(62 days)
755nm
The GentleMAX Family of Laser Systems is indicated for temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias.
1064nm
The GentleMAX Family of Laser Systems is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to. lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
The Candela GentleMAX Family of Laser Systems contains two separate laser heads (Alexandrite and Nd: YAG), which produce laser light outputs of 755 nm and 1064 nm, respectively. The output of each laser head is optically combined on the laser rail, so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system which can output either 755 nm or 1064 nm wavelengths. The laser system creates a beam of high intensity light that penetrates deep into the skin tissue where it delivers a controlled amount of therapeutic heat. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.
1. Acceptance Criteria and Device Performance:
| Feature/Characteristic | Acceptance Criteria (Predicate Device K112715) | Reported Device Performance (Modified GentleMax Pro Laser System) |
|---|---|---|
| Wavelength | 755nm/1064nm | 755nm/1064nm |
| Pulse duration | 0.25 - 300 ms | 0.25 - 100 ms |
| Maximum fluence | 53 J/cm² (ALEX), 80 J/cm² (YAG) | 53 J/cm² (ALEX), 80 J/cm² (YAG) |
| Spot size | 1.5, 3, 6, 8, 10, 12, 15, 18mm | 1.5, 3, 6, 8, 10, 12, 15, 18mm |
| Pulse repetition rate | 10 Hz, maximum | 10 Hz, maximum |
| Pulsing control | Fingerswitch or footswitch | Fingerswitch or footswitch |
| Indications for Use | Same as predicate device (detailed below) | Same as predicate device (detailed below) |
Detailed Indications for Use (both predicate and modified device):
- 755nm:
- Temporary hair reduction.
- Stable long-term or permanent hair reduction (defined as long-term, stable reduction in the number of hairs regrowing at 6, 9, and 12 months post-treatment) on all skin types (Fitzpatrick I-VI) including tanned skin.
- Treatment of benign pigmented lesions.
- Treatment of wrinkles.
- Photocoagulation of dermatological vascular lesions (e.g., port-wine stains, hemangiomas, telangiectasias).
- 1064nm:
- Removal of unwanted hair, stable long-term or permanent hair reduction, and treatment of PFB (defined as long-term, stable reduction in the number of hairs regrowing at 6, 9, and 12 months post-treatment) on all skin types Fitzpatrick I-VI including tanned skin.
- Photocoagulation and hemostasis of pigmented and vascular lesions such as port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins.
- Coagulation and hemostasis of soft tissue.
- Benign pigmented lesions (e.g., lentigos, solar lentigos, café au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos - significant reduction in black and/or blue-black tattoos, and plaques).
- Indicated for pigmented lesions to reduce lesion size, for patients benefiting from aggressive treatment, and for patients not responding to other laser treatments.
- Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
- Treatment of wrinkles.
2. Sample Size and Data Provenance:
- Sample size for the test set: Not applicable. The submission states "No clinical data was required for these modifications." The evaluation was based on non-clinical performance data and a comparison to a predicate device.
- Data provenance: Not explicitly stated as clinical data was not required. However, the non-clinical tests (EMC, electrical safety, software verification and validation) would have been performed by the manufacturer, Candela Corporation. This indicates prospective testing of the modified device's performance against established specifications. The original predicate device's performance would be historical (retrospective) for comparison.
3. Number of Experts and Qualifications for Ground Truth:
- Not applicable for this submission. No clinical studies requiring expert ground truth were conducted for these modifications. The evaluation focused on comparing technical specifications and performance data to a predicate device, not on clinical outcomes assessed by experts.
4. Adjudication Method for the Test Set:
- Not applicable as no clinical test set requiring adjudication by experts was used. The assessment appears to be based on engineering and safety testing.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No. The document explicitly states, "No clinical data was required for these modifications." Therefore, an MRMC study comparing human readers with and without AI assistance was not performed.
6. Standalone Performance Study:
- Yes, in a non-clinical context. The "Performance Data" section indicates that "All performance testing demonstrated that the GentleMax Pro Laser System performs according to specifications and functions as intended." This refers to electromagnetic compatibility (EMC), electrical safety, and software verification and validation tests run on the device itself. This can be considered a standalone performance evaluation of the device's technical functionality, distinct from clinical efficacy.
7. Type of Ground Truth Used:
- For the non-clinical testing of the modified device, the "ground truth" was the pre-defined engineering specifications, safety standards (e.g., EMC, electrical safety), and validated software requirements.
- For the overall substantial equivalence claim, the "ground truth" or reference was the performance and safety profile of the legally marketed predicate device (K112715). The modified device was deemed substantially equivalent because its non-clinical tests confirmed it performed "as well as the predicate device" and raised no new safety or effectiveness issues.
8. Sample Size for the Training Set:
- Not applicable. This submission describes modifications to an existing laser system and relies on non-clinical testing and comparison to a predicate device, not the development or training of an AI algorithm or statistical model that would require a training set of data.
9. How Ground Truth for the Training Set Was Established:
- Not applicable, as no training set was used.
{0}------------------------------------------------
Section 7 - 510(k) Summary
510(k) Summary GentleMAX Family of Laser Systems
General Information:
This Special 510(k) is to provide notification of substantial equivalence for the modified GentleMAX Family of Laser Systems (GentleMax Pro Laser System) to the previously cleared GentleMAX Family of Laser Systems manufactured by Candela Corporation.
| Submitted by: | Candela Corporation530 Boston Post RoadWayland, MA 01778-1886 |
|---|---|
| Contact Person: | Sam Wade,Global Vice President, Regulatory AffairsTel: 508-358-7400 x330Fax: 508-358-5602 |
| Date prepared: | October 23, 2013 |
| Trade Name: | GentleMAX Family of Laser Systems |
| Common Name: | Dermatology Laser System |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery anin dermatology (21 CFR 878.4810, Product Code GEX) |
| Predicate Devices: | GentleMAX Family of Laser Systems (K112715) |
Intended Use / Indications for Use:
The GentleMAX Family of Laser Systems is indicated for the following:
755nm
The GentleMAX Family of Laser Systems is indicated for temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias.
GentleMAX Family of Laser Systems
7-1
{1}------------------------------------------------
Section 7 - 510(k) Summary
1064nm
The GentleMAX Family of Laser Systems is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to. lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
Description:
The Candela GentleMAX Family of Laser Systems contains two separate laser heads (Alexandrite and Nd: YAG), which produce laser light outputs of 755 nm and 1064 nm, respectively. The output of each laser head is optically combined on the laser rail, so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system which can output either 755 nm or 1064 nm wavelengths. The laser system creates a beam of high intensity light that penetrates deep into the skin tissue where it delivers a controlled amount of therapeutic heat. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.
Technological Characteristics:
The Candela GentleMAX Family of Laser Systems delivers laser energy through an optical fiber handpiece delivery system, which can output either 755 nm or 1064 nm wavelengths. The output of this laser is delivered to the area of treatment by means of a lens coupled user replaceable optical fiber with a treatment handpiece attached to its distal end. A trigger switch (fingerswitch or footswitch) is used to control the delivery of laser pulses. The Dynamic Cooling Device provides a short burst of cryogen spray prior to firing the laser pulse. The laser output energy is delivered via an optical fiber to a handpiece, which produces circular beams with diameters of 1.5, 3, 6, 8, 10, 12, 15 and 18 millimeters on the skin. The cryogen, which is housed within the laser enclosure, is delivered via a hose to a nozzle located in the handpiece.
{2}------------------------------------------------
Section 7 – 510(k) Summary
| Product | Modified GentleMAX Family ofLaser Systems (GentleMax ProLaser System) | GentleMAX Family of LaserSystems | |
|---|---|---|---|
| 510(k) | KXXXXXXX | K112715 | |
| Manufacturer | Candela Corporation | Candela Corporation | |
| Product Code | GEX | GEX | |
| RegulationNumber. | 21 CFR 878.4810 | 21 CFR 878.4810 | |
| Device Class | II | II | |
| Indications forUse | 755nm | 755nm | |
| The GentleMAX Family of LaserSystems is indicated for temporaryhair reduction. Stable long-term orpermanent reduction throughselective targeting of melanin inhair follicles. Permanent hairreduction is defined as long-termstable reduction in the number ofhairs regrowing after a treatmentregime. On all skin types(Fitzpatrick I- VI) including tannedskin. Permanent hair reduction isdefined as the long-term, stablereduction in the number of hairsregrowing when measured at 6, 9,and 12 months after the completionof a treatment regime.Treatment of benign pigmentedlesions.Treatment of wrinkles.The photocoagulation ofdermatological vascular lesions(such as port-wine stains,hemangiomas, telangiectasias.1064nmThe GentleMAX Family of LaserSystems is indicated for theremoval of unwanted hair, forstable long term or permanent hairreduction and for treatment of PFB.Permanent hair reduction is definedas the long-term, stable reduction inthe number of hairs regrowingwhen measured at 6 9 and 12 | The GentleMAX Family of LaserSystems is indicated for temporaryhair reduction. Stable long-term orpermanent reduction throughselective targeting of melanin inhair follicles. Permanent hairreduction is defined as long-termstable reduction in the number ofhairs regrowing after a treatmentregime. On all skin types(Fitzpatrick I- VI) including tannedskin. Permanent hair reduction isdefined as the long-term, stablereduction in the number of hairsregrowing when measured at 6, 9,and 12 months after the completionof a treatment regime.Treatment of benign pigmentedlesions.Treatment of wrinkles.The photocoagulation ofdermatological vascular lesions(such as port-wine stains,hemangiomas, telangiectasias.1064nmThe GentleMAX Family of LaserSystems is indicated for theremoval of unwanted hair, forstable long term or permanent hairreduction and for treatment of PFB.Permanent hair reduction is definedas the long-term, stable reduction inthe number of hairs regrowingwhan measured at 6 9 and 12 | ||
| months after the completion of atreatment regime. The lasers areindicated on all skin typesFitzpatrick I-VI including tannedskin. Photocoagulation andhemostasis of pigmented andvascular lesions such as but notlimited to port wine stains,hemangioma, warts, telangiectasia,rosacea, venus lake, leg veins andspider veins. Coagulation andhemostasis of soft tissue. Benignpigmented lesions such as, but notlimited to, lentigos (age spots),solar lentigos (sun spots), café aulait macules, seborrheic keratosis,nevi, chloasma, verrucae, skin tags,keratosis, tattoos (significantreduction in the intensity of blackand/or blue-black tattoos) andplaques.The laser is indicated for pigmentedlesions to reduce lesion size, forpatients with lesions that wouldpotentially benefit from aggressivetreatment, and for patients withlesions that have not responded toother laser treatments.The laser is also indicated for thereduction of red pigmentation inhypertrophic and keloid scarswhere vascularity is an integral partof the scar.Treatment of wrinkles. | months after the completion of atreatment regime. The lasers areindicated on all skin typesFitzpatrick I-VI including tannedskin. Photocoagulation andhemostasis of pigmented andvascular lesions such as but notlimited to port wine stains,hemangioma, warts, telangiectasia,rosacea, venus lake, leg veins andspider veins. Coagulation andhemostasis of soft tissue. Benignpigmented lesions such as, but notlimited to, lentigos (age spots),solar lentigos (sun spots), café aulait macules, seborrheic keratosis,nevi, chloasma, verrucae, skin tags,keratosis, tattoos (significantreduction in the intensity of blackand/or blue-black tattoos) andplaques.The laser is indicated for pigmentedlesions to reduce lesion size, forpatients with lesions that wouldpotentially benefit from aggressivetreatment, and for patients withlesions that have not responded toother laser treatments.The laser is also indicated for thereduction of red pigmentation inhypertrophic and keloid scarswhere vascularity is an integral partof the scar.Treatment of wrinkles. | ||
| Laser type | Flashlamp-excited, Solid stateAlexandrite and Nd:YAG laser | Flashlamp-excited, Solid stateAlexandrite and Nd:YAG laser | |
| Wavelength | 755nm/1064nm | 755nm/1064nm | |
| Pulse duration | 0.25 - 100 ms | 0.25 - 300 ms | |
| Maximumfluence | 53 J/cm² (ALEX)80 J/cm² (YAG) | 53 J/cm² (ALEX)80 J/cm² (YAG) | |
| Spot size | 1.5, 3, 6, 8, 10, 12, 15, 18mm | 1.5, 3, 6, 8, 10, 12, 15, 18mm | |
| Pulse repetitionrate | 10 Hz, maximum | 10 Hz, maximum | |
| Pulsing control | Fingerswitch or footswitch | Fingerswitch or footswitch | |
| Productdimensions | 42" x 18" x 27" | 45" x 26.5" x 38" | |
| (HxWxL) | |||
| ProductWeight | 260 lbs | 341 lbs |
GentleMAX Family of Laser Systems
7-3
{3}------------------------------------------------
{4}------------------------------------------------
Section 7 - 510(k) Summary
Performance Data:
Minor software and hardware modifications were made to support the changes for the modified GentleMAX Family of Laser Systems (GentleMax Pro Laser System). The testing performed to evaluate these modifications included electromagnetic compatibility (EMC), electrical safety, and software verification and validation. All performance testing demonstrated that the GentleMax Pro Laser System performs according to specifications and functions as intended.
Clinical Data:
No clinical data was required for these modifications.
Summary of Substantial Equivalence:
The modifications made to the GentleMAX Family of Laser Systems do not affect the indications for use or alter the fundamental scientific technology of the device, nor does it affect the mode of use. There are no labeling changes that affect the indications for use of the device. The modifications made to the GentleMAX Family of Laser Systems raises no new issues of safety or effectiveness. The modified GentleMAX Family of Laser Systems have the samc intended uses, utilizes similar operating principles, and matches key design aspects, including similar spot size, the same wavelength and the same maximum delivered fluence as the predicate devices. The non-clinical tests that were performed that our device is as safe, as effective, and performs as well as the predicate device. On the basis of similarities in methods of assembly, method of operation, and intended uses, as confirmed by the performance testing, Candela Corporation believes that the modified GentleMAX Family of Laser Systems is substantially equivalent to the predicate device.
GentleMAX Family of Laser Systems
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Cemer - WO66-G609 Silver Spring, MD 20993-0002
December 26, 2013
Candela Corporation Mr. Sam Wade Global Vice President, Regulatory Affairs 530 Boston Post Road Wayland, Massachusetts 01778
Re: K133283
Trade/Device Name: Gentlemax Family of Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: December 6, 2013 Received: December 9, 2013
Dear Mr. Wade:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Mr. Sam Wade
ﻪ ﻣﺮﺩﺍﺩ ﺍ
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Section 6 - Indications for Use Statement
Indications for Use Statement
510(k) Number (if known):
GentleMAX Family of Laser Systems Device Name:
Indications for Use:
755mm
.. - ---
The GentleMAX Family of Laser Systems is indicated for temporary hair reduction. Stable longterm or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias)
1064nm
The GentleMAX Family of Laser Systems is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
GentleMAX Family of Laser Systems
| Jitendra V. Virani -S | |
|---|---|
| 2013.12.26 11:08:06 -05'00' | 6-1 |
| Division Sign-Off | |
| Division of Surgical Devices | A2 - 1 |
| 510(k) Number K133283 |
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.