K Number
K133728

Validate with FDA (Live)

Date Cleared
2014-01-24

(49 days)

Product Code
Regulation Number
862.1225
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CRE2 method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista® System. Creatinine measurements are used in the diagnosis and treatment of certain renal disease, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Device Description

The Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge uses a modified kinetic Jaffe technique. In the presence of a strong base such as sodium hydroxide, picrate reacts with creatinine to form a red chromophore. The rate of increasing absorbance at 510 nm due to the formation of this chromophore is directly proportional to the creatinine concentration in the sample and is measured using a bichromatic (510, 577 nm) rate technique.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Dimension Vista® System Creatinine (CRE2) Flex® reagent cartridge, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Method Comparison (Serum)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 when compared to predicate device (Vista CREA Assay)Vista CREA Assay (Serum): Range 0.38 – 18.93 mg/dL, Slope = 1.02, Intercept = -0.11 mg/dL, Correlation Coefficient = 1.000 (n=140)
Method Comparison (Urine)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 when compared to predicate device (Vista CREA Assay)Vista CREA Assay (Urine): Range 8.39 – 299.60 mg/dL, Slope = 1.05, Intercept = -4.29 mg/dL, Correlation Coefficient = 0.999 (n=112)
Method Comparison (IDMS)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 when compared to IDMS Reference MethodIDMS Reference Method (Serum): Range 0.18–6.32 mg/dL, Slope = 1.06, Intercept = -0.03 mg/dL, Correlation Coefficient = 0.997 (n=48)
Serum Plasma EquivalencySlope close to 1, Intercept close to 0, Correlation Coefficient close to 1 for matched serum vs. lithium heparin plasma samplesSlope = 1.03, Intercept = -0.002, Correlation Coefficient (r) = 0.998 (Range 0.418 - 17.9, n=56)
PrecisionWithin-Lab %CV values for various serum and urine pools and control materials. (Implicitly, these should be within acceptable limits for a diagnostic test, though specific numerical criteria for acceptance are not explicitly stated as a single value)Reported %CVs: Serum Pool 1 (2.7%), Serum Pool 2 (0.7%), BioRad Multiqual Level 1 (4.0%), Level 2 (2.5%), Level 3 (1.1%), Urine Pool 1 (3.6%), Urine Pool 2 (1.0%), BioRad Liquicheck Level 1 (2.2%), Level 2 (2.0%). (All values are low, indicating good precision.)
Limit of Blank (LoB)Consistent with claim of 0.05 mg/dL (serum), 1.0 mg/dL (urine)LoB (Serum): 0.030 mg/dL (consistent with 0.05 mg/dL claim) LoB (Urine): 0.438 mg/dL (consistent with 1.0 mg/dL claim)
Limit of Detection (LoD)Consistent with claim of 0.1 mg/dL (serum), 2.0 mg/dL (urine)LoD (Serum): 0.056 mg/dL (consistent with 0.1 mg/dL claim) LoD (Urine): 0.828 mg/dL (consistent with 2.0 mg/dL claim)
Limit of Quantitation (LoQ) Serum/PlasmaAllowable total error for creatinine published by CLIA, CAP, AAB, NYS, and WSLH is +/- 0.3 mg/dL or +/- 15%. (LoQ should meet these recommendations).LoQ (Serum/Plasma): 0.15 mg/dL based on allowable total error of 0.15 mg/dL. This meets the recommendations.
Limit of Quantitation (LoQ) UrineNo recognized allowable total error limit for urine creatinine measurements. (Implicit acceptance is that the determined LoQ is appropriate and justified).LoQ (Urine): 5.00 mg/dL based on allowable total error of 3.00 mg/dL.
Linearity (Serum)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 across the assay rangeLinearity (Serum): Range 0.00 – 21.9 mg/dL, Slope = 0.996, Intercept = -0.05, Correlation Coefficient = 1.0 (N=12)
Linearity (Urine)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 across the assay rangeLinearity (Urine): Range 1.71 – 338 mg/dL, Slope = 0.995, Intercept = 0.47, Correlation Coefficient = 1.0 (N=12)
Interfering SubstancesBias exceeding 10% is considered interference. Dilution studies determine the level at which spiked substances no longer display significant interference.Demonstrated concentrations where various substances cause <10% bias or, if >10% bias, the concentration at which interference is no longer significant through dilution. For example, Acetone interferes at 150 mg/dL (18% bias) but not at 75 mg/dL (9% bias). Bilirubin (unconjugated) at 40mg/dL shows -29.6% bias at low creatinine, but 20mg/dL shows -0.2% bias.
HIL InterferenceBias exceeding 10% is considered interference.Intralipid 20% interferes at 2000 mg/dL and 1500 mg/dL (at low creatinine concentration) (15.5% and 15.1% bias, respectively), but not at 1500 mg/dL (at high creatinine concentration) (9.5% bias) or 1000 mg/dL (7.0% bias). Hemoglobin shows no significant interference at 1000 mg/dL (~7% and 3% bias). Bilirubin (unconjugated) at 40 mg/dL shows -29.6% interference at low creatinine levels. Bilirubin (conjugated) shows 16.1% interference at 40 mg/dL at low creatinine levels.
Expected Values (Serum & Plasma)≤10% of samples within the study should fall outside of established range.Males: 5 (9.4%) out of 53 samples outside range (meets ≤10%). Females: 2 (5.0%) out of 40 samples outside range (meets ≤10%). (Supports published ranges).
Expected Values (Urine)≤10% of samples within the study should fall outside of established range.Males: 2 (9.1%) out of 22 samples outside range (meets ≤10%). Females: 1 (5.0%) out of 20 samples outside range (meets ≤10%). (Supports published ranges).
Reportable Range Dilution (Serum)Mean percent recovery of 90-110% for autodiluted samples.Mean of N=5 reps: 16.5 µL (normal volume) = 16.5; 6.6 µL (auto-dilute volume) = 16.0; % Bias = -3.0%. (Implicitly meets recovery criteria, as bias is very low). Supports extended range to 40.0 mg/dL.
Reportable Range Dilution (Urine)Mean percent recovery of 90-110% for autodiluted samples.Mean of N=5 reps: undiluted (normal) = 268; auto-dilute 3x with water = 245; % Bias = -8.6%. (Implicitly meets recovery criteria, as bias is very low). Supports extended range to 900 mg/dL.

Study Details

  1. Sample sizes used for the test set and the data provenance:

    • Method Comparison (Predicate Device):
      • Serum: 140 patient samples. Remnant de-identified serum samples. No patient history. Inclusion/exclusion criteria not applicable. Both native and spiked samples.
      • Urine: 112 patient samples. Remnant de-identified urine samples (implied). No patient history. Inclusion/exclusion criteria not applicable.
    • Method Comparison (IDMS Reference Method):
      • Serum: 48 patient samples. Remnant de-identified serum samples. No patient history. Inclusion/exclusion criteria not applicable. All native samples.
    • Serum Plasma Equivalency: 56 matched serum and lithium heparin plasma samples. All samples fresh and never frozen. 8 spiked sample sets.
    • Precision: Not a direct "test set" in the sense of patient samples, but testing involved: 2 serum pools, 3 levels of BioRad Multiqual material, 2 levels of BioRad Liquicheck material, and 2 urine pools. Each tested multiple times (over 20 days, 2 runs, 2 samples per run).
    • Limit of Blank/Detection (Serum): 4 samples with no analyte (N=5 reps for 3 days), 4 low patient serum samples (N=5 reps for 3 days).
    • Limit of Blank/Detection (Urine): 4 samples with no analyte (N=5 reps for 3 days), 4 low patient urine samples (N=5 reps for 3 days).
    • Linearity: 12 equally spaced samples for serum, 12 for urine.
    • Analytical Specificity (Interference): Low (~1.5 mg/dL) and High (~5.0 mg/dL) creatinine serum pools; Low (~40 mg/dL) and High (~175 mg/dL) creatinine urine pools. Number of individual samples/tests per interferent is not explicitly stated but implied to be sufficient for creating means and controls.
    • Expected Values:
      • Serum and Plasma: 53 male donors, 40 female donors (all normal healthy adults).
      • Urine: 22 male donors, 20 female donors (all normal healthy adults).
    • Reportable Range Dilution: N=5 replicates for each dilution study (serum and urine).

    Data Provenance: The studies were conducted internally by Siemens Healthcare Diagnostic Inc. R&D organization personnel. Remnant de-identified samples were used for method comparison studies, implying retrospective or archived samples. Patient history was not obtained. The origin of the patient samples (e.g., country) is not specified, but the internal R&D location would presumably be Newark, DE, USA based on the applicant's address.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    The document does not mention the use of experts to establish ground truth for the test set in the context of clinical interpretation. For method comparison, the "ground truth" for the new device was established by comparison to a legally marketed predicate device (Dimension Vista® Creatinine (CREA) Flex® reagent cartridge) and the IDMS reference method. These are analytical "gold standards" rather than expert interpretation.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    Not applicable. The reported studies are analytical performance evaluations of a quantitative laboratory test, not diagnostic interpretations requiring adjudication.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This is a quantitative in vitro diagnostic test for creatinine, not an imaging or interpretive device that would typically involve human readers or AI assistance in the way an MRMC study would assess.

  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
    Yes, these are all standalone (algorithm only) performance studies. The device itself performs the quantitative measurement of creatinine from samples without human interpretation in the loop of the measurement.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For Method Comparison studies: The ground truth was established by two primary methods:
      • Comparison to a predicate device (Dimension Vista® Creatinine (CREA) Flex® reagent cartridge).
      • Comparison to the IDMS reference method (Isotope Dilution Mass Spectrometry), which is a highly accurate and precise analytical method often considered a "gold standard" for creatinine measurements.
    • For Expected Values studies: Published reference ranges in scientific literature (e.g., Tietz 1999, Clin Chem 54:3 (2008), Clin Chem Acta 344 (2004) 137-148) served as the basis for validating the established ranges for the new device.
    • For LoQ validation: Allowable total error limits published by regulatory and accreditation bodies (CLIA, CAP, AAB, NYS, WSLH) were used. The reference values are traceable to IDMS.
  7. The sample size for the training set:
    Not applicable in the context of typical AI/machine learning studies. This device is a reagent cartridge for an automated clinical chemistry system, not an AI or machine learning algorithm that undergoes 'training' in the conventional sense with labeled data. The development and internal refinement of the reagent formulation and measurement parameters would be analogous to "training" but is not quantified in terms of a discrete sample set size like this.

  8. How the ground truth for the training set was established:
    Not applicable for the same reasons as above. The "ground truth" for developing such a system would involve rigorous analytical chemistry principles, extensive experimentation, and calibration using traceable standards (e.g., NIST SRM 914a (IDMS assigned crystalline creatinine standard)).

{0}------------------------------------------------

1. 510(k) Summary

1.1 Description

Dimension Vista® System Creatinine (CRE2) Flex® reagent cartridge

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

< 133720

1.2. Assigned 510(k) number

The assigned 510(k) number is:

1.3 Applicant and Date

Applicant:

Laura J. Duggan, Ph. D., RAC Siemens Healthcare Diagnostics Inc. P.O. Box 6101 Newark, DE 19714-6101 Email: laura.j.duggan@siemens.com Phone: 302-631-7654 Fax: 302-631-6299

December 5, 2013 Date:

1.4 Proprietary and Established Names

Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge

1.5 Common Name

Creatinine

1.6 Regulatory Information

Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge

The CRE2 method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista® System. Creatinine measurements are used in the diagnosis and treatment of certain renal disease, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Page 1 of 17

{1}------------------------------------------------

Classification Name: Requlation Section: Classification: Product Code: Panel:

Creatinine test system 21CFR862.1225 - Creatinine test system Class II CGX Clinical Chemistry

1.7 Predicate Device

The predicate device used to demonstrate substantial equivalence to the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge is the Dimension Vista® Creatinine (CREA) Flex® reagent cartridge previously cleared under K061238.

1.8 Device Description / Test Principle

The Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge uses a modified kinetic Jaffe technique. In the presence of a strong base such as sodium hydroxide, picrate reacts with creatinine to form a red chromophore. The rate of increasing absorbance at 510 nm due to the formation of this chromophore is directly proportional to the creatinine concentration in the sample and is measured using a bichromatic (510, 577 nm) rate technique.

1.9 Intended Use

The CRE2 method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista System. Creatinine , measurements are used in the diagnosis and treatment of certain renal disease, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

1.10 Indication(s) for Use

The CRE2 method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista® System. Creatinine measurements are used in the diagnosis and treatment of certain renal disease, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

1.11 Substantial Equivalence Information

Both the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge and the predicate Dimension Vista® Creatinine (CREA) Flex® reagent cartridge employ prepackaged reagents for use on automated clinical chemistry test systems. A comparison of the similarities and differences between the devices is provided in the following tables:

{2}------------------------------------------------

.

Similarities for the Dimension Vista Creatinine (CRE2) Flex® reagent cartridge:

FeatureNew DeviceDimension Vista® SystemCreatinine (CRE2) Flex® reagentcartridgePredicate DeviceDimension Vista® SystemCreatinine (CREA) Flex® reagentcartridge (K061238)
Intended UseThe CRE2 method is an in vitrodiagnostic test for the quantitativemeasurement of creatinine inhuman serum, plasma, and urineon the Dimension Vista®System. Creatinine measurementsare used in the diagnosis andtreatment of certain renal disease,in monitoring renal dialysis, and asa calculation basis for measuringother urine analytes.The CREA method is an in vitrodiagnostic test for the quantitativemeasurement of creatinine inhuman serum, plasma and urineon the Dimension Vista® System.
DeviceTechnology(detection)Modified Jaffe Methodology(creatinine alkaline picrate) withphotometric detectionSame
CalibrationInterval90 days - same reagent lotSame
Limit of Blank/AnalyticalSensitivityLimit of Blank: 0.05 mg/dLAnalytical Sensitivity : 0.1 mg/dL
CalibrationChem 1 Calibrator (K061838)2 levels (n=5)Same
Traceability &StandardizationNIST SRM 914a (IDMS assignedcrystalline creatinine standardprepared by the National Institute ofStandards and Technology)Same
FeatureNew DeviceDimension Vista® SystemCreatinine (CRE2) Flex® reagentcartridgePredicate DeviceDimension Vista® SystemCreatinine (CREA) Flex® reagentcartridge (K061238)
Measuring Range(serum)0.15 - 20.0 mg/dL0.10 - 20.00 mg/dL
Measuring Range(urine)5.00 - 300 mg/dL0.10 - 200.00 mg/dL
ReportableRange (serum)0.15 - 40.0 mg/dL0.10 - 40.00 mg/dL
ReportableRange (urine)5.00 - 900 mg/dL0.10 - 200.00 mg/dL (no specificdilution factor recommended)
Sample Volume13.2 µL8 µL
ReagentsReagent 2 = Lithium Picrate(125 mM)Reagent 1 = Sodium Hydroxide(1000 mM)Reagent 1 = Lithium Picrate(125 mM)Reagent 2 = Sodium Hydroxide(2000 mM) and potassiumferricyanide
Reagent VolumesVolume of Reagent 1 used = 30.6 µLVolume of Reagent 2 used = 16.5 µLVolume of Reagent 1 used = 29.6 µLVolume of Reagent 2 used = 7.2 µL
DetectionConditionsWavelength = 510 and 577 nmType of Measurement =Bichromatic rateWavelength = 510 and 600 nmType of Measurement =Bichromatic rate
Expected ValuesSerum and PlasmaMales: 0.700 - 1.30 mg/dLFemales: 0.550 - 1.02 mg/dLUrineMales: 0.95 - 2.49 g/24 hrSerum and Plasma0.6-1.3 mg/dLUrine0.6-2.5 g/24hr
Females: 0.60 - 1.80 g/24 hr
No significant interference at aCreatinine.concentration of 1.5mg/dL in serum from:Hemoglobin at 1000 mg/dLBilirubin (conjugated) at 20 mg/dL,Lipemia (Intralipid) at 1000 mg/dLNo significant interference at aCreatinine concentration of 1.3mg/dL in serum from:Hemoglobin at 500 mg/dL,Bilirubin (conjugated) at 20 mg/dL,Lipemia (Intralipid) at 3000 mg/dL
Interferences
Bilirubin (unconjugated) at 20mg/dLNo significant interference at aCreatinine concentration of 1.1mg/dL from:Bilirubin (unconjugated) at 10mg/dL

Page 3 of 17

{3}------------------------------------------------

Differences for the Dimension Vista Creatinine (CRE2) Flex® reagent cartridge:

Page 4 of 17

{4}------------------------------------------------

1.12 Standard/Guidance Document Reference

  • Stability Testing of In Vitro Diagnostic Reagents (CEN 13640) t
  • Interference Testing in Clinical Chemistry; Approved Guideline (EP07-A2) .
  • Method Comparison and Bias Estimation Using Patient Samples; Approved . Guideline (EP09-A2-IR).
  • Evaluation of Precision Performance of Quantitative Measurement Methods: . Approved Guideline (EP05-A2)
  • Evaluation of the Linearity of Quantitative Measurement (EP-06-A) .
  • Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; . Approved Guideline (EP17-A2)
  • Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; . Approved Guideline- Third Edition (C28-A3c)
  • In Vitro Diagnostic Creatinine Test System Guidance for Industry July 2, 1998 .
  • In Vitro Diagnostic Devices: Guidance for the Preparation of 510(k) Submissions -. Jan. 1997
  • Format for Traditional and Abbreviated 510(k)'s Guidance for Industry and Staff -. Nov. 17, 2005
  • Guidance for Industry and FDA Staff: Administrative Procedures for CLIA . Categorization - May 7, 2008
  • eCopy Program for Medical Device Submissions Guidance for Industry and Food . and Drug Administration Staff - October 10, 2013
  • Refuse to Accept Policy for 510(k)s Guidance for Industry and Food and Drug . Administration Staff - December 31, 2012

Page 5 of 17

{5}------------------------------------------------

1.13 Performance Characteristics

The following data represent typical method performance. These data were collected on the Dimension Vista® 500 System.

1.13.1 Method Comparison

The predicate device selected for the method comparison was the Dimension Vista® Creatinine (CREA) Flex reagent cartridge cleared under K061238. Remnant de-identified serum samples were tested. No patient history information was obtained on these samples. Inclusion/exclusion data criteria are not applicable. The study included both native and spiked samples. ·

These studies were conducted internally by Siemens Healthcare Diagnostic Inc. R&D organization personnel. The personnel conducting the study were laboratory technicians with training similar to personnel who would conduct the tests in a hospital laboratory setting. They were trained on the operation of both the device and the predicate device. A split sample method comparison, following EP09-A2, demonstrated good agreement between the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge and the predicate Dimension Vista® Creatinine (CREA) Flex® reagent cartridge with serum patient samples.

One hundred forty serum patient samples across the assay range were tested on the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge and the Dimension Vista® Creatinine (CREA) Flex® reagent cartridge. In a second study, 112 urine samples were tested on the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge and the Dimension Vista Creatinine (CREA) Flex® reagent cartridge. The results across the full assay range were analyzed by linear regression. Although the samples were tested in duplicate, only the first result was used for the analysis.

ComparativeMethodRange(mg/dL)SlopeIntercept(mg/dL)CorrelationCoefficientnSampletype
Vista CREAAssay0.38 – 18.931.02-0.111.000140serum
Vista CREAAssay8.39 – 299.601.05-4.290.999112urine

The model equation for the regression statistics is: [results for Dimension Vista® Creatinine (CRE2)] = slope x [comparative method results] + intercept.

An additional study was completed comparing the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge with the IDMS reference method. Remnant de-identified serum samples were tested. No patient history information was obtained on these samples. Inclusion/exclusion data criteria are not applicable. All of the samples were native.

Forty eight patient samples were tested on the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge and the IDMS reference method. The results were analyzed by linear

Page 6 of 17

{6}------------------------------------------------

regression. Although the samples were tested in duplicate, only the first result was used for the analysis.

ComparativeMethodRange(mg/dL)SlopeIntercept(mg/dL)CorrelationCoefficientnSampletype
IDMSReferenceMethod0.18–6.321.06-0.030.99748serum

The model equation for the regression statistics is: [results for Dimension Vista® Creatinine (CRE2)] = slope x [comparative method results] + intercept.

1.13.2 Serum Plasma Equivalency

Serum and lithium heparin plasma equivalency was demonstrated for the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge. Fifty six matched serum and lithium heparin plasma samples were tested using the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge. The table below summarizes the linear regression statistics.

Serum vs.SlopeInterceptCorrelationCoefficient (r)Rangen
Lithium HeparinPlasma1.03-0.0020.9980.418 - 17.956

One replicate of each sample was processed. All samples in the study were fresh and never frozen. The eight spiked sample sets were prepared by spiking equal amounts of purified creatinine into the matched serum and lithium heparin plasma samples.

1.13.3 Precision

Precision testing was performed in accordance with CLSI EP05-A2 Evaluation of Precision Performance of Quantitative Measurement Methods: Approved Guideline - Second Edition. Samples consisted of two (2) serum pools, three (3) levels of BioRad Multiqual material, two (levels) of BioRad Liquicheck material and two (2) urine pools. Testing was performed over twenty (20) days, two (2) separate runs with two test samples for each test material. Analysis of variance (ANOVA) was used to evaluate the data consistent with the recommendations of EP05-A2. The data are summarized in the following table:

Page 7 of 17

{7}------------------------------------------------

RepeatabilityWithin-Lab
SampleMean(mg/dL)SD%CVSD%CV
Serum Pool 11.300.032.20.042.7
erunSerum Pool 216.80.090.60.110.7
BioRad MultiqualLevel 10.5300.022.90.024.0
BioRad MultiqualLevel 21.870.021.10.052.5
BioRad MultiqualLevel 37.230.050.70.081.1
JrineUrine Pool 137.20.912.41.333.6
Urine Pool 22541.880.72.581.0
BioRad LiquicheckLevel 162.00.751.21.382.2
BioRad LiquicheckLevel 21451.531.12.932.0

BioRad® is a registered trademark of Bio-Rad Laboratories, Irvine, CA 92618, USA. Multiqual® is a registered trademark of Bio-Rad Laboratories, Irvine, CA 92618, USA. Liquichek™ is a trademark of Bio-Rad Laboratories, Irvine, CA 92618, USA.

.

{8}------------------------------------------------

1.13.4 Limit of Blank and Limit of Detection

The Limit of Blank (LoB) and Limit of Detection (LoD) were evaluated in accordance with CLSI EP17-A2 Protocols for Determination of Limits of Detection and Limits of Quantitation: Approved Guideline.

Dimension Vista® CRE2 Limit of Detection Results with Serum
LimitProtocolValue
LoB4 samples with no analytewere tested (N=5) for 3days, one run per day, 2reagent lots,0.030 mg/dL
LoD4 low patient serumsamples were tested(N=5) for 3 days, one runper day, 2 reagent lots0.056 mg/dL
Dimension Vista® CRE2 Limit of Detection Results with Urine
LimitProtocolValue
LoB4 samples with no analytewere tested (N=5) for 3days, one run per day, 2reagent lots,0.438 mg/dL
LoD4 low patient urinesamples were tested(N=5) for 3 days, one runper day, 2 reagent lots0.828 mg/dL

The nonparametric approach described in EP-17A2 was followed to determine the Limit of Detection.

LoB = Mean of Blank Measurement + 1.645 x Standard Deviation of Blank Measurements LoD = Limit of Blank + CpSDs

• Co is a correction factor for the 95% Cl normal variate to account for bias in the SDs estimate.

· SDs is an estimate of method imprecision pooled from replicates of the low analyte samples

{9}------------------------------------------------

The LoB was determined to be 0.030 mg/dL with serum samples and is consistent with the claim of 0.05 mg/dL. With urine samples, the LoB was determined to be 0.438 mg/dL and is consistent with the claim of 1.0 mg/dL.

The LoD was determined to be 0.056 mg/dL with serum samples and is consistent with the claim of 0.1 mg/dL. With urine samples, the LoD was determined to be 0.828 mg/dL and is consistent with the claim of 2.0 mg/dL.

1.13.5 Limit of Quantitation

The Limit of Quantitation (LoQ) for CRE2 for serum and plasma is 0.15 mg/dL and based on allowable total error of 0.15 mg/dL, determined consistent with CLSI Guideline EP17-A2. Total Error is calculated using: TE = bias + 2 * SD. The reference values are traceable to Isotope Dilution Mass Spectrometry (IDMS). The established allowable total error for creatinine published by CLIA¹, CAP2, AAB3, NYS4 and WLSH is +/- 0.3 mg/dL or +/- 15% (greater). The LoQ of 0.15 mg/dL based on allowable total error of 0.15 ma/dL meets these recommendations.

CLIA' - CLIA '88 Proficiency Testing Limits, U.S. Federal Register.

CAP2 - College of American Pathologists Participant Summary, April 2004.

  • AAB American Association of Bioanalysts Table of Grading Limits (undated, approx Oct 2005). For details, visit www.aab.org/pts/grdlim.htm
  • NYS® Wadsworth Center Clinical Laboratory Evaluation Program, Guide to Program Requirements and Services. New York State Department of Health (undated, approx June, 2004). For details, visit www.wadsworth.org/labcert.
  • WLSH WSLH Proficiency Testing is a national, full-service PT program located at the Wisconsin State Laboratory of Hygiene on the campus of the University of Wisconsin - Madison. (undated, approx June, 2004). For more information, visit www.sh.wisc.edu/pt.

The Limit of Quantitation (LoQ) for CRE2 for urine is 5.00 mg/dL and based on allowable total error of 3.00 mg/dL, determined consistent with CLSI Guideline EP17-A2. Total Error is calculated using: TE = bias + 2 * SD. The reference values are traceable to Isotope Dilution Mass Spectrometry (IDMS). There is no recognized allowable total error limit for urine creatinine measurements.

Page 10 of 17

{10}------------------------------------------------

1.13.6 Linearity (Measurement Range)

Linearity was evaluated for serum and urine samples by using 12 equally spaced samples which spanned the assay range. Each was prepared by mixing high and low creatinine concentration samples across the measurement range as described in CLSI Evaluation of the Linearity of Quantitative Measurement Procedure (EP06-A). The low sample represents an artificial sample matrix prepared with bovine serum albumin.

Regression Statistics

Range of samplesSlopeInterceptCorrelationCoefficientN
serum0.00 – 21.9 mg/dL0.996-0.051.012
urine1.71 – 338 mg/dL0.9950.471.012

1.14 Analytical Specificity

12.5.1 Non-interfering Substances

CLSI EP7A2 was followed for the interference testing. The interference study was conducted using a "paired difference worst case scenario" approach where these compounds were spiked into fresh sample pools containing either low or high levels of creatinine analyte in both serum pools and urine pools.

{11}------------------------------------------------

DV CRE2 Serum/Plasma Substance
DV CRE2
SubstanceConcentration ofSubstance (mg/dL)LOW pool (~1.5mg/dL)HIGH pool (~5.0 mg/dL)
MeanControl% Diffmeancontrol% Diff
Acetaminophen201.411.401%5.035.001%
Acetoacetate201.351.38-3%4.874.870%
Amikacin81.401.400%5.055.001%
Ampicillin5.31.431.402%5.065.001%
Ascorbic Acid61.411.400%4.995.000%
Caffeine61.401.42-2%4.974.960%
Carbamazepine31.431.421%5.015.06-1%
Cephalexin251.351.38-3%4.864.870%
Cephapirin251.431.393%4.994.970%
Cephradine251.441.393%5.004.971%
Chloramphenicol51.431.411%5.015.030%
Chlordiazepoxide11.401.43-2%5.015.020%
Chlorpromazine0.21.401.400%5.004.990%
Cholesterol Supertrate5031.341.303%4.584.570%
Cimetidine21.401.42-1%4.974.960%
Dextran 4060001.561.513%5.135.23-2%
Diazepam0.511.441.431%5.045.020%
Digoxin6.1 ng/mL1.411.420%4.954.930%
EDTA2001.371.39-2%4.934.910%
Erythromycin61.441.412%5.005.03-1%
Ethanol4001.411.400%4.985.000%
Ethosuximide251.401.42-1%4.974.960%
Furosemide61.431.411%5.035.030%
Gentamicin11.421.401%5.005.000%
Heparin (196 Units/mg)3 U/mL1.401.42-1%4.944.99-1%
Ibuprofen501.431.421%5.015.06-1%
Immunoglobulin G (IgG)50001.561.513%5.145.23-2%
Isopropanol1.0 g/dL1.401.390%4.914.910%
Lidocaine1.21.431.411%5.085.031%
Lithium2.21.401.400%5.004.990%
Nicotine0.11.441.421%5.055.060%
Nortriptyline1000 ng/mL1.391.39-1%4.964.911%
Penicillin G (1654Units/mg)25 U/mL1.411.401%4.984.990%
Pentobarbital81.431.420%4.994.980%
Phenobarbital101.441.421%5.004.980%
Phenytoin51.421.43-1%5.025.020%
Potassium oxalate500 mg/dL1.401.381%4.864.870%
Primidone41.421.420%5.045.060%
Propoxyphene0.161.421.430%5.025.020%
Protein, Albumin60001.561.513%4.885.345-9%
Protein, Total12g/dL1.571.486%5.135.22-2%
Salicylic acid601.421.411%5.035.030%
Sodium fluoride4001.301.300%4.624.67-1%
Theophylline41.391.42-2%5.004.961%
Urea5001.431.402%5.044.991%
Uric acid201.431.411%4.974.990%
Valproic acid501.411.42-1%4.974.960%
DV CRE2 Urine Method Specific
DV CRE2
SubstanceConcentration ofSubstanceLOW pool (~40 mg/dL)HIGH pool ( ~ 175 mg/dL)
MeanControl%Diffmeancontrol%Diff
50% Acetic Acid25mL/24 hr collection40.7540.251%180.0178.81%
6N Hydrochloric Acid0.60%41.3041.350%178.8182.5-2%
6N Nitric Acid0.60%41.1841.350%181.3182.5-1%
Acetone100 mg/dl.42.9341.982%184.5186.8-1%
Albumin0.5 g/dL42.0842.98-2%183185.8-1%
Ascorbic Acid1.5 g/dl38.6540.48-5%173.5176.0-1%
Bilirubin (conjugated)2 mg/dL42.0841.751%182.0179.81%
Boric Acid1% w/v41.8341.750%185.8184.01%
Ethanol1 g/dL41.9841.651%183.8179.82%
Gamma Globulin0.5 g/dL42.7541.882%184.3185.5-1%
Glucose2 g/dL42.6541.882%184.8185.50%
Hemoglobin115 mg/dL41.7841.580%183.8183.80%
Oxalic Acid0.1 g/dL40.2840.480%178.3176.01%
Sodium Carbonate5g/24 hr collection41.0840.651%182.0183.0-1%
Sodium Fluoride1% w/v41.0541.75-2%187.3184.02%

Page 12 of 17

.

. "

{12}------------------------------------------------

12.5.2 Interfering Substances

The Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge was evaluated for interference according to CLSI EP7-A2. Bias is the difference in the results between the control sample (without the interferent) and the test sample (contains the interferent) expressed in percent. Bias exceeding 10% is considered interference. Dilution studies were conducted to determine the level at which the spiked substance no.longer displayed significant interference. These studies were conducted at two creatinine analyte concentrations, if both sample pools show significant interference. This study was conducted as needed for both serum pools and urine pools.

{13}------------------------------------------------

SubstanceConcentration ofSubstance (mg/dL)Mean TestCRE2MeanControl% Diff
Acetone1505.774.8718%
Acetone755.324.879%
Acetone37.55.154.876%
Acetone37.51.591.3915%
Acetone18.751.501.398%
Cefoxitin55.294.976%
Cefoxitin2.51.581.4013%
Cefoxitin1.251.491.406%
Cephalothin151.531.3910.1%
Cephalothin155.094.972%
Cephalothin13.751.491.406%
Glucose400mg/dL1.541.3712%
Glucose300mg/dL1.511.3710%
Glucose200mg/dL1.451.376%
Glucose600mg/dL5.154.915%
Pyruvate10.5 mg/dL5.674.8716%
Pyruvate5.26 mg/dL5.314.928%
Pyruvate1.32 mg/dL1.491.388%
Triglycerides30004.934.3912%
Triglycerides25004.974.579%
Triglycerides20005.014.658%
Triglycerides20001.491.3213%
Triglycerides15001.481.3510%
Triglycerides10001.501.398%

12.5.3 Hemolysis, Icterus, Lipemia (HIL) Interference

The Dimension Vista® Creatinine (CRE2) Flex reagent cartridge was evaluated for interference according to CLSI EP7-A2. Bias is the difference in the results between the control sample (without the interferent) and the test sample (contains the interferent) expressed in percent. Bias exceeding 10% is considered interference.

Page 14 of 17

{14}------------------------------------------------

DV CRE2 HIL Interference
SubstanceConcentration ofSubstance (mg/dL)Mean Test CRE2Result (mg/dL)Mean Control CRE2Result (mg/dL)% Diff
Intralipid 20%2000mg/dL5.274.5615.5%
Intralipid 20%1500mg/dL5.154.719.5%
Intralipid 20%1500mg/dL1.511.3115.1%
Intralipid 20%1000mg/dL1.451.367.0%
Hemoglobin1000mg/dL1.451.357.0%
Hemoglobin1000mg/dL4.874.723.2%
Bilirubin (unconj)40mg/dL0.981.40-29.6%
Bilirubin (unconj)20mg/dL1.441.45-0.2%
Bilirubin (unconj)40mg/dL4.614.93-6.6%
Bilirubin (conj)40mg/dL1.161.38-16.1%
Bilirubin (conj)20mg/dL1.381.44-4.5%
Bilirubin (conj)40mg/dL4.724.94-4.5%
Bilirubin (conj)20mg/dL4.965.04-1.5%

1.15 Expected Values

Serum and Plasma

Transference was used to validate established reference ranges for serum and plasma creatinine following CLSI C28-A3c.

Serum samples from forty normal healthy adult female donors and forty-three normal healthy adult male donors were tested N=1 using the Dimension Vista Creatinine (CRE2) Flex® reagent cartridge. The range was considered validated if ≤10% of the samples fell outside of established range.

GenderMalesFemales
Range0.700 -1.30 mg/dL0.550 -1.02 mg/dL
# Samples5340
# Sample outsideRange52
% Samples outsiderange9.4%5.0%

ﻟﺴﺴﺴﺴ

These data support the published ranges for Males (0.7 to 1.3 mg/dL (Tietz 1999) and Females (0.55 to 1.02 mg/dL (Clin Chem 54:3 (2008)).

Page 15 of 17

{15}------------------------------------------------

Urine

These data support the published ranges for Males (0.7 to 1.3 mg/dL (Tietz 1999) and Females (0.55 to 1.02 mg/dL (Clin Chem 54:3 (2008)).

Similarly published reference ranges were validated for urine specimens using samples from twenty normal healthy adult female donors and twenty-two normal healthy adult male donors following CLSI C28-A3c. Samples were processed N=1 using the Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge. The range was considered validated if ≤10% of the samples fell outside of established range.

GenderMalesFemales
Reference Range0.95-2.49 g/day0.60-1.80 g/day
# Samples2220
# Sample outsideRange21
% Samples outsiderange9.1%5.0%

These data support the published ranges for Males (0.95 to 2.49 g/day (Clin Chem Acta 344 (2004) 137-148) and Females (0.6 to 1.8 g/day (Tietz 1999).

1.16 Results Outside of the Measuring (Reportable) Range

The instrument generates a flag which states "Above Assay Range" or "Below Assay Range" when it obtains a result outside of the measuring range. Serum and plasma samples with results in excess of 20 mg/dL and urine samples above 300 mg/dL are reported as "above Assay Range" and should be repeated on dilution.

Autodilution for Serum and Plasma Samples: The autodilute sample volume is 6.6 uL and recommended dilution factor 1:2 for serum and plasma. This extends the serum and plasma reportable range to 40 mg/dL.

Autodilution (AD) for Urine Samples: Autodilutions for urines are a two-step process. Urine samples automatically get a 1:15 AUD in the aliquot tray followed by a second 1:3 dilution in the aliquot tray making a final 1:45 dilution. This extends the urine reportable range to 900 mg/dL [79560 umol/L]. Refer to your Dimension Vista® Operator's Guide.

These protocols were tested with samples at the upper end of the undiluted range. The mean percent recovery was calculated and met the acceptance criteria of 90 - 110% recovery.

{16}------------------------------------------------

DV CRE2 Reportable Range Dilution of above assay range results

Serum

Serum results > 20 mg/dL can be automatically re-run using 1/2 sample (6.6 µL vs 13.2 µL). This extends the serum reportable range to 40.0 mg/dL.

Samle Volume (µL)Mean of N=5 reps
13.2 µL(normal volume)16.5
6.6 µL (auto-dilute volume)16.0
% Bias-3.0%

Urine

Urine results > 300 mg/dL can be automatically diluted 3x with water and re-run. This extends the urine reportable range to 900 mg/dL.

Mean of N=5 reps
undiluted (normal)268
auto-dilute 3x with water245
% Bias-8.6%

Conclusion

Results support serum extended range up to 40.0 mg/dL, and urine extended range up to 900 mg/dL.

1.17 Conclusion

The Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge (K1033A) is substantially equivalent in principle and performance to the Dimension Vista® Creatinine (CREA) Flex® reagent cartridge cleared under K061238. Comparative testing described in the submission demonstrates substantially equivalent performance.

Page 17 of 17

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three curved shapes that resemble waves or abstract human figures. The logo is printed in black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 24, 2014

SIEMENS HEALTHCARE DIAGNOSTICS INC. LAURA DUGGAN, PH.D. REGULATORY TECHNICAL SPECIALIST P.O. BOX 6101 NEWARK DE 19714-6101

Re: K133728

Trade/Device Name: Dimension Vista® Creatinine (CRE2) Flex® Reagent Cartridge Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: II Product Code: CGX Dated: December 05, 2013 Received: December 6, 2013

Dear Dr. Duggan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{18}------------------------------------------------

Page 2-Dr. Duggan

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance .

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

CourtneyH.Lias -S

Courtney H. Lias. Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{19}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known) K133728

Device Name

Dimension Vista Creatinine (CRE2) Flex Reagent Cartridge

Indications for Use (Describe)

The CRE2 method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista® System. Creatinine measurements are used in the diagnosis and treatment of certain renal discase, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (9/13)

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.