K Number
K133359

Validate with FDA (Live)

Device Name
ENDORINGS
Manufacturer
Date Cleared
2014-02-11

(103 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EndoRings is intended to be attached to the distal end of the endoscope to facilitate endoscopic therapy, to be used for the following: Keeping the suitable depth of endoscope's view field

Device Description

The EndoRings is a flexible silicone rubber add-on device designed to attach to of the designated endoscope. The product features a short tube like shape with three flexible circular layers of wings that easily fold backwards during intubation and scope forward movement, and fold forward during scope withdrawal, to maintain the field of view and scope centering position.

AI/ML Overview

This submission describes a medical device, the EndoRings, and its comparison to a predicate device, the Arc EndoCuff, for demonstrating substantial equivalence. The information provided is primarily focused on the device's technical characteristics, biocompatibility, and a feasibility clinical study. There is no mention of an AI/ML component in this 510(k) summary. Therefore, many of the requested categories related to AI/ML performance (e.g., test set, training set, ground truth methods, MRMC study, standalone performance) are not applicable.

Here's an analysis of the provided text based on the requested criteria:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Study Objectives / Endpoints)Reported Device Performance
Primary Endpoint: Reaching the cecumAchieved in 100% of cases.
Secondary Endpoint: Incidence of complicationsNo reports of adverse events, colon tissue damage, or other complications during the procedure or in the 24 hours post-procedure follow-up.
Secondary Endpoint: Ability to complete therapeutic interventionsAll applicable answers were scored as "Good" to "Excellent". All interventions were successfully completed without operational difficulties. The device did not interfere with inserting tools through the working channel or maneuvering the scope distal tip.
Secondary Endpoint: Procedure timeNot explicitly stated as a numerical finding. However, the procedure was conducted "without any interruption."
Secondary Endpoint: Ease of scope insertion, advancement, and withdrawal"None of the users rated any figure of merit as 'unacceptable' or 'difficult'. All answers varied between 'excellent' to 'acceptable', while the vast majority of users' responses related to the device were rated as 'excellent'."
Secondary Endpoint: Ability to center the scope inside the gastrointestinal tractCovered under the overall user feedback: "None of the users rated any figure of merit as 'unacceptable' or 'difficult'. All answers varied between 'excellent' to 'acceptable', while the vast majority of users' responses related to the device were rated as 'excellent'."
Secondary Endpoint: Subjective evaluation of the additional area screened by the physicianNot explicitly stated as a distinct finding, but implied by the device's function to "maintain the endoscopes field of view" and user feedback on device performance.
Secondary Endpoint: Patient wellbeingNo adverse events, complications or complaints of patient discomfort were observed during the procedures or reported in the 24 hours follow up.
Device Malfunction/Reliability FailureNo report of device malfunction or reliability failure.
User Rating of Performance (General)None of the users rated any figure of merit as "unacceptable" or "difficult". Majority of users' responses rated as "excellent"; all varied between "excellent" to "acceptable".

2. Sample size used for the test set and the data provenance

  • Sample Size: 60 patients
  • Data Provenance: Prospective clinical study conducted at Elisha Medical Center, Haifa, Israel. The population was described as "diverse adult population in terms of gender, ethnicity and age."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

N/A. This study is a clinical feasibility trial of a physical device, not an AI/ML algorithm requiring expert ground truth for classification or detection tasks. The performance metrics are based on clinical outcomes, user feedback, and observations during endoscopic procedures by the medical professionals conducting the colonoscopies.

4. Adjudication method for the test set

N/A. As this is a clinical feasibility study of a physical device, there is no mention of an adjudication method for data. Clinical outcomes and user experience were evaluated directly during the procedures.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

N/A. This is a 510(k) submission for a physical medical device (EndoRings), not an AI/ML-driven diagnostic tool. Therefore, an MRMC comparative effectiveness study comparing human readers with and without AI assistance is not applicable and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

N/A. The EndoRings is a physical accessory to an endoscope, not a standalone algorithm.

7. The type of ground truth used

The "ground truth" for this study is based on direct clinical observation, objective measurements (e.g., cecum reach), and subjective assessments by the clinicians performing the procedures, as well as patient follow-up for adverse events. This includes:

  • Observational Data: Successful cecum reach, absence of complications.
  • User Feedback: Clinicians' subjective scoring of ease of use, ability to complete interventions, scope centering, etc.
  • Outcome Data: Absence of adverse events or patient discomfort during and after the procedure.

8. The sample size for the training set

N/A. There is no AI/ML component; therefore, no training set is applicable. The study involved clinical testing of a final production unit.

9. How the ground truth for the training set was established

N/A. Not applicable as there is no training set for an AI/ML algorithm.

{0}------------------------------------------------

510(k) SUMMARY

EndoAid EndoRings

FEB 1 1 2014

Applicant Information:

EndoAid, Ltd. ARAN Building P.O. Box 3067 43 Haeshel Street Caesarea 38900 Israel

Phone:(972) 4-6239014
Facsimile:(972) 4-6273260
Contact Person:Dan Rottenberg
Date Prepared:January 23, 2014

Device Information:

Trade Name:EndoRings
Common or Usual Name:Endoscopic Distal Attachment
Classification:Class II per 21 CFR 876.1500
Product Code:FED
Predicate Device:Arc EndoCuff AEC 110; AEC 120; AEC 140 (K122565)

Device Description:

The EndoRings is a flexible silicone rubber add-on device designed to attach to of the designated endoscope. The product features a short tube like shape with three flexible circular layers of wings that easily fold backwards during intubation and scope forward movement, and fold forward during scope withdrawal, to maintain the field of view and scope centering position.

Intended Use / Indications for Use:

The EndoRings is intended to be attached to the distal end of the endoscope to facilitate endoscopic therapy, to be used for the following:

  • Keeping the suitable depth of endoscope's view field

{1}------------------------------------------------

Technological Characteristics:

The technological characteristics of EndoAid EndoRings are equivalent to the characteristics of the predicate device.

Materials

The EndoRings silicone rubber material has been chosen as conforming to USP class VI and tested for biocompatibility in accordance with ISO 10993.

Size

The EndoRings length is 24mm, while the predicate Arc EndoCuff (K122565) length is 23.8mm. The main body tube outer diameter is 15.7mm for the EndoRings, and 17.2mm for the EndoCuff. There is a difference in the thin flexible wings /projections maximal free/open diameter which is 50mm for the EndoRings, and 34.8mm for the Arc EndoCuff. The wings fold easily when touching the colon walls, and the diameter of the wings can reduce to 22mm for the EndoRings compared to 17.4mm for the EndoCuff projections.

Non-clinical Data

The EndoRings is used to maintain the endoscopes field of view during endoscopy therapy by manipulating colonic folds, maintaining a central position in the lumen and avoiding sudden slippage. The nominated predicate device maximizes the field of view in a similar way.

The bench testing demonstrates that the EndoRings design combines some of the benefits of both being a slim and stabilizing cylindrical device (like the Arc EndoCuff), being able to stabilize and view the mucosa effectively on a greater diameter, without interfering with the standard field of view, since the wings are designed in a way they do not interfere with the endoscopic image.

The design of the device main cylindrical tube made from flexible polymer (silicone rubber) is the means of attachment / retention to the distal end of an endoscope (like the Endoculf).

The EndoAid EndoRings and the Arc EndoCuff both contact the patient in the same way and for the same period of time. Both are made from flexible polymers materials and both are sterilized.

The performance data showed that the sterilization method selected (EndoRings is Ethylene Oxide sterilized and Arc EndoCuff is irradiated K122565) has not introduced any additional risks and the patient contacting materials have been tested for biocompatibility with reports demonstrating no negatives in safety or effectiveness.

The bench testing and clinical study has demonstrated that the device does not introduce any additional risks when undertaking endoscopic therapy and meeting the intended use.

All of these tests demonstrated that the EndoRings meets its intended performance specifications.

{2}------------------------------------------------

Clinical Data:

The device performance was validated by testing of a final production unit in clinical conditions. The procedures were performed at Elisha Medical Center, Haifa, Israel on a diverse adult population in terms of gender, ethnicity and age. Sixty (60) patients were enrolled in conformance with the device labeling.

The purpose of this prospective clinical study was to establish the feasibility and usability of the ENDORINGSTM when used during standard colonoscopy procedures.

The primary endpoint was reaching the cecum of the colon with the ENDORINGS.

Secondary endpoints were:

    1. Incidence of complications.
    1. Ability to complete therapeutic interventions as biopsies, polypectomies, APC etc.
    1. Procedure time
    1. Ease of scope insertion, advancement and withdrawal.
    1. Ability to center the scope inside the gastrointestinal tract.
    1. Subjective evaluation of the additional area screened by the physician.
    1. Patient wellbeing

Referring to the criteria above, the study objectives were met:

  • . Reaching the cecum: In 100 % of the cases, the primary endpoint, reaching the cecum, was achieved.
  • Complications : Regarding secondary safety endpoints, there were no reports of adverse . events, colon tissue damage, or other complications during the procedure or in the 24 hours post procedure follow-up.
  • Therapeutic interventions: The ability to insert tool through the working channel and . maneuver the scope distal tip to take specimen were tested: all the applicable answers were scored as "Good" to "Excellent". All of the interventions were successfully completed without any reports of operational difficulties.

{3}------------------------------------------------

  • Device performance: The entire procedure was conducted without any interruption. There . was no report of device malfunction or reliability failure.
  • . Scoring Analysis: None of the users rated any figure of merit as "unacceptable" or "difficult". All answers varied between "excellent" to "acceptable", while the vast majority of users' responses related to the device were rated as "excellent".
  • Patient wellbeing: No adverse events, complications or complaints of patient discomfort . were observed during the procedures or reported in the 24 hours follow up

The above test results and analysis, demonstrated that the EndoRings meets the user needs in clinical environment, under defined conditions. The test pass criteria were met. Thus, the clinical study supports the substantial equivalence of the EndoRings to its predicate device.

Substantial Equivalence:

After a review of the bench and clinical testing, the company concluded that the EndoRings is as safe and effective as the predicate device for facilitating endoscopic therapy.

The EndoRings and the predicate device have the same indication for use and very similar technological characteristics, and principles of operation. The minor technological differences between the EndoRings and its predicate device raise no new types of safety or effectiveness questions. In vitro verification testing and clinical study demonstrates that the EndoRings performs as intended and meets all design specifications with respect to their mechanical, functional and handling characteristics, and that its materials are biocompatible.

Thus, the EndoAid EndoRings is substantially equivalent to the Arc EndoCuff.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 11, 2014

EndoAid, Ltd. % Janice M. Hogan Partner Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103

Re: K133359

Trade/Device Name: EndoRings Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: January 23, 2014 Received: January 23, 2014

Dear Janice M. Hogan,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{5}------------------------------------------------

Page 2 - Janice M. Hogan

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K133359

Device Name EndoRings

Indications for Use (Describe)

The EndoRings is intended to be attached to the distal end of the endoscope to facilitate endoscopic therapy, to be used for the following:

Keeping the suitable depth of endoscope's view field

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FOR FOR FOR FOR FOR USE ONLY Sandard .. -Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/6/Picture/12 description: The image shows the name "Herbert P. Lerner" and the date and time "2014.02.11 11:22:04 -05:00". The text is in a clear, sans-serif font and is easily readable. The date and time are formatted in a specific way, with the year, month, and day separated by periods, and the time including hours, minutes, seconds, and timezone.

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.