K Number
K140436
Date Cleared
2014-12-23

(305 days)

Product Code
Regulation Number
862.1117
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Reagent Kit:

The ARCHITECT Galectin-3 assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of galectin-3 in human serum and EDTA plasma.

The ARCHITECT Galectin-3 assay may be used in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure (HF). The ARCHITECT Galectin-3 assay is used with the ARCHITECT i System with STAT protocol capability.

Calibrators:

The ARCHITECT Galectin-3 Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of galectin-3 in human serum and EDTA plasma.

Controls:

The ARCHITECT Galectin-3 Controls are for the verification of the ARCHITECT i System when used for the quantitative determination of galectin-3 in human serum and EDTA plasma.

Device Description

The ARCHITECT Galectin-3 assay is a two-step immunoassay for the guantitative determination of galectin-3 in human serum or EDTA plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex.

In the first step, sample and M3/38 anti-galectin-3 coated paramagnetic microparticles are combined. Galectin-3 present in the sample binds to the anti-galecin-3 coated microparticles. After washing, 87B5 anti-galectin-3 acridinium-labeled conjugate is added to create a reaction mixture in the second step. Following another wash cycle, pre-trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs).

A direct relationship exists between the amount of galectin-3 in the sample and the RLUs detected by the ARCHITECT i System optics. The kit is composed of the following:

ARCHITECT Galectin-3 Reagent Kit (5P03)
ARCHITECT Galectin-3 Calibrators (5P03-01)
ARCHITECT Galectin-3 Controls (5P03-10)

AI/ML Overview

Here's the information about the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (from text)Reported Device Performance (from text)
Precision≤ 10% Total CV for samples 4.0 to 114.0 ng/mLTotal precision for one lot: ≤ 5.8% for all samples. Total reproducibility for two lots: ≤ 4.7% for all samples. (Meets criteria)
Linearity/Reportable RangeBased on +/-10% deviation from linearity (%DL)Linear range: 5.5 to 103.1 ng/mL. (Meets criteria)
Detection Limit (LoQ)≤ 4.0 ng/mLLoB: 1.0 ng/mL; LoD: 1.1 ng/mL; LoQ: 2.8 ng/mL. (Meets criteria, as LoQ is ≤ 4.0 ng/mL)
Analytical SpecificityPercent cross-reactivity with other human recombinant galectin and collagen proteins < 0.5%All individual patient samples supplemented with various human recombinant galectin and collagen proteins resulted in a percent cross-reactivity of ≤ 0.3%. (Meets criteria)
Therapeutic InterferenceN/A (no explicit quantitative criteria stated for therapeutic interference, but implicitly, minimal interference is expected)Percent differences ranging from -2.9% to 4.0% were observed across various therapeutics. (Deemed acceptable by the study)
Endogenous Substance InterferenceN/A (no explicit quantitative criteria stated for endogenous substance interference, but implicitly, minimal interference is expected)Mean % Difference ranging from -4.7% to 4.7% across various endogenous substances. (Deemed acceptable by the study)
Clinical Performance (Prognosis in HF)"Statistically significant higher risk for the occurrence of the primary endpoint in subjects with galectin-3 levels in the high risk category (> 17.8 ng/mL), compared with galectin-3 levels in the low risk category (< 17.8 ng/mL)."Hazard Ratio for Primary Endpoint (High Risk Category > 17.8 ng/mL) = 1.753 (95% CI: 1.265 – 2.427), which is statistically significant (p-value implicitly <0.05). (Meets criteria)

2. Sample Size Used for the Test Set and Data Provenance

  • Precision (Analytical):

    • Test Set Description: 3 levels of controls and 5 levels of human serum and plasma panels.
    • Sample Size: Minimum of two replicates at two separate times per day for 20 different days (for one lot study) and for 10 different days (for two lots study), for a total of 40 replicates for each lot in the reproducibility study.
    • Data Provenance: Human serum and plasma samples. Retrospective (implied, as samples were collected and then tested for the study). Country of origin is not specified but implied to be US since the submission is to FDA.
  • Linearity/Assay Reportable Range:

    • Test Set Description: Two serum samples (one low, one high) and two plasma samples (one low, one high).
    • Sample Size: Neat and diluted panels tested in replicates of three.
    • Data Provenance: Serum and plasma samples. Retrospective (implied). Country of origin not specified.
  • Detection Limit:

    • Test Set Description: 7 normal human serum samples in ARCHITECT Galectin-3 Calibrator A, and Low Level Panel Set.
    • Sample Size: Replicates of five, twice per day over three days on two systems with two reagent lots (N= 120).
    • Data Provenance: Normal human serum samples. Retrospective (implied). Country of origin not specified.
  • Analytical Specificity (Cross-Reactivity):

    • Test Set Description: Four serum and four plasma samples.
    • Sample Size: The 8 prepared samples were tested in replicates of three.
    • Data Provenance: Human serum and plasma samples. Retrospective (implied). Country of origin not specified.
  • Analytical Specificity (Therapeutic Interference):

    • Test Set Description: Serum and plasma samples with various therapeutic agents spiked in.
    • Sample Size: Each galectin-3 sample pool was split into 2 equal aliquots for each therapeutic tested (test and control).
    • Data Provenance: Serum and plasma samples. Retrospective (implied). Country of origin not specified.
  • Analytical Specificity (Endogenous Substance Interference):

    • Test Set Description: Serum and plasma samples with various endogenous substances spiked in.
    • Sample Size: Each galectin-3 sample was split into two aliquots for each endogenous substance tested (test and control).
    • Data Provenance: Serum and plasma samples. Retrospective (implied). Country of origin not specified.
  • Clinical Study (Prognosis in HF):

    • Test Set Description: Specimens from a multicenter cohort of outpatients with chronic heart failure (HF).
    • Sample Size: 405 serum specimens.
    • Data Provenance: Human serum specimens from a multicenter cohort. Retrospective clinical study ("obtained at time of study entry"). Country of origin is not specified but the submission is to the US FDA.
  • Expected Values/Reference Range:

    • Test Set Description: Plasma samples from apparently healthy subjects without known heart disease.
    • Sample Size: 274 plasma samples (133 females, 141 males).
    • Data Provenance: Plasma samples from an observational study. Retrospective (implied). Country of origin not specified.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • No information is provided about experts establishing ground truth for the analytical performance studies. These studies typically use pre-defined concentrations or reference materials.
  • For the clinical study evaluating prognosis in HF: The ground truth (primary endpoint: hospitalization due to worsening HF, ventricular assist device placement, cardiac transplantation, or all-cause mortality) is based on outcome data. This would typically be recorded clinical events, not established by an adjudication panel of experts in the same way an imaging study would be. The clinical evaluation and diagnosis of heart failure, and classification of events, would be performed by treating physicians/specialists, but the specific process for confirming these outcomes for the study is not detailed as involving a separate panel of experts for ground truth establishment.

4. Adjudication Method for the Test Set

  • Not applicable for the analytical performance studies as they rely on quantitative measurements against known standards or reference methods.
  • For the clinical study, an explicit adjudication method (e.g., 2+1, 3+1) is not described for the primary endpoint events. The primary endpoint (hospitalization due to worsening HF, ventricular assist device placement, cardiac transplantation, or all-cause mortality) are events recorded during patient follow-up. While clinical judgments are involved in diagnosing and reporting these events, the document doesn't indicate a separate, independent adjudication panel for confirming these outcomes specifically for the study data.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) assay that quantitatively measures galectin-3 from patient samples. MRMC studies are typically used for imaging devices or AI algorithms that assist human readers in interpreting complex images or data. The ARCHITECT Galectin-3 assay provides a numerical value, which is then used in conjunction with clinical evaluation, rather than assisting a human reader in a diagnostic task that involves pattern recognition.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Yes, the ARCHITECT Galectin-3 assay operates as a standalone algorithm (device only). It produces a quantitative value (ng/mL) of galectin-3 from a patient sample. The performance characteristics described (precision, linearity, detection limit, analytical specificity) are all based on the device's ability to accurately and precisely provide this numerical output. The "human-in-the-loop" aspect comes during the interpretation of the result in conjunction with clinical evaluation, but the assay itself functions independently to generate the measurement. The clinical study demonstrates its effectiveness in its standalone capacity to provide a prognostic indicator.

7. The Type of Ground Truth Used

  • Analytical Studies: The ground truth for analytical performance studies (precision, linearity, detection limit, analytical specificity) generally relies on reference materials, known concentrations, or established standard methods. For instance, precision is measured against the device's own reported values, linearity against expected dilutions, and detection limits against known low-concentration samples.
  • Clinical Study: The ground truth for the clinical validation study was outcomes data. Specifically, the primary endpoint was a composite of hospitalization due to worsening HF, ventricular assist device placement, cardiac transplantation, or all-cause mortality. These are objective clinical events recorded during patient follow-up.

8. The Sample Size for the Training Set

  • This is an in vitro diagnostic (IVD) assay, not typically an AI/ML algorithm that requires a distinct "training set" in the conventional sense of supervised machine learning.
  • The "training" equivalent for an IVD kit would be the internal development and optimization work, including the choice of reagents, antibodies, and manufacturing processes. The document does not specify a "training set" size. The calibrators and controls (6 levels for calibrators, 3 levels for controls) are used for routine calibration and quality control of the instrument and assay, not a training set for an AI model.

9. How the Ground Truth for the Training Set Was Established

  • As explained above, there isn't a "training set" for an AI/ML model for this type of IVD device.
  • The calibrators' concentrations (0.0 ng/mL to 114.0 ng/mL) and controls' target concentrations (9.1 ng/mL, 20.5 ng/mL, 74.1 ng/mL) are established. The text states that "The primary calibrators for the ARCHITECT Galectin-3 assay were manufactured gravimetrically and produced based on protein content." This means their concentrations are determined by a highly precise gravimetric (weighing) method and based on the amount of recombinant human galectin-3 protein added. This serves as the "ground truth" for calibrating the assay.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

FUJIREBIO DIAGNOSTICS, INC. STACEY DOLAN MANAGER, REGULATORY AFFAIRS 201 GREAT VALLEY PKWY. MALVERN PA 19355-1307

December 23,2014

Re: K140436

Trade/Device Name: ARCHITECT Galectin-3 Reagent Kit ARCHITECT Galectin-3 Calibrators ARCHITECT Galectin-3 Controls Regulation Number: 21 CFR 862.1117 Regulation Name: B-type natriuretic peptide test system Regulatory Class: II Product Code: OSX, JIT, JJX Dated: November 12, 2014 Received: November 13, 2014

Dear Ms. Stacey Dolan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140436

Device Name ARCHITECT Galectin-3

Indications for Use (Describe) Reagent Kit:

The ARCHITECT Galectin-3 assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of galectin-3 in human serum and EDTA plasma.

The ARCHITECT Galectin-3 assay may be used in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure (HF). The ARCHITECT Galectin-3 assay is used with the ARCHITECT i System with STAT protocol capability.

Calibrators:

The ARCHITECT Galectin-3 Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of galectin-3 in human serum and EDTA plasma.

Controls:

The ARCHITECT Galectin-3 Controls are for the verification of the ARCHITECT i System when used for the quantitative determination of galectin-3 in human serum and EDTA plasma.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

A. 510(k) Number:

K140436

B. Purpose for Submission:

New device

C. Measurand:

Galectin-3

D. Type of Test:

Quantitative, Automated chemiluminescence immunoassay on the ARCHITECT i Systems

E. Applicant:

Address:

Fujirebio Diagnostics, Inc.
201 Great Valley Parkway
Malvern, PA 19355
Contact person:Stacey Dolan(610) 240-3843dolans@fdi.com
-------------------------------------------------------------------

Summary preparation date: November 12, 2014

F. Proprietary and Established Names:

ARCHITECT Galectin-3 Reagent Kit ARCHITECT Galectin-3 Calibrators ARCHITECT Galectin-3 Controls

G. Regulatory Information:

    1. Regulation section: 21 CFR § 862.1117, Test, Natriuretic Peptide 21 CFR § 862.1150, Calibrator 21 CFR § 862.1660, Quality Control Material (Assayed and Unassayed)
    1. Classification: Class II
    1. Product code: OSX, Galectin-3 In Vitro Diagnostic Assay

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and white abstract shape on the left, followed by the company name in blue and black text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc."

JIT, Calibrator, Secondary JJX, Single (Specified) Analyte Controls (Assayed and Unassayed)

    1. Panel: 75, Chemistry

H. Intended Use:

    1. Intended use(s): See indications for use below.
    1. Indication(s) for use:

Reagent Kit

The ARCHITECT Galectin-3 assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of galectin-3 in human serum and EDTA plasma. The ARCHITECT Galectin-3 assay may be used in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure (HF). The ARCHITECT Galectin-3 assay is used with the ARCHITECT i System with STAT protocol capability.

Calibrators

The ARCHITECT Galectin-3 Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of galectin-3 antigen in human serum and EDTA plasma.

Controls

The ARCHITECT Galectin-3 Controls are for the verification of the accuracy and precision of the ARCHITECT i System when used for the quantitative determination of galectin-3 in human serum and EDTA plasma.

    1. Special conditions for use statement(s): Prescription use only
    1. Special instrument requirements: Abbott ARCHITECT i 2000 SR System analyzers in STAT mode

-Device Description:

The ARCHITECT Galectin-3 assay is a two-step immunoassay for the guantitative determination of galectin-3 in human serum or EDTA plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex.

In the first step, sample and M3/38 anti-galectin-3 coated paramagnetic microparticles are combined. Galectin-3 present in the sample binds to the anti-galecin-3 coated microparticles. After washing, 87B5 anti-galectin-3 acridinium-labeled conjugate is added to create a reaction mixture in the second step. Following another wash cycle, pre-trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs).

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a blue and white abstract shape, and the company name is written in blue and black text. The word "FUJIREBIO" is in blue, and the words "Diagnostics, Inc." are in black.

A direct relationship exists between the amount of galectin-3 in the sample and the RLUs detected by the ARCHITECT i System optics. The kit is composed of the following:

ARCHITECT Galectin-3 Reagent Kit (5P03)

The ARCHITECT Galectin-3 Reagent Kit consists of 100 (1 x 100) tests. Each kit contains paramagnetic microparticles coated with the anti-galectin-3 Rat monoclonal antibody M3/38, and an acridinium-labeled anti-galectin-3 mouse monoclonal antibody 87B5 conjugate. Reagents can be stored on-board the ARCHITECT i System in accordance with assay-specific instructions.

MICROPARTICLES| 1 Bottle (6.6 mL) Anti-galectin-3 (rat, monoclonal) coated microparticles in PBS buffer with protein (bovine) stabilizers. Minimum concentration: 0.08% solids. Preservative: ProClin 300.

CONJUGATE 1 Bottle (5.9 mL) Anti-galectin-3 (mouse, monoclonal) acridinium-labeled conjugate in PBS buffer with protein (bovine) stabilizers. Minimum concentration: 800 ng/mL. Preservative: ProClin 300.

ARCHITECT Galectin-3 Calibrators (5P03-01)

The ARCHITECT Galectin-3 Calibrators are used for the calibration of the ARCHITECT i System. Each calibrator kit contains one bottle each of Calibrators A, B, C, D, E, and F. The calibrator kit is packaged separately.

CALIBRATORS 6 Bottles (4.0 mL each) of ARCHITECT Galectin-3 Calibrators. Calibrators A through F are prepared with artificial matrix. Calibrators B through F contain recombinant human galectin-3. Preservatives: EDTA and ProClin 950.

CalibratorConcentration(ng/mL)
CAL A0.0
CAL B5.7
CAL C11.4
CAL D22.8
CAL E68.4
CAL F114.0

The calibrators are at the following concentrations:

ARCHITECT Galectin-3 Controls (5P03-10)

The ARCHITECT Galectin-3 Controls are used for the verification of the accuracy and precision of the ARCHITECT i System. Each control kit contains one bottle each of low, medium, and high control. The control kit is packaged separately.

CONTROLS 3 Bottles (8.0 mL each) of ARCHITECT Galectin-3 Controls. The Low, Medium, and High Controls are prepared with artificial matrix and contain recombinant human galectin-3. Preservatives: EDTA and ProClin 950.

{6}------------------------------------------------

The calibrators are at the following concentrations:

ControlConcentrationTarget(ng/mL)ConcentrationRange(ng/mL)
Low Control9.16.4 - 11.8
Medium Control20.514.4 - 26.7
High Control74.151.9 - 96.3

Other Materials/Equipment Required (not Provided):

  • MULTI-ASSAY MANUAL DILUENT 1 Bottle (100 mL) ARCHITECT i Multi-Assay Manual O Diluent containing phosphate buffered saline solution. Preservative: antimicrobial agent.
  • PRE-TRIGGER SOLUTION Pre-Trigger Solution containing 1.32 % (w/v) hydrogen O peroxide.
  • TRIGGER SOLUTION Trigger Solution containing 0.35N sodium hydroxide. O
  • WASH BUFFER| Wash Buffer containing phosphate buffered saline solution. Preservative: O Antimicrobial Agents.
  • ARCHITECT i2000SR System in STAT mode O
  • ARCHITECT i System |ASSAY CD-ROM US Addition A (3K50) O
  • ARCHITECT i System e-Assay CD-ROM at www.abbottdiagnostics.com O
  • ARCHITECT i REACTION VESSELS O
  • ARCHITECT i SAMPLE CUPS o
  • ARCHITECT iSEPTUM O
  • ARCHITECT / REPLACEMENT CAPS O
  • Pipettes / Pipette Tips o

J. Substantial Equivalence Information:

    1. Predicate device name(s): BG Medicine, Inc. Galectin-3 Assay
    1. Predicate 510(k) number(s): K093758
    1. Comparison with predicate:

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized blue and green graphic to the left of the company name. The company name is written in blue, with "Diagnostics, Inc." written in a smaller font size below the main name.

Similarities
ARCHITECT Galectin-3(Proposed Device)BG Medicine Galectin-3 Assay(Predicate Device)K093758
Device TypeIn vitro diagnosticIn vitro diagnostic
ClassificationClass IIClass II
CFR section862.117862.117
Product CodeOSXOSX
Product UsageClinical and Hospital laboratoriesClinical and Hospital laboratories
Intended UseQuantitative determination ofgalectin-3 in human serum andEDTA plasma.ARCHITECT Galectin-3 assaymay be used in conjunction withclinical evaluation as an aid inassessing the prognosis ofpatients diagnosed with chronicheart failure (HF).Quantitatively measures galectin-3in serum or EDTA plasma byenzyme-linked immunosorbentassay (ELISA) on a microtiter plateplatform, to be used in conjunctionwith clinical evaluation as an aid inassessing the prognosis ofpatients diagnosed with chronicheart failure (HF).
Assay Range5.5 — 103.1 ng/mL1.4 – 94.8 ng/mL
Type of SpecimenHuman Serum or EDTA plasmaHuman Serum or EDTA plasma
Specimen CollectionMethodRoutine Phlebotomy TechniquesRoutine Phlebotomy Techniques
Capture AntibodyMouse monoclonal (M3/38)Mouse monoclonal (M3/38)
Detection AntibodyMouse monoclonal (87B5)Mouse monoclonal (87B5)
AnalyteHuman galectin-3Human galectin-3

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized blue and white graphic on the left, followed by the company name in blue and black text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc."

Differences
ARCHITECT Galectin-3(Proposed Device)BG Medicine Galectin- 3 Assay(Predicate Device)K093758
Instrument SystemARCHITECT i SystemNone
Principle of OperationChemiluminscent MicroparticleImmunoassay (CMIA)Manual Enzyme LinkedImmunosorbent Assay (ELISA)
Calibrators6 Levels (0.0 - 114.0 ng/mL)Ready to Use-Liquid-Supplied as separate kit1 Level (12 ng/vial) Serial dilutedprior to use-Lyophilized-Supplied with Kit
Controls-3 Levels (9.1, 20.5, and 74.1ng/mL)-Liquid-Supplied as separate kit-2 Levels (18.4 and 69 ng/mL)-Lyophilized-Supplied with Kit
Interpretation of ResultsCalibrator CurveStandard Curve
Cut-offgalectin-3 risk categories:• galectin-3 greater than 17.8ng/mL•galectin-3 less than or equal to17.8 ng/mLgalectin-3 risk categories:• galectin-3 greater than 25.9ng/mL• galectin-3 between 17.8 and 25.9ng/mL•galectin-3 less than or equal to17.8 ng/mL

K. Standard/Guidance Document Referenced (if applicable):

  • . CLSI C28-A3c: Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline
  • . CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach: Approved Guideline
  • . CLSI EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline
  • CLSI EP14-A2: Evaluation of Matrix Effects: Approved Guideline ●
  • CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
  • . CLSI EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline
  • CEN13640: Stability Testing of In Vitro Diagnostic Reagents ●
  • . Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff
  • . Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers. Document issued on: November 30, 2000

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo features a stylized blue and green graphic to the left of the company name. The company name is in blue, with the word "FUJIREBIO" in a larger font than "Diagnostics, Inc." There is a registered trademark symbol next to the word "FUJIREBIO".

L. Test Principle

The ARCHITECT Galectin-3 assay is a two-step immunoassay for the guantitative determination of galectin-3 in human serum or EDTA plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex. In the first step, sample and M3/38 antigalectin-3 coated paramagnetic microparticles are combined. Galectin-3 present in the sample binds to the anti-galecin-3 coated microparticles. After washing, 87B5 anti-galectin-3 acridinium-labeled conjugate is added to create a reaction mixture in the second step. Following another wash cycle, pre-trigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs). A direct relationship exists between the amount of galectin-3 in the RLUs detected by the ARCHITECT i System optics.

The ARCHITECT Galectin-3 Calibrators are for the ARCHITECT i System when used for the quantitative determination of galectin-3 antigen in human serum and EDTA plasma.

The ARCHITECT Galectin-3 Controls are for the verification of the accuracy and precision of the ARCHITECT i System when used for the quantitative determination of galectin-3 in human serum and EDTA plasma.

M. Performance Characteristics (if/when applicable):

Data were generated using the ARCHITECT i 2000SR System with the STAT protocol.

    1. Analytical performance:
    • a. Precision/Reproducibility:

Precision

The ARCHITECT Galectin-3 assay is designed to have an imprecision of ≤ 10% Total CV for samples with galectin-3 concentrations ranging from 4.0 to 114.0 ng/mL.

A study was performed based on guidance from the National Committee for Clinical Laboratory Standards (NCCLS) document EP5-A2. Testing was conducted using one (1) lot of ARCHITECT Galectin-3 Reagents, Calibrators, and Controls, and one (1) ARCHITECT i 2000sg instrument. Three (3) levels of controls and five (5) levels of human serum and plasma panels were assayed with a minimum of two (2) replicates at two (2) separate times per day for twenty (20) different days. A single calibration curve was used throughout the study.

The precision analyses determined that the total precision for one (1) lot individually is ≤ 5.8% for all samples in this study.

System Reproducibility

A study was performed based on guidance from the NCCLS document EP5-A2. Testing was conducted using two (2) lots of ARCHITECT Galectin-3 Reagents, Calibrators, and Controls on

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a blue and white abstract shape. The text "FUJIREBIO" is in large, bold, blue letters, and the text "Diagnostics, Inc." is in smaller, black letters below the company name.

two (2) different ARCHITECT i 2000SR instruments. Each reagent lot was matched with a different lot of calibrators and controls. Three (3) levels of controls and five (5) levels of human serum and plasma panels were assaved with a minimum of two (2) replicates at two (2) separate times per day for ten (10) different days for a total of forty (40) replicates for each lot. Each reagent lot used a single calibration curve throughout the study.

The precision analyses determined that the total reproducibility for two (2) lots is ≤ 4.7% for all samples in this study.

  • b. Linearity/assay reportable range:
    A study was performed based on guidance from the CLSI document EP6-A. Two (2) serum samples (one low and one high for galectin-3 levels) and two (2) plasma samples (one low and one high for galectin-3 levels) were used for this study.

The high serum and high plasma samples were diluted with their respective low sample. The neat and diluted panels were tested in replicates of three (3) using the ARCHITECT Galectin-3 assay on the ARCHITECT i2000SR System in STAT mode.

Based on a +/-10% deviation from linearity (%DL), the data supports a linear range of 5.5 to 103.1 ng/mL for the ARCHITECT Galectin-3 assay.

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):
    The ARCHITECT Galectin-3 Calibrators are for the ARCHITECT i System when used for the quantitative determination of galectin-3 antigen in human serum and EDTA plasma. 6 Bottles (4.0 mL each) are supplied for the ARCHITECT Galectin-3 Calibrators. Calibrators A through F are prepared with artificial matrix. Calibrators B through F contain recombinant human galectin-3. Preservatives: EDTA and ProClin 950. The calibrators are at the following concentrations:
CalibratorConcentration(ng/mL)
CAL A0.0
CAL B5.7
CAL C11.4
CAL D22.8
CAL E68.4
CAL F114.0

There is currently no known internationally recognized consensus reference method or reference material for standardization. Galectin-3 assay values are expressed as ng/mL. This value is related to a Fujirebio Diagnostics maintained reference preparation. The primary calibrators for the ARCHITECT Galectin-3 assay were manufactured gravimetrically and produced based on protein content.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and green abstract shape on the left, followed by the company name in blue, bold letters. Below the company name, the words "Diagnostics, Inc." are written in a smaller, non-bold font.

The stability data supports the current shelf life assignment for the ARCHITECT Galectin-3 Calibrators of 12 months.

The ARCHITECT Galectin-3 Controls are for the verification of the accuracy and precision of the ARCHITECT i System when used for the quantitative determination of galectin-3 in human serum and EDTA plasma. 3 Bottles (8.0 mL each) are supplied for the ARCHITECT Galectin-3 Controls. The Low, Medium, and High Controls are prepared with artificial matrix and contain recombinant human galectin-3. Preservatives: EDTA and ProClin 950. The controls are at the following concentrations:

ControlConcentrationTarget(ng/mL)ConcentrationRange(ng/mL)
Low Control9.16.4 - 11.8
Medium Control20.514.4 - 26.7
High Control74.151.9 - 96.3

The stability data supports the current shelf life assignment for the ARCHITECT Galectin-3 Controls of 12 months.

  • d. Detection limit:
    The ARCHITECT Galectin-3 assay is designed to have a Limit of Quantitation (LoQ) of ≤ 4.0 ng/mL. Based on guidance from the CLSI document EP17-A2, a study was performed by diluting 7 normal human serum samples in ARCHITECT Galectin-3 Calibrator A. Two (2) lots of ARCHITECT Galectin-3 Calibrator A (run as the zero-level panel) and Low Level Panel Set were run in replicates of five (5), twice per day over three (3) days on two (2) ARCHITECT I2000SR Systems in STAT mode with two (2) ARCHITECT Galectin-3 assay reagent lots (N= 5x2x3x2x2 = 120).

The Limit of Blank (LoB). Limit of Detection (LoD). and Limit of Quantitation (LoQ) determined for the ARCHITECT Galectin-3 assay was 1.0 ng/mL. 1.1 ng/mL. and 2.8 ng/mL respectively.

The LoQ met the predetermined acceptance criteria of LoD ≤ 4.0 ng/mL. These data support the performance claims of the assay.

  • e. Analytical specificity:
    A study was conducted to evaluate the potential interference from other recombinant human galectin and collagen proteins in the ARCHITECT Galectin-3 assay.

Four (4) serum and four (4) patient plasma samples were used for this study. Three (3) serum samples and two (2) plasma samples were augmented with human galectin-3 antigen targeting a 'normal' concentration of galectin-3 (10-13 ng/mL) and an elevated concentration of galectin-3 (=26 ng/mL). The other 3 samples (One (1) serum and two (2) plasma) were at the target concentrations and did not require augmentation. The eight (8) samples were supplemented with the various cross reactants each prepared at a concentration of 10 µg/mL in PBS.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic to the left of the company name. The company name is written in a sans-serif font, with "FUJIREBIO" in a larger, bolder font than "Diagnostics, Inc."

The prepared samples were tested in replicates of three (3) using the ARCHITECT Galectin-3 assay on the ARCHITECT i2000SR System in STAT mode.

All individual patient samplemented with the various human recombinant galectin and collagen proteins resulted in a percent cross reactivity with the ARCHITECT Galectin-3 assay of ≤ 0.3% and support the performance claims of the assay (below 0.5%).

A study was performed based on guidance from the CLSI document EP07-A2. Potentially interfering drugs were evaluated to determine whether galectin-3 concentrations were affected when using the ARCHITECT Galectin-3 assay.

Serum and plasma samples selected based on their endogenous galectin-3 values which fell within or near the galectin-3 target ranges were used in this study. The predetermined galectin-3 targets were a 'normal' concentration of galectin-3 (10-13 ng/mL) and an elevated concentration near the upper end of the reference range for galectin-3 (≈26 ng/mL).

Each qalectin-3 sample pool was split into 2 equal aliquots for each therapeutic tested. One aliquot was spiked with the therapeutic agent stock for use as the test sample. The samples were not diluted more than 5% by the agent. The other aliquot (control) was spiked with an equal volume of the same solvent used for therapeutic agent preparation for each respective therapeutic agent.

The data are summarized in the following table.

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic element on the left, consisting of blue and white shapes. To the right of the graphic is the company name, "FUJIREBIO" in a bold, blue font, with "Diagnostics, Inc." appearing in a smaller, dark gray font below it.

TherapeuticTherapeuticConcentrationPercentDifference (%)
Acetaminophen (4-Acetamidophenol)1324 µmol/L-2.9%
Acetylsalicylic Acid3.62 mmol/L-0.7%
Amlodipine besylate10.6 µmol/L-0.1%
Ampicillin152 µmol/L-0.1%
L-Ascorbic Acid342 µmol/L1.9%
Atenolol37.6 µmol/L-0.4%
Caffeine308 µmol/L0.7%
Carvedilol74 µmol/L2.4%
Captopril23 µmol/L-1.9%
Chloramphenicol155 µmol/L1.1%
Diclofenac sodium salt169 µmol/L-0.7%
Digoxin7.8 nmol/L0.8%
(+)-cis-Diltiazem hydrochloride576.5 µmol/L-0.5%
Disopyramide29.5 µmol/L-2.5%
Dopamine hydrochloride5.87 µmol/L4.0%
Enalaprilat dihydrate0.86 µmol/L1.6%
Furosemide181 µmol/L1.0%
Hydrochlorothiazide20.2 µmol/L0.0%
Ibuprofen2425 µmol/L-0.6%
Indomethacin100 µmol/L-1.7%
Lidocaine51.2 µmol/L0.1%
Lisinopril • 2H2O0.74 µmol/L-0.9%
Losartan potassium130 µmol/L-1.1%
Lovastatin (Simvastatin)191 µmol/L1.7%
Methyldopa71 µmol/L-2.3%
(±)-Metoprolol (+)-tartrate salt18.7 µmol/L-2.8%
Naproxen2170 µmol/L-1.0%
Nifedipine1156 nmol/L0.4%
Quinidine37 µmol/L-0.7%
Ramipril14.4 µmol/L0.0%
Spironolactone1.44 µmol/L-2.0%
Theophylline222 µmol/L-2.9%
Verapamil hydrochloride244 µmol/L0.6%
Warfarin32.5 µmol/L-2.1%
Trasylol/Aprotinin100 KIE/mL0.0%

A study was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP07-A2. Potentially interfering substances were evaluated to determine whether galectin-3 concentrations were affected when using the ARCHITECT Galectin-3 assay.

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and green abstract design on the left, followed by the company name in blue, bold letters. Below the company name, the words "Diagnostics, Inc." are written in a smaller, black font.

Serum and plasma samples selected based on their endogenous galectin-3 values which fell within or near the galectin-3 target ranges were used in this study. The predetermined galectin-3 targets were a 'normal' concentration of galectin-3 (10-13 ng/mL) and an elevated concentration near the upper end of the reference range for galectin-3 (≈26 ng/mL).

Each galectin-3 sample was split into two (2) aliquots for each endogenous substance tested. One (1) aliquot was spiked with an endogenous substance for use as the test sample. One (1) aliquot was spiked with an equal volume of ARCHITECT PBS (phosphate-buffered solution) for use as a control samples were not diluted more than 20% by the agent.

The data are summarized in the following table.

Potentially InterferingSubstanceInterferentConcentrationMean %Difference
Bilirubin (Unconjugated)≥ 40 mg/dL-3.8
Bilirubin (Conjugated)≥ 40 mg/dL0.7
Hemoglobin≥ 250 mg/dL4.6
Triglycerides≥ 3000 mg/dL4.4
Human Serum Albumin≥ 12 g/dL-4.7
Human Gamma Globulin≥ 5 g/dL-3.2
Cholesterol≥ 500 mg/dL3.7
Whole Blood Lysate b5 mg/dL4.7
Creatinine≥ 5 mg/dL0.5
Rheumatoid Factor800 IU/mL3.5
Human Anti-MouseAntibodies≥1000 ng/mL1.1
  • f. Assay cut-off:
    See Clinical Cutoff in M (4) below

    1. Comparison studies: N/A
    1. Clinical studies:
    • a. Clinical sensitivity:

See 3(c) below

  • b. Clinical specificity:

See 3(c) below

  • Other clinical supportive data (when a. and b. are not applicable): C.

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized blue and green graphic to the left of the company name. The company name is written in a bold, sans-serif font, with "FUJIREBIO" on the top line and "Diagnostics, Inc." on the bottom line.

The clinical performance of the ARCHITECT Galectin-3 assay was evaluated in a clinical validation study with specimens from a multicenter cohort of outpatients with chronic heart failure (HF). A total of 405 serum specimens were obtained at time of study entry from patients with varying degrees of chronic HF, encompassing all New York Heart Association (NYHA) classification categories. The median follow-up time for the study population was approximately 652 days.

Of the 405 baseline samples, 143 (35.3%) were obtained from females and 262 (64.7%) were obtained from males. The mean subject age was 58 years. The specimens were tested using the ARCHITECT Galectin-3 assay and categorized based on galectin-3 value as high risk (>17.8ng/mL) or low risk (≤17.8 ng/mL). A total of 226 subjects were in the high risk category and the remaining 179 subjects were in the low risk category. A multivariable Cox proportional hazards regression model was used to evaluate the association of qalectin-3 category with the primary endpoint of hospitalization due to worsening HF, ventricular assist device placement, cardiac transplantation, or all-cause mortality.

The multivariable Cox proportional hazards regression model was adjusted for the parameters of age, gender, NYHA class, left ventricular ejection fraction (LVEF), diabetes status, and smoking status. The results of the regression model demonstrated a statistically significant higher risk for the occurrence of the primary endpoint in subjects with galectin-3 levels in the high risk category (> 17.8 nq/mL), compared with galectin-3 levels in the low risk category (< 17.8 ng/mL).

Hazard Ratio for Primary Endboint (hospitalization due to worsening heart failure, ventricular assist device placement. cardiac transplantation, or all-cause mortality; first to occur) for Chronic Heart Failure Subjects

Hazard Ratiob(95% Confidence Interval)
High Risk Categorya
(> 17.8 ng/mL)1.753 (1.265 – 2.427)

adjusted for baseline risk factors: age, gender, NYHA functional classification, left ventricular ejection fraction, diabetes status, and smoking status.

6 Hazard ratio is relative to the galectin-3 low risk category (≤ 17.8 ng/mL)

Elevated baseline levels of galectin-3 (>17.8 ng/mL) in chronic HF patients were shown to be significantly and independently associated with a higher risk of hospitalization due to worsening HF, ventricular assist device placement, cardiac transplantation, or all-cause mortality (first to occur).

The figure below presents the Kaplan-Meier Survival Function Estimates time-to-event (hospitalization due to worsening HF, ventricular assist device placement, cardiac transplantation, or all-cause mortality) based on the galectin-3 cutoff value of 17,8 ng/ml., where the probability of survival is plotted with the elapsed time in days from baseline (i.e., follow-up time).

{16}------------------------------------------------

Additional Information (AI) Request Premarket Notification (510(k)) ARCHITECT Galectin-3

Image /page/16/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized blue and green graphic on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc."

Image /page/16/Figure/2 description: This image is a survival probability graph that compares two groups: Galectin less than or equal to 17.8 ng/mL and Galectin greater than 17.8 ng/mL. The x-axis represents follow-up time in days, ranging from 0 to 1500. The y-axis represents survival probability, ranging from 0.0 to 1.0. The graph shows that the group with Galectin less than or equal to 17.8 ng/mL has a higher survival probability than the group with Galectin greater than 17.8 ng/mL.

Follow-up time (days)

The bar graph and table below depict the cumulative probability of the primary endpoint (hospitalization due to worsening HF, ventricular assist device placement, cardiac transplantation, or all-cause mortality) by galectin-3 category, at selected time points.

Image /page/16/Figure/5 description: The image is a bar graph that shows the cumulative probability of an event over time, broken down by two groups: those with <= 17.8 ng/mL and those with > 17.8 ng/mL. The x-axis represents time in months (6, 12, 24, and 36 months), and the y-axis represents the cumulative probability of the event. At 36 months, the cumulative probability of the event is 43.4% for the <= 17.8 ng/mL group and 64.4% for the > 17.8 ng/mL group.

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, followed by the company name in blue text. Below the company name, "Diagnostics, Inc." is written in smaller, black text. The graphic on the left is a combination of blue and green shapes, with a white teardrop shape in the center.

Cumulative Probability (95% ConfidenceInterval) of Event for Primary Endpoint, byGalectin-3 Category, at Selected TimePoints (in percent)
Galectin-3Category6Months12Months24Months36Months
≤ 17.8 ng/mL11.7(7.0 -16.4)15.7(10.4 -21.0)23.4(16.9 -29.9)43.4(33.3 -53.6)
> 17.8 ng/mL21.2(15.9 -26.6)33.9(27.7 -40.1)43.2(36.5 -49.9)64.4(56.4 -72.4)

Interpretation

ARCHITECT Galectin-3 assay results should be interpreted in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure. In the clinical validation study, chronic heart failure patients with qalectin-3 levels over 17.8 ng/mL were found to have a higher risk of adverse outcomes, defined as hospitalization for HF, placement of left ventricular assist device, cardiac transplantation and mortality, when compared to patients with levels below 17.8 ng/mL.

Interpretation Relative to Natriuretic Peptides

Galectin-3 and natriuretic peptides, such as NT-proBNP and BNP, are measures of separate and distinct biological processes. Galectin-3 and natriuretic peptides provide independent and complementary information on the prognosis of patients with chronic heart failure.

Elevated baseline levels of both galectin-3 and a natriuretic peptide have been associated with poorer prognosis in chronic heart failure patients, when compared to patients with low baseline levels of both markers. Elevated levels of one or the other marker, but not both, have been associated with a poorer prognosis in chronic heart failure when compared to patients with low levels of both markers, and with a better prognosis when compared to patients with elevated baseline levels of both markers.

The table below illustrates this for N-terminal pro B-type natriuretic peptide (NT-proBNP) in the clinical validation study by evaluating primary endpoint event rates by categories of galectin-3 and NT-proBNP levels.

{18}------------------------------------------------

Image /page/18/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a blue and white design with curved lines and a teardrop shape. The company name is written in blue, with "FUJIREBIO" in a larger font and "Diagnostics, Inc." in a smaller font below it.

Event Rates by Joint Galectin-3 and NT-proBNP Category. Median NT-proBNP value in study is 1222 pg/mL.

Subjects In Each Category Experiencing a Hospitalizationfor Heart Failure, Ventricular Assist Device Placement,Cardiac Transplantation, or Death
Within OverallStudyHazard Ratio(95% ConfidenceInterval)P-value
Galectin-3 ≤ 17.8 ng/mL andNT-proBNP≤ median(N=99)24.2%1(referent category)NA
Galectin-3 ≤ 17.8 ng/mL andNT-proBNP> median(N=64)48.4%2.077(1.216-3.549)0.0075
Galectin-3 > 17.8 ng/mL andNT-proBNP≤ median(N=83)48.2%1.835(1.104-3.050)0.0191
Galectin-3 > 17.8 ng/mL andNT-proBNP> median(N=115)66.1%4.014(2.533-6.362)<0.0001

The figure below is a plot of Kaplan-Meier survival function estimates by joint galectin-3/NTproBNP categories, for the study endpoint.

{19}------------------------------------------------

Image /page/19/Picture/1 description: The image contains the logo for FUJIREBIO Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a blue and white abstract shape. The company name is written in blue, with "FUJIREBIO" on the top line and "Diagnostics, Inc." on the bottom line in a smaller font.

Image /page/19/Figure/2 description: The image contains text describing Kaplan-Meier curves. The curves are for the endpoint of the clinical validation study. The curves are separated by galectin-3 and NT-proBNP levels. The Logrank test p-value is less than 0.001.

    1. Clinical cut-off:
      See 3(c) above

{20}------------------------------------------------

Image /page/20/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized blue and green graphic to the left of the company name. The company name is written in blue, with "Diagnostics, Inc." written in a smaller, dark gray font below the company name.

5. Expected values/Reference range:

An observational study was performed to determine the reference distribution for the ARCHITECT Galectin-3 assay. Galectin-3 levels were measured in 274 plasma samples from a population of apparently healthy subjects without known heart disease but that otherwise resemble the heart failure patient population by age and gender distribution.

Specimens were from women between the ages of 60 and 80 years (n = 133) and men between the ages of 55 and 80 (n = 141). The reference population consisted of the following race/ethnic groups:

  • 55 African-American .

  • 189 Caucasian

  • 9 Other ●

  • . 9 Asian

  • . 12 Hispanic
    All subjects had detectable galectin-3 levels (minimum - maximum: 6.8 - 54.4 ng/mL) within the measuring interval of the ARCHITECT Galectin-3 assay. The 97.5% of the galectin-3 reference population was 27.5 ng/mL. The data are summarized in the following table.

Percentile(%)Galectin-3 Reference Population(ng/mL)
2.58.2
59.3
2512.4
5014.8
7518.5
9022.4
9525.7
97.527.5

6. Conclusion

The results of these analytical (nonclinical) and clinical studies demonstrate that the ARCHITECT Galectin-3 assay is substantially equivalent to the performance of the BG Medicine Galectin-3 assay.

N. Proposed Labeling:

The labeling satisfies the requirements of 21 CFR Part 809.10.

§ 862.1117 B-type natriuretic peptide test system.

(a)
Identification. The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.(b)
Classification. Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”