K Number
K140189
Date Cleared
2014-06-05

(132 days)

Product Code
Regulation Number
876.1500
Panel
SU
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EndoWrist® Vessel Sealer is a bipolar electrosurgical instrument for use with the da Vinci Xi Surgical System and the ERBE VIO dV electrosurgical generator. It is intended for grasping and for bipolar coagulation and mechanical transection of vessels up to 7 mm in diameter and tissue bundles that fit in the jaws of the instrument. The EndoWrist Vessel Sealer has not been shown to be effective for tubal sterilization or tubal coagulation procedures, and should not be used for these procedures.

Device Description

The EndoWrist Vessel Sealer is a sterile, single-use (disposable), 8 mm instrument with an integrated cord that connects to the instrument housing and an Erbe VIO dV generator. The EndoWrist Vessel Sealer device consists of a distal wristed end effector and a proximal housing connected by a tubular shaft. The housing contains mechanisms to actuate the end effector when attached to the da Vinci Model 154000 Surgical System. An integrated cord attached to the housing is connected to a receptacle in the IESU. An electrode sealing surface and a cutting blade within the jaws of the instrument enable sealing of vessels and cutting of sealed vessels and other tissues. The sealing and cutting functions are controlled using the da Vinci Model IS4000 Surgical System foot pedals.

A slider on the housing allows the user to manually open and close the instrument jaws for intraoperative cleaning and grip release. A lock button on the housing allows the user to hold the jaws open for intraoperative cleaning. Two release buttons on the housing allow removal of the Vessel Sealer from the sterile adapter on the da Vinci Model IS4000 System instrument arm.

The ERBE VIO dV generator provides the high frequency (radio frequency) electrical current for tissue sealing.

AI/ML Overview

This document describes the EndoWrist® Vessel Sealer, a surgical instrument intended for grasping, bipolar coagulation, and mechanical transection of vessels up to 7mm in diameter and tissue bundles. The device is for use with the da Vinci Xi Surgical System and the ERBE VIO dV electrosurgical generator.

1. A table of acceptance criteria and the reported device performance

The provided text does not explicitly define specific numerical acceptance criteria. Instead, it states that performance was evaluated against "design output meets the design input requirements" and demonstrated "substantial equivalence" to the predicate device, the EndoWrist® One Vessel Sealer.

However, based on the descriptions of the tests, we can infer some performance aspects that were evaluated:

Performance AspectAcceptance Criteria (Inferred from testing)Reported Device Performance
Form/Fit/FunctionMeet architectural and functional requirements (e.g., physical dimension, range of motion, cutting/dissecting capabilities, creepage and clearance, force withstand, intuitive motion, electrical, programming, and general features)."Architectural and functional requirements...were tested." (Implied successful achievement since substantial equivalence was claimed).
Durability (Single-use)Withstand use over its useful life (single use)."Reliability testing was conducted on the subject device to confirm the requirement for the disposable instrument to withstand use over its useful life (single use)." (Implied successful achievement).
Seal Performance (Vessels/Tissue Bundles)On tissue bundles, have seal performance comparable to the predicate device."Burst pressure testing was conducted on the device to demonstrate that the subject device...has seal performance on tissue bundles comparable to the predicate device...and that the subject device effectively transects tissue bundles." (Implied successful achievement).
Transection Performance (Vessels/Tissue Bundles)Effectively transect tissue bundles."Burst pressure testing...demonstrate...that the subject device effectively transects tissue bundles." (Implied successful achievement).
Clinical Efficacy (Tissue Interaction/Sealing in vivo)Meet user needs and intended use, including evaluation of tissue effect and tissue interaction in a simulated clinical environment. Seal and transection performance on tissue bundles in vivo comparable to the predicate device."Clinical validation to confirm the device meets the user needs and intended use." "In-vivo tissue bundle testing was also conducted to assess the device seal and transection performance on tissue bundles and to demonstrate that the subject device has seal performance comparable to the predicate device." (Implied successful achievement).
Usability/SafetyEnable intended users to perform essential and high-risk tasks safely and effectively; validate that use-related risks are mitigated to acceptable levels; assess overall ease of use; evaluate for unknown use-related risks."A summative usability validation study was conducted with sixteen (16) teams of users (Surgeons and Operating Room (OR) staff)... The study assessed the following: ... ensure intended users could perform essential and high risk tasks in the expected use environments in a safe and effective manner; validate that use-related risks have been mitigated to acceptable levels of residual risk; assess the overall ease of use and usability of the EndoWrist Vessel Sealer; and evaluate whether the design introduced any previously unknown use-related risks." (Implied successful achievement based on the overall substantial equivalence determination).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Bench Testing: No specific sample size is given for architectural, functional, or reliability testing. Burst pressure testing also lacks a specified sample size.
  • Animal and Cadaver Testing:
    • Clinical validation (porcine model): No specific sample size (number of animals) is stated.
    • In-vivo tissue bundle testing (living animal model - porcine): No specific sample size (number of animals or tissue bundles) is stated.
  • Human Factors Evaluation:
    • Summative usability validation study: Sixteen (16) teams of users (Surgeons and Operating Room (OR) staff).
  • Data Provenance: Not explicitly stated, but the submission is to the U.S. FDA, suggesting the studies were likely conducted in the US or in a manner compliant with US regulatory guidelines. The studies appear to be prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Bench Testing: The ground truth for bench tests (e.g., physical dimensions, range of motion, electrical properties) is typically established by engineering specifications and objective measurements, not by experts in the same way clinical ground truth is established.
  • Animal and Cadaver Testing: For clinical validation and in-vivo tissue bundle testing, the "ground truth" would be the observed acute effects (e.g., successful sealing, transection, tissue effect) as assessed by the researchers and potentially veterinary surgeons involved in the studies. The text mentions "clinical validation to confirm the device meets the user needs and intended use," implying expert assessment of clinical outcomes, but specific numbers or qualifications of these experts are not provided.
  • Human Factors Evaluation: The evaluators/observers in the usability study would assess the user performance against predefined tasks and safety criteria. The participants themselves (16 teams of Surgeons and OR staff) represent the "experts" whose interaction with the device is being evaluated, rather than establishing a ground truth for a diagnostic output.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe any formal adjudication method (like 2+1 or 3+1 consensus) for establishing ground truth in any of the testing types. It suggests that performance was observed and evaluated by the study teams.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This device is a surgical instrument (vessel sealer), not a diagnostic or AI-assisted interpretation device, so this type of study is not applicable. The human factors study focused on user interaction and usability of the instrument itself, not on improving human "reading" or diagnostic accuracy.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm performance study was not done. The device is a physical surgical instrument, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Bench Testing: Engineering specifications, objective physical and electrical measurements, and mechanical performance standards.
  • Animal and Cadaver Testing: Direct observation of surgical outcomes (e.g., successful sealing, lack of burst, effective transection), tissue effects (likely assessed visually and potentially histopathologically, though not explicitly stated), and clinical performance in simulated procedures. This is a form of direct observation/outcomes data in a preclinical model.
  • Human Factors Evaluation: Observational data of user performance on critical tasks, error rates, and subjective feedback regarding ease of use, all assessed against predefined usability and safety criteria.

8. The sample size for the training set

This is not applicable as the device is a mechanical surgical instrument, not an AI or machine learning algorithm requiring a training set.

9. How the ground truth for the training set was established

Not applicable (see point 8).

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.