(30 days)
Not Found
No
The document explicitly states "Mentions AI, DNN, or ML: Not Found" and the device description focuses on standard imaging and pressure measurement technologies. While it mentions "semiautomatically visualize boundary features and perform spectral analysis of RF ultrasound signals," this does not necessarily indicate AI/ML and could be achieved through traditional signal processing techniques.
No.
The device is used for diagnostic imaging and pressure measurement, not for treatment. Its stated uses include evaluating vascular morphology, providing images of vessel lumen and wall structures, and measuring intravascular blood pressure.
Yes
The device is explicitly stated to be used for "qualitative and quantitative evaluation of vascular morphology," "adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures," "qualitative blood flow information," and "diagnostic ultrasound imaging examinations." The pressure feature is also intended to "measure intravascular blood pressure during diagnostic angiography and/or interventional procedures." These uses clearly indicate a diagnostic purpose.
No
The device description explicitly details hardware components such as a CPU, connection box, cables, patient interface modules (PIMs), and a control pad with a touch screen display. It also mentions different configurations (tower/portable and integrated) and operating modes (IVUS imaging and FFR pressure), all of which involve physical hardware. The performance studies also include tests for electrical safety, electromagnetic compatibility, packaging, temperature/humidity, drop tests, and mounting load, further indicating the presence of hardware.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The Volcano S5™ series Intravascular Ultrasound Imaging and Pressure Systems are used for imaging and pressure measurement within blood vessels. This is a direct measurement and visualization of structures and physiological parameters in vivo (within the living body), not an analysis of a sample taken in vitro (outside the living body).
- Intended Use: The intended use clearly describes the evaluation of vascular morphology and pressure within the arteries and vessels. This is a diagnostic procedure performed directly on the patient.
Therefore, the device falls under the category of medical devices used for imaging and physiological measurement in vivo, not In Vitro Diagnostics.
N/A
Intended Use / Indications for Use
The Volcano S5™/ s5i® Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.
ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.
VH® IVUS intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of RF ultrasound signals of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations.
The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.
Rotational 45MHz feature is intended for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vasculature. As an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. The pullback feature of the PIMr withdraws the imaging core within the protective sheath for a maximum of 15 cm.
Product codes (comma separated list FDA assigned to the subject device)
IYO, DSK, DSA
Device Description
The Volcano s5 Series Intravascular Ultrasound Imaging and Pressure Systems are currently available in 2 configurations: (1) a tower or a portable model, (2) an integrated model.
The Volcano S5 Intravascular Ultrasound Imaging and Pressure Systems are the integrated configurations that are integrated in the catheterization (cath) laboratory, meaning that the CPU is located outside the cath lab and the controls and accessories are cabled in a trench under the floor into the cath lab for use on the patient. Cables from the CPU enter the cath lab through trench and are consolidated through the Connection Box located in the cath lab which then distributes connections to all the S5 accessories and bedside peripherals.
The S5 tower systems are the tower/portable (roll-around or mobile) versions of the integrated system. These systems can be rolled into the cath lab itself and the accessories and bedside peripherals directly connect to the system.
There are two (2) operating modes available on both the integrated as well as the tower models of the Volcano Intravascular Ultrasound Imaging and Pressure Systems, namely: (1) the Intravascular Ultrasound (IVUS) imaging mode and (2) the Fractional Flow Reserve (FFR) pressure mode.
When operating the IVUS mode, the system console gathers and displays high-resolution intraluminal images that can be analyzed both quantitatively. When operating in pressure mode, the system acquires intraluminal data from a pressure guidewire while simultaneously taking aortic pressure data from the established ECG/EKG catheterization lab equipment. Catheters and guidewires are connected to the system via the Patient Interface Modules (PIMs).
As an accessorv to the S5 Systems, the Volcano CORE™ Control Pad is intended to be a secondary controller in the Volcano S5 Intravascular Ultrasound Imaging and Pressure Systems (integrated and mobile systems). It is intended to be used in the exam room, in the sterile field at the bedside in the exam room, in the control room, or on the mobile cart. Images, data, and case navigation controls are relayed to and from the CORE™ Control Pad display via the system central processing unit (CPU). These images, data and controls are presented in a graphical user interface (GUI) displayed on the touch screen of the CORE™ Control Pad. The GUI displayed on the screen of the CORE™ Control Pad will represent, but may not duplicate exactly, the intravascular ultrasound images displayed on the main system monitor. The relayed case navigation controls are intended to allow the user to navigate IVUS and FFR cases and to make measurements on intravascular ultrasound images that are presented on the primary system display. The CORE™ Control Pad touch screen display is an adjunct to the main display on the Volcano CORE™ Series Intra-vascular Imaging and Pressure System. The CORE™ Control Pad is not intended as a standalone diagnostic tool.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Intravascular Ultrasound (IVUS)
Anatomical Site
coronary arteries and vessels of the peripheral vasculature; all blood vessels, including coronary and peripheral arteries.
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Applicable testing was performed as require by the Quality System to evaluate the modification to the Volcano S5 Systems. The following tests were conducted:
- Software Verification and Validation
- Simulated Use Validation
- Electrical Safety
- Electromagnetic Compatibility
- Packaging Validation
- Extreme Temperature and Humidity
- Drop Test
- Mounting Load
- Tensile Load CCP Pig Tail Cable
- Acoustic Noise Level Test CCP
- Reliability HALT (Highly Accelerated Life Testing)
- Mean Time Between Failure (MTBF)
The test results were found to be acceptable by the respective test plans and protocols.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1560 Ultrasonic pulsed echo imaging system.
(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
510(k) SUMMARY
The 510(k) Summary is submitted as required by section 807.92(a)
| SPONSER: | Volcano Corporation
3721 Valley Centre Dr., Suite 500
San Diego, CA 92130 | | |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|--|
| CONTACT/
SUBMITTER: | Marcus Garcia
Regulatory Affairs Specialist
Volcano Corporation
1 Fortune Dr.
Billerica, MA 01821 | Phone: 978-439-3312
Fax: 978-262-0035
email: mgarcia@volcanocorp.com | |
| DATE SUBMITTED: February 4, 2014 | | | |
| DEVICE: | s5™ / s5i® Intravascular Ultrasound Imaging and Pressure Systems | | |
| Trade Name: | Volcano CORETM Control Pad, Accessory to the Volcano s5™/s5i®
Intravascular Ultrasound Imaging and Pressure Systems | | |
| Common Name: | Ultrasonic pulsed echo imaging system | | |
Classification and Product Codes:
CFR Number | Class | Product Code |
---|---|---|
892.1560 Ultrasonic pulsed echo imaging system | II | IYO |
870.1110 Blood Pressure Computer | II | DSK |
870.2900 Patient Transducer and Electrical Cable | II | DSA |
PREDICATE DEVICE: K133641, Volcano CORE™ and CORE™ Mobile Series Precision Guided Therapy Systems
DEVICE DESCRIPTION:
The Volcano s5 Series Intravascular Ultrasound Imaging and Pressure Systems are currently available in 2 configurations: (1) a tower or a portable model, (2) an integrated model.
The Volcano S5 Intravascular Ultrasound Imaging and Pressure Systems are the integrated configurations that are integrated in the catheterization (cath) laboratory, meaning that the CPU is located outside the cath lab and the controls and accessories are cabled in a trench under the floor into the cath lab for use on the patient. Cables from the CPU enter the cath lab through trench and are consolidated through the Connection Box located in the cath lab which then distributes connections to all the S5 accessories and bedside peripherals.
1
K140291 page 2 of 4
The S5 tower systems are the tower/portable (roll-around or mobile) versions of the integrated system. These systems can be rolled into the cath lab itself and the accessories and bedside peripherals directly connect to the system.
There are two (2) operating modes available on both the integrated as well as the tower models of the Volcano Intravascular Ultrasound Imaging and Pressure Systems, namely: (1) the Intravascular Ultrasound (IVUS) imaging mode and (2) the Fractional Flow Reserve (FFR) pressure mode.
When operating the IVUS mode, the system console gathers and displays high-resolution intraluminal images that can be analyzed both quantitatively. When operating in pressure mode, the system acquires intraluminal data from a pressure guidewire while simultaneously taking aortic pressure data from the established ECG/EKG catheterization lab equipment. Catheters and guidewires are connected to the system via the Patient Interface Modules (PIMs).
As an accessorv to the S5 Systems, the Volcano CORE™ Control Pad is intended to be a secondary controller in the Volcano S5 Intravascular Ultrasound Imaging and Pressure Systems (integrated and mobile systems). It is intended to be used in the exam room, in the sterile field at the bedside in the exam room, in the control room, or on the mobile cart. Images, data, and case navigation controls are relayed to and from the CORE™ Control Pad display via the system central processing unit (CPU). These images, data and controls are presented in a graphical user interface (GUI) displayed on the touch screen of the CORE™ Control Pad. The GUI displayed on the screen of the CORE™ Control Pad will represent, but may not duplicate exactly, the intravascular ultrasound images displayed on the main system monitor. The relayed case navigation controls are intended to allow the user to navigate IVUS and FFR cases and to make measurements on intravascular ultrasound images that are presented on the primary system display. The CORE™ Control Pad touch screen display is an adjunct to the main display on the Volcano CORE™ Series Intra-vascular Imaging and Pressure System. The CORE™ Control Pad is not intended as a standalone diagnostic tool.
INDICATIONS FOR USE
The Volcano S5™/ s5i® Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and
2
vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.
ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.
VH® IVUS intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of RF ultrasound signals of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations.
The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.
Rotational 45MHz feature is intended for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vasculature. As an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. The pullback feature of the PIMr withdraws the imaging core within the protective sheath for a maximum of 15 cm.
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:
The proposed device is identical to the currently marketed device except for the addition of a new optional accessory, the Volcano CORE™ Control Pad. The Volcano CORE™ Control Pad acts as secondary controller in the Volcano S5 Intravascular Ultrasound Imaging and Pressure Systems (integrated and mobile systems). The technological characteristics, fundamental scientific technology, and indications for use remain unchanged.
PERFORMANCE DATA:
Applicable testing was performed as require by the Quality System to evaluate the modification to the Volcano S5 Systems. The following tests were conducted:
- . Software Verification and Validation
- . Simulated Use Validation
- . Electrical Safety
3
K140291 page 4 of 4
- Electromagnetic Compatibility .
- Packaging Validation .
- . Extreme Temperature and Humidity
- . Drop Test
- . Mounting Load
- Tensile Load CCP Pig Tail Cable .
- Acoustic Noise Level Test CCP .
- Reliability HALT (Highly Accelerated Life Testing) .
- Mean Time Between Failure (MTBF) .
The test results were found to be acceptable by the respective test plans and protocols.
Biocompatibility and sterilization testing was not required as the proposed accessory does not come in contact with the patient or any fluid path.
Conclusion:
Completion of these tests concluded that the proposed Volcano S5 Systems are substantially equivalent to the predicate device.
4
Public Health Service
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Image /page/4/Picture/3 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three overlapping shapes that resemble a person with outstretched arms.
March 7, 2014
Volcano Corporation Marcus Garcia 1 Fortune Dr Billerica, MA 01821 US
Re: K140291
Trade/Device Name: Volcano Core Control Pad, Accessory to the Volcano S5/S5i Intravascular Ultrasound Imaging & Pressure Systems Regulation Number: 21 CFR 892.1560 Regulation Name: System, Imaging, Pulsed Echo, Ultrasonic Regulatory Class: Class II Product Code: IYO Dated: February 4, 2014 Received: February 5, 2014
Dear Marcus Garcia,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
5
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
for
Sincerely yours,
Linda J. Ricci-S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
6
K140291
Indications for Use
510(k) Number (if known):
Device Name: Volcano s5TM/s5i® Intravascular Ultrasound Imaging and Pressure Systems
Indications for Use:
The Volcano S5™ sories Intravascular Ultrasound Imaging and Pressure Systems is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.
ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.
VH® IVUS intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semiautomatically visualize boundary features and perform spectral analysis of RF ultrasound signals of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations.
The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.
Rotational 45MHz feature is intended for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vasculature. As an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. The pullback feature of the PMr withdraws the imaging core within the protective sheath for a maximum of 15 cm.
Over-the Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Page 1 of