K Number
K133914
Manufacturer
Date Cleared
2014-04-03

(101 days)

Product Code
Regulation Number
892.5050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VeriSuite patient position verification and correction of the patient's position during a radiotherapy treatment with external beams or charged particles. It is based on stereoscopic X-ray images and DRRs calculated from a CT image series of the treatment region of the patient and information from the treatment planning.

Device Description

VeriSuite is an image processing system for verification and correction of the patient position during a radiation therapy treatment. The verification or correction is performed by a comparison of X-ray images that are acquired during the treatment with DRRs (digital reconstructed radiographs) calculated from a CT image series of the patient and information from the radiation therapy planning. The correction can also be based on fiducial, radio-opaque markers that are implanted in the patient.

With the fluoroscopy extension it is possible to observe the position of the patient and the treatment area during the treatment. Mainly the movements of the patient caused by breathing are in the focus. This information can be used to interrupt the treatment by the user.

VeriSuite is a system of devices consisting of the VeriSuite software and a number of hardware devices.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device (VeriSuite) and primarily focuses on regulatory approval and substantial equivalence to predicate devices, rather than detailed performance studies or specific acceptance criteria. Therefore, much of the requested information regarding acceptance criteria, study design, and performance metrics is not explicitly present in this document.

However, I can extract what is available and indicate what is missing.

1. Table of Acceptance Criteria and Reported Device Performance

This information is not provided in the document. The submission focuses on functional equivalence to predicate devices and software verification/validation, but not specific quantitative performance acceptance criteria or results of such tests in a clinical setting.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document. The document mentions "Non-clinical verification and validation software tests" but does not detail the size or nature of the test set used for these tests, nor the provenance of any data used.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not provided in the document.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

This information is not provided in the document. There is no mention of a comparative effectiveness study involving human readers with and without AI assistance.

6. Standalone (Algorithm Only) Performance Study

The document mentions "Non-clinical verification and validation software tests were conducted to confirm that VeriSuite 1.8 with fluoroscopy meets its intended use and is safe and effective." This implies foundational testing of the algorithm itself. However, specific standalone performance metrics (e.g., accuracy, precision) in a clinical context are not reported in this summary. The device's function is a "patient position verification system" and "calculation of a correction vector," which, by its nature, involves algorithm-only operation to produce outputs for human review.

7. Type of Ground Truth Used

This information is not provided in the document. For the "Non-clinical verification and validation software tests," the nature of the ground truth would depend on the specific tests conducted (e.g., simulated data with known positions, comparison to known anatomical landmarks in a phantom). Explicit details are missing.

8. Sample Size for the Training Set

This information is not provided in the document. As a "patient position verification system" that calculates DRRs from CT data and identifies correction vectors, the system's core functionality likely relies on computational geometry and image processing rather than a "training set" in the machine learning sense for image recognition. If there were any machine learning components, the training set size is not disclosed.

9. How the Ground Truth for the Training Set Was Established

This information is not provided in the document, and likely less relevant given the description of the device's function.


Summary of available information from the document:

  • Device Function: VeriSuite is an image processing system for verification and correction of patient position during radiation therapy. It compares X-ray images with DRRs calculated from CT series and can also use fiducial markers. The fluoroscopy extension allows observation of patient movement (e.g., breathing) to interrupt treatment.
  • Safety and Effectiveness: "Non-clinical verification and validation software tests were conducted to confirm that VeriSuite 1.8 with fluoroscopy meets its intended use and is safe and effective."
  • Substantial Equivalence: The submission concludes that the device is "as safe and effective and substantially equivalent to the predicate devices described herein." This is the primary "proof" offered in a 510(k) summary for new features.
  • Predicate Devices:
    • VeriSuite 1.8 (K092653) for the core system.
    • EXACTRAC 5.5 / ExacTrac X-RAY 6D (K072506) for the fluoroscopy functionality.
  • Differences from Predicate (EXACTRAC 5.5): VeriSuite does not support optical prepositioning or internal respiratory gating signal display. Minor differences in hardware (higher resolution flat panels, X-ray image, and pixel size for VeriSuite).

In essence, this 510(k) summary acts as a declaration of substantial equivalence based on the stated intended use, technological comparison, and internal verification/validation, rather than a detailed report of a clinical efficacy study with specific statistical acceptance criteria. For such details, one would typically need to review the full 510(k) submission, which is not provided here.

{0}------------------------------------------------

Page 6 of 44

K133914
Page 1 of 4

V. 510(k) Summary of Safety and Effectiveness

A. Submitter

Name:MedCom GmbH
Address:Rundeturmstr. 12Darmstadt, HE 64283Germany
Establishment Reg.:3006579682
Telephone:+49 (6151) 95147-18
Fax:+49 (6151) 95147-20
Contact:Mr. Luca Salvatorelsalvatore@medcom-online.de
Date:December 20, 2013
B. Device
Trade Name:VeriSuite also marketed as VeriSuite 1.8 and VeriSuite-Particle and VeriSuite-Particle 1.8
Common name:Patient position verification system
Classification:Regulatory Class:II
Product Code:LHN / Classification Namesystem, radiation therapy,charged-particle, medical
Subsequent Product Code:IYE / Classification Name:accelerator, linear, medical
CFR Section:892.5050
Panel:Radiology
C. Predicate Devices
Device trade name:510(k) number:VeriSuite 1.8K092653

892.5050

Class II

LHN

MedCom GmbH

Company name:

Classification:

Product code:

Classification Number:

{1}------------------------------------------------

IYE Subsequent code:

Device trade name: EXACTRAC 5.5 / ExacTrac X-RAY 6D 510(k) number: K072506 Company name: BRAINLAB AG 892.5050 Classification Number: Class II Classification: Product code: IYE

D. Reason for Submission

Changed device: Fluoroscopy extension

E. Standards

  1. IEC 61217 (2008), Radiotherapy equipment - Coordinates, movements, and scales Consolidated Edition 1.2. (Radiology)

  2. ISO 14971:2007, Medical devices - Application of risk management to medical devices. (General I (QS/RM))

  3. IEC 62304 First edition 2006-05, Medical device software - Software life cycle processes. (Software/Informatics)

F. Description

VeriSuite is an image processing system for verification and correction of the patient position during a radiation therapy treatment. The verification or correction is performed by a comparison of X-ray images that are acquired during the treatment with DRRs (digital reconstructed radiographs) calculated from a CT image series of the patient and information from the radiation therapy planning. The correction can also be based on fiducial, radio-opaque markers that are implanted in the patient.

With the fluoroscopy extension it is possible to observe the position of the patient and the treatment area during the treatment. Mainly the movements of the patient caused by breathing are in the focus. This information can be used to interrupt the treatment by the user.

VeriSuite is a system of devices consisting of the VeriSuite software and a number of hardware devices:

DeviceType510(k) / registration number
Beam limitingcollimator deviceRalco 302K946320
X-ray generatorSedecal SHF 835 (RF)9617251
X-ray tubesVarian A277 or A2721717855
Flat panel digitalimagerVarian 4030E,K024147

{2}------------------------------------------------

Image /page/2/Figure/2 description: The image shows a system overview diagram. The diagram includes components such as Collimators Ralco R302 1, 2, and 3, Xray Tubes 1, 2, and 3 Varian A277, and Digital Flat Panels Varian 4030E 1 and 2. The diagram also includes a VeriSuite Workstation running Windows XP Professional, with components such as VeriSuite Application, Varian DLLs PaxScan API L04, and Network Cards Intel PRO 1000 GT.

All these hardware devices are legally marketed in the US as listed in the table above.

G. Intended Use

VeriSuite is an active therapeutic medical device for verification of the patient position and calculation of a correction vector for the treatment of tumors during a radiation therapy with photons, electrons (from a linear accelerator) or particles (protons, heavy ions).

The VeriSuite system calculates digitally reconstructed radiographs (DRRs) based on a high-resolution CT data set for a treatment position. With these DRRs and Xray images acquired during the performance of the position verification procedure a correction vector for the patient position can be calculated.

The VeriSuite® system with fluoroscopy extension provides the possibility to a medical user to observe the position of the patient and the treatment area during the treatment, mainly the movements of the patient caused by breathing. This information can be used to interrupt the treatment.

An authorized person must evaluate the correctness of the calculation and approve the result for further usage. The system shall only be used after correct installation in appropriate treatment rooms by trained personnel. Legal regulations especially requlation for the operation of X-ray devices must be regarded.

VeriSuite must not be used for diagnostic purposes.

H. Technological Comparison to Predicate Devices

The changed device VeriSuite 1.8 (build number B641.4, subject of this submission) is substantially equivalent to the predicate device VeriSuite 1.8 (K092653) except the fluoroscopy extension.

{3}------------------------------------------------

Image /page/3/Picture/1 description: This image shows handwritten text on a white background. The text includes "K133914" at the top, followed by "Page 4 of 4" in a cursive style. Additionally, there is a printed line of text at the bottom left that reads "Page 9 of 44". The handwriting appears to be done with a dark ink pen.

The fluoroscopy extension of VeriSuite is substantially equivalent to the fluoroscopy functionality of the predicate device EXACTRAC 5.5 / ExacTrac X-RAY 6D (K072506) in terms of intended use and technology.

In contrast to EXACTRAC 5.5, VeriSuite does not support optical prepositioning and internal respiratory gating signal display. Because of the reproducibility of the patient preposition due to the fixation of the patient, optical prepositioning is not required in the VeriSuite environment. Respiratory gating can be displayed using external devices.

Minor differences exist in the hardware used in terms of resolution of the flat panels, x-ray image and pixel size, which are higher for VeriSuite.

Refer to section XII for a detailed predicate device comparison.

l. Non-clinical Performance Data

Non-clinical verification and validation software tests were conducted to confirm that VeriSuite 1.8 with fluoroscopy meets its intended use and is safe and effective. See section XVI.I for details.

J. Conclusion

Based on the information provided in this Premarket Notification MedCom concludes that VeriSuite 1.8 with fluoroscopy extension is as safe and effective and substantially equivalent to the predicate devices described herein.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 3, 2014

MedCom GmbH % Mr. Luca Salvatore Quality Manager Rundeturmstrabe 12 64283 Darmstadt GERMANY

Re: K133914

Trade/Device Name: VeriSuite 1.8, VeriSuite-Particle 1.8, VeriSuite, VeriSuite-Particle Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: January 6, 2014 Received: January 8, 2014

Dear Mr. Salvatore:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2-Mr. Salvatore

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.lda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D. O'Hara
for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known)

K133914

Device Name VeriSuite

Indications for Use (Describe)

The VeriSuite patient position verification and correction of the patient's position during a radiotherapy treatment with external beams or charged particles. It is based on stereoscopic X-ray images and DRRs calculated from a CT image series of the treatment region of the patient and information from the treatment planning.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Mechal D. O'Hara

FORM FDA 3881 (9/13)

Ser nees ( 80) : 441-6746 Pat Iraka

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond fo, a collection of information unless it displays a currently valid OMB number."

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.