(20 days)
The InToneMV device is a non-implanted electrical stimulator indicated for use in the treatment of female urinary incontinence. It applies electrical stimulation to the pelvic floor musculature and surrounding structures. It is intended for acute and ongoing treatment of mixed urinary incontinence where the following results may improve urinary control: strengthening of pelvic floor muscles and inhibition of the detrusor muscle through reflexive mechanisms. The biofeedback feature can be used for muscle re-education purposes.
The InToneMV device includes three parts: an insertion unit, a hand-held control unit, and a PCbased software application.
- The insertion unit includes (1) an inflation bulb, (2) the inflatable probe, and (3) the flexible tubing connecting the inflation bulb and the probe. The inflatable probe is inserted into the vagina and manually inflated by the patient to ensure a customized fit. Electrical stimulation is delivered via stainless steel electrodes on the probe to induce a contraction of the pelvic floor muscles.
- The external hand-held control unit includes user keys to initiate and control treatment sessions. The control unit is designed to record and store results of the electrical stimulation and patient generated pelvic floor exercises.
- The PC-based software application is utilized by the clinician to program the hand-held control unit, store and display the results of electrical stimulation and patient generated pelvic floor exercises. The application allows the clinician to select and lock in the appropriate amount of electrical stimulation necessary to stimulate pelvic floor contraction.
The provided document is a 510(k) summary for a medical device called InToneMV. This document describes the device, its intended use, and its substantial equivalence to a predicate device (InTone K110179).
However, the document does not contain acceptance criteria or a study that proves the device meets specific acceptance criteria in terms of clinical performance or effectiveness metrics. Instead, it focuses on demonstrating substantial equivalence through modifications to an existing device and safety testing.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not provide specific performance acceptance criteria (e.g., sensitivity, specificity, accuracy, or clinical outcome measures) or reported device performance against such criteria. The submission is a Special 510(k), which typically involves modifications to an existing device where the original performance characteristics are assumed to be maintained if the modifications don't impact safety or effectiveness.
The document states: "The data collected and documented throughout this submission provides objective evidence that InToneMV performs as well as or better than the predicate device for the treatment of female urinary incontinence." This is a general claim of performance equivalence rather than a report against specific acceptance criteria for the InToneMV itself.
2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not describe a clinical test set or sample size for evaluating the InToneMV's performance. It explicitly states: "Clinical performance of InToneMV is further detailed in Section 14." and "There have been no modifications to the clinical performance." However, Section 14 is not included in the provided snippets. Given that this is a Special 510(k) for device modifications, it's highly probable that new clinical data was not collected specifically for InToneMV's clinical performance because its clinical performance is considered identical to the predicate.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not applicable, as no new clinical test set or ground truth establishment is described for the InToneMV device's clinical performance.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable, as no new clinical test set or ground truth establishment is described for the InToneMV device's clinical performance.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The InToneMV is a medical device (pelvic floor muscle stimulator), not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The InToneMV is a medical device, not an algorithm. Its performance relies on the device's electrical stimulation and biofeedback capabilities for the patient, potentially overseen by a clinician.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no new clinical test set or ground truth establishment is described for the InToneMV device's clinical performance. The submission relies on the predicate device's established effectiveness.
8. The sample size for the training set
Not applicable. The InToneMV is a physical medical device, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable.
Summary of what the document does describe regarding validation:
- Bench Testing: "Bench testing of InToneMV is further detailed in Section 13." This would relate to the physical and electrical safety and performance of the modified device components (e.g., smaller probe, flexible cord).
- Electrical Safety Tests: "IEC60601 electrical safety tests were performed on the modifications. Results identified InToneMV as electrically safe."
- Software: "The software in InToneMV is identical to the predicate device, InTone (K110179). There have been no modifications to the existing, validated software."
- Materials: "The patient contact materials used in InToneMV are identical to the predicate device, InTone. There have been no modifications to the materials."
- Design Controls and Risk Management: Modifications were completed according to internal procedures, and risks were evaluated per ISO 14971, with an acceptable risk/benefit analysis.
The core argument for the InToneMV's performance and safety is its substantial equivalence to the predicate device (InTone K110179) and the fact that modifications only apply to the insertion unit, which were validated through non-clinical testing (bench and electrical safety) and adherence to design controls. The clinical performance and effectiveness are considered to be unchanged from the predicate device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for InControl Medical. Below the logo is the address 3225 Gateway Road, Ste. 250, Brookfield, WI 53045. The logo is in a serif font and the address is in a sans-serif font.
IN TONEMV JAN - 6 2014
7. 510(k) Summary
Submission Date
December 17th, 2013
Submitter Information
Jessica Andreshak Director of Quality Assurance and Regulatory Affairs InControl Medical, LLC 3225 Gateway Road, Ste. 250, Brookfield, WI 53045 USA Phone: (262) 373.0422 Fax: (262) 373.0463 Email: jandreshak@incontrolmedical.com
Device Information
Table 4. Device Information
| Type of 510(k): | Special 510(k) |
|---|---|
| Common Name: | Pelvic Floor Muscle Stimulator |
| Trade Name (proprietary name): | InToneMV™ |
| Classification name: | Stimulator, Electrical, Non-Implantable, For Incontinence |
| Classification Regulation: | 21 CFR 876.5320 |
| Class: | Class II |
| Product Code: | KPI |
Legally Marketed Device for Substantial Equivalence
| Table 5. Predicate Device Information | |
|---|---|
| -- | --------------------------------------- |
| 510(k) | Name | Product Code | Manufacturer |
|---|---|---|---|
| K110179 | InTone | KPI | InControl Medical, LLC3225 Gateway Road, Ste. 250Brookfield, WI 53045 USA |
Device Summary
InToneMV is substantially equivalent to the predicate device, InTone (K110179). InToneMV is a modification of the predicate device.
Confidential
{1}------------------------------------------------
3225 Gateway Road, Ste
Special 510(k) Submission IN TONEMV
Brookfield, WI 53045
The InToneMV device includes three parts: an insertion unit, a hand-held control unit, and a PCbased software application. Each of these parts are summarized below, along with a summary of modifications:
- . The insertion unit includes (1) an inflation bulb, (2) the inflatable probe, and (3) the flexible tubing connecting the inflation bulb and the probe. The inflatable probe is inserted into the vagina and manually inflated by the patient to ensure a customized fit. Electrical stimulation is delivered via stainless steel electrodes on the probe to induce a contraction of the pelvic floor muscles. This portion of the device has been modified from the predicate device, InTone. The modifications included a smaller probe and a flexible cord between the probe and the inflation bulb.
- . The external hand-held control unit includes user keys to initiate and control treatment sessions. The control unit is designed to record and store results of the electrical stimulation and patient generated pelvic floor exercises. There are no modifications to the hand-held control unit; therefore it is identical between InToneMV and the predicate device, InTone.
- . The PC-based software application is utilized by the clinician to program the hand-held control unit, store and display the results of electrical stimulation and patient generated pelvic floor exercises. The application allows the clinician to select and lock in the appropriate amount of electrical stimulation necessary to stimulate pelvic floor contraction. There are no modifications to the PC-based software application; therefore it is identical between InToneMV and the predicate device, InTone.
A further description of InToneMV is provided in Section 12 and compares the predicate device, InTone (K110179).
Intended Use
The InToneMV device is a non-implanted electrical stimulator indicated for use in the treatment of female urinary incontinence. It applies electrical stimulation to the pelvic floor musculature and surrounding structures. It is intended for acute and ongoing treatment of mixed urinary incontinence where the following results may improve urinary control: strengthening of pelvic floor muscles and inhibition of the detrusor muscle through reflexive mechanisms. The biofeedback feature can be used for muscle re-education purposes.
This intended use is identical to the predicate device, InTone (K110179).
Equivalence Comparison to the Predicate
This Special 510(k) submission is for modifications to an existing device, InTone (K110179). Modifications only apply to the insertion unit of the device (no changes are being made to the hand-held control unit or the clinician PC application). The changes to the insertion unit include reducing the size of the probe and adding a flexible cord between the probe and the inflation bulb. InToneMV is substantially equivalent to the predicate device, InTone. The intended use statement of InToneMV is identical to the predicate device, InTone.
{2}------------------------------------------------
3225 Gateway Road. Brookfield, WI 53045 Special 510(k) Submission N TONE MV
The technical characteristics shared between the two devices is further compared in Section 12.
Testing Summary
The software in InToneMV is identical to the predicate device, InTone (K110179). There have been no modifications to the existing, validated software.
A gap assessment was conducted by Underwriter Laboratories and IEC60601 electrical safety tests were performed on the modifications. Results identified InToneMV as electrically safe.
The patient contact materials used in InToneMV are identical to the predicate device, InTone. There have been no modifications to the materials.
Animal testing was not applicable for InToneMV.
Bench testing of InToneMV is further detailed in Section 13.
Clinical Performance Summary
InToneMV and the predicate device, InTone (K110179), share the same technology and treatment protocol. There have been no modifications to the clinical performance.
Clinical performance of InToneMV is further detailed in Section 14.
Design Control Summary
The modifications of the predicate device, InTone (K110179), have been completed according to InControl Medical's internal procedures and each design change has been verified and/or validated to support InToneMV.
The modifications of the predicate device, InTone, have been evaluated for risks according to InControl Medical's internal procedures based on ISO 14971. The risks associated with InToneMV were reduced to as low as possible and the risk/benefit analysis was acceptable.
The risk management and design controls for InToneMV is further detailed in Section 14.
Conclusion
The modifications to the predicate device, InTone do not impact the safety or effectiveness. The data collected and documented throughout this submission provides objective evidence that InToneMV performs as well as or better than the predicate device for the treatment of female urinary incontinence.
Confidential
11 | Page
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2014
InControl Medical, LLC Jessica Andreshak Director of Quality Assurance and Regulatory Affairs 3225 Gateway Road, Suite 250 Brookfield, WI 53045
Re:
Trade/Device Name: InToneMV Regulation Number: 21 CFR§ 876.5320 Regulation Name: Nonimplanted electrical continence device Regulatory Class: II Product Code: KPI Dated: December 16, 2013 Received: December 17, 2013
Dear Jessica Andreshak,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Jessica Andreshak
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
225 Gateway Road. S Brookfield, WI 53045
Special 510(k) Submission IN TONE MV
- Statement of Indications for Use
510(k) Number (if known) K133826
Device Name
InToneMV
Indications for Use
The InToneMV device is a non-implanted electrical stimulator indicated for use in the treatment of female urinary incontinence. It applies electrical stimulation to the pelvic floor musculature and surrounding structures. It is intended for acute and ongoing treatment of mixed urinary incontinence where the following results may improve urinary control: strengthening of pelvic floor muscles and inhibition of the detrusor muscle through reflexive mechanisms. The biofeedback feature can be used for muscle re-education purposes.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Lerner -S 2014.01.06 15:51:18 -05'00'
§ 876.5320 Nonimplanted electrical continence device.
(a)
Identification. A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).(b)
Classification. Class II (performance standards).