K Number
K140517
Date Cleared
2014-04-28

(59 days)

Product Code
Regulation Number
878.4010
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SURGISEAL Topical Skin Adhesive is intended for topical applications only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, simple, thoroughly cleansed, trauma induced lacerations.

SURGISEAL may be used in conjunction with, but not in place of, deep dermal sutures.

Device Description

SURGISEAL® Topical Skin Adhesive is a sterile, professional liquid skin adhesive containing a monomeric (2-octyl cyanoacrylate) formulation and the colorant D&C Violet #2. It incorporates the Teardrop applicator that consists of a thermoformed blister tray with a heat sealed lid with an attached applicator sponge tip. The Stylus applicator incorporates the identical adhesive formula but incorporates the Stylus applicator that consists of a plastic ampoule, which houses the adhesive, contained with the longer plastic sleeve within a longer plastic sleeve with an attached applicator tip. When SURGISEAL is applied to the skin with either Applicator, it polymerizes in minutes.

In vitro studies have shown that SURGISEAL acts as a physical barrier to microbial penetration as long as the adhesive film remains intact. Clinical studies were not conducted to demonstrate microbial barrier properties and a correlation between microbial barrier properties and a reduction in infection have not been established.

AI/ML Overview

The provided document describes a 510(k) submission for SURGISEAL® Topical Skin Adhesive and SURGISEAL Stylus™ Topical Skin Adhesive. The submission is for a labeling change to include a "Microbial Barrier description." The document explicitly states: "In vitro studies have shown that SURGISEAL acts as a physical barrier to microbial penetration as long as the adhesive film remains intact. Clinical studies were not conducted to demonstrate microbial barrier properties and a correlation between microbial barrier properties and a reduction in infection have not been established." Furthermore, it clarifies that the 510(k) submission is for a labeling change to characterize the ease, safety, or effective use of the product based on extensive bench performance testing, which has demonstrated to be safe & efficacious.

Therefore, the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of clinical studies, ground truth, sample sizes, and expert adjudication cannot be extracted accurately for the microbial barrier claim, as no clinical studies were performed. The "Performance Testing" section mentions "Microbial Barrier Tests and Film Thickness determinations for each of the two applicators" as bench tests.

However, based on the information provided about the bench performance testing for the microbial barrier property, here's what can be inferred and stated:

1. Table of acceptance criteria and the reported device performance:

Acceptance Criteria (Bench Test for Microbial Barrier)Reported Device Performance (Bench Test for Microbial Barrier)
Not explicitly stated in the document, but implicitly tied to the general intent of demonstrating a "physical barrier to microbial penetration.""In vitro studies have shown that SURGISEAL acts as a physical barrier to microbial penetration as long as the adhesive film remains intact."

2. Sample size used for the test set and the data provenance:

  • Sample size: Not specified in the provided text for the in vitro microbial barrier studies.
  • Data provenance: "In vitro studies," meaning laboratory-based studies. The location/country of the lab is not specified, but it's a retrospective assessment of prior in vitro work.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. The ground truth for in vitro microbial barrier studies would typically be established through standardized microbiological testing protocols, not human expert consensus.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable as it was an in vitro laboratory test, not a subjective assessment requiring human adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. The submission is for a topical skin adhesive, not an AI-assisted diagnostic or interpretive device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • No, not applicable. This is for a medical device (topical skin adhesive), not an algorithm.

7. The type of ground truth used:

  • For the microbial barrier claim: The ground truth was based on the results of in vitro microbiological testing demonstrating the physical barrier property.

8. The sample size for the training set:

  • Not applicable. This is not a machine learning or algorithm-based device requiring a training set in that context. The "training" here refers to the development and testing of the physical adhesive product.

9. How the ground truth for the training set was established:

  • Not applicable in the context of a "training set" for an algorithm. The development of the adhesive product was likely based on various bench tests and physicochemical characterizations to achieve desired properties, which would serve as "ground truth" for product development.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the date "APR 28 2014" in a bold, sans-serif font. The letters and numbers are uniformly sized and spaced. The date is likely printed or stamped onto a document or surface.

Image /page/0/Picture/2 description: The image shows the logo for Adhezion Biomedical. The logo consists of a stylized graphic to the left of the company name. The graphic is made up of curved lines that form a circular shape. The company name is written in a sans-serif font, with the word "Adhezion" in a larger font size than the word "Biomedical" which is written below it.

510(k) Summary

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92 is below:

Submitter Information:
NameAdhezion Biomedical, LLC
AddressOne Meridian BoulevardSuite 1B02Wyomissing, PA 19610
Phone Number(610) 241-7191
Fax Number(610) 373-2081
EstablishmentRegistration3006385287
Name of contactpersonRichard Jones, Regulatory and Quality Assurance Consultant
Date preparedFebruary 26, 2014
Name of Device(s):
Trade or proprietarynamesSURGISEAL® Topical Skin AdhesiveSURGISEAL Stylus™ Topical Skin Adhesive
Common or usualnameDevice, Tissue Adhesive for Topical Approximation
Classification nameClass II
Classification PanelGeneral and Plastic Surgery Devices
RegulationClass II, under 21 CFR 878.4010
Product Code(s)MPN
Legally Marketeddevice(s) to whichequivalence isclaimedSURGISEAL® Topical Skin Adhesive (K082993, K123936, K130329)SURGISEAL Stylus™ Topical Skin Adhesive (K130474)Dermabond® Topical Skin Adhesive (P960052/S003)Indermil™ Tissue Adhesive (P010002/S004)LiquiBand® Flow Control (K122446)
Reason for 510(k)submissionLabeling Change to include Microbial Barrier description
Device DescriptionSURGISEAL® Topical Skin Adhesive is a sterile, professional liquidskin adhesive containing a monomeric (2-octyl cyanoacrylate)formulation and the colorant D&C Violet #2. It incorporates theTeardrop applicator that consists of a thermoformed blister tray with aheat sealed lid with an attached applicator sponge tip. The Stylusapplicator incorporates the identical adhesive formula but incorporates
the Stylus applicator that consists of a plastic ampoule, which housesthe adhesive, contained with the longer plastic sleeve within a longerplastic sleeve with an attached applicator tip. When SURGISEAL isapplied to the skin with either Applicator, it polymerizes in minutes.
In vitro studies have shown that SURGISEAL acts as a physical barrierto microbial penetration as long as the adhesive film remains intact.Clinical studies were not conducted to demonstrate microbial barrierproperties and a correlation between microbial barrier properties and areduction in infection have not been established.
Indications for useSURGISEAL is intended for topical applications only to hold closedeasily approximated skin edges of wounds from surgical incisions,including punctures from minimally invasive surgery, simple,thoroughly cleansed, trauma induced lacerations.SURGISEAL may be used in conjunction with, but not in place of, deep
dermal sutures.
TechnologicalCharacteristicsThe technological characteristics of both SURGISEAL Topical SkinAdhesive (K082993, K123936 & K130329) and SURGISEALSTYLUS Skin Adhesive (K130474) are identical. SURGISEALadhesive consists of a monomeric (2-octyl cyanoacrylate) liquidadhesive formulation packaged in a single-use applicator. The device isa low viscosity formulation to allow for varied layered applications ofthe adhesive to the intended area and allow for either a single consistentintact film thickness, continuous layer or two thin layers of the adhesiveto the wound area.
SURGISEAL is used for topical applications only to hold closed, easilyapproximated skin edges of wounds while maintaining woundapproximation.
In vitro studies have shown that SURGISEAL acts as a physical barrierto microbial penetration as long as the adhesive film remains intact.Clinical studies were not conducted to demonstrate microbial barrierproperties and a correlation between microbial barrier properties and areduction in infection have not been established.
SubstantialEquivalenceBiocompatibility:
Biocompatibility testing was previously conducted on the currentlymarketed devices, SURGISEAL Teardrop (K082993, K123936 &K130329) and SURGISEAL Stylus (K130474) per the InternationalStandard ISO-10993, "Biological Evaluation of Medical Devices, PartI : Evaluation and Testing". The existing testing is deemed supportiveof the proposed labeled devices. Based on the results from thosestudies, both proposed labeled devices are considered to be non-toxic,non-irritating, non-sensitizing and biocompatible. No additional
biocompatibility testing was necessary as the subject devices areidentical those identified in K082993 and K130474.
Performance Testing:
Bench tests in this submission include: Microbial Barrier Tests andFilm Thickness determinations for each of the two applicators.
Sterilization and Shelf-Life
SURGISEAL® Topical Skin Adhesive is terminally sterilized byelectron beam irradiation as well as by gamma irradiation ; both inaccordance with ISO 11137-2:2006
SURGISEAL Stylus™ Topical Skin Adhesive is sterilized inaccordance with the order of operation of the assembly. The filledampoule containing the adhesive is sterilized by gamma irradiation inaccordance with ISO 11137-2:2006. Then the finished bulk applicatorin the secondary packaging is sterilized by ethylene oxide in accordancewith ISO 11135-1:2008 and ISO 11135-2:2008.
The proposed labeling modification sterilization does not impact the 24month (2 year) expiration date (shelf-life) proposed for these productsand will therefore remain to be labeled with a two (2) year expirationdate.

{1}------------------------------------------------

{2}------------------------------------------------

The labeling change clarifies the performance of the device in order to characterize the ease, safety, or effective use of the product. Based on extensive bench performance testing, the modified labeled devices, SURGISEAL Teardrop and SURGISEAL Stylus, have demonstrated to be safe & efficacious.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a person embracing or protecting another person, or a bird-like figure. The symbol is composed of three curved lines that converge at the bottom.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WOGG-G609 Silver Spring, MI) 20993-0002

April 28, 2014

Adhezion Biomedical, LLC Mr. Richard Jones Regulatory and Quality Assurance Consultant One Meridian Boulevard, Suite 1B02 Wyomissing, Pennsylvania 19610

Re: K140517

Trade/Device Name: SURGISEAL™ Topical Skin Adhesive, SURGISEAL Stylus" Topical Skin Adhesive Regulation Number: 21 CFR 878,4010 Regulation Name: Tissue adhesive Regulatory Class: Class II Product Code: MPN Dated: February 26, 2014 Received: February 28, 2014

Dear Mr. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Mr. Richard Jones

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.lda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincercly yours,

David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S.

Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section 7: Indications for Use

510(k) Number (if known): TBD K140517

Device Names: SURGISEAL® Topical Skin Adhesive SURGISEAL Stylus™ Topical Skin Adhesive

Indications for Use:

SURGISEAL Topical Skin Adhesive is intended for topical applications only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, simple, thoroughly cleansed, trauma induced lacerations.

SURGISEAL may be used in conjunction with, but not in place of, deep dermal sutures

Prescription Use X_ AND/OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Krause -S

Page 1 of

Abbreviated 510(k) Submission SURGISEAL* Topical Skin Adhesive Adhezion Biomedical, LLC Page 7-1

§ 878.4010 Tissue adhesive.

(a)
Tissue adhesive for the topical approximation of skin —(1)Identification. A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.(2)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
Tissue adhesive for non-topical use —(1)Identification. A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
Classification. Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.