K Number
K133534
Date Cleared
2014-04-18

(151 days)

Product Code
Regulation Number
870.1425
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

McKesson Cardiology ECG Mobile is a software application used for accessing and displaying ECG data and related patient information previously stored, analyzed or retrieved by the Horizon Cardiology ECG Management device. The McKesson Cardiology ECG Mobile software application is intended to be used from a mobile device to perform the following:

  • View ECG test results, such as waveforms, synopsis measurements and diagnosis statements, as well as other relevant current or historical patient information originally stored, analyzed or retrieved by the Horizon Cardiology ECG Management device.
  • View the ECG lead traces using different display settings.
  • . Compare the results of current ECG tests with any previous ECG test results stored on the Horizon Cardiology ECG Management device.
  • . Perform ECG-related measurements and edit synopsis measurement values stored on the Horizon Cardiology ECG Management device.
  • Enter new diagnosis statements and/or modify diagnosis records stored on the Horizon . Cardioloģy ECG Management device.
  • . Confirm ECG procedures and results.
  • . Communicate information with the Horizon Cardiology ECG Management device such as login credentials and user settings.

The McKesson Cardiology ECG Mobile application is intended to be used under the direct supervision of a licensed healthcare practitioner and by trained operators.

The McKesson Cardiology ECG Mobile application is not intended for real-time monitoring.

Device Description

McKesson Cardiology ECG Mobile provides secure access to ECG records and related information contained on the Horizon Cardiology ECG Management device. The McKesson Cardiology ECG Mobile software functions as an accessory to the Horizon Cardiology ECG Management device, and requires a WiFi or cellular connection to a pre-installed and properly configured Horizon Cardiology ECG Management device. Through providing remote access to the ECG data stored on the McKesson Cardiology ECG Management device, the user is able to review current and previous ECG tests results, perform measurements on the ECG waveforms, and edit and confirm the ECG test results. McKesson Cardiology ECG Mobile does not store ECG or patient related information on the mobile device, does not directly communicate with cardiographs or other waveform acquisition devices, and does not use any automatic electronic data processing and pattern recognition methods to derive measurements (e.g. intervals and amplitudes) or provide diagnostic statements from the ECG data. The device does not allow modification of the original ECG traces (waveforms) stored on the Horizon Cardiology ECG Management device.

The McKesson Cardiology ECG Mobile application is not intended to replace the functionalities provided by the Horizon Cardiology ECG Management desktop client but to extend those to make selected functionalities described above available via mobile devices.

The McKesson Cardiology ECG Mobile functions as a non-real time system and is not intended for real time monitoring.

McKesson Cardiology ECG Mobile can be used both inside of medical facilities except in areas where cellular phone or wireless device use is prohibited.

McKesson Cardiology ECG Mobile is intended to operate on Apple®iPad running iOS mobile operating system from Apple Inc.

AI/ML Overview

This 510(k) summary describes the McKesson Cardiology™ ECG Mobile, a software application designed to provide remote access to ECG data and related patient information from a mobile device.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly define quantitative acceptance criteria for the McKesson Cardiology™ ECG Mobile. Instead, the performance data focuses on qualitative outcomes and ensuring the device functions as intended and is substantially equivalent to predicate devices.

Acceptance Criteria (Implied)Reported Device Performance
Met all design specifications and requirements"McKesson Cardiology™ ECG Mobile met all design specifications and requirements."
Operability with the predicate Horizon Cardiology ECG Management device"Unit and system level testing included assurance of operability with the predicate Horizon Cardiology ECG Management device."
User accuracy qualification of ECG waveform and related information representation"User accuracy qualification of ECG waveform and related information representation in a simulated user test environment."
Display and measurements acquisition capabilities verified in clinical qualification"Clinical Qualification testing performed verified the McKesson Cardiology ECG Mobile display and measurements acquisition capabilities."
Adequate user readability and image quality on a small screen"Testing was performed to ensure adequate user readability and image quality on a small screen."
Users with varying qualifications could use the device (Human Factors)"Testing was performed to ensure... that users with varying qualifications could use the device."
Functions as intended by design requirements"In all instances, the McKesson Cardiology™ ECG Mobile functioned as intended by the design requirements."
Substantial equivalence to predicate devices"The observed results demonstrated substantial equivalence with the predicate devices."

2. Sample Sizes Used for the Test Set and Data Provenance

  • Clinical Qualification Testing: Sample cases were "selected by a clinician based on technical characteristics and relevancy to the intended function." The exact number of samples is not specified.
  • Usability Testing (Human Factors): Conducted by "three (3) U.S. medical facilities." The number of individual users involved in this testing is not specified, but it implies a prospective approach within the US.
  • Data Provenance: The usability testing was conducted in "U.S. medical facilities," indicating prospective data collection from the US. The origin of the sample cases for clinical qualification is not explicitly stated but implies relevance to clinical practice.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Clinical Qualification Testing: "Sample cases selected by a clinician." The number of clinicians is not specified. Their qualifications are not explicitly detailed beyond being a "clinician."
  • Usability Testing: The document mentions "users with varying qualifications" without specifying if these users were deemed "experts" for ground truth establishment. The focus of usability testing is on human factors and device interaction rather than establishing medical ground truth.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The "clinical qualification" involved a clinician selecting sample cases and assessing the device's display and measurement capabilities against "the original data as presented by Horizon Cardiology ECG Management." This suggests a direct comparison against a reference system rather than a consensus-based adjudication among multiple experts.

5. Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader, multi-case (MRMC) comparative effectiveness study is mentioned. The device is purely for viewing and editing existing ECG data, not for independent diagnostic interpretation or AI-assisted interpretation, therefore, comparing human readers with and without AI assistance is not applicable to its intended use.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This device is a software application for viewing and editing ECG data previously stored and analyzed by another device (Horizon Cardiology ECG Management). It "does not use any automatic electronic data processing and pattern recognition methods to derive measurements (e.g. intervals and amplitudes) or provide diagnostic statements from the ECG data." Therefore, a standalone performance study of an algorithm without human-in-the-loop performance is not applicable to this device, as it does not contain such algorithms.

7. Type of Ground Truth Used

The ground truth used for performance assessment (specifically clinical qualification) was "the original data as presented by Horizon Cardiology ECG Management." This implies that the Horizon Cardiology ECG Management system serves as the reference standard for waveforms and data representation.

8. Sample Size for the Training Set

The document does not mention a "training set" as the device does not employ machine learning or AI algorithms that would require such data.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

510(k) SUMMARY McKesson Israel Ltd.'s McKesson Cardiology™ ECG Mobile

McKesson Israel Ltd. 4 HaNechoshet Street Tel Aviv lsrael 69710 Contact Person: Tomer Levy, VP Engineering Phone: +972 (0)3 7698000 Facsimile: +972 (0)3 6478593 Date Prepared: April 2, 2014 Name of the device: McKesson Cardiology™ ECG Mobile Common Name: McKesson Cardiology™ ECG Mobile Classification Name: Programmable diagnostic computer Classification Regulation: 21 CFR § 870.1425 Product code: DQK Device Class: Class II Predicate Device: McKesson Israel Ltd., Horizon Cardiology ECG Management (K113515). AirStrip Technologies, LP, AirStrip Remote Patient Monitoring (RPM) Remote Data Viewing (K122133)

Intended Use / Indications for Use

McKesson Cardiology ECG Mobile is a software application used for accessing and displaying ECG data and related patient information previously stored, analyzed or retrieved by the Horizon Cardiology ECG Management device. The McKesson Cardiology ECG Mobile software application is intended to be used from a mobile device to perform the following:

  • View ECG test results, such as waveforms, synopsis measurements and diagnosis statements, as well as other relevant current or historical patient information originally stored, analyzed or retrieved by the Horizon Cardiology ECG Management device.
  • View the ECG lead traces using different display settings.
  • . Compare the results of current ECG tests with any previous ECG test results stored on the Horizon Cardiology ECG Management device.
  • . Perform ECG-related measurements and edit synopsis measurement values stored on the Horizon Cardiology ECG Management device.
  • Enter new diagnosis statements and/or modify diagnosis records stored on the Horizon . Cardioloģy ECG Management device.
  • . Confirm ECG procedures and results.
  • . Communicate information with the Horizon Cardiology ECG Management device such as login credentials and user settings.

McKesson Cardiology ECG Mobile 510(K) Summary

APR 1 8 2014

{1}------------------------------------------------

McKesson Cardiology™ ECG Mobile application is intended to be used under the direct supervision of a licensed healthcare practitioner and by trained operators. McKesson Cardiology™ ECG Mobile is not intended for real time monitoring.

Technological Characteristics

McKesson Cardiology ECG Mobile provides secure access to ECG records and related information contained on the Horizon Cardiology ECG Management device. The McKesson Cardiology ECG Mobile software functions as an accessory to the Horizon Cardiology ECG Management device, and requires a WiFi or cellular connection to a pre-installed and properly configured Horizon Cardiology ECG Management device. Through providing remote access to the ECG data stored on the McKesson Cardiology ECG Management device, the user is able to review current and previous ECG tests results, perform measurements on the ECG waveforms, and edit and confirm the ECG test results. McKesson Cardiology ECG Mobile does not store ECG or patient related information on the mobile device, does not directly communicate with cardiographs or other waveform acquisition devices, and does not use any automatic electronic data processing and pattern recognition methods to derive measurements (e.g. intervals and amplitudes) or provide diagnostic statements from the ECG data. The device does not allow modification of the original ECG traces (waveforms) stored on the Horizon Cardiology ECG Management device.

The McKesson Cardiology ECG Mobile application is not intended to replace the functionalities provided by the Horizon Cardiology ECG Management desktop client but to extend those to make selected functionalities described above available via mobile devices.

The McKesson Cardiology ECG Mobile functions as a non-real time system and is not intended for real time monitoring.

McKesson Cardiology ECG Mobile can be used both inside of medical facilities except in areas where cellular phone or wireless device use is prohibited.

McKesson Cardiology ECG Mobile is intended to operate on Apple®iPad running iOS mobile operating system from Apple Inc.

Performance Data

Non-clinical testing performed included software verification, and security testing to ensure that the McKesson Cardiology™ ECG Mobile met all design specifications and requirements. Unit and system level testing included assurance of operability with the predicate Horizon Cardiology ECG Management device and user accuracy qualification of ECG waveform and related information representation in a simulated user test environment.

Performance testing additionally included clinical qualification and usability testing (human factors testing) involving intended McKesson Cardiology ECG Mobile clinical setting for reviewing, comparing, and rating the use and display quality of the ECG Mobile device.

Clinical Qualification testing performed verified the McKesson Cardiology ECG Mobile.display and measurements acquisition capabilities using sample cases selected by a clinician based on technical characteristics and relevancy to the intended function of the McKesson Cardiology ECG Mobile application. The testing considered parameters such as gender, age groups and clinical characteristics,

McKesson Cardiology ECG Mobile 510(K) Summary

{2}------------------------------------------------

different display manipulations (e.g. zoom, comparing the morphology, editing the synopsis value, navigating between screens), display quality and accuracy when compared to the original data as presented by Horizon Cardiology ECG Management.

Usability testing was conducted by three (3) U.S. medical facilities and in environment simulating a real use of the application for validating that human factors in using the device were sufficiently addressed. In particular, the testing was performed to ensure adequate user readability and image quality on a small screen and that users with varying qualifications could use the device.

In all instances, the McKesson Cardiology™ ECG Mobile functioned as intended by the design requirements, and the observed results demonstrated substantial equivalence with the predicate devices.

Substantial Equivalence

McKesson Cardiology™ ECG Mobile has the same intended use, similar indications, technological characteristics and principles of operation as the predicate devices. All of the devices display ECG patient and procedure data, and allow for editing, interpretation, and confirmation by an over-reading physician.

Any differences between the predicate devices and McKesson Cardiology™ ECG Mobile do not raise any new types of safety or efficacy questions. In summary McKesson Israel, Ltd. is of the opinion that the McKesson Cardiology™ ECG Mobile is substantially equivalent to the company's Horizon Cardiology™ ECG Management (K113515), as well as AirStrip Technologies, LP, Airstrip Remote Patient Monitoring (RPM) Remote Data Viewing (K122133).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings outstretched, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The logo is black and white and appears to be a scanned image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-00002

May 2, 2014

McKesson Israel Ltd. Paul Sumner 5995 Windward Parkway Alpharetta, GA 30005 US

Re: K133534

Trade/Device Name: McKesson Cardiology ECG Mobile Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK Dated: March 20, 2014 Received: March 25, 2014

Dear Paul Sumner,

This letter corrects our substantially equivalent letter of April 18, 2014.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Paul Sumner

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Athel Blair

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication for Use Statement

510(k) Number (if known):

Device Name: McKesson Cardiology™ ECG Mobile

Indications for Use:

McKesson Cardiology ECG Mobile is a software application used for accessing and displaying ECG data and related patient information previously stored, analyzed or retrieved by the Horizon Cardiology ECG Management device. The McKesson Cardiology ECG Mobile software application is intended to be used from a mobile device to perform the following:

  • View ECG test results, such as waveforms, synopsis measurements and diagnosis . statements, as well as other relevant current or historical patient information originally stored, analyzed or retrieved by the Horizon Cardiology ECG Management device.
  • View the ECG lead traces using different display settings.
  • Compare the results of current ECG tests with any previous ECG test results stored on the . Horizon Cardiology ECG Management device.
  • Perform ECG-related measurements and edit synopsis measurement values stored on the ● Horizon Cardiology ECG Management device.
  • . Enter new diagnosis statements and/or modify diagnosis records stored on the Horizon Cardiology ECG Management device.
  • . Confirm ECG procedures and results.
  • Communicate information with the Horizon Cardiology ECG Management device such as . login credentials and user settings.

The McKesson Cardiology ECG Mobile application is intended to be used under the direct supervision of a licensed healthcare practitioner and by trained operators.

The McKesson Cardiology ECG Mobile application is not intended for real-time monitoring.

X ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------Prescription Use Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Date: 2014.06.06 14:29:36 -04'00'

§ 870.1425 Programmable diagnostic computer.

(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).