K Number
K140430
Device Name
740 SELECT
Date Cleared
2014-06-13

(114 days)

Product Code
Regulation Number
870.2300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 740 SELECT series of monitors is indicated for use as a portable, multi-parameter, variable aculty device for use by health care professionals, clinicians and medically qualified personnel, in a variety of clinical environments and hospital departments, for non-invasive spot checking, and/or continuous monitoring and recording of:

  • Blood pressure and pulse rate of adult, pediatric and neonatal patients; .
  • Functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric and . neonatal patients;
  • Additionally the Masimo Rainbow SET technology provides noninvasive monitoring of . carboxyhemoglobin saturation (SpCO) and/or respiration (RRa). Other information displayed on the 740 SELECT with the Masimo Rainbow SET option includes: Signal IQ Waveform, Low Signal IQ (Low SIQ), Perfusion Index (PI), and/or Pleth Variability Index (PVI) indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused;
  • Carbon Dioxide concentration of the expired and inspired breath and respiration rate of adult, . pediatric, infant or neonatal patient and included, with the Oridion option, an Integrated Pulmonary Index (IPI):
  • Intermittent predictive body temperature (oral, axillary, rectal) of adult, pediatric and neonatal ● patients;
  • Infrared (over the temporal artery) measurement of body temperature of people of all ages. .
Device Description

The 740 SELECT is a rugged portable non-invasive multi-parameter device used for spot checking or continuous monitoring of blood pressure, pulse rate, functional oxygen saturation (%Sp02), body temperature and CO2 in various modes of operation.

AI/ML Overview

The CASMED 740 SELECT is a multi-parameter physiological monitor. The provided document does not contain an "acceptance criteria" table with specific performance metrics and acceptance thresholds for the device as a whole. Instead, it states that the device has undergone clinical and non-clinical testing to demonstrate substantial equivalence to predicate devices, and that the OEM's clinical work for the pulse oximeters, CO2, and temperature functions is cited.

Therefore, the acceptance criteria are implicitly tied to demonstrating substantial equivalence, meaning the 740 SELECT performs comparably to the identified predicate devices for each parameter (NIBP, SpO2, Temperature, CO2).

Here's an analysis of the provided information based on your request, with limitations due to the nature of the available document (a 510(k) summary, not a detailed study report):

1. Table of Acceptance Criteria and Reported Device Performance

As noted, a formal table of acceptance criteria and specific reported device performance values is not provided in this 510(k) summary. The summary focuses on demonstrating substantial equivalence. The document indicates that for the NIBP, SpO2, CO2, and Temperature functions, the 740 SELECT incorporates technologies from other manufacturers (OEMs) that have already undergone validation.

The implicit acceptance criteria for each parameter would be that its performance is equivalent to, or within the accepted ranges of, its respective predicate device or the OEM component's validated performance.

2. Sample Sizes and Data Provenance

  • Test Set Sample Sizes:
    • NIBP: The document states that CAS has "conducted a full program of individual hardware, software and systems verification and validation studies." It also mentions "clinical validation reports for the NIBP." However, specific sample sizes for the NIBP clinical validation are not provided.
    • Pulse Oximeters, CO2, Temperature: The document explicitly states that the 740 SELECT incorporates OEM components (Masimo, Nellcor, Exergen, Phasein, Oridion). For these parameters, the "OEM's clinical work" is cited. This implies that the validation data for these components predates the 740 SELECT and was done by the original equipment manufacturers. No specific sample sizes for these OEM components' clinical work are provided in this document pertaining to the 740 SELECT's submission.
  • Data Provenance: The document does not specify the country of origin for any clinical data. Given the submission is to the FDA, it is likely some data would involve studies conducted or accepted in the US, but this is not explicitly stated. It refers to "clinical validation reports for the NIBP" (presumably conducted by CAS Medical Systems) and "the OEM's clinical work" for other parameters. It is retrospective in the sense that the OEM data was already established.

3. Number of Experts and Qualifications for Ground Truth - Test Set

  • NIBP: Not specified. "Clinical validation reports" are mentioned, which would typically involve qualified medical professionals, but their number and specific qualifications are not detailed.
  • Pulse Oximeters, CO2, Temperature: Not specified. For the OEM components, the ground truth establishment would have been part of their original validation studies, which are not detailed here.

4. Adjudication Method - Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for establishing ground truth in the clinical validation studies, either for the NIBP component or for the OEM components.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done as described (human readers improving with AI vs. without AI assistance).
  • The device being submitted (740 SELECT) is a physiological monitor, not an AI-assisted diagnostic tool that would typically involve human "readers" interpreting data with and without AI assistance for comparative effectiveness. The focus is on the accuracy and reliability of the physiological measurements themselves.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

  • Yes, in essence, standalone performance was assessed for the measurement algorithms.
  • The document implies that the measurement algorithms for NIBP, SpO2, CO2, and Temperature are intended to operate independently to provide accurate physiological readings. The "clinical validation reports for the NIBP" and the "OEM's clinical work for the pulse oximeters, CO2 and the temperature functions" refer to studies evaluating the accuracy of these measurement technologies.
  • The 740 SELECT itself is a "device" with algorithms embedded to perform these measurements. The "clinical validation" and "OEM's clinical work" would have primarily focused on the accuracy of these algorithms and sensors in determining physiological parameters against a reference standard.

7. Type of Ground Truth Used

  • NIBP: While not explicitly stated, NIBP validation typically uses invasive arterial line measurements as the gold standard ground truth for comparing non-invasive readings.
  • Functional Oxygen Saturation (SpO2): For pulse oximeters, the ground truth is often established by co-oximetry of arterial blood samples.
  • CO2: For CO2 monitors (Capnography), the ground truth is typically derived from direct measurement of CO2 concentration in exhaled gas using a laboratory reference method.
  • Temperature: For thermometry, the ground truth is usually established by highly accurate reference thermometers measuring core body temperature (e.g., rectal, esophageal, or arterial temperature probes).

8. Sample Size for the Training Set

  • Not specified. The document does not provide details on training sets for any of the algorithms. Since the device integrates established technologies, the initial development and "training" (if applicable to the algorithm type) would have occurred during the development of these OEM components, not explicitly for the 740 SELECT submission.

9. How the Ground Truth for the Training Set Was Established

  • Not specified. As mentioned above, details about training sets and their ground truth establishment are not present in this 510(k) summary. This information would typically be found in detailed design and validation documentation for the individual components by their respective OEM manufacturers.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the word "CASMED" in bold, black letters. The letters are large and fill most of the frame. A small registered trademark symbol is located to the right of the letter "D".

JUN 1 3 2014

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Submitter:CAS Medical Systems, Inc.
Address:44 East Industrial Rd. Branford CT. 06405 USA
Contact:Ron Jeffrey - Director, Regulatory AffairsPhone - (203) 488-6056Fax - (203) 488-9438Email - rjeffrey@casmed.com
Prepared:February 14, 2014
Trade Name:740 SELECT
Common Name:Monitor, Physiological, Patient
Classification Name:Cardiovascular Monitoring Device 870.2300 (MWI)

Substantially Equivalent Device(s):

The 740 SELECT is equivalent to the following devices:

  • CASMED 740 SELECT (Phase 1) Monitor (K130411);
  • Masimo Corporation Radical 7 Pulse Co-Oximeter (K120657);
  • Covidien (Nellcor) Bedside Respiratory Patient Monitoring System (K 130320); .
  • Exergen Corporation TemporalScanner Thermometer (K011291; ●
  • PHASEIN AB (Masimo) Infrared Sidestream Gas Analyzer (ISA) (K103604); .
  • PHASEIN AB (Masimo) Infrared Mainstream Gas Analyzer (IRMA) (K 123403);
  • Oridion (Covidien) Capnostream20 (K094012)

DESCRIPTION

The 740 SELECT is a rugged portable non-invasive multi-parameter device used for spot checking or continuous monitoring of blood pressure, pulse rate, functional oxygen saturation (%Sp02), body temperature and CO2 in various modes of operation.

The non-invasive blood pressure (NIBP) parameter automatically inflates an occluding cuff and, using the oscillometric step-deflation measurement technique, determines systolic, diastolic and mean arterial pressure as well as pulse rate. Measurement results along with user prompts and error messages are displayed on the front panel. The frequency of NIBP determination can be selected by the user in

{1}------------------------------------------------

varied times between one and ninety minutes. The auto and manual operating modes cover a variety of clinical uses.

The pulse oximeter parameter (%SpO2) determines arterial oxyhemoglobin saturation by measuring the absorption of red and infrared light passing through the tissue. Changes in absorption caused by pulsations of blood in the vascular bed are used to determine arterial saturation and pulse rate. The oximeter requires no routine calibration or maintenance. Oxygen saturation and heart rate are displayed on a light emitting diode (LED) digital display. On each detected pulse, the perfusion LED indicates patient perfusion signals. This bar graph gives the user a pulse-by-pulse visual indication of waveform signal quality. An audio "beep" can be enabled that is generated each time the SpO2 module detects a pulse. In addition to %SpO2 the SELECT (equipped with the Masimo rainbow® Pulse CO-Oximetry™ and Masimo SET ® oximetry) can monitor Pulse Rate (PR), Carboxyhemoglobin (SpC0®), Acoustic Respiration Rate (RRa) and Pleth Variability Index (PVI®). Alternately, the customer can choose a Nellcor Pulse Oximeter at this time of ordering.

The predictive temperature parameter has the capability of taking temperature in either normal (predictive) or monitor mode. In the normal mode, the thermometer's microprocessor "predicts" body temperature in about four (4) seconds for oral temperatures, about ten (10) seconds for axillary temperatures and about fifteen (15) seconds for rectal temperatures. The default setting used by the monitor for Temperature determinations is the Predictive (normal) mode. Alternately, the customer can choose an Exergen TemporalScanner (infrared) Thermometer at the time of ordering. The Exergen TemporalScanner Thermometer has the capability of taking Temporal Artery (TA) Temperature in about 3 seconds. The Exergen Temp scanner has the ability to measure the patient's Arterial temperature in °F or °C.

The 740 SELECT has (2) optional choices for the parameter of CO2 monitoring. The customer may order either the Masimo - Phasein™ Infrared Sidestream Gas Analyzer (ISA™) or the Infrared Mainstream Gas Analyzer (IRMATM). Alternately, the customer can select the option of the Oridion Sidestream MicroPod™ EtCO2. Each of these CO2 options is externally attached to the SELECT.

The 740 SELECT Intended Use

The 740 SELECT series of monitors is indicated for use as a portable, multi-parameter, variable acuity device for use by health care professionals, clinicians and medically qualified personnel, in a variety of clinical environments and hospital departments, for non-invasive spot checking, and/or continuous monitoring and recording of:

  • . Blood pressure and pulse rate of adult, pediatric and neonatal patients;
  • . Functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric and neonatal patients;
  • . Additionally the Masimo rainbow® Pulse CO-Oximetry™ and Masimo SET technology provides noninvasive monitoring of carboxyhemoglobin saturation (SpCO®) and/or respiration (RRa), Other information displayed on the 740 SELECT with the Masimo Rainbow SET option includes: Signal IQ Waveform, Low Signal IQ (Low SIQ), Perfusion Index (PI), and/or Pleth Variability Index (PVI®) indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused:
  • . Carbon Dioxide concentration of the expired and inspired breath and respiration rate and FICO2 (fractional concentration of inspired CO2) of adult, pediatric, infant or neonatal patient and included, with the Oridion option, an Integrated Pulmonary Index (IPI);

{2}------------------------------------------------

  • . Intermittent predictive body temperature (oral, axillary, rectal) of adult, pediatric and neonatal patients:
  • . Infrared (over the temporal artery) measurement of body temperature of people of all ages.

740 SELECT Monitor Technologies Compared to Predicate Devices

The 740 SELECT compares substantially to the cited predicate device(s) in that it uses fundamentally the same technologies, adopted from the parameters manufacturer, without alteration. The non-invasive blood pressure component is the CASMED MAXIQ™ an oscillometric instrument having neither a stethoscope nor a microphone and does not need the quiet environment necessary to detect the auscultatory sounds. They work on the principle that when the artery opens during a portion of the pressure cycle, an oscillation is superimposed on the pressure inside the cuff due to a tiny enlargement of the circumference of the limb caused by the surge of blood under the cuff. The amplitude of the oscillometric signals change over the course of the deflation of the cuff. Oscillometric devices look for oscillation amplitude of certain percentages of the maximum amplitude at mean arterial pressure (MAP), defining one percentage as the systolic point and another as the diastolic point. Alternatively, a combination of the amplitude, the slope of the increase or decrease, and some other complex factors are used to find these points.

The 740 SELECT pulse oximeter can include either of two industry leading components specific to the customer's requirement. Masimo rainbow Pulse CO-Oximetry™ and Masimo SET® or Covidien (Nellcor's) OxiMax technology. The pulse oximeter parameter (%SpO2) determines arterial oxyhemoglobin saturation by measuring the absorption of red and infrared light passing through the tissue. Changes in absorption caused by pulsations of blood in the vascular bed are used to determine arterial saturation and pulse rate. Additional leading edge functions, as offered by the pulse oximeters manufacturer, are made available to the customer with the 740 SELECT.

The FILAC™ 3000 temperature parameter has the capability of taking temperature in either normal (predictive) or monitor mode. In the normal mode, the thermometer's microprocessor "predicts" body temperature in about four (4) seconds for Oral temperatures, about ten (10) seconds for Axillary temperatures and in about fifteen (15) seconds for Rectal temperatures. Exergen's TemporalScanner is a hand held device that measures the skin temperature over the temporal artery. The Exergen TemporalScanner Thermometer has the capability of taking Temporal Artery (TA) Temperature in about 3 seconds. The Exergen Temp scanner has the ability to measure the patient's Arterial temperature in ºF or °C.

For the CO2 parameter, the 740 SELECT allows for the external attachment of a choice of two manufacturers solutions. Offered are the Masimo -Phaseln™ (Masimo) Sidestream Gas Analyzer (ISA) or the Mainstream Gas Analyzer (IRMA). The other choice is the Oridion Micro Pod™ capnography module.

Non-Clinical Performance Testing to Demonstrate Substantial Equivalence

The 740 SELECT has successfully undergone performance, safety, electromagnetic, and environmental testing to ensure it has been found to be substantially equivalent to the primary and secondary predicate devices. In addition to the above laboratory tests, CAS has conducted a full program of individual hardware, software and systems verification and validation studies of the monitor and accessories.

{3}------------------------------------------------

Clinical Testing to Show Substantial Equivalence

The 740 SELECT has successfully undergone clinical validation for the indicated use. The premarket notification cites clinical validation reports for the NIBP, and the OEM's clinical work for the pulse oximeters, CO2 and the temperature functions.

Conclusions Drawn from Clinical and Non-Clinical Testing

Clinical evaluation, safety testing, software and systems validation demonstrate the 740 SELECT is substantially equivalent to the predicate device(s).

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, overlaid on a stylized human figure. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 13, 2014

CAS Medical Systems, Inc. Ron Jeffrey Director, Regulatory Affairs 44 East Industrial Rd. Branford, Connecticut 06405

K140430 Re:

Trade/Device Name: 740 Select Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (With Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: May 5, 2014 Received: May 6, 2014

Dear Mr. Jeffrey,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{5}------------------------------------------------

Page 2 - Ron Jeffrey

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health ... ..

Enclosure

{6}------------------------------------------------

K140430

Indications for Use

510(k) Number:

Device Name: 740 SELECT

The 740 SELECT series of monitors is indicated for use as a portable, multi-parameter, variable aculty device for use by health care professionals, clinicians and medically qualified personnel, in a variety of clinical environments and hospital departments, for non-invasive spot checking, and/or continuous monitoring and recording of:

  • Blood pressure and pulse rate of adult, pediatric and neonatal patients; .
  • Functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric and . neonatal patients;
  • Additionally the Masimo Rainbow SET technology provides noninvasive monitoring of . carboxyhemoglobin saturation (SpCO) and/or respiration (RRa). Other information displayed on the 740 SELECT with the Masimo Rainbow SET option includes: Signal IQ Waveform, Low Signal IQ (Low SIQ), Perfusion Index (PI), and/or Pleth Variability Index (PVI) indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused;
  • Carbon Dioxide concentration of the expired and inspired breath and respiration rate of adult, . pediatric, infant or neonatal patient and included, with the Oridion option, an Integrated Pulmonary Index (IPI):
  • Intermittent predictive body temperature (oral, axillary, rectal) of adult, pediatric and neonatal ● patients;
  • Infrared (over the temporal artery) measurement of body temperature of people of all ages. .
Prescription Use ✔AND/OROver-the-Counter Use _
(Part 21 CFR 801 Subpart D)(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Date:
2014.06.13
09:46:52 -04'00'

Page 1 of 1

0.41001

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).