxsuprineD-Propoxyphene
CannabinolKetamineQuinidine
ChloralhydrateKetoprofenQuinine
ChloramphenicolLabetalolRanitidine
ChlordiazepoxideLevorphanolSalicylic acid
ChlorothiazideLoperamideSecobarbital
(±) ChlorpheniramineMaprotilineSerotonin
ChlorpromazineMeperidineSulfamethazine
ChlorquineMeprobamateSulindac
CholesterolMethadoneTemazepam
ClomipramineMethylphenidateTetracycline
ClonidineMorphine-3-DglucuronideTetrahydrocortisone
Cocaine hydrochlorideNalidixic acidTetrahydrozoline
CodeineNaloxoneΔ9-THC-COOH
CortisoneNaltrexoneThebaine
(-) CotinineNaproxenThiamine
CreatinineNiacinamideThioridazine
DeoxycorticosteroneNifedipineD,L-Thyroxine
DextromethorphanNorcodeinTolbutamine
DiazepamNorethindroneTriamterene
DiclofenacD-NorpropoxypheneTrifluoperazine
DiflunisalNoscapineTrimethoprim
DigoxinD,L-OctopamineTrimipramine
DiphenhydramineOxalic acidTryptamine
DoxylamineOxazepamD, L-Tyrosine
Ecgonine hydrochlorideOxolinic acidUric acid
Ecgonine methylesterOxycodoneVerapamil
(IR,2S)-(-)-EphedrineOxymetazolineZomepirac
l-EphedrinePapaverineAcebutolol

f. Specificity

To test the specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device for all formats. Compounds that produced positive results are listed below. There were no differences observed for different formats.

{7}------------------------------------------------

THCResult% Cross-Reactivity
(11-nor-Δ9-THC-9-COOH, Cutoff=50 ng/mL)Positive at 50 ng/mL100%
Delta-9-TetrahydrocannabinolPositive at 50,000 ng/mL0.1%
11-nor-delta-9-THC-carboxyglucuronidePositive at 75 ng/mL67%
(-)-11-nor-9-carboxy-delta9-THCPositive at 75 ng/mL67%
11-Nor-Δ9-TetrahydrocannabinolPositive at 50 ng/mL100%
11-Hydroxy-Δ9-TetrahydrocannabinolPositive at 5,000 ng/mL1%
11-Nor-Δ8-TetrahydrocannabinolPositive at 50 ng/mL100%
Δ8-THC-COOHPositive at 50,000 ng/mL0.1%
METResult% Cross-Reactivity
(D-Methamphetamine, Cutoff=1000 ng/mL)Positive at 1000 ng/mL100%
(+/-) 3,4-Methylenedioxy-n-ethylamphetamine(MDEA)Positive at 20,000 ng/mL5%
Procaine (Novocaine)Positive at 60,000 ng/mL1.7%
TrimethobenzamidePositive at 20,000 ng/mL5%
MethamphetaminePositive at 1000 ng/mL100%
Ranitidine (Zantac)Positive at 50,000 ng/mL2%
(+/-) 3,4-Methylenedioxymethamphetamine (MDMA)Positive at 2500 ng/mL40%
ChloroquinePositive at 50,000 ng/mL2%
EphedrinePositive at 100,000 ng/mL1%
FenfluraminePositive at 50,000 ng/mL2%
p-HydroxymethamphetaminePositive at 10,000 ng/mL10%

g. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, the urine samples, with 1.0001.035 specific gravity or urine samples with pH 49 were spiked with target drugs at 25% below and 25% above Cut-Off level, respectively. These samples were tested using three batches of each device for all formats. Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different formats.

    1. Comparison Studies
      The method comparison for the One Step Marijuana Test, and the One Step Methamphetamine Test was performed in-house with three laboratory assistants for each format of the device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results. The results are presented in the table below:
StripformatNegativeTHC
LowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Betweenthe cutoffand +50%)HighPositive byGC/MS(greaterthan +50%)

{8}------------------------------------------------

Viewer0001326
Viewer APositive0001326
Negative10161610
Viewer BPositive0001226
Negative10161620
Viewer CPositive0001226
Negative10161620

Discordant Results of THC Strip

ViewerSample NumberGC/MS ResultStrip Format Viewer Results
Viewer A22352Negative
Viewer B20753Negative
Viewer B22352Negative
Viewer C20753Negative
Viewer C22352Negative
CassetteformatNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Betweenthe cutoffand +50%)HighPositive byGC/MS(greaterthan +50%)
ViewerAPositive0001226
Negative10161620
ViewerBPositive0001126
Negative10161630
ViewerCPositive0001126
Negative10161630

Discordant Results of THC Casette

ViewerSample NumberGC/MS ResultStrip Format Viewer Results
Viewer A22352Negative
Viewer A24753Negative
Viewer B20753Negative
Viewer B24753Negative
Viewer B22352Negative
Viewer C20753Negative
Viewer C24753Negative
Viewer C22352Negative
CupformatNegativeLowNegative byGC/MS(less than -50%)Near CutoffNegative byGC/MS(Between -50% andNear CutoffPositive byGC/MS(Between the cutoffHighPositive byGC/MS(greater than +50%)

·

{9}------------------------------------------------

cutoff)and +50%)
ViewerPositive0001226
ANegative10161620
ViewerPositive0001126
BNegative10161630
ViewerPositive0001226
CNegative10161620

Discordant Results of THC Cup

ViewerSample NumberGC/MS ResultStrip Format Viewer Results
Viewer A22352Negative
Viewer A24753Negative
Viewer B20753Negative
Viewer B24753Negative
Viewer B22352Negative
Viewer C20753Negative
Viewer C24753Negative
DipCardformatNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Betweenthe cutoffand +50%)HighPositive byGC/MS(greaterthan +50%)
ViewerAPositive0001326
ViewerANegative10161610
ViewerBPositive0001226
ViewerBNegative10161620
ViewerCPositive0001226
ViewerCNegative10161620

Discordant Results of THC Dip Card

ViewerSample NumberGC/MS ResultStrip Format Viewer Results
Viewer A22352Negative
Viewer B20753Negative
Viewer B22352Negative
Viewer C20753Negative
Viewer C24753Negative

{10}------------------------------------------------

StripformatNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Betweenthe cutoffand +50%)HighPositive byGC/MS(greaterthan +50%)
ViewerPositive0001227
ANegative10191510
ViewerPositive0001127
BNegative10191520
ViewerPositive0001227
CNegative10191510

:

Discordant Results of MET Strip

ViewerSample NumberGC/MS ResultStrip Format Viewer Results
Viewer A1331008Negative
Viewer B1231003Negative
Viewer B1331008Negative
Viewer C1331008Negative
CassetteformatNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Betweenthe cutoffand +50%)HighPositive byGC/MS(greaterthan +50%)
ViewerAPositive0001127
Negative10191520
ViewerBPositive0001227
Negative10191510
ViewerCPositive0001127
Negative10191520

Discordant Results of MET Cassette

:

ViewerSample NumberGC/MS ResultStrip Format ViewerResults
Viewer A1231003Negative
Viewer A1331008Negative
Viewer B1231003Negative
Viewer C1231003Negative
Viewer C1331008Negative

.

:

{11}------------------------------------------------

DipCardformatNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Betweenthe cutoffand +50%)HighPositive byGC/MS(greaterthan +50%)
ViewerAPositive0001127
Negative10191520
ViewerBPositive0001127
Negative10191520
ViewerCPositive0001127
Negative10191520

Discordant Results of MET Dip Card

ViewerSample NumberGC/MS ResultStrip Format ViewerResults
Viewer A1231003Negative
Viewer A1331008Negative
Viewer B1231003Negative
Viewer B1331008Negative
Viewer C1231003Negative
Viewer C1331008Negative
Cup formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cutoff)Near Cutoff Positive by GC/MS (Between the cutoff and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001227
Viewer ANegative10191510
Viewer BPositive0001227
Viewer BNegative10191510
Viewer CPositive0001127
Viewer CNegative10191520

Discordant Results of MET Cup

ViewerSample NumberGC/MS ResultStrip Format ViewerResults
Viewer A1231003Negative
Viewer B1231003Negative
Viewer C1231003Negative
Viewer C1331008Negative

Lay-user study

A lay user study was performed at three intended user sites with 140 lay persons. For a typical Strip device study, participants were 54 females and 86 males tested the Marijuana samples, and

{12}------------------------------------------------

54 females and 86 males tested the Methamphetamine samples. They had diverse educational and professional backgrounds and ranged in age from 21 to >50. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled samples and a device. The typical results are summarized below. There are no statistic differences of layperson results between different formats of the devices.

% of CutoffNumber of samplesTHC Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100%Cutoff200No. of PositiveNo. of Negative100%
-75%Cutoff2012.5020100%
-50% Cutoff2025020100%
-25% Cutoff2037.511995%
+25% Cutoff2062.519195%
+50% Cutoff2075200100%
+75% Cutoff2087.5200100%

Typical Comparison between GC/MS and Lay Person Results (THC)

Typical Comparison between GC/MS and Lay Person Results (MET)

% of CutoffNumber of samplesMET Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100%Cutoff200020100%
-75%Cutoff20250020100%
-50% Cutoff20500020100%
-25% Cutoff2075021890%
+25% Cutoff20125019195%
+50% Cutoff201500200100%
+75% Cutoff201750200100%
    1. Clinical Studies
      Not applicable
    1. Conclusion
      Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity and method comparison of the devices, it's concluded that the Healgen THC One Step Marijuana Test, and Healgen mAMP One Step Methamphetamine Test are substantially equivalent to the predicate.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features the HHS symbol, which is a stylized representation of a human figure embracing the world. The symbol is composed of three curved lines that form the shape of a person with outstretched arms. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 6, 2014

HEALGEN SCIENTIFIC, LLC C/O JOE SHIA 504 EAST DIAMOND AVE. SUITE F GAITHERSBURG MD 20877

Rc: K140546

Trade/Device Name: Healgen THC One Step Marijuana Test Healgen mAMP One Step Methamphetamine Test Regulation Number: 21 CFR 862.3870 Regulation Name: Cannabinoid test system Regulatory Class: II Product Code: LDJ, LAF Dated: April 17, 2014 Received: April 18, 2014

Dear Mr. Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for the stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and conment of control the device, subject to the general controls provisions of the Act. The I ou may on on on of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

1f your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Paris 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vatual the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{14}------------------------------------------------

Page 2-Mr. Joe Shia

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{15}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510{k} Number (if known) K140546

Device Name

Healgen THC One Step Marijuana Test Healgen mAMP One Step Methamphetamine Test

Indications for Use (Describe)

Healgen THC One Step Marijuana Test is an immunochromatographic assay for the qualitative determination of II-nor-49-THC-9-COOH in human urine at a Cut-Off concentration of 50 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Healgen mAMP One Step Methamphetamine Test is an immunochromatographic assay for the qualitative determination of methamphetamine in human urine at a Cut-Off concentration of 1000 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

EF

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{16}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

ﺎ ﺍﻟﻌﺎﻡ ﺍﻟﻤﺴﺘﻘﺒﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤ

§ 862.3870 Cannabinoid test system.

(a)
Identification. A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds includedelta -9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.(b)
Classification. Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).