(93 days)
K 130584
No
The document describes a multi-parameter patient monitor with standard physiological monitoring capabilities and software improvements related to alarms and waveform scaling. There is no mention of AI or ML in the intended use, device description, or the sections specifically checked for these terms.
No.
The device is a patient monitor, designed to observe and record physiological parameters, not to treat or cure a medical condition.
No
This device is described as a "multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters." It provides continuous monitoring of vital signs and includes features for alarm generation, but it does not state that it provides a diagnosis or aids in making a diagnosis.
No
The device description explicitly mentions hardware components such as a 12.1 inch LCD display, mounting plate, and interfaces with optional extension modules, indicating it is not solely software.
Based on the provided text, the Monitor B40 is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: In Vitro Diagnostics are devices intended for use in the collection, preparation, and examination of specimens taken from the human body (such as blood, urine, or tissue) to obtain information for diagnostic or monitoring purposes.
- Monitor B40 Function: The Monitor B40 is described as a "portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients." It directly measures physiological signals from the patient's body (ECG, blood pressure, SpO2, temperature, respiration, airway gases, Entropy).
- Lack of Specimen Handling: The description does not mention the collection, preparation, or analysis of any specimens taken from the patient. The measurements are taken on the patient.
Therefore, the Monitor B40 falls under the category of a patient monitoring device, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The Monitor B40 is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
The Monitor B40 is intended for use under the direct supervision of a licensed health care practitioner.
The Monitor B40 is not intended for use during MRI.
The Monitor B40 can be a stand-alone monitor or interfaced to other devices via a network.
The Monitor B40 monitors and displays: ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring.
Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and Entropy.
Product codes (comma separated list FDA assigned to the subject device)
MHX, BZQ, CBR, CBS, CBQ, CCK, CCL, DXN DQA, DRT, DSB, DSK,GWQ, FLL,NHO, NHP, NHQ
Device Description
The proposed Monitor B40V2.1 still is a multi-parameter patient monitor. It retains the features of the predicate Monitor B40V2 (K130584) and now includes supporting for an additional optional extension module Airway Gas Option (N-CAiO), and few software improvements by adding alarm breakthrough, extending the upper limit of ECG PVC (Premature Ventricular Contraction) alarm and providing four waveform scale options for Masimo SpO2 and Nellcor SpO2. Same as the predicate Monitor B40V2 (K130584), the proposed Monitor B40V2.1 continues interfacing with following optional extension modules: E-MiniC module (K052582), CARESCAPE Respiratory modules (E-sCO and E-sCAiO) (K123195) and E-Entropy module (K061907). The compatibility with CARESCAPE Respiratory modules (E-sCOV and E-sCAiOV) (K123195) is also provided but with spirometry function disabled.
The proposed Monitor B40V2.1 still has a 12.1 inch LCD display but is from different LCD manufacturer and LCD backlight is changed from CCFL to LED due to RoHS compliance.
As with the predicate Patient Monitor B40V2 (K130584), the proposed Patient Monitor B40V2.1 still includes features and subsystems that are optional or configurable. The proposed Patient Monitor B40V2.1 will continue interfacing to a variety of existing central station systems via a cabled network interface.
As with the predicate Patient Monitor B40V2 (K130584), the proposed Patient Monitor B40V2.1 keeps a mounting plate on the bottom of the monitor. The monitor can be mounted in a variety of ways (e.g. shelf, countertop, table, wall, pole, or head/foot board) using existing mounting accessories.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adult, pediatric, and neonatal patients
Intended User / Care Setting
licensed health care practitioner, hospital environment and during intra-hospital transport.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The proposed Monitor B40V2.1 and its applications comply with voluntary standards as detailed in this premarket submission. The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- Requirements Reviews .
- . Design Reviews
- Testing on unit level (Module verification) .
- Integration testing (System verification) .
- . Final acceptance testing (Validation)
- Performance testing (Verification) .
- Safety testing (Verification) .
The subject of this premarket submission, The proposed Monitor B40V2.1 did not require clinical studies to support substantial equivalence.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K 130584 Monitor B40
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
0
FFR 2 1 2014
8200 West Tower Avenue Milwaukee, Wisconsin, 53223
GE Medical Systems Information Technologies gemedicalsystems.com
510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided: Date: Nov. 20, 2013
Submitter: Sun YanLi Regulatory Affairs Manager GE MEDICAL SYSTEMS CHINA CO., LTD. No. 19 Changjiang road National Hi-Tech Dev. Zone Wuxi, Jiangsu, China 214028
Primary Contact Person:
Robert Casarsa Regulatory Affairs Leader GE Medical Systems Information Technologies, Inc. Telephone: 414-362-3063 Fax at 414-362-2585 E-mail: Robert.casarsa@ge.com
Secondary Contact Douglas Kentz Person: Regulatory Affairs Director GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, Wisconsin 53223 Phone: 414 362-2038 Fax: 414-262-2585 E-mail: Douglas.kentz@ge.com
Trade Monitor B40 Device: Name: Common/Usual Multi-parameter patient monitor Name:
1
21 CFR 870.1025 monitor, physiological, patient(with arrhythmia Classification Names: detection or alarms) 21 CFR 868.2375 monitor, breathing frequency 21 CFR 868.1700 analyzer, gas, nitrous-oxide, gaseous phase (anesthetic conc.) 21 CFR 868.1620 analyzer, gas, halothane, gaseous-phase (anesthetic conc.) 21 CFR 868.1500 analyzer, gas, enflurane, gaseous-phase (anesthetic concentration) 21 CFR 868.1400 analyzer, gas, carbon-dioxide, gaseous-phase 21 CFR 868.1720 analyzer, gas, oxygen, gaseous-phase 21 CFR 870.1130 system, measurement, blood-pressure, noninvasive 21 CFR 870.2700 oximeter 21 CFR 870.2300 monitor, cardiac (incl. cardiotachometer & rate alarm) 21 CFR 870.2770 plethysmograph, impedance 21 CFR 870.1110 computer, blood-pressure 21 CFR 882.1400 full-montage standard electroencephalograph 21 CFR 880.2910 thermometer, electronic, clinical 21 CFR 868.1500 analyzer, gas, desflurane, gaseous-phase (anesthetic concentration) 21 CFR 868.1500 analyzer, gas, sevoflurane, gaseous-phase (anesthetic concentration) 21 CFR 868.1500 analyzer, gas, isoflurane, gaseous-phase (anesthetic concentration) Product Code: MHX, BZQ, CBR, CBS, CBQ, CCK, CCL, DXN DQA, DRT, DSB, DSK,GWQ, FLL,NHO, NHP, NHQ Predicate Device(s): K 130584 Monitor B40 The proposed Monitor B40V2.1 still is a multi-parameter patient Device Description: monitor. It retains the features of the predicate Monitor B40V2 (K130584) and now includes supporting for an additional optional extension module Airway Gas Option (N-CAiO), and few software improvements by adding alarm breakthrough, extending the upper limit of ECG PVC (Premature Ventricular Contraction) alarm and providing four waveform scale options for Masimo SpO2 and Nellcor SpO2. Same as the predicate Monitor B40V2 (K130584), the proposed Monitor B40V2.1 continues interfacing with following optional extension modules: E-MiniC module (K052582), CARESCAPE Respiratory modules (E-sCO and E-sCAiO) (K123195) and E-Entropy module (K061907). The compatibility with
2
CARESCAPE Respiratory modules (E-sCOV and E-sCAiOV) (K123195) is also provided but with spirometry function disabled.
The proposed Monitor B40V2.1 still has a 12.1 inch LCD display but is from different LCD manufacturer and LCD backlight is changed from CCFL to LED due to RoHS compliance.
As with the predicate Patient Monitor B40V2 (K130584), the proposed Patient Monitor B40V2.1 still includes features and subsystems that are optional or configurable. The proposed Patient Monitor B40V2.1 will continue interfacing to a variety of existing central station systems via a cabled network interface.
As with the predicate Patient Monitor B40V2 (K130584), the proposed Patient Monitor B40V2.1 keeps a mounting plate on the bottom of the monitor. The monitor can be mounted in a variety of ways (e.g. shelf, countertop, table, wall, pole, or head/foot board) using existing mounting accessories.
Intended Use: The Monitor B40 is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
·
The Monitor B40 is intended for use under the direct supervision of a licensed health care practitioner.
The Monitor B40 is not intended for use during MRI.
The Monitor B40 can be a stand-alone monitor or interfaced to other devices via a network.
The Monitor B40 monitors and displays: ECG (including ST segment. arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring.
Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/S kin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and
3
respiratory rate) and Entropy.
| Technology: | The proposed Monitor B40V2.1 employs the same functional
scientific technology as the predicate Monitor B40V2 (K120598). |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Determination of
Substantial Equivalence: | Changes from the predicate Monitor B40V2(K130584) |
| | Supporting Optional Extension Module: Airway Gas Option
(N-CAIO) |
| | The Airway Gas Option (N-CAIO) is identical to
CARESCAPE Respiration module (E-sCAIO) (K123195)
except different color and base material of the front bezel.
There are three functions from the N-CAIO module that are
not supported by the B40V2.1. The B40V2.1 disables the N-
CAIO's Secondary Anesthetic Agent, MACage and Balance
Gas parameters. |
| | Alarm breakthrough
Provided new Alarm Breakthrough feature that allows the
proposed Monitor B40V2.1 to give alarm sound for certain
predefined alarm conditions even if alarm sounds are turned
off in the monitor. |
| | ECG PVC Alarm Inner Limit Change |
- ECG PVC Alarm Upper Limit Change . Extended ECG Premature Ventricular Contraction the upper limit adjustable range from 30/min to 100/min to provide more flexibility for user to set alarm limit: "1-100/ min" from lower to upper limits.
- . SpO2 Waveform Scale Change for Masimo and Nellcor SpO2 Changed the SpO2 waveform scale for Masimo SpO2 and Nellcor SpO2 from fixed 2X to four options (1X, 2X, 4X and 8X).
- . LCD Display Change Replaced the existing 12 inch Sharp LCD display with 12 inch TianMa LCD display that uses the LED instead of CCFL backlight for RoHS compliance purpose.
- . Instruction Manual Change Added instruction for extension Module Airway Gas Option (N-CAiO). Added instruction for alarm breakthrough. Modified instruction for ECG PVC alarm limits, SpO2 waveform scale and new LCD display.
4
Summary of Non-Clinical Tests:
The proposed Monitor B40V2.1 and its applications comply with voluntary standards as detailed in this premarket submission. The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- Requirements Reviews .
- . Design Reviews
- Testing on unit level (Module verification) .
- Integration testing (System verification) .
- . Final acceptance testing (Validation)
- Performance testing (Verification) .
- Safety testing (Verification) .
Summary of Clinical Tests:
The subject of this premarket submission, The proposed Monitor B40V2.1 did not require clinical studies to support substantial equivalence.
- The design changes made to the proposed Monitor B40V2.1 have Conclusion: no effect on the device's ability to obtain patient measurements as there are no changes to the parameter measuring hardware. To assess if the changes had any significant impact to the device, all related risks were re-evaluated and found to be unchanged. GE Healthcare considers the proposed Monitor B40V2.1 to be as safe, as effective, and performance is substantially equivalent to the predicate device.
5
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract representation of a human figure or a caduceus.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 21, 2014
GE Medical Systems China Co., Ltd. Robert Casarsa 8200 West Tower Ave Milwaukee, WI 53223 US
Re: K133576
Trade/Device Name: Monitor b40 Regulation Number: 21 CFR 870.1025 Regulation Name: Multiparameter Patient Monitor Regulatory Class: Class II Product Code: MHX, BZQ, CBR, CBS, CBQ, CCK, CCL, DXN DQA, DRT, DSB, DSK, GWQ, FLL, NHO, NHP, NHQ Dated: January 24, 2014 Received: January 21, 2014
Dear Robert Casarsa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
6
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Owen P. Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
7
Indications for Use
5 l 0(k) Number (if known): _K13 3 5 7 6
Device Name: Monitor B40
Indications for use:
The Monitor B40 is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
The Monitor B40 is intended for use under the direct supervision of a licensed health care practitioner.
The Monitor B40 is not intended for use during MRI.
The Monitor B40 can be a stand-alone monitor or interfaced to other devices via a network.
The Monitor B40 monitors and displays: ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring.
Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myo cardial/Core/Surface temperature, impedance respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and Entropy.
Over-The-Counter Use Prescription Use × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW TIJIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by
Owen P. Faris -S
Date: 2014.02.21
10:47:24-05'00'
Page 1 of 1