(234 days)
The MICRODENTAL, Inc. MicrO2 device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
The MicrO2 OSA device series consists of maxillary and mandibular devices that when interfaced together alleviate snoring and mild to moderate sleep apnea by holding the mandible forward during sleep, preventing the tongue and soft tissues of the throat from collapsing into the airway. These separate upper and lower arch devices are designed with twin-mated posts and are CAD/CAM generated specifically for each prescription. Designed as a patient-specific device, the MicrO2 series consists of one or multiple lower device(s) together with one or multiple mated-post upper device(s) that are manufactured to the dentist prescriber's requested advancement positions to provide a selection of gentle adjustments according to patient comfort and need. As such, prescribed advancements can be achieved by simply removing the current upper or lower device and inserting the next upper or lower device in the mandibular advancing series. The MicrO2 does not have any adjustment mechanisms to modify or maintain the mandibular position such as pistons, straps or repositioning elastics. The maximum protrusion of a MicrO2 archform is 6mm, which is comparable to the predicate device's 6mm advancement screw, measuring from plate to plate. This also negates the need for patients utilizing external controlling components such as keys, screwdrivers or ligature ties. The maximum protrusion of a MicrO2 archform is 6mm, which is comparable to the predicate device. The MicrO2's twin-mated post and lingualess design yields a small and a comfortable patient-specific mandibular advancement device. The design of the device maximizes tongue space and mandibular movement resulting in the ability to open and close during wear. The Micr02 OSA device is identical to the Somnomed MAS RXA (Classic) except for the difference in the adjustment mechanism. Any differences introduced by these modifications when compared to the predicate device, do not introduce any new safety concerns.
The provided text describes a 510(k) summary for the MicrO2 Obstructive Sleep Apnea Device. This document focuses on demonstrating substantial equivalence to a predicate device through bench testing. It does not contain information about clinical studies with human participants to directly prove the device meets specific acceptance criteria related to its efficacy in reducing snoring or sleep apnea in patients.
Therefore, many of the requested sections (2, 3, 4, 5, 6, 7, 8, 9) cannot be filled based on the provided text.
Here's an analysis of the available information:
1. A table of acceptance criteria and the reported device performance
The acceptance criteria are implicitly defined by performing "as well or better than the predicate device." The "reported device performance" is a qualitative assessment of "Yes" for improved or equal performance.
| Test Design | Acceptance Criteria (against predicate device) | Reported Device Performance (MicrO2 OSA) |
|---|---|---|
| AP Direction Fin Test | Equal or improved Mean Shear Strength (lbf) | Yes (Improved/Equal) |
| Lateral Direction Fin Test | Equal or improved Mean Shear Strength (lbf) | Yes (Improved/Equal) |
| Cross-Arch Compression Test | Equal or improved Mean Shear Strength (lbf) | Yes (Improved/Equal) |
| Leachability Test (ug/g) | Equal or improved leachability | Yes (Improved/Equal) |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the text. The testing described is bench testing, not clinical testing on a patient population.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable as the provided text describes bench testing (mechanical and chemical property comparison), not a study requiring expert ground truth for clinical outcomes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as the provided text describes bench testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The device is a physical intraoral device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable. The device is a physical intraoral device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the bench testing, the "ground truth" was established by comparing the mechanical and chemical properties of the MicrO2 OSA device directly against those of the predicate device (Somnomed MAS RXA). The predicate device's established performance served as the benchmark.
8. The sample size for the training set
This information is not applicable. There is no mention of a "training set" as this is not an AI/machine learning study. The document describes bench tests on physical devices.
9. How the ground truth for the training set was established
This information is not applicable for the reasons stated above.
{0}------------------------------------------------
JUL 2 4 2014
510(k) Summary 6
MicroDental Laboratories is hereby submitting a 510(k) Summary as per 21 CFR 807.92.
| 1. | Date Summary Prepared: | November 15, 2013 |
|---|---|---|
| 2. | Submitter Information: | MICRODENTAL, Inc. |
| 5601 Arnold Rd | ||
| Dublin, CA 94568 | ||
| 3. | Contact Person: | Laura Sheppard |
| Senior Director of Compliance and Regulatory | ||
| 4. Device Name: | ||
| Proprietary/Trade name: | MicrO2 Obstructive Sleep Apnea Device | |
| Common name: | Mandibular Advancement Device | |
| Classification name: | Intraoral devices for snoring and mild to moderate obstructivesleep apnea | |
| Device Classification: | II |
21 CFR 872.5570
LAK
5. Identification of Predicate Device(s)
Regulation number:
Product code:
Device Name: Somnomed MAS RXA Manufacturer: SomnoMed, Ltd. 510(k) Number: K050592
The predicate device has the same intended use and uses similar materials as the MicrO2 OSA device.
6. Device Description:
The MicrO2 OSA device series consists of maxillary and mandibular devices that when interfaced together alleviate snoring and mild to moderate sleep apnea by holding the mandible forward during sleep, preventing the tongue and soft tissues of the throat from collapsing into the airway. These separate upper and lower arch devices are designed with twin-mated posts and are CAD/CAM generated specifically for each prescription. Designed as a patient-specific device, the MicrO2 series consists of one or multiple lower device(s) together with one or multiple mated-post upper device(s) that are manufactured to the dentist prescriber's requested advancement positions to provide a selection of gentle adjustments according to patient comfort and need. As such, prescribed advancements can be achieved by simply removing the current upper or lower device and inserting the next upper or lower device in the mandibular advancing series. The MicrO2 does not have any adjustment mechanisms to modify or
{1}------------------------------------------------
maintain the mandibular position such as pistons, straps or repositioning elastics. The maximum protrusion of a MicrO2 archform is 6mm, which is comparable to the predicate device's 6mm advancement screw, measuring from plate to plate. This also negates the need for patients utilizing external controlling components such as keys, screwdrivers or ligature ties. The maximum protrusion of a MicrO2 archform is 6mm, which is comparable to the predicate device. The MicrO2's twin-mated post and lingualess design yields a small and a comfortable patient-specific mandibular advancement device. The design of the device maximizes tongue space and mandibular movement resulting in the ability to open and close during wear. The Micr02 OSA device is identical to the Somnomed MAS RXA (Classic) except for the
difference in the adjustment mechanism. Any differences introduced by these modifications when compared to the predicate device, do not introduce any new safety concerns.
7. Indications for Use:
The MICRODENTAL, Inc. MicrO2 device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
8. Bench Testing in Support of Substantial Equivalence:
The results of bench testing and pre-clinical testing support the above indications for use as well as the claim of Substantial Equivalence. The substantial equivalence was demonstrated through the materials that were used by the predicate devices as well as the technical and performance characteristics compared to the predicate. Compression and shear testing are important performance characteristics as they approximate the mechanical force applied by during normal use of the device. Bench testing results are listed in the table below. Multiple tests were performed to address fin fracture and arch fracture risks, evaluating the strength of the Micro2 device when a patient applies muscular forces during wear or physical forces during patient mishandling of the device. These tests were identified by Risk Analysis conducted in accordance with ISO-14971.
The MicrO2 OSA device uses the same materials as those in the predicate devices and is a well known material for dental applications. The material manufacturing processes between Micr02 OSA device and the predicate device are identical. No new chemicals or formulas have been introduced. As such, additional biocompatibility testing was not conducted. Monomer leachablity testing was performed and compared to the predicate device.
22
{2}------------------------------------------------
9. Conclusions from Bench Testing
The MicrO2 OSA device has completed all of the internal and standards testing above with acceptable results, demonstrating substantial equivalence to the referenced predicate. The subject under test performed as well or better than the predicate device, the Somnomed MAS RXA.
| Test Design | Devices | Equal or Improved based onMean Shear Strength (lbf) |
|---|---|---|
| AP Direction Fin Test | Somnomed MAS RXA | |
| Micr02 OSA | Yes | |
| Lateral Direction FinTest | Somnomed MAS RXA | |
| Micr02 OSA | Yes | |
| Cross-Arch CompressionTest | Somnomed MAS RXA | |
| Micr02 OSA | Yes | |
| Leachability Test**Based on ug/g | Somnomed MAS RXA | |
| Micr02 OSA | Yes |
- Substantial Equivalence Conclusion
| Substantial Equivalent Table | MicrO2 OSA Device | Somnomed MAS RXA |
|---|---|---|
| Intended Use | ||
| Intended as an intraoral device | YES | YES |
| Intended to be worn during sleep only | YES | YES |
| Intended to reduce snoring or help alleviate snoring | YES | YES |
| Intended to reduce or help alleviate mild to moderate obstructive sleep apnea | YES | YES |
| For single patient, multi-use | YES | YES |
| For personal use at home or in sleep laboratories | YES | YES |
| For adult patient populations | YES | YES |
| Prescription device | YES | YES |
| Cleaned daily | YES | YES |
| Design | ||
| Separate upper & lower tray pleces | YES | YES |
| All rigid tray pieces for more stable fit | YES | YES |
| Patient Specific | YES | YES |
{3}------------------------------------------------
| Advances the lower jaw | YES | YES |
|---|---|---|
| Advancement position maintained by ramp design | YES | YES |
| Can be adjusted, repaired and/or refit | YES | YES |
| Ease of Insertion | YES | YES |
| Ease of removal | YES | YES |
| Allows the patient to open & close mouth during wear | YES | YES |
| Permits patient to breathe through the mouth | YES | YES |
| Material | ||
| Hard PMMA material | YES | YES |
| Ability to be manufactured from alternate materials based on patient needs | YES | YES |
| Non-sterile | YES | YES |
In summary, Indications for Use, patient population, Use environment and the performance data between the MICRODENTAL, Inc. MicrO2 OSA device, and the predicate device listed in Item #5 above shows nearly identical data.
From a technical design perspective, both of the devices improve the flow of air through the patient's pharyngeal space during sleep by repositioning the mandible and the use of upper and lower trays. While the device mechanics are slightly different, they both advance the lower jaw forward in order to increase the pharyngeal space. The mechanism of each device allows for adjustment of the amount of advancement based on dentist prescription. The dentist prescribes the amount of advancement increments according the physician professional assessment in the same way that the dentist adjusts the mandibular position with the pistons, screws, straps or repositioning elastics of the predicate devices.
Both devices use primarily the same material - PMMA - to provide a rigid tray.
There are no new questions of safety or efficacy raised by the MicrO2 OSA device; therefore, the device supports a claim of Substantial Equivalence.
24
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and three horizontal bars extending from the eagle's body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 24, 2014
MICRODENTAL, Inc. Ms. Laura Sheppard Senior Director, Compliance and Regulatory Affairs 5601 Arnold Rd Dublin, CA 94568
Re: K133683
Trade/Device Name: MicrO2 Obstructive Sleep Apnea Device Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II Product Code: LRK Dated: June 20, 2014 Received: June 24, 2014
Dear Ms. Sheppard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Sheppard
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mary Publisher-S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use 5
Indications for Use
To be assigned K133683 510(k) Number (if known):
Device Name: MicrO2 OSA Device
Indications for Use:
The MICRODENTAL, Inc. Micr02 device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Prescription Use __ YES_ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use NO_ AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ of
N 2014.07
§ 872.5570 Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea.
(a)
Identification. Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.(b)
Classification. Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”