(25 days)
The Medos hilite 2800 & 2400 LT Hollow Fiber Oxygenators are indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from human blood. It is designed to operate at a blood flow rate of 0.5 to 2.8 liters per minute for the hilite 2800 and 0.5 to 2.4 liters per minute for the hilite 2400 LT for periods of up to six (6.0) hours.
The Medos Hilite 2800 & 2400 LT Oxygenators consist of a hollow fiber membrane oxygenator and extracorporeal heat exchanger. The Medos Hilite 2800 hollow fiber membrane consists of a polypropylene gas permeable mat. The Medos Hilite 2400 LT hollow fiber membrane consists of a polymethylpentene plasma tight mat. The unique mat design increases the interaction between blood and gas, creating a highly efficient blood oxygenator. The heat exchanger consists of a polyester non-porous hollow fiber configured heat exchanger as the primary element to affect heat exchange. This element is encased by a polycarbonate housing, which directs the blood around the outside of the fibers while water flows through the inner lumen of the fibers and therefore effects heat exchange while minimizing priming volume.
The device allows for trapping and removal of air. Oxygenated blood is delivered to the patient through the tubing and appropriate cannula. Blood flow is driven by a roller pump or centrifugal pump connected through the tubing. The Medos Hilite 2800 and 2400 LT Oxygenators may be purchased separately or pre-connected with tubing and other components of an extracorporeal circuit.
The provided text is a 510(k) summary for the Medos Hilite 2800 & 2400 LT Oxygenators. It describes the device, its intended use, and argues for its substantial equivalence to a predicate device. However, it does not contain the kind of detailed information about acceptance criteria and a specific study that proves the device meets those criteria, which would be typically found for an AI/ML-based medical device.
Therefore, it is not possible to fill in all the requested information as the document is for a traditional medical device (blood oxygenator) and not an AI/ML diagnostic tool.
Here's an attempt to address the points based on the available information, noting where an AI/ML context does not apply:
Acceptance Criteria and Device Performance:
The document describes "Performance Testing" which is generally for demonstrating functional requirements and safety. It doesn't present a table of specific numerical acceptance criteria with corresponding reported device performance values in the way you'd expect for, say, sensitivity or specificity of a diagnostic algorithm.
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria Category | Reported Device Performance | Comments |
|---|---|---|
| Material & Integrity Tests (Due to Change in Potting Material) | The document lists tests performed but does not provide specific numerical criteria or results for each. These are presented as tests performed to ensure compliance after a material change. | |
| Blood Pathway Integrity | Performed | Outcome implied to be satisfactory for substantial equivalence. |
| Heat Exchanger Fluid Pathway Integrity | Performed | Outcome implied to be satisfactory for substantial equivalence. |
| Chemical Analysis | Performed on specific materials (Macroplast CR 3502/CR 4100, Hilite 7000, Hilite 7000 LT) | Outcome implied to be satisfactory for substantial equivalence. |
| Biological Safety - Toxicology | ||
| Evaluation of Biological Safety - toxicology | Performed on Macroplast CR 3502/CR 4100 | Outcome implied to be satisfactory for substantial equivalence. |
| Cytotoxicity (L 929-Proliferation) | Performed on Macroplast CR 3502/CR 4100, EO sterilized | Outcome implied to be satisfactory for substantial equivalence. |
| Hemolysis (elution method) | Performed on Macroplast CR 3502/CR 4100, EO sterilized | Outcome implied to be satisfactory for substantial equivalence. |
| Cytotoxicity | Performed | Outcome implied to be satisfactory for substantial equivalence. |
| Sensitization | Performed | Outcome implied to be satisfactory for substantial equivalence. |
| Intracutaneous Reactivity | Performed | Outcome implied to be satisfactory for substantial equivalence. |
| Acute Systemic Toxicity | Performed | Outcome implied to be satisfactory for substantial equivalence. |
| Pyrogen | Performed | Outcome implied to be satisfactory for substantial equivalence. |
| Hemocompatibility | Performed | Outcome implied to be satisfactory for substantial equivalence. |
Study Details (as they pertain to a traditional medical device submission for substantial equivalence):
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not applicable in the context of an AI/ML test set. The "tests" listed are bench performance and biocompatibility tests on the device materials and components, not a diagnostic test on a data set. No human or animal subject sample size is specified for these tests, nor is there a "test set" in the AI/ML sense.
- Data Provenance: Not applicable. The document focuses on demonstrating physical and biological equivalence, not data analysis from specific patient populations.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This is not a diagnostic device relying on expert-labeled ground truth for algorithm performance validation. The "ground truth" for the performance tests would be established by validated test methodologies and industry standards for medical device materials and function.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This is not a diagnostic device requiring adjudication of human-labeled classifications.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI-assisted diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an algorithm-based device. The "performance" described is the functional and biological safety performance of the physical oxygenator device itself.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Not applicable for AI/ML ground truth. For the physical device, the "ground truth" is implied by adherence to established engineering specifications, material standards, and biocompatibility guidelines. The testing confirms the device's physical and chemical properties and interaction with blood.
8. The sample size for the training set
- Not applicable. This is not an AI/ML device that requires a training set.
9. How the ground truth for the training set was established
- Not applicable.
{0}------------------------------------------------
Special 510(k) Submission Section 2, 510(k) Summary
SPECIAL 510(K) 510(k) Summary For the Medos Medizintechnik AG Medos Hilite 2800 & 2400 LT Oxygenators
FEB 18 2014
l. SUBMITTER/510(k) HOLDER:
Medos Medizintechnik AG Obere Steinfurt 8-10 5222 Stolberg, Germany Telephone: +49 7131 2706 150 Telefax: +49 7131 2706 250 Head of Regulatory Affairs: Heiko Frerichs Email: h.frerichs@medos-ag.com FDA Registration: 3010223691
II. CONTACT PERSON
Leann Christman Novalung, Inc. 3526 West Liberty Rd., Suite 100 Ann Arbor, MI 48103 Telephone: (734) 995-9089 ext. 232 Fax: (734) 995-0840 Email: leann.christman@Novalung.com
III. DEVICE NAME
Proprietary Trade Name: Medos Hilite 2800 & 2400 LT Oxygenators Common/Usual Name: Blood Oxygenator
IV. DEVICE CLASS:
Class II
V. CLASSIFICATION NAME AND CITATION:
Classification Name: Oxygenator, Cardiopulmonary Bypass Classification regulation: Sec. 870.4350
VI. PRODUCT CODES:
DTZ
VII. PREDICATE DEVICES
K090450, Medos Hilite 2800 & 2400 LT Oxygenators
PRODUCT DESCRIPTION VIII.
{1}------------------------------------------------
The Medos Hilite 2800 & 2400 LT Oxygenators consist of a hollow fiber membrane oxygenator and extracorporeal heat exchanger. The Medos Hilite 2800 hollow fiber membrane consists of a polypropylene gas permeable mat. The Medos Hilite 2400 LT hollow fiber membrane consists of a polymethylpentene plasma tight mat. The unique mat design increases the interaction between blood and gas, creating a highly efficient blood oxygenator. The heat exchanger consists of a polyester non-porous hollow fiber configured heat exchanger as the primary element to affect heat exchange. This element is encased by a polycarbonate housing, which directs the blood around the outside of the fibers while water flows through the inner lumen of the fibers and therefore effects heat exchange while minimizing priming volume.
The device allows for trapping and removal of air. Oxygenated blood is delivered to the patient through the tubing and appropriate cannula. Blood flow is driven by a roller pump or centrifugal pump connected through the tubing. The Medos Hilite 2800 and 2400 LT Oxygenators may be purchased separately or pre-connected with tubing and other components of an extracorporeal circuit.
INDICATIONS OF USE IX.
The Medos Hilite 2800 & 2400 LT Oxygenators are indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from human blood. It is designed to operate at a blood flow rate of 0.5 to 2.8 liters per minute for the Hilite 2000 and 0.5 to 2.4 liters per minute for the Hilite 2400 LT for periods of up to six (6.0) hours.
X. TECHNOLOGICAL CHARACTERISITICS AND SUBSTANTIAL EQUIVALENCE
The Medos Hilite 2800 & 2400 LT Oxygenators are identical to the predicate device in terms of intended use, indications for use, levels of attachment, fundamental scientific technology, materials and surgical technique. Based on the information provided herein, the subject Medos Hilite 2800 & 2400 LT Oxygenators have been demonstrated to be substantially equivalent to the previously cleared Medos Hilite 2800 & 2400 LT Oxygenators (K090450). Please refer to the Table 9-1 for a comparison of the predicate and subject Medos Hilite 2800 & 2400 LT Oxygenators regarding substantial equivalence.
XI. PERFORMANCE TESTING
The Medos Hilite 2800 & 2400 LT Oxygenators have been subjected to extensive safety, performance, and validations prior to release. Final testing for the systems includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications. Safety tests have further been performed to ensure the device complies with applicable industry and safety standards. The following tests were performed due to the change in potting material:
- . Blood Pathway Integrity
- Heat Exchanger Fluid Pathway Integrity .
- Chemical analysis .
- o Macroplast CR 3502 / CR 4100, EO sterilized
- Hilite 7000, PUR Macroplast CR3505/CR4605 o
- Hilite 7000 LT. PUR Macroplast CR3505/CR4605 o
{2}------------------------------------------------
- Hilite 7000, PUR Macroplast CR3505/CR4100 o
- . Evaluation of Biological Safety - toxicology
- o Macroplast CR 3502 / CR 4100, PUR posting material for hollow fibers of oxygenators
- Cytotoxicity, L 929-Proliferation .
- o Macroplast CR 3502 / CR 4100, EO sterilized
- Hemolysis (elution method) .
- o Macroplast CR 3502 / CR 4100, EO sterilized
- . Cytotoxicity
- Sensitization .
- . Intracutaneous Reactivity
- . Acute Systemic Toxicity
- Pyrogen .
- . Hemocompatibility
SUMMARY AND CONCLUSIONS XII.
Medos Medizintechnik AG makes the claim that the Medos Hilite 2800 & 2400 LT Oxygenators are substantially equivalent to the cited predicate in terms of intended use, indications for use, fundamental technology, design characteristics, generic materials of construction, and operational characteristics. As shown in Table 9-1 and the discussion above, the differences between the Medos Hilite 2800 & 2400 LT Oxygenators and cited predicate are minor and raise no new issues of safety or effectiveness. The Medos Hilite 2800 & 2400 LT Oxygenators meet design specifications.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 18, 2014
Medos Medizintechnik AG c/o Leann Christman Novalung, Inc. 3526 West Liberty, Suite 100 Ann Arbor, Mich. 48103
Re: K140177
Trade/Device Name: Medos hilite 2800 & 2400 LT Hollow Fiber Oxygenators Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ, DTR Dated: January 22, 2014 Received: January 24, 2014
Dear Ms. Christman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Leann Christman
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
M. A. Daman
for
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known) K 140177
Device Name: Medos hillte 2800 & 2400 LT Oxygenators
Indications for use:
The Medos hilite 2800 & 2400 LT Hollow Fiber Oxygenators are indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from human blood. It is designed to operate at a blood flow rate of 0.5 to 2.8 liters per minute for the hilite 2800 and 0.5 to 2.4 liters per minute for the hilite 2400 LT for periods of up to six (6.0) hours.
Prescription Device:
Federal Law (US) restricts this device to sale by or on the order of a physician.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: Yes
OR
Over-The-Counter Use: No
ML. Alleman
§ 870.4350 Cardiopulmonary bypass oxygenator.
(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”