K Number
K134021
Manufacturer
Date Cleared
2014-06-10

(162 days)

Product Code
Regulation Number
872.3275
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EsTemp Implant, EsTemp Clear is a non-eugenol temporary cement for luting implantretained crowns, bridges, inlays, onlay and splints.

Device Description

Estemp temporary resin cement series consists of EsTemp Implant & EsTemp Clear. Each is designed to be suitable for various applications such as cementing temporary crowns, bridges, inlays and splints. They are available in an automix dual barrel syringe. It is light-curing temporary filling materials

AI/ML Overview

The provided text describes the 510(k) summary for the SPIDENT EsTemp Implant and EsTemp Clear dental cements. The information focuses on establishing substantial equivalence to a predicate device rather than presenting a study with specific acceptance criteria and detailed device performance metrics in the format requested.

Here's an analysis based on the provided text, while acknowledging that a direct "study" as typically understood for AI-driven devices (with acceptance criteria, test sets, ground truth, etc.) is not present for this type of medical device (dental cement).

The submission relies on demonstrating substantial equivalence to a predicate device (Premier Implant Cement, K033309) through various performance characteristics, primarily mechanical and chemical properties, and biocompatibility.

Absence of Information: The provided text does not include information on:

  • A test set size or data provenance in the context of an AI/algorithm study.
  • Number of experts for ground truth or their qualifications.
  • Adjudication method for a test set.
  • MRMC comparative effectiveness study.
  • Standalone (algorithm only) performance.
  • Sample size for a training set or how ground truth was established for it.

1. Table of Acceptance Criteria and Reported Device Performance

Instead of traditional "acceptance criteria" and "device performance" in the context of an AI study, the document presents performance characteristics and biocompatibility test results, indicating passing relative to established standards or equivalence to the predicate device.

Performance CharacteristicAcceptance Criteria (Implied by Predicate Equivalence/Standards)Reported Device Performance (EsTemp Implant, EsTemp Clear)
Chemistry of setting reactionEquivalent to predicateEquivalent to predicate
Proportion of powder to liquidEquivalent to predicateEquivalent to predicate
Maximum solubilityEquivalent to predicateEquivalent to predicate
DisintegrationEquivalent to predicateEquivalent to predicate
Dimensional ChangeEquivalent to predicateEquivalent to predicate
Working and setting TimeEquivalent to predicateEquivalent to predicate
Bonding StrengthEquivalent to predicateEquivalent to predicate
Optimum Film thicknessEquivalent to predicateEquivalent to predicate
ConductivityEquivalent to predicateEquivalent to predicate
Amount of heat generatedEquivalent to predicateEquivalent to predicate
Compressive StrengthEquivalent to predicateEquivalent to predicate
Biocompatibility TestingAdherence to ISO 10993 and FDA Blue Book Memorandum #95-1
1. CytotoxicityPassPass
2. SensitizationPassPass
3. Irritation/Intracutaneous reactionPassPass
4. Subacute toxicityPassPass
5. GenotoxicityPassPass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable and not provided in the document, as this is not a study of an AI/algorithm. The performance testing refers to physicochemical and biological evaluations of the material itself.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and not provided. The "ground truth" for dental cement performance is established through standardized laboratory testing protocols (e.g., ISO, ASTM standards for dental materials) and validated methods, not expert consensus in the medical image interpretation sense. The biocompatibility tests follow ISO 10993 standards, conducted by qualified labs.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and not provided. Data adjudication methods are typically used in clinical or image-based studies to resolve discrepancies among expert readers, which is not relevant for this type of material performance testing.

5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable and not provided. This device is a dental cement, not an AI diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable and not provided. This device is a dental cement, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the performance evaluation of the dental cement is based on:

  • Standardized laboratory tests: Results of tests for properties like solubility, dimensional change, working/setting time, bonding strength, film thickness, conductivity, heat generation, and compressive strength, evaluated against recognized dental material standards or in comparison to the predicate device.
  • Biocompatibility standards: Confirmation that the device meets the applicable requirements of the FDA Blue Book Memorandum #95-1 and ISO 10993 Biological Evaluation of Medical Devices Part -1.

8. The sample size for the training set

This information is not applicable and not provided. There is no "training set" in the context of this traditional medical device submission.

9. How the ground truth for the training set was established

This information is not applicable and not provided. There is no "training set" for this type of device.

{0}------------------------------------------------

SPIDENT

SPIDENT Co., Ltd. 203&312. Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821

Tel: +82(32)819-4570 Fax : +82(32)819-4572

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c).

Date: November 22, 2013

  1. Company and Correspondent making the submission:
Company
NameSPIDENT Co., Ltd.
Address203&312, Korea Industrial Complex, 722,Gojan-Dong, Namdong-Gu, Incheon, Korea405-821
PhoneFax+82(32)819-4571+82(32)819-4572
ContactJemo Ahn
    1. Device:
      Proprietary Name - EsTemp Implant. EsTemp Clear Common Name - Dental Cement Classification Name - Cement, Dental
    1. Predicate Device:
      Premier Implant Cement, K033309
    1. Classifications Names & Citations: EMA, 872.3275
    1. Description:

Estemp temporary resin cement series consists of EsTemp Implant & EsTemp Clear. Each is designed to be suitable for various applications such as cementing temporary crowns, bridges, inlays and splints. They are available in an automix dual barrel syringe. It is light-curing temporary filling materials

The EsTemp Implant, EsTemp Clear is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. They are substantially equivalent in design, function and intended use to the predicate devices.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "SPIDENT" in bold, black letters. The letters are enclosed in a square border that is open at the top and bottom. The font is sans-serif and appears to be a logo or brand name. The image is simple and clean, with a focus on the text.

6. Indication for use:

EsTemp Implant, EsTemp Clear is a non-eugenol temporary cement for luting implantretained crowns, bridges, inlays, onlay and splints.

7. Performance

Testing performed including Chemistry of the setting reaction. Proportion of powder to liquid. Maximum solubility and disintegration, Dimensional Change, Working and setting Time, Bonding Strength, Optimum Film thickness, Conductivity & Amount of heat generated during setting. Compressive Strength use which demonstrated that the EsTemp Implant, EsTemp Clear is equivalent to the predicates in specifications and performance characteristics.

Therefore, it was considered that the subject device was as effective as and performs as good as the predicate device.

In conclusion, it can be said that the effectiveness and performance of the subject device are substantially equivalent to those of the predicate device.

8. Biocompatibility

Biocompatibility testing confirmed that the device meets the applicable requirements of the FDA Blue Book Memorandum #95-1 entitled Use of International Standards ISO 10993 Biological Evaluation of Medical Devices Part -1: Evaluation and Testing, and is biocompatible. Accordingly, it was considered that the subject device was substantially equivalent in safety to the predicate device.

No.Testing BiocompatibilityResults
1CytotoxicityPass
2SensitizationPass
3Irritation or Intracutaneous reactivityPass
4Subacute toxicityPass
5GenotoxicityPass

9. Review:

EsTemp Implant, EsTemp Clear is the similar device characteristics as the predicate device, the Premier Implant Cement; intended use, material, chemical composition, design and use concept are similar.

EsTemp Implant, EsTemp Clear have the similar mechanical properties as the predicate device; Dimensional Change, Working and setting Time, Solubility, Film thickness and Compressive Strength.

{2}------------------------------------------------

EsTemp Implant, EsTemp Clear has been subjected to extensive safety, performance, and product validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the US regulations and ISO 4049.

Based on the comparison of intended use and technical features, the EsTemp Implant, EsTemp Clear is substantially equivalent to the predicate devices.

    1. Conclusions:
      In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, FDA's "Guidance for the Preparation of Premarket notifications for Dental Composite" and based on the information provided in this premarket notification SPIDENT Co., Ltd. concludes that the EsTemp Implant, EsTemp Clear is safe and effective and substantially equivalent to predicate devices as described herein.
    1. SPIDENT Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA.
      END

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure embracing a three-pronged shape, possibly representing support or assistance.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 10, 2014

SPIDENT Company. Limited. c/o Ms. Lena Pak SPIDENT. USA. Incorporated 2115 Linwood Avenue, 50 Floor Fort Lee, NJ 07024

Re: K134021

Trade/Device Name: EsTemp Implant, EsTemp Clear Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: 872.3275 Product Code: EMA Dated: March 5, 2014 Received: March 7, 2014

Dear Ms. Pak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Pak

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

MarySBohner-S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Submission ~ EsTemp implant, EsTemp Clear

510(k) Number K 134021

Device Name: EsTemp Implant, EsTemp Clear

Indication for use: EsTemp Implant, EsTemp Clear is a non-eugenoi temporary cement for luting implantretained crowns, bridges, inlays, onlay and splints.

AND/OR Over-The-Counter Use _ Y Prescription Use______________________________________________________________________________________________________________________________________________________________ (Per 21CFR807 Subpart C) (Per 21CFR801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sheena A Green
2014.06.10 08:39:25 -04'00'

0.00023

SPIDENT CO., LTD.

§ 872.3275 Dental cement.

(a)
Zinc oxide-eugenol —(1)Identification. Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.(b)
Dental cement other than zinc oxide-eugenol —(1)Identification. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class II.