K Number
K133175
Date Cleared
2014-11-13

(392 days)

Product Code
Regulation Number
862.1113
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Jaundice Meter (JM-105) is a non-invasive transcutaneous bilirubinometer. It measures yellowness of subcutaneous tissue in newborn infants. The unit pro-vides a visual digital measurement that has been shown to correlate with serum bilirubin in newborn infants.

The device is intended for use in hospitals, clinics or doctor's offices under a physicians supervision / direction to assist clinicians in monitoring of newborn infants. The device is not intended as a standalone for diagnosis of hyperbilirubinemia. It is to be used in conjunction with other clinical signs and laboratory measurements.

Newborn infants whose JM-105 Jaundice Meter test results are indicative of hy-perbilirybinemia should be evaluated by their physician(s) for appropriate patient management. Specific neonatal patient Bilirubin levels should be confirmed by other methods, such as serum bilirubin, prior to treatment determinations.

The JM 105 is a prescription medical device

  • The JM 105 is not intended for home use.
    The JM 105 may only be used at the sternum measurement site for Physician's office applications.
Device Description

The Jaundice Meter is a non-invasive transcutaneous bilirubinometer. It measures yellowness of subcutaneous tissue in newborn infants. The unit provides a visual digital measurement that has been shown to correlate with serum bilirubin in newborn infants.

The JM-105 is a portable, hand held, battery powered device that includes a docking station with a built in reading checker. The JM-105 batteries can be charged using a battery charger or an optional USB cable.

The basic functionality including measurement of the JM-105 is equivalent to the JM-103. The display of the JM-105 has been improved (larger screen, touchscreen) and data storage and transmission functionality was added. The measuring probe, hardware, and software used to process the measurements are identical and therefore use the same measuring principle. The JM-103 and JM-105 determines the yellowness of subcutaneous tissue by using two optical paths to measure the optical density difference at two wavelengths. The measuring principle is further described in the "Principles of Operations" section of the Instructions for Use.

In addition to the features offered with the JM-103, the JM-105 provides the following.

  • Internal memory up to 100 patient files
  • Data transfer via HL7
  • Easily mark & ID babies that need special attention with patient flagging
  • Cost-efficient screening

There are no sterile or single-use components or accessories for the JM-105.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the Draeger Medical Systems Jaundice Meter JM-105, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (JM-103, as JM-105 is derived)Reported Device Performance (JM-105, same as JM-103)
Measuring range: 0.0 to 20.0 mg/dL or 0 to 340 umol/LMeasuring range: 0.0 to 20.0 mg/dL or 0 to 340 umol/L
Accuracy (Clinical Data Standard Error of Estimate): ±1.5 mg/dL or ±25.5umol/L (>35 weeks gestation)Accuracy (Clinical Data Standard Error of Estimate): ±1.5 mg/dL or ±25.5umol/L (>35 weeks gestation)

Notes on Acceptance Criteria: The document primarily refers to the JM-103's performance criteria as the basis for substantial equivalence for the JM-105, stating that the measuring principle and core functionality are identical. The accuracy mentioned is specifically for infants > 35 weeks gestation.

2. Sample Size Used for the Test Set and Data Provenance

  • Hospital Studies (JM-103):

    • Sample Size: Not explicitly stated as a number of patients or measurements. The document mentions "All patients meeting the above criteria were included in the study."
    • Data Provenance: Not explicitly stated, but the context of an FDA submission for a US market device generally implies studies either conducted in or accepted by the US regulatory framework. It's retrospective inference, as the studies were performed previously for the predicate JM-103.
    • Retrospective/Prospective: The initial studies for the JM-103 would have been prospective, as new data was collected to demonstrate performance. The JM-105 relies on this previously collected data.
  • Doctor's Office Studies (JM-103):

    • Sample Size: Not explicitly stated as a number of patients or measurements.
    • Data Provenance: Not explicitly stated, but implies clinical sites (doctor's offices) where the device was intended for use.
    • Retrospective/Prospective: Same as hospital studies, likely prospective for the JM-103, now serving as retrospective for JM-105.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Number of Experts: Not explicitly stated.
  • Qualifications of Experts: Not explicitly stated. The ground truth method (serum bilirubin) implies that clinical laboratory professionals would be involved in generating the comparison data, and physicians would be involved in determining the need for serum bilirubin tests.

4. Adjudication Method for the Test Set

  • No specific adjudication method (e.g., 2+1, 3+1) is mentioned or implied for either the transcutaneous bilirubin readings or the serum bilirubin ground truth. The comparison is directly between the device measurement and the laboratory measurement.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study is mentioned. The device, a Jaundice Meter, provides a direct measurement and is not an imaging or diagnostic AI requiring human interpretation or assistance for its primary function. Its use is to assist clinicians in monitoring, not to replace their diagnostic decision-making or enhance their reading of complex medical images. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not directly apply in this context.

6. Standalone Performance Study (Algorithm only without human-in-the-loop performance)

  • Yes, the clinical performance data described for the JM-103 (and relied upon for the JM-105) is a standalone performance study. The device provides a direct measurement (transcutaneous bilirubin), which is then compared against the established ground truth (serum bilirubin). There is no "human-in-the-loop" interaction for interpreting the device's reading itself; the reading is digital and absolute. The clinicians then use this reading in conjunction with other clinical signs.

7. Type of Ground Truth Used

  • Ground Truth: Laboratory measured total serum bilirubin (TSB).

8. Sample Size for the Training Set

  • The document does not describe a "training set" in the context of an AI/ML algorithm. The JM-103/JM-105 is a device based on an optical measurement principle, not a machine learning model that undergoes a training phase. Its internal parameters would be set during design and calibration, not through iterative training on a dataset in the modern AI sense.

9. How the Ground Truth for the Training Set Was Established

  • As noted above, there is no mention of a "training set" for an AI/ML algorithm. The device's measurement principle (optical density difference at two wavelengths) is based on established biophysical principles, not an AI model that learns from data.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine. The caduceus is made up of three intertwined snakes and a staff.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 13, 2014

DRAEGER MEDICAL SYSTEMS, INC. BRYAN OVERTON DIRECTOR, QA AND COMPLIANCE 3135 QUARRY RD TELFORD PA 18969

Re: K133175

Trade/Device Name: Jaundice Meter JM-105 Regulation Number: 21 CFR 862.1113 Regulation Name: Bilirubin (total and unbound) in the neonate test system Regulatory Class: I, reserved Product Code: MOM Dated: September 26, 2014 Received: September 26, 2014

Dear Mr. Overton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Courtney

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K133175

Device Name Jaundice Meter JM-105

Indications for Use (Describe)

The Jaundice Meter (JM-105) is a non-invasive transcutaneous bilirubinometer. It measures yellowness of subcutaneous tissue in newborn infants. The unit pro-vides a visual digital measurement that has been shown to correlate with serum bilirubin in newborn infants.

The device is intended for use in hospitals, clinics or doctor's offices under a physicians supervision / direction to assist clinicians in monitoring of newborn infants. The device is not intended as a standalone for diagnosis of hyperbilirubinemia. It is to be used in conjunction with other clinical signs and laboratory measurements.

Newborn infants whose JM-105 Jaundice Meter test results are indicative of hy-perbilirybinemia should be evaluated by their physician(s) for appropriate patient management. Specific neonatal patient Bilirubin levels should be confirmed by other methods, such as serum bilirubin, prior to treatment determinations.

The JM 105 is a prescription medical device

  • The JM 105 is not intended for home use.
    The JM 105 may only be used at the sternum measurement site for Physician's office applications.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

acc. to 807.92

Submitter's Name and Address:Draeger Medical Systems, Inc3135 Quarry RoadTelford, PA 18969 USA
Contact Person:Bryan OvertonPhone: 215.660.2135Fax: 215.721.5424
Date submission was prepared:November 6, 2014
Device Name:Proprietary Name: Jaundice Meter JM-105Common Name: Transcutaneous BilirubinometerClassification Name: Bilirubin (total and unbound) in the neonate test systemRegulation Number: 21 CFR 862.1113Product Code: MQM
Class:I, reserved

Predicate Devices:

510(k) NumberDevice Name
K042522Draeger JM-103 Jaundice Meter

Intended Use/Indications for Use

The Jaundice Meter (JM-105) is a non-invasive transcutaneous bilirubinometer. It measures yellowness of subcutaneous tissue in newborn infants. The unit provides a visual digital measurement that has been shown to correlate with serum bilirubin in newborn infants.

The device is intended for use in hospitals, clinics or doctor's offices under a physicians supervision / direction to assist clinicians in monitoring of newborn infants. The device is not intended as a standalone for diagnosis of hyperbilirubinemia. It is to be used in conjunction with other clinical signs and laboratory measurements.

Newborn infants whose JM-105 Jaundice Meter test results are indicative of hyperbilirubinemia should be evaluated by their physician(s) for appropriate patient management. Specific neonatal patient Bilirubin levels should be confirmed by other methods, such as serum bilirubin, prior to treatment determinations.

The JM 105 is a prescription medical device The JM 105 is not intended for home use.

{4}------------------------------------------------

510(k) Summarv JM-105

Image /page/4/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue. The word is the logo for the Dräger company.

The JM 105 may only be used at the sternum measurement site for Physician's office applications.

Further information on application

Newborn infants whose Jaundice Meter test results are indicative of hyperbilirubinemia should be evaluated by their physicians for appropriate patient management. Specific neonatal patient bilirubin levels should be confirmed by other methods, such as serum bilirubin, before treatment determinations.

Limitations (Doctors Office Use)

Use only on infants up to 14 days of age.

For doctors' office application, use only the sternum location when taking measurements.

Please be aware, performance in doctors' offices may vary from performance in hospitals.

Device Description:

The Jaundice Meter is a non-invasive transcutaneous bilirubinometer. It measures yellowness of subcutaneous tissue in newborn infants. The unit provides a visual digital measurement that has been shown to correlate with serum bilirubin in newborn infants.

The JM-105 is a portable, hand held, battery powered device that includes a docking station with a built in reading checker. The JM-105 batteries can be charged using a battery charger or an optional USB cable.

The basic functionality including measurement of the JM-105 is equivalent to the JM-103. The display of the JM-105 has been improved (larger screen, touchscreen) and data storage and transmission functionality was added. The measuring probe, hardware, and software used to process the measurements are identical and therefore use the same measuring principle. The JM-103 and JM-105 determines the yellowness of subcutaneous tissue by using two optical paths to measure the optical density difference at two wavelengths. The measuring principle is further described in the "Principles of Operations" section of the Instructions for Use.

In addition to the features offered with the JM-103, the JM-105 provides the following.

  • Internal memory up to 100 patient files
  • Data transfer via HL7
  • Easily mark & ID babies that need special attention with patient flagging
  • Cost-efficient screening

There are no sterile or single-use components or accessories for the JM-105.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Dräger" in blue. The font is bold and sans-serif. The two dots above the "a" are also blue. The word is centered in the image.

JM-105 Comparison to JM-103 Predicate

Draeger JM-105Draeger JM-103
Intended UseThe Jaundice Meter (JM-105) isa non-invasive transcutaneousbilirubinometer. It measuresyellowness of subcutaneoustissue in newborn infants.Same
Special condition for useThe Jaundice Meter is indicatedfor use in neonatal patients born>35 weeks gestation who havenot undergone transfusion orphototherapy treatment.Same
Measuring MethodDetermines the yellowness ofsubcutaneous tissue by usingtwo optical paths to measure theoptical density difference at twowavelengthsSame
Measuring range0.0 to 20.0 mg/dL or 0 to 340umol/LSame
Accuracy (Clinical DataStandard Error of Esti-mate)±1.5 mg/dL or ±25.5umol/L(>35 weeks gestation)Same
Light SourcePulse Xenon Arc LampSame
SensorsSilicon photodiodesSame
Power Source2.4V, Special Ni-MH batterySame
Protection type & levelInternally powered instrument,BF typeSame
Dimensions56mm(W)×168mm(H)×45mm(D)48mm(W)x154mm(H)x32mm (D)
Weight203g including Ni-MH battery)150g including Ni-MH battery)
Averaging functionSingle, Avg. 1, 2, 3, 4 or 5Same
Data Storage and Trans-missionYesNo

Substantial Equivalence:

The JM-105 and the above referenced predicate device have the same intended use. These devices are used by health care professionals as a screening device used on newborn infants to detect hyperbilirubinemia.

The measurement principle used in JM-105 is equivalent to that of the JM-103, and has been proven to be effective in screening infants > 35 weeks gestation.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The word is written in all capital letters except for the "ä", which has two dots above it. The font is sans-serif and the letters are closely spaced together. The background is white.

Summary of Nonclinical Testing

This submission includes the results of testing for the JM-105 to demonstrate compliance to the product requirements. This includes electrical, mechanical, and software testing.

Testing was also performed to assess compliance to the following recognized consensus standards:

  • IEC60601-1, Medical Electrical Equipment, Part 1: General Requirements . for Safetv
  • IEC60601-1-2, Medical Electrical Equipment, Part 1-2: General Require-. ments for Safety Collateral Standard: Electromagnetic Compatibility

The test reports and results are located in section 10 of the submission.

Clinical Performance

The clinical performance data is provided in the JM-103 510(k) premarket notification (see K042522). The objective of the studies were to determine the accuracy and precision of transcutaneous bilirubin (TcB) measurements in infants neonates.

Hospital Study Design

Selection criteria

The patient selection criteria used for the studies included infants less than 30 days old and weighing greater than 1000 grams. Although the selection criteria was established as "less than 30 days of age," the infants in the hospital studies were primarily NICU and newborn infants unless their medical condition required a longer duration of care. The test was performed on infants who were determined by their physician to require a serum bilirubin test.

Demographics of patient population

All patients meeting the above criteria were included in the study. There was significant effort to ensure sufficient representation of all skin pigmentation to verify that the JM-103 could be used across all populations with consistent results. The demographics of the patient population included Caucasian, African-American, East-Asian, Indian-Pakistani, and Hispanic infants.

Doctor's Office Study Design

Studies were performed at two doctors' office sites comparing JM-103 Total Calculated Bilirubin (TcB) to laboratory measured total serum bilirubin (TSB).

Selection criteria

The ages of the infants in the study ranged from approximately 24 hours to 7-10 days, with a mean of 3 days (at site 1) and 5 days (at site 2). The

1374 :

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the word "Dräger" in blue. The font is bold and sans-serif. The two dots above the "a" are also blue. The word is likely a logo for the Dräger company, which is a German company that makes medical and safety technology.

test was performed on infants who were determined by their physician to require a serum bilirubin test.

Conclusions

Based on the data and information presented in this submission, the JM-105 is substantially equivalent to the currently legally marketed to predicate device referenced above.

§ 862.1113 Bilirubin (total and unbound) in the neonate test system.

(a)
Identification. A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus).(b)
Classification. Class I.