K Number
K133740

Validate with FDA (Live)

Date Cleared
2014-03-05

(86 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.

The SAUFLON CLARITI TORIC (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.

The SAUFLON CLARITI MULTIFOCAL (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may require a reading addition of +4.00 Diopters or less.

The SAUFLON CLARITI MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and that may require a reading addition of +4.00 Diopters or less.

The lenses may be prescribed for daily wear for up to one month with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion as recommended by the eye care professional.

Sauflon Clariti (somofilcon A) Soft (hydrophilic) Contact lens with UV blocker and handling tint helps protect against transmission of harmful UV radiation to the cornea and into the eye. The lens has a tint added to make the lens more visible for handling.

Device Description

The Sauflon Clariti (somofilcon A) Soft (Silicone Hydrogel) Contact Lens with UV Blocker and handling tint is available as a single vision, toric and multifocal lens. The lens material (somofilcon A) is a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with Alkyl bis methacrylate. When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used to block UV radiation. A D&C Green 6 tint is also included in the composition of the lens for ease of handling and visibility.

The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380nm

The Sauflon Clariti (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and Handling Tint is spherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:

• Chord Diameter: 13.0 to 18.0mm
• Centre Thickness: 0.03 to 0.70mm
• Base Curve: 7.5 to 9.5mm
• Powers: -30.00 DS to +30.00 DS
• Toric cylinder options: Between -0.75 and -9.75
• Toric Axis options: 10° to 180° (5° steps)
• Multifocal ADD: Add power of up to +4.00, labelled with indicative add strength to be read in conjunction with the fitting guide.

The physical/optical properties of the lenses are:

· Refractive Index: 1.4008
· %Transmittance @ 590nm: 98.30%
. %Transmittance @ 280-315nm: 0.71 .
. %Transmittance @ 316-380nm: 20.62
. Surface Character: Hydrophilic
Water Content: . 56%
60 x 10-11 (cm2/sec) (ml O2/ml x mmHg) Oxygen Permeability (DK): . at 35°C (Fatt Method for determination of oxygen permeability).
1.17 Specific Gravity:

AI/ML Overview

The provided document is a 510(k) summary for a contact lens, which primarily focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets distinct acceptance criteria through a standalone study with a defined ground truth.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in the format of defined numerical thresholds that the new device must meet independently. Instead, it relies on demonstrating that the physical, optical, and chemical properties of the subject device are substantially equivalent to its predicate devices. The reported device performance is presented through a comparison table outlining these properties for the subject device and the predicates.

PropertySubject Device Performance (Sauflon Clariti (somofilcon A) with UV Blocker and Handling Tint)Predicate Device 1 (Sauflon Clariti with UV Blocker, K130342)Predicate Device 2 (Air Optix Aqua, K033919/K073459)
MaterialSomofilcon A Silicone HydrogelSomofilcon A Silicone HydrogelLotrafilcon B Silicone Hydrogel
Indications for UseDaily wear monthly replacementDaily wear monthly replacementDaily wear monthly replacement
Manufacturing ProcessCast MouldingCast MouldingCast Moulding
Water Content56%56%33%
Refractive Index1.401.401.42
Light Transmittance≥96%≥96%≥96%
Dk @35°C (Edge Corrected)60 (polarographic method)60 (polarographic method)110 (Coulometric method)
Color/TintBlue Visibility Tint (D&C Green 6)No Visibility TintBlue Visibility Tint (Copper phthalocyanine)
UV BlockerBenzophenoneBenzophenoneNone
Modulus (MPa)0.550.550.92
Tensile Strength (MPa)1.051.050.9
Elongation at Break %163163205
Packaging MaterialsInjected molded polypropylene blisters or Sabic PP PCGH19 covered by aluminium foil laminate and blister strips are packed into printed cartonsInjected molded polypropylene blisters or Sabic PP PCGH19 covered by aluminium foil laminate and blister strips are packed into printed cartonsInjected molded polypropylene blisters covered by aluminium foil laminate and blister strips are packed into printed cartons
Packaging SolutionBorate buffered saline solution containing 0.005% poloxamer 407.Borate buffered saline solution containing 0.005% poloxamer 407.Phosphate buffered saline solution
Packaging MethodHermetically sealed blister packHermetically sealed blister packHermetically sealed blister pack

Summary of Device Performance against "Acceptance Criteria" (Substantial Equivalence): The document claims "substantial equivalence" based on these comparisons. For most properties, the subject device directly matches one of the predicate devices. Where differences exist (e.g., tint, water content, Dk, UV blocker, Modulus, Tensile Strength, Elongation at Break, packaging materials, packaging solution), the differences are argued to not raise new questions of safety or effectiveness.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "A clinical study was conducted to evaluate the safety and efficacy of SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) with UV Blocker and handling tint by comparison with Air Optix Aqua hydrophilic contact lenses (Ciba Vision Inc.)."

  • Sample Size for Test Set: The specific number of subjects in this clinical study (test set) is not provided in the summary.
  • Data Provenance: The study involved a comparison with Air Optix Aqua lenses. The summary does not explicitly state the country of origin for the data or whether it was retrospective or prospective, but clinical trials for regulatory submissions are typically prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is usually relevant for studies involving human interpretation (e.g., image analysis). For a contact lens clinical study evaluating safety and efficacy, the "ground truth" is generally derived from direct clinical observations (e.g., visual acuity, adverse events, fit) performed by qualified clinicians (optometrists, ophthalmologists). The document does not specify the number or qualifications of experts involved in establishing ground truth for the clinical study.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method (such as 2+1, 3+1, or none) for the clinical study. This is less common to report for contact lens clinical studies compared to, for example, studies involving subjective image interpretation where expert discrepancies need resolution.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not conducted as described in the provided text. MRMC studies are typically used to assess the performance of diagnostic devices interpreted by multiple human readers, often with and without AI assistance, which is not applicable to a contact lens evaluation in this context. The clinical study was a direct comparison of the subject contact lens to a predicate contact lens.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable. The device is a physical contact lens, not an algorithm or AI system. The "standalone performance" of the contact lens is assessed through its physical, optical, chemical properties, and human clinical trials (with humans wearing the device, not interacting with an algorithm).

7. The Type of Ground Truth Used

For the clinical study:

  • Derived from: Clinical observations (e.g., visual acuity, comfort, lens fit, adverse events) as measured during the study by clinical investigators. This is a form of outcomes data and expert clinical assessment.
  • For the physicochemical studies: Laboratory measurements of defined properties (e.g., water content, Dk, transmittance, modulus, tensile strength).

8. The Sample Size for the Training Set

This question is not applicable. The document describes a physical medical device (contact lens), not an AI/algorithm-based device that requires a training set. The manufacturing process for contact lenses (cast moulding) does not involve a "training" phase in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable, as there is no training set for this device.

{0}------------------------------------------------

510(K) SUMMARY AS REQUIRED BY SECTION 807.92(c)

1.- SUBMITTER INFORMATION:

Company Name:Sauflon Pharmaceuticals Ltd.
Address:49 - 53 York StreetTwickenhamMiddlesexTW1 3LP
Phone:020 8322 4200
Fax:0845 051 8701
Contact Person:Dr Christopher Smejkal
DATE SUMMARY PREPARED:24th January 2014
DEVICE NAME:
Trade Name:Sauflon Clariti (somofilcon A) Soft (hydrophilic) ContactLens with UV Blocker and Handling tint

Common Name: Soft Contact Lens

Classification:CLASS II (21 CFR 886.5925) CODE -LPL,
SOFT (HYDROPHILIC) CONTACT LENS

Duration of Use:

Sauflon Clariti (somofilcon A) Soft (hydrophilic) Contact lens with UV Blocker and handling tint is a daily wear, frequent replacement lens for surface contact of the eye, and intended for daily removal and less than 24 hours contact duration.

2 .- SUBSTANTIAL EQUIVALENCE:

The sponsor considers the Sauflon Clariti (somofilcon A) Soft (hydrophilic) with UV Blocker and handling tint to be substantially equivalent to the Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) Lens for Daily Wear which has been approved pursuant to K130342, and Air Optix Aqua (Lotrafilcon B) Soft (Hydrophilic) Visibility

{1}------------------------------------------------

Tinted Contact Lens for Daily Wear which has been approved pursuant to K033919/K073459.

3.- DESCRIPTION of the DEVICE:

The Sauflon Clariti (somofilcon A) Soft (Silicone Hydrogel) Contact Lens with UV Blocker and handling tint is available as a single vision, toric and multifocal lens. The lens material (somofilcon A) is a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with Alkyl bis methacrylate. When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used to block UV radiation. A D&C Green 6 tint is also included in the composition of the lens for ease of handling and visibility.

The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380nm

The Sauflon Clariti (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and Handling Tint is spherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:

• Chord Diameter:13.0 to 18.0mm
• Centre Thickness:0.03 to 0.70mm
• Base Curve:7.5 to 9.5mm
• Powers:-30.00 DS to +30.00 DS
• Toric cylinder options:Between -0.75 and -9.75
• Toric Axis options:10° to 180° (5° steps)
• Multifocal ADD:

Add power of up to +4.00, labelled with indicative add strength to be read in conjunction with the fitting guide.

The physical/optical properties of the lenses are:

· Refractive Index:1.4008
· %Transmittance @ 590nm:98.30%
  • . %Transmittance @ 590nm:
  • %Transmittance @ 280-315nm: 0.71 .
  • . %Transmittance @ 316-380nm: 20.62
  • . Surface Character: Hydrophilic
  • Water Content: . 56%

{2}------------------------------------------------

  • 60 x 10-11 (cm2/sec) (ml O2/ml x mmHg) Oxygen Permeability (DK): . at 35°C (Fatt Method for determination of oxygen permeability).
    • 1.17 Specific Gravity:

4.- INDICATIONS FOR USE

.

The SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.

The SAUFLON CLARITI TORIC (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes that may exhibit astigmatism up to 10.00 Diopters.

The SAUFLON CLARITI MULTIFOCAL (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may require a reading addition of +4.00 Diopters or less.

The SAUFLON CLARITI MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and that may require a reading addition of +4.00 Diopters or less.

The lenses may be prescribed for daily wear with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion as recommended by the eye care professional.

Sauflon Clariti (somofilcon A) Soft (hydrophilic) Contact lens with UV blocker and handling tint helps protect against transmission of harmful UV radiation to the cornea and into the eve.

{3}------------------------------------------------

The sponsor considers the SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and handling tint to be substantially equivalent to the Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear which has been approved pursuant to K130342, and Air Optix Aqua (Lotrafilcon B) Soft (Hydrophilic) Visibility Tinted Contact Lens for Daily Wear which has been approved pursuant to K033919/K073459.

The following table summarises the primary features for this comparison, illustrating the similarities and differences.

{4}------------------------------------------------

Table 6.1. Comparison of Physical Properties for the SAUFLON CLARITI (Somofilion A) Soft (Hydrophility) Contact Lens with

Blocker and Handling Tint versus Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) and Air Optix Aqua (Lotrafilcon B) Soft (Hydrophilic) VisibilityTinted Contact Lenses for Daily Wear
PREDICATE DEVICE - SAUFLONCLARITI WITH UV BLOCKER(K130342)PREDICATE DEVICE – AIR OPTIX(K033919/K073459)SUBJECT DEVICE - SAUFLON CLARITI WITHUV BLOCKER AND HANDLING TINT
LENS MATERIALSomofilcon A Silicone HydrogelLotrafilcon B Silicone HydrogelSomofilcon A Silicone Hydrogel
INDICATIONS FORUSEDaily wear monthly replacementDaily wear monthly replacementDaily wear monthly replacement
MANUFACTURINGPROCESSCast MouldingCast MouldingCast Moulding
WATER CONTENT56%33%56%
REFRACTIVEINDEX1.401.421.40
LIGHTTRANSMITTANCE≥96%≥96%≥96%
DK @35°C (EDGECORRECTED)60 (polarographic method)110 (Coulometric method)60 (polarographic method)
COLOURNo Visibility TintBlue Visibility TintBlue Visibility Tint
TINT
TINTNo Visibility TintCopper phthalocyanineD&C Green 6
UV BLOCKERBenzophenoneNoneBenzophenone
MODULUS (MPa)0.550.920.55
TENSILE STRENGTH(MPa)1.050.91.05
ELONGATION ATBREAK %163205163
PACKAGINGMATERIALSInjected molded polypropylene blisters orSabic PP PCGH19 covered by aluminiumfoil laminate and blister strips are packedinto printed cartonsInjected molded polypropylene blisters covered byaluminium foil laminate and blister strips arepacked into printed cartonsInjected molded polypropylene blisters or Sabic PPPCGH19 covered by aluminium foil laminate and blisterstrips are packed into printed cartons
PACKAGINGSOLUTIONBorate buffered saline solution containing0.005% poloxamer 407.Phosphate buffered saline solutionBorate buffered saline solution containing 0.005%poloxamer 407.
PACKAGINGMETHODHermetically sealed blister packHermetically sealed blister packHermetically sealed blister pack

{5}------------------------------------------------

SAUFLON CLARITI (SOMOFILCON A) SOFT (HYDROPHILIC) CONTACT LENS WITH UV BLOCKER AND HANDLING
TINT

{6}------------------------------------------------

6.- PHYSICOCHEMICAL STUDIES

The physical, optical and chemical properties of the lenses as assessed by various test methods show substantial equivalence with the predicate devices as illustrated in the preceding table. Studies were also conducted to verify that leachable substances were either low or below measurable levels to assuage any concerns for its intended use.

7.- TOXICOLOGY STUDIES

Toxicology testing demonstrates that Clariti (somofilcon A) Soft (Hydrophilic) with UV blocker and handling tint is non-toxic in all cytotoxicity, ocular irritation and acute systemic toxicity tests. This lens is therefore equivalent in terms of biocompatibility to the predicate devices specified.

8.- EVALUATION OF DYE LEACHABILITY

Dye leachability was evaluated and no dye was observed to leach from Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) Contact Lenses with UV Blocker and handling tint after 14 days soaking in saline at 37°C.

9.- HUMAN CLINICAL STUDIES

A clinical study was conducted to evaluate the safety and efficacy of SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) with UV Blocker and handling tint by comparison with Air Optix Aqua hydrophilic contact lenses (Ciba Vision Inc.). Subjects used OptiFree Replenish solution (Alcon Laboratories Inc.) for daily lens maintenance, care and storage. The results of this study showed the safety, acceptability and substantial equivalence of the Sauflon CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and handling tint to the predicate device for its intended use.

10.- CONCLUSIONS

Based on the above evaluations the SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and handling tint is substantially equivalent to the predicate, marketed lenses. Based on these evaluations the SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and handling tint has been shown to be safe and effective for its intended use.

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

March 5, 2014

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Sauflon Pharmaceuticals Ltd. Dr. Christopher Smeikal Head of Regulatory and Technical Affairs 49-53 York Street Twickenham. Middlesex TW1 3LP, United Kingdom

Re: K133740

Trade Name: Sauflon Clariti (somofilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker and Handling Tint Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: II Product Code: LPL Dated: January 24, 2014 Received: January 29, 2014

Dear Dr. Smejkal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{8}------------------------------------------------

Page 2 - Dr. Christopher Smejkal

You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

and and the comments of the comments of the

Sincerely yours,

Kesia Y. Alexander -S

for Malvina B. Eydelman, M.D.

  • Director

..............................................................................................................................................................................

and the commended to the comments of 1000 - 1000

Division of Ophthalmic and Ear,

Nose and Throat Devices . . . . . . .

. . . . . . . .

Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

,

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K133740

Device Name

Sauflon Clariti (somofileon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint.

Indications for Use (Describe)

The SAUFLON CLARITI (somofilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.

The SAUFLON CLARITI TORIC (somofile) Contact Leas with UV blocker and handling tint is indicated for daily woar for the optical corrective anteropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.

The SAUFLON CLARITI MULTIFOCAL (somofileon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive anetropia (myopia and or presbyopia in phakic or aphakic persons with non-diseased eyes that may require a reading addition of +4.00 Diopters or less.

The SAUFLON CLARITI MULTIFOCAL TORIC (somofilion A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and that may require a reading addition of +4.00 Diopters or less.

The lenses may be prescribed for daily wear for up to one month with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion as recommended by the eye care professional.

Saullon Claili (somofilcon A) Soll (hydrophilic) Contact lens with UV blocker and handling tint helps protect against transmission of harmful UV radiation to the cornea and into the eye. The lens has a itint added to make the lens more visible for handling.

Type of Use (Select one or both, as applicable)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ADDITIONAL SHEET

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Kealth (CDRH) (Signature)

Prescription Use (Part 21 CFR 801 Subpart D)

Marc W. Robboy -S 2014.02.18 09:13:27 -05'00'

FORM FDA 3881 (1/14)

::

Page 1 of 2

{10}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the deta needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.